Reproducibility of Quantitative Cerebral T2 Relaxometry, Diffusion Tensor Imaging, and 1H Magnetic Resonance Spectroscopy at 3.0 Tesla
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
OBJECTIVES: The reproducibility of quantitative cerebral T2 relaxometry, diffusion tensor imaging, and H magnetic resonance (MR) spectroscopic imaging was assessed on a clinical 3.0 T MR system. MATERIALS AND METHODS: Repeated measurements in 10 healthy volunteers were used to establish the reproducibility of quantitative measures derived from different quantitative MR techniques, namely the T2 relaxation time, the apparent diffusion coefficient (ADC), the fractional anisotropy (FA), and metabolite concentrations of N-acetyl-aspartate (NAA), creatine (Cr), choline (Cho), and myo-inositol (mI). Results were compared with previously reported reproducibility measures from 1.5 T. RESULTS: The coefficient of variation (CV) was < or =1.6% for T2, < or =1.6% for ADC, and < or =5.3%, for FA in the cerebrum. For metabolites the CV was < or =8.0% in the frontal lobe and < or =20.4% in the temporal lobe. CONCLUSIONS: The reproducibility of quantitative brain MRI at 3.0 T is better than or at least comparable to the reproducibility at 1.5 T.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
T1, T2 relaxation and magnetization transfer in tissue at 3T
Stanisz, G. J., Odrobina, E. E., Pun, J., Escaravage, M., Graham, S. J., Bronskill, M. J., & Henkelman, R. M. (2005) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20605 · Cited by 1,266 · cited 1×
“The average T2 within the cerebrum for white and gray matter combined was 78–84 milliseconds, which is in accordance with the T2s derived by Stanisz et al of 69 and 99 milliseconds for white matter and gray matter, respectively.(26) The average ADC and FA values measured within the cerebrum were 786–827 10 6 mm(2)/s a…”
Cited by
The B‐matrix must be rotated when correcting for subject motion in DTI data
Leemans, A., & Jones, D. K. (2009) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21890 · Cited by 1,281
Vollmar, C., O'Muircheartaigh, J., Barker, G. J., Symms, M. R., Thompson, P., Kumari, V., Duncan, J. S., Richardson, M. P., & Koepp, M. J. (2010) · NeuroImage · doi:10.1016/j.neuroimage.2010.03.046 · Cited by 259
Fox, R. J., Sakaie, K., Lee, J.-C., Debbins, J. P., Liu, Y., Arnold, D. L., Melhem, E. R., Smith, C. H., Philips, M. D., Lowe, M., et al. (2012) · American Journal of Neuroradiology · doi:10.3174/ajnr.a2844 · Cited by 137
“Scanrescan reproducibility of DTI measures with 3T scanners has been previously published and has been satisfactory for FA and ADC values ( 5%).(20,21) In our study, we have additionally determined regional values for diffusivity measures and found them to be comparable.”
MethodPalacios, E. M., Martin, A. J., Boss, M. A., Ezekiel, F., Chang, Y. S., Yuh, E. L., Vassar, M. J., Schnyer, D. M., MacDonald, C. L., Crawford, K. L., et al. (2017) · American Journal of Neuroradiology · doi:10.3174/ajnr.a5025 · Cited by 110
Reproducibility of tract‐based white matter microstructural measures using the ENIGMA ‐ DTI protocol
Acheson, A., Wijtenburg, S. A., Rowland, L. M., Winkler, A., Mathias, C. W., Hong, L. E., Jahanshad, N., Patel, B., Thompson, P. M., McGuire, S. A., et al. (2017) · Brain and Behavior · doi:10.1002/brb3.615 · Cited by 43
Metadata
- Authors
- Jansen, J. F. A., Kooi, M. E., Kessels, A. G. H., Nicolay, K., & Backes, W. H.
- Venue
- Investigative Radiology
- Published
- 2007-06
- Volume
- 42
- Issue
- 6
- Pages
- 327-337
- ISSN
- 0020-9996 (ISSN-L)
- OpenAlex
- W2037616978
- Cited by
- 60
- References
- 2
- FWCI
- 3.65
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Reproducibility of Quantitative Cerebral T2 Relaxometry, Diffusion Tensor Imaging, and 1H Magnetic Resonance Spectroscopy at 3.0 Tesla. https://app.cochat.ai/papers/10.1097/01.rli.0000262757.10271.e5
- The paper
- Jansen, J. F. A., Kooi, M. E., Kessels, A. G. H., Nicolay, K., & Backes, W. H. (2007). Reproducibility of Quantitative Cerebral T2 Relaxometry, Diffusion Tensor Imaging, and 1H Magnetic Resonance Spectroscopy at 3.0 Tesla. Investigative Radiology. https://doi.org/10.1097/01.rli.0000262757.10271.e5
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.