Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol

Coleman, Andrew, Steel, Stephen, Freeman, Paul, de Greeff, Annemarie, Shennan, Andrew · 2008 · Blood Pressure Monitoring

Abstract

BACKGROUND: The Omron M7 (HEM-780-E) is an automated oscillometric upper arm blood pressure monitor for the professional and home use markets. The aim of this study was to validate the accuracy of this device according to the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI) SP10 validation criteria. METHODS: Participants were recruited until a total of 85 were obtained that filled the blood pressure ranges specified by the BHS protocol. Recruitment to the study was from the general medical and specialist clinics and from the staff at Guy's and St Thomas' Hospital in London, UK. Nine sequential same-arm blood pressure readings were taken from each participant by two trained observers, alternating between a mercury reference sphygmomanometer and the Omron M7 (HEM-780-E). The differences between the reference and test device readings, for both systolic and diastolic pressures, were compared with BHS and AAMI criteria to determine the outcome of the study. RESULTS: The Omron M7 (HEM-780-E) is graded 'A' for systolic and 'A' for diastolic blood pressures according to the BHS criteria. The mean (standard deviation) of the difference between the observer and the device measurements was 0.75+/-6.5) mmHg for systolic and 1.33+/-5 mmHg for diastolic pressures. The device, therefore, also satisfies the AAMI SP10 standard for the study population, which requires differences of <+/-5 (8) mmHg. CONCLUSION: The Omron M7 (HEM-780-E) achieved an 'A/A' performance classification under the BHS criteria and passed the AAMI requirements for the study population. It can be recommended for professional and home-use in this population.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0b013e3282cb57b6

Cites

13 of 13 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 13 references

Cited by

See all 259

Metadata

Authors
Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A.
Venue
Blood Pressure Monitoring
Published
2008-02
Volume
13
Issue
1
Pages
49-54
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2107967797
Cited by
259
References
13
FWCI
10.34
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. https://app.cochat.ai/papers/10.1097/mbp.0b013e3282cb57b6
The paper
Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A. (2008). Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e3282cb57b6

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.