Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study

Stergiou, George S., Yiannes, Nikolaos G., Rarra, Vayia C.·2006·Blood Pressure Monitoring

Abstract

BACKGROUND: Limited evidence exists on the accuracy of oscillometric devices for blood pressure measurement in children. This study validated the Omron 705 IT monitor (Omron Healthcare Europe BV, Hoofddorp, The Netherlands) in normotensive children and adolescents. METHODS: Simultaneous blood pressure measurements were taken by two observers (connected mercury sphygmomanometers) four times, sequentially with three measurements by using the tested device. Absolute device-observer blood pressure differences were classified into three zones (within 5, 10 and 15 mmHg) and assessed using the European Society of Hypertension International Protocol criteria. The number of readings with a difference within 5 mmHg was calculated for each individual. The American Association for the Advancement of Medical Instrumentation criterion (mean difference+ or -SD <5+ or -8 mmHg) was also applied. RESULTS: A total of 197 study participants (591 readings) were included (99 boys, mean age+ or -SD 10.6+ or -2.4 years (range 6-16 years). The device produced 378 (64%), 532 (90%) and 579 (98%) measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 415 (70%), 525 (89%) and 577 (98%) measurements, respectively, for diastolic blood pressure. In all, 136 study participants (69%) had at least two systolic blood pressure differences within 5 mmHg and 16 (8%) had no differences within 5 mmHg (for diastolic blood pressure 147 (75%) and nine (5%) participants, respectively). Mean systolic blood pressure difference was 4.0+ or -4.8 mmHg and diastolic blood pressure -2.1+ or -5.9 mmHg and was <5+ or -8 mmHg in quartiles of participants divided by age, height, body mass index, arm circumference, blood pressure or pulse rate. CONCLUSIONS: The Omron 705 IT appears to be an accurate device for blood pressure measurement in normotensive children and adolescents.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/01.mbp.0000209074.38331.16

Cites

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 192

Metadata

Authors
Stergiou, G. S., Yiannes, N. G., & Rarra, V. C.
Venue
Blood Pressure Monitoring
Published
2006-08
Volume
11
Issue
4
Pages
229-234
ISSN
1359-5237 (ISSN-L)
OpenAlex
W1975069589
Cited by
192
References
0
FWCI
4.81
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study. https://app.cochat.ai/papers/10.1097/01.mbp.0000209074.38331.16
The paper
Stergiou, G. S., Yiannes, N. G., & Rarra, V. C. (2006). Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study. Blood Pressure Monitoring. https://doi.org/10.1097/01.mbp.0000209074.38331.16

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.