Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients*

Brunner, Richard, Kitzberger, Reinhard, Miehsler, Wolfgang, Herkner, Harald, Madl, Christian, Holzinger, Ulrike · 2011 · Critical Care Medicine
Cited by 93

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Continuous glucose monitoring has been proposed to optimize glucose control in critically ill patients. To achieve strict glucose regulation, accurate and reliable continuous glucose-monitoring systems are essential. OBJECTIVE: Evaluation of a subcutaneous continuous glucose-monitoring system for use in critically ill patients. DESIGN: Pooled-data analysis of two prospective, randomized, controlled trials. SETTING: An eight-bed medical intensive care unit of a university hospital. PATIENTS: A total of 174 critically ill patients on intensive insulin therapy. INTERVENTIONS: Subcutaneous continuous glucose monitoring. MEASUREMENTS: Two thousand forty-five continuous glucose-monitoring system sensor glucose values were compared with arterial reference blood glucose levels, determined by a blood gas analyzer. Continuous glucose monitoring data were recorded continuously for up to 72 hrs by using a subcutaneous continuous glucose-monitoring sensor. The correlation of both methods and differences between continuous glucose-monitoring systems and reference values were calculated, as well as the conformity of continuous glucose-monitoring values with the International Organization for Standardization criteria (<0.83 mmol/L [15 mg/dL] difference for glucose values ≤ 4.12 mmol/L [≤ 75 mg/dL] and <20% difference for glucose values >4.12 mmol/L [>75 mg/dL]). RESULTS: The Pearson correlation coefficient was 0.92, showing strong correlation between the two methods. The intraclass correlation coefficient was 0.92, indicating that 92% of the variability is due to subjects and measurement occasions. Mean difference between continuous glucose-monitoring system and reference values was -0.10 mmol/L (confidence interval: -0.13 to -0.07) (-2 mg/dL [confidence interval: -2 to -1]) (continuous glucose-monitoring system minus reference) and absolute difference 0.44 mmol/L (confidence interval: 0.41-0.47) (8 mg/dL [confidence interval: 7-8]). According to the insulin titration error grid analysis, 99.1% of continuous glucose-monitoring system values were in the acceptable treatment zone. No continuous glucose-monitoring system measurements were found in the life-threatening zone, and 92.9% of the continuous glucose-monitoring system glucose values met the International Organization for Standardization criteria. CONCLUSION: The subcutaneous continuous glucose-monitoring system is reliable for use in critically ill patients and showed glucose values with a strong correlation to arterial reference blood glucose levels, determined by a blood gas analyzer.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/ccm.0b013e318206bf2e

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 93

Metadata

Authors
Brunner, R., Kitzberger, R., Miehsler, W., Herkner, H., Madl, C., & Holzinger, U.
Venue
Critical Care Medicine
Published
2011-04
Volume
39
Issue
4
Pages
659-664
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2317808362
Cited by
93
References
2
FWCI
10.12
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients*. https://app.cochat.ai/papers/10.1097/ccm.0b013e318206bf2e
The paper
Brunner, R., Kitzberger, R., Miehsler, W., Herkner, H., Madl, C., & Holzinger, U. (2011). Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients*. Critical Care Medicine. https://doi.org/10.1097/ccm.0b013e318206bf2e

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.