Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol

Cuckson, Alexandra C., Reinders, Annemarie, Shabeeh, Husain, Shennan, Andrew H. · 2002 · Blood Pressure Monitoring
Cited by 123

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

BACKGROUND: The market for devices for the self-measurement of blood pressure is growing, and as accuracy is of prime importance, there is increasing pressure for manufacturers to provide evidence of independent testing. Recent reviews have shown that only five automated upper arm devices for self-measurement of blood pressure have been recommended for use. We tested the Microlife BP 3BTO-A, a lightweight, upper arm, automated oscillometric device, according to a modified version of the British Hypertension Society protocol and also analysed the computer-generated oscillograms for possible causes of inaccuracy. METHODS: One hundred and twenty-six subjects were recruited from general medical and specialist clinics and from amongst the staff at Guy's and St Thomas' Hospital, London, UK. Only 85 of these were included in the final analysis. Nine sequential readings were taken by two trained observers alternating between the mercury sphygmomanometer and the device. The last seven readings were analysed according to the British Hypertension Society protocol. Modifications to the protocol were: (1) the exclusion of patients whose blood pressure varied by more than 15 mmHg between sequential observer readings and (2) limited testing in the low systolic pressure range. RESULTS: The Microlife achieved a grade A for both systolic and diastolic pressure according to the British Hypertension Society protocol. The mean differences (standard deviation) between the observers and the device were -1.6 (7.7) mmHg and -2.1 (6.3) mmHg for systolic and diastolic blood pressure, respectively, therefore also fulfilling the criteria set by the Association for the Advancement of Medical Instrumentation. Sub-analysis for different pressure ranges showed that the device was less accurate in the high-pressure range (>160/100 mmHg). CONCLUSION: The Microlife can be recommended for clinical use in an adult population.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00126097-200212000-00005

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 123

Metadata

Authors
Cuckson, A. C., Reinders, A., Shabeeh, H., & Shennan, A. H.
Venue
Blood Pressure Monitoring
Published
2002-12
Volume
7
Issue
6
Pages
319-324
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2161305236
Cited by
123
References
2
FWCI
1.45
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol. https://app.cochat.ai/papers/10.1097/00126097-200212000-00005
The paper
Cuckson, A. C., Reinders, A., Shabeeh, H., & Shennan, A. H. (2002). Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol. Blood Pressure Monitoring. https://doi.org/10.1097/00126097-200212000-00005

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.