Abstract
The need to provide rigid bony fixation in the surgical treatment of craniofacial deformities has inspired an on-going evolution of surgical innovations and implants. Because of the young age of many treated craniosynostosis patients and the unique pattern of cranial vault growth, the extensive implantation of metal devices is potentially problematic. The use of resorbable plate and screw devices offers all of the benefits of rigid fixation without many of their potential risks. Since the introduction of resorbable plate and screw devices in 1996, tens of thousands of craniofacial patients have received implants, but long-term results from a large series have yet to be reported. A combined prospective and retrospective analysis was done on 1883 craniosynostosis patients under 2 years of age treated by 12 surgeons from seven different geographic locations over a 5-year period who used the same type of resorbable bone fixation devices (poly-L-lacticpolyglycolic copolymer). Specifically, the incidence of postoperative infection, fixation device failure, occurrence of delayed foreign-body reactions, and the need for reoperation resulting from device-related problems were determined. Technical difficulties and trends in device use were also noted. From this series, significant infectious complications occurred in 0.2 percent, device instability primarily resulting from postoperative trauma occurred in 0.3 percent, and self-limiting local foreign-body reactions occurred in 0.7 percent of the treated patients. The overall reoperation rate attributable to identifiable device-related problems was 0.3 percent. Improved bony stability was gained by using the longest plate geometries/configurations possible and bone grafting any significant gaps across plated areas that were structurally important. The specific types of plates and screws used evolved over the study period from simple plates, meshes, and threaded screws to application-specific plates and threadless push screws whose use varied among the involved surgeons. This report documents the safety and long-term value of the use of resorbable (LactoSorb) plate and screw fixation in pediatric craniofacial surgery in the infant and young child. Device-related complications requiring reoperation occurred in less than 0.5 percent of the implanted patients, which is less frequent than is reported for metallic bone fixation. Resorbable bone fixation for the rapidly growing cranial vault has fewer potential complications than the traditional use of metal plates, screws, and wires.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Rubin, P. J., & Yaremchuk, M. J. (1997) · Plastic and Reconstructive Surgery · doi:10.1097/00006534-199710000-00043 · Cited by 328
Combined Report of Problems and Complications in 793 Craniofacial Operations
Whitaker, L. A., Munro, I. R., Salyer, K. E., Jackson, I. T., Ortiz-Monasterio, F., & Marchac, D. (1979) · Plastic and Reconstructive Surgery · doi:10.1097/00006534-197908000-00011 · Cited by 267
Resorbable Plate Fixation in Pediatric Craniofacial Surgery
Eppley, B. L., Sadove, A. M., & Havlik, R. J. (1997) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-199707000-00001 · Cited by 180
Self-citationResorbable Plate Fixation in Pediatric Craniofacial Surgery: Long-term Outcome
Imola, M. J., Hamlar, D. D., Shao, W., Chowdhury, K., & Tatum, S. (2001) · Archives of Facial Plastic Surgery · doi:10.1001/archfaci.3.2.79 · Cited by 163
Observations on the Use of Rigid Fixation for Craniofacial Deformities in Infants and Young Children
Fearon, J. A., Munro, I. R., & Bruce, D. A. (1995) · Plastic and Reconstructive Surgery · doi:10.1097/00006534-199504000-00002 · Cited by 139
A Comparison of Resorbable and Metallic Fixation in Healing of Calvarial Bone Grafts
Eppley, B. L., & Sadove, A. M. (1995) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-199508000-00009 · Cited by 111
Self-citationBioabsorbable Plates and Screws in Pediatric Craniofacial Surgery
