Pedicle Screw Placement Using Augmented Reality Surgical Navigation With Intraoperative 3D Imaging: A First In-Human Prospective Cohort Study
Open accessThe DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
Study Design.: Prospective observational study. Objective.: The aim of this study was to evaluate the accuracy of pedicle screw placement using augmented reality surgical navigation (ARSN) in a clinical trial. Summary of Background Data.: Recent cadaveric studies have shown improved accuracy for pedicle screw placement in the thoracic spine using ARSN with intraoperative 3D imaging, without the need for periprocedural x-ray. In this clinical study, we used the same system to place pedicle screws in the thoracic and lumbosacral spine of 20 patients. Methods.: The study was performed in a hybrid operating room with an integrated ARSN system encompassing a surgical table, a motorized flat detector C-arm with intraoperative 2D/3D capabilities, integrated optical cameras for augmented reality navigation, and noninvasive patient motion tracking. Three independent reviewers assessed screw placement accuracy using the Gertzbein grading on 3D scans obtained before wound closure. In addition, the navigation time per screw placement was measured. Results.: One orthopedic spinal surgeon placed 253 lumbosacral and thoracic pedicle screws on 20 consenting patients scheduled for spinal fixation surgery. An overall accuracy of 94.1% of primarily thoracic pedicle screws was achieved. No screws were deemed severely misplaced (Gertzbein grade 3). Fifteen (5.9%) screws had 2 to 4 mm breach (Gertzbein grade 2), occurring in scoliosis patients only. Thirteen of those 15 screws were larger than the pedicle in which they were placed. Two medial breaches were observed and 13 were lateral. Thirteen of the grade 2 breaches were in the thoracic spine. The average screw placement time was 5.2 ± 4.1 minutes. During the study, no device-related adverse event occurred. Conclusion.: ARSN can be clinically used to place thoracic and lumbosacral pedicle screws with high accuracy and with acceptable navigation time. Consequently, the risk for revision surgery and complications could be minimized. Level of Evidence: 3
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Accuracy of Pedicular Screw Placement In Vivo
GERTZBEIN, S. D., & ROBBINS, S. E. (1990) · Spine · doi:10.1097/00007632-199001000-00004 · Cited by 1,381
Gelalis, I. D., Paschos, N. K., Pakos, E. E., Politis, A. N., Arnaoutoglou, C. M., Karageorgos, A., Ploumis, A., & Xenakis, T. A. (2011) · European Spine Journal · doi:10.1007/s00586-011-2011-3 · Cited by 735 · cited 1×
“Most grade 2 screws (13/15) in this study were lateral, in agreement with previously published data.(14) The intraoperative screw revision rate of 1.”
ResultSilbermann, J., Riese, F., Allam, Y., Reichert, T., Koeppert, H., & Gutberlet, M. (2011) · European Spine Journal · doi:10.1007/s00586-010-1683-4 · Cited by 214
Intraoperative image-guided spinal navigation: technical pitfalls and their avoidance
Rahmathulla, G., Nottmeier, E. W., Pirris, S. M., Deen, H. G., & Pichelmann, M. A. (2014) · Neurosurgical FOCUS · doi:10.3171/2014.1.focus13516 · Cited by 195 · cited 1×
“State-of-the-art i3Dn systems use passive optical infrared cameras with a dynamic reference frame attached to the spine for patient tracking.(11) Intraoperative 3D augmented reality surgical navigation (ARSN) is a novel concept of navigation using optical video cameras to augment the surgical field with 3D intraoperat…”
BackgroundElmi-Terander, A., Skulason, H., Söderman, M., Racadio, J., Homan, R., Babic, D., van der Vaart, N., & Nachabe, R. (2016) · Spine · doi:10.1097/brs.0000000000001830 · Cited by 160
Self-citationRivkin, M. A., & Yocom, S. S. (2014) · Neurosurgical FOCUS · doi:10.3171/2014.1.focus13499 · Cited by 100
Chan, A., Parent, É., Narvacan, K., San, C., & Lou, E. (2017) · The Spine Journal · doi:10.1016/j.spinee.2017.04.001 · Cited by 97 · cited 1×
“This is especially important in scoliosis treatment where most of the screws are placed in the thoracic spine where the pedicles are narrow.(7,8) Consequently, i3Dn decreases the risk of revision surgery to almost half of that associated with the free-hand method.”
