Diagnostic Imaging of Spinal Deformities: Reducing Patients Radiation Dose With a New Slot-Scanning X-ray Imager
Abstract
STUDY DESIGN: Clinical trial comparing image quality and entrance dose between Biospace EOS system, a new slot-scanning radiographic device, and a Fuji FCR 7501S computed radiography (CR) system for 50 patients followed for spinal deformities. OBJECTIVE: Based on their physical properties, slot-scanners show the potential to produce image quality comparable to CR systems using less radiation. This article validates this assertion by comparing a new slot-scanner to a CR system through a wide-ranging evaluation of dose and image quality for scoliosis examinations. SUMMARY OF BACKGROUND DATA: For each patient included in this study, lateral and posteroanterior images were acquired with both systems. For each system, entrance dose was measured for different anatomic locations. METHODS: Dose and image quality being directly related, comparable images were obtained using the same radiograph tube voltage on both systems while tube currents were selected to match signal-to-noise ratios on a phantom. Different techniques were defined with respect to patient's thickness about the iliac crests. Given dose amplitudes expected for scoliosis examinations, optically stimulated luminescence dosimeters were chosen as optimal sensors. Two radiologists and 2 orthopedists evaluated the images in a randomized order using a questionnaire targeting anatomic landmarks. Visibility of the structures was rated on a 4 level scale. Image quality assessment was analyzed using a Wilcoxon signed-rank tests. RESULTS: Average skin dose was reduced from 6 to 9 times in the thoracoabdominal region when using the slot-scanner instead of CR. Moreover, image quality was significantly better with EOS for all structures in the frontal view (P < 0.006) and lateral view (P < 0.04), except for lumbar spinous processes, better seen on the CR (P < 0.003). CONCLUSION: We established that the EOS system offers overall enhanced image quality while reducing drastically the entrance dose for the patient.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Computed Tomography — An Increasing Source of Radiation Exposure
Brenner, D. J., & Hall, E. J. (2007) · New England Journal of Medicine · doi:10.1056/nejmra072149 · Cited by 8,748
Breast Cancer Mortality After Diagnostic Radiography: Findings From the U.S. Scoliosis Cohort Study
Morin Doody, M., Lonstein, J. E., Stovall, M., Hacker, D. G., Luckyanov, N., & Land, C. E. (2000) · Spine · doi:10.1097/00007632-200008150-00009 · Cited by 577
Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative
Brenner, D. J. (2002) · Pediatric Radiology · doi:10.1007/s00247-002-0671-1 · Cited by 554
Breast Cancer in Women With Scoliosis Exposed to Multiple Diagnostic X Rays
Hoffman, D. A., Lonstein, J. E., Morin, M. M., Visscher, W., Harris, B. S. H., & Boice, J. D. (1989) · JNCI Journal of the National Cancer Institute · doi:10.1093/jnci/81.17.1307 · Cited by 280
Cited by
Shere, C., & Clark, E. M. (2022) · Archives of Orthopaedic and Trauma Surgery · doi:10.1007/s00402-022-04508-z · Cited by 11
Metadata
- Authors
- Deschênes, S., Charron, G., Beaudoin, G., Labelle, H., Dubois, J., Miron, M.-C., & Parent, S.
- Venue
- Spine
- Published
- 2010-04
- Volume
- 35
- Issue
- 9
- Pages
- 989-994
- ISSN
- 0362-2436 (ISSN-L)
- OpenAlex
- W2045607856
- Cited by
- 359
- References
- 4
- FWCI
- 13.26
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Diagnostic Imaging of Spinal Deformities: Reducing Patients Radiation Dose With a New Slot-Scanning X-ray Imager. https://app.cochat.ai/papers/10.1097/brs.0b013e3181bdcaa4
- The paper
- Deschênes, S., Charron, G., Beaudoin, G., Labelle, H., Dubois, J., Miron, M.-C., & Parent, S. (2010). Diagnostic Imaging of Spinal Deformities: Reducing Patients Radiation Dose With a New Slot-Scanning X-ray Imager. Spine. https://doi.org/10.1097/brs.0b013e3181bdcaa4
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.