Kumar, A. V., Staffenberg, D. A., Petronio, J. A., & Wood, R. J. (1997) · Journal of Craniofacial Surgery · doi:10.1097/00001665-199703000-00006 · Cited by 100
Self-citationReasons for Removal of Rigid Internal Fixation Devices in Craniofacial Surgery
Orringer, J. S., Barcelona, V., & Buchman, S. R. (1998) · Journal of Craniofacial Surgery · doi:10.1097/00001665-199801000-00009 · Cited by 97
Critical Review of Microfixation in Pediatric Craniofacial Surgery
Goldberg, D. S., Bartlett, S. P. P., Yu, J. C.-K., Hunter, J. V., & Whitáker, L. A. (1995) · Journal of Craniofacial Surgery · doi:10.1097/00001665-199507000-00008 · Cited by 77
Infections in Craniofacial Surgery: A Combined Report of 567 Procedures from Two Centers
Fearon, J. A., Yu, J., Bartlett, S. P., Munro, I. R., & Whitaker, L. (1997) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-199709001-00006 · Cited by 74
Role of Resorbable Plates and Screws in Craniofacial Surgery
Pensler, J. M. (1997) · Journal of Craniofacial Surgery · doi:10.1097/00001665-199703000-00013 · Cited by 61
Self-citationBioresorbable Fixation for Congenital Pediatric Craniofacial Surgery: A 2-Year Follow-Up
Kurpad, S. N., Goldstein, J. A., & Cohen, A. R. (2000) · Pediatric Neurosurgery · doi:10.1159/000055976 · Cited by 52
Self-citationOutcome Analysis for Correction of Single Suture Craniosynostosis Using Resorbable Fixation
Losken, A., Williams, J. K., Burstein, F. D., Cohen, S. R., Hudgins, R., Boydston, W., Reisner, A., & Simms, C. (2001) · Journal of Craniofacial Surgery · doi:10.1097/00001665-200109000-00009 · Cited by 46
Self-citationResorbable Coupling Fixation in Craniosynostosis Surgery: Experimental and Clinical Applications
Eppley, B. L., & Sadove, A. M. (1995) · Journal of Craniofacial Surgery · doi:10.1097/00001665-199511000-00011 · Cited by 41
Self-citationCranial Plate and Screw Fixation in Infancy
Parsing, J. A., Posnick, J. C., Magge, S. N., Spinelli, H. M., Wolfe, S. A., Munro, I., & Mulliken, J. B. (1996) · Journal of Craniofacial Surgery · doi:10.1097/00001665-199607000-00003 · Cited by 30
The use of a resorbable plating system for treatment of craniosynostosis
Surpure, S., Smith, K. S., Sullivan, S. M., & Francel, P. C. (2001) · Journal of Oral and Maxillofacial Surgery · doi:10.1053/joms.2001.27497 · Cited by 27
Cited by
Poly-lactic acid synthesis for application in biomedical devices — A review
Lasprilla, A. J. R., Martinez, G. A. R., Lunelli, B. H., Jardini, A. L., & Filho, R. M. (2012) · Biotechnology Advances · doi:10.1016/j.biotechadv.2011.06.019 · Cited by 1,174
Polylactic acid: Synthesis and biomedical applications
Singhvi, M. S., Zinjarde, S. S., & Gokhale, D. V. (2019) · Journal of Applied Microbiology · doi:10.1111/jam.14290 · Cited by 883
Biocompatibility and safety of PLA and its copolymers
Ramot, Y., Haim-Zada, M., Domb, A. J., & Nyska, A. (2016) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2016.03.012 · Cited by 576
Poly (Lactic Acid) Production for Tissue Engineering Applications
Lopes, M. S., Jardini, A. L., & Filho, R. M. (2012) · Procedia Engineering · doi:10.1016/j.proeng.2012.07.534 · Cited by 457
Mathijssen, I. M. J. (2015) · Journal of Craniofacial Surgery · doi:10.1097/scs.0000000000002016 · Cited by 232
“The reported rate of complications is low: infection 8/1883 patients(135) and 5/146 patients,(136) delayed foreign-body reaction in 12/1883 patients,(135) and postoperative resorbable device failure with loss of bony position in 5/1883 patients.(135)”
Background
Metadata
- Authors
- Eppley, B. L., Morales, L., Wood, R., Pensler, J., Goldstein, J., Havlik, R. J., Habal, M., Losken, A., Williams, J. K., Burstein, F., Rozzelle, A. A., & Sadove, A. M.
- Venue
- Plastic and Reconstructive Surgery
- Published
- 2004-09
- Volume
- 114
- Issue
- 4
- Pages
- 850-856
- ISSN
- 0032-1052 (ISSN-L)
- OpenAlex
- W2087696832
- Cited by
- 250
- References
- 16
- FWCI
- 4.88
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.