BackgroundLearning curve of 3D fluoroscopy image–guided pedicle screw placement in the thoracolumbar spine
Ryang, Y., Villard, J., Obermüller, T., Friedrich, B., Wolf, P., Gempt, J., Ringel, F. A., & Meyer, B. (2014) · The Spine Journal · doi:10.1016/j.spinee.2014.10.003 · Cited by 93
Leković, G. P., Potts, E. A., Karahalios, D. G., & Hall, G. D. R. (2007) · Journal of Neurosurgery Spine · doi:10.3171/spi-07/10/393 · Cited by 82
Ling, J. M., Dinesh, S. K., Pang, B. C., Chen, M. W., Lim, H. L., Louange, D. T., Yu, C. S., & Wang, C. M. E. (2014) · Journal of Clinical Neuroscience · doi:10.1016/j.jocn.2013.02.034 · Cited by 52
Pedicle Screw Placement With O-arm and Stealth Navigation
Patil, S. S., Lindley, E. M., Burger, E. L., Yoshihara, H., & Patel, V. V. (2012) · Orthopedics · doi:10.3928/01477447-20111122-15 · Cited by 44
Liu, H., Chen, W., Liu, T., Meng, B., & Yang, H. (2017) · Journal of orthopaedic surgery · doi:10.1177/2309499017718901 · Cited by 36
Cited by
Morimoto, T., Kobayashi, T., Hirata, H., Otani, K., Sugimoto, M., Tsukamoto, M., Yoshihara, T., Ueno, M., & Mawatari, M. (2022) · Journal of Clinical Medicine · doi:10.3390/jcm11020470 · Cited by 281
Augmented and virtual reality in spine surgery, current applications and future potentials
Ghaednia, H., Fourman, M. S., Lans, A., Detels, K., Dijkstra, H., Lloyd, S., Sweeney, A., Oosterhoff, J. H. F., & Schwab, J. H. (2021) · The Spine Journal · doi:10.1016/j.spinee.2021.03.018 · Cited by 237
Applicability of augmented reality in orthopedic surgery – A systematic review
Jud, L., Fotouhi, J., Andronic, O., Aichmair, A., Osgood, G., Navab, N., & Farshad, M. (2020) · BMC Musculoskeletal Disorders · doi:10.1186/s12891-020-3110-2 · Cited by 198
Virtual Reality and Augmented Reality—Translating Surgical Training into Surgical Technique
McKnight, R. R., Pean, C. A., Buck, J. S., Hwang, J. S., Hsu, J. R., & Pierrie, S. N. (2020) · Current Reviews in Musculoskeletal Medicine · doi:10.1007/s12178-020-09667-3 · Cited by 175
Accuracy of augmented reality-assisted pedicle screw placement: A systematic review
Youssef, S., McDonnell, J. M., Wilson, K. V., Turley, L., Cunniffe, G., Morris, S., Darwish, S., & Butler, J. S. (2024) · European Spine Journal · doi:10.1007/s00586-023-08094-5 · Cited by 36
“Of note, one study was omitted from further analysis in terms of collectively evaluating overall accuracy and respective breaches as the ARSN patient cohort was included in an additional study [18, 26].”
Metadata
- Authors
- Elmi-Terander, A., Burström, G., Nachabe, R., Skulason, H., Pedersen, K., Fagerlund, M., Ståhl, F., Charalampidis, A., Söderman, M., Holmin, S., Babic, D., Jenniskens, I., Edström, E., & Gerdhem, P.
- Venue
- Spine
- Published
- 2019-04-01
- Volume
- 44
- Issue
- 7
- Pages
- 517-525
- ISSN
- 0362-2436 (ISSN-L)
- 1528-1159
- OpenAlex
- W2895212268
- Cited by
- 229
- References
- 12
- FWCI
- 20.50
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.