Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol
Abstract
BACKGROUND: The Omron M7 (HEM-780-E) is an automated oscillometric upper arm blood pressure monitor for the professional and home use markets. The aim of this study was to validate the accuracy of this device according to the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI) SP10 validation criteria. METHODS: Participants were recruited until a total of 85 were obtained that filled the blood pressure ranges specified by the BHS protocol. Recruitment to the study was from the general medical and specialist clinics and from the staff at Guy's and St Thomas' Hospital in London, UK. Nine sequential same-arm blood pressure readings were taken from each participant by two trained observers, alternating between a mercury reference sphygmomanometer and the Omron M7 (HEM-780-E). The differences between the reference and test device readings, for both systolic and diastolic pressures, were compared with BHS and AAMI criteria to determine the outcome of the study. RESULTS: The Omron M7 (HEM-780-E) is graded 'A' for systolic and 'A' for diastolic blood pressures according to the BHS criteria. The mean (standard deviation) of the difference between the observer and the device measurements was 0.75+/-6.5) mmHg for systolic and 1.33+/-5 mmHg for diastolic pressures. The device, therefore, also satisfies the AAMI SP10 standard for the study population, which requires differences of <+/-5 (8) mmHg. CONCLUSION: The Omron M7 (HEM-780-E) achieved an 'A/A' performance classification under the BHS criteria and passed the AAMI requirements for the study population. It can be recommended for professional and home-use in this population.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Blood pressure measuring devices: recommendations of the European Society of Hypertension
O'Brien, E., Waeber, B., Parati, G., Staessen, J., & Myers, M. G. (2001) · BMJ · doi:10.1136/bmj.322.7285.531 · Cited by 903
OʼBrien, E., Pickering, T., Asmar, R., Myers, M., Parati, G., Staessen, J., Mengden, T., Imai, Y., Waeber, B., Palatini, P., et al. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200202000-00002 · Cited by 744
O??Brien, E., Petrie, J., Little, W., de Swiet, M., Padfield, P. L., Altma, D. G., Bland, M., Coats, A., & Atkins, N. (1993) · Journal of Hypertension · doi:10.1097/00004872-199306000-00013 · Cited by 519
El Assaad, M. A., Topouchian, J. A., & Asmar, R. G. (2003) · Blood Pressure Monitoring · doi:10.1097/00126097-200306000-00006 · Cited by 224
Coleman, A., Freeman, P., Steel, S., & Shennan, A. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000189788.05736.5f · Cited by 179
Self-citationState of the Market: A Review of Ambulatory Blood Pressure Monitoring Devices
O’Brien, E., Atkins, N., & Staessen, J. (1995) · Hypertension · doi:10.1161/01.hyp.26.5.835 · Cited by 158
A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor
Jones, C., Taylor, K., Chowienczyk, P., Poston, L., & Shennan, A. H. (2000) · Blood Pressure Monitoring · doi:10.1097/00126097-200008000-00007 · Cited by 129
Self-citationCuckson, A. C., Reinders, A., Shabeeh, H., & Shennan, A. H. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200212000-00005 · Cited by 121
Self-citation
Cited by
Green, B. B. (2008) · JAMA · doi:10.1001/jama.299.24.2857 · Cited by 659
McManus, R. J., Mant, J., Bray, E. P., Holder, R., Jones, M. I., Greenfield, S., Kaambwa, B., Banting, M., Bryan, S., Little, P., et al. (2010) · The Lancet · doi:10.1016/s0140-6736(10)60964-6 · Cited by 597
Smeets, T., Cornelisse, S., Quaedflieg, C. W. E. M., Meyer, T., Jelicic, M., & Merckelbach, H. (2012) · Psychoneuroendocrinology · doi:10.1016/j.psyneuen.2012.04.012 · Cited by 376
Kallioinen, N., Hill, A., Horswill, M. S., Ward, H. E., & Watson, M. O. (2017) · Journal of Hypertension · doi:10.1097/hjh.0000000000001197 · Cited by 357
Nocturnal Hypertension: New Technology and Evidence
Kario, K. (2018) · Hypertension · doi:10.1161/hypertensionaha.118.10971 · Cited by 298
Metadata
- Authors
- Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A.
- Venue
- Blood Pressure Monitoring
- Published
- 2008-02
- Volume
- 13
- Issue
- 1
- Pages
- 49-54
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2107967797
- Cited by
- 259
- References
- 13
- FWCI
- 10.34
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. https://app.cochat.ai/papers/10.1097/mbp.0b013e3282cb57b6
- The paper
- Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A. (2008). Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e3282cb57b6
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.