Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
Abstract
OBJECTIVE: The measurement of blood pressure in atrial fibrillation is considered as difficult and uncertain, and current guidelines recommend the use of the auscultatory method. The accuracy of automated blood pressure monitors in atrial fibrillation remains controversial. METHOD: A systematic review and meta-analysis was performed of studies comparing automated (oscillometric or automated Korotkov) versus manual auscultatory blood pressure measurements (mercury or aneroid sphygmomanometer) in patients with sustained atrial fibrillation. RESULTS: Twelve validations were analyzed (566 patients; five home, three ambulatory and three office devices). Pooled correlation coefficients between automated and manual blood pressure measurements were stronger for SBP than DBP (r = 0.89 versus 0.76, P < 0.001). Automated measurements were higher than manual measurements [pooled average SBP difference 0.5 mmHg, 95% confidence interval (CI) -0.9, 1.9; DBP 2.5 mmHg, 95%CI -0.6, 5.7). The mean difference was within 5 mmHg in six and four (SBP and DBP, respectively) of six validations. The SD of mean difference was within 8 mmHg in two and three (SBP and DBP, respectively) of four validations. The proportion of absolute automated-manual differences within 5 mmHg was at least 65% in four and two (SBP and DBP, respectively) of eight validations. Three studies showed no impact of heart rate on the automated-manual blood pressure differences. CONCLUSION: There is limited evidence and significant heterogeneity in the studies that validated automated blood pressure monitors in atrial fibrillation. These monitors appear to be accurate in measuring SBP but not DBP. Given that atrial fibrillation is common in the elderly, in whom systolic hypertension is more common and important than diastolic hypertension, automated monitors appear to be appropriate for self-home but not for office measurement.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“the clinical application and the accuracy of HBPM may be limited by certain conditions related either to the individual [63–66], to the procedure itself [4,67], to the oscillometric technique for BP measurement or to cuff-related issues [4,65,68–74].”
In the citing paper
“Monitoring was repeated as necessary until ≥ 70% of the measured values obtained during both daytime and nighttime were satisfactory [20].”
From The effect of the physical activity on polymorphic premature ventricular complexes in chronic kidney disease (2017) · cites this paper as [20]
Citations
Cites
Bias in meta-analysis detected by a simple, graphical test
Egger, M., Smith, G. D., Schneider, M., & Minder, C. (1997) · BMJ · doi:10.1136/bmj.315.7109.629 · Cited by 56,722
Operating characteristics of a rank correlation test for publication bias
Begg, C. B., & Mazumdar, M. (1994) · Biometrics · doi:10.2307/2533446 · Cited by 16,955
Benjamin, E. J. (1994) · JAMA · doi:10.1001/jama.1994.03510350050036 · Cited by 2,613
Kannel, W. B., Wolf, P. A., Benjamin, E. J., & Levy, D. (1998) · The American Journal of Cardiology · doi:10.1016/s0002-9149(98)00583-9 · Cited by 2,033
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationDanaei, G., Finucane, M. M., Lin, J. K., Singh, G. M., Paciorek, C. J., Cowan, M. J., Farzadfar, F., Stevens, G. A., Lim, S. S., Riley, L. M., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(10)62036-3 · Cited by 1,239
Brien, E. O., Petrie, J. C., Littler, W. A., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D. G., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 764
Bland, M., & Altman, D. G. (1990) · Computers in Biology and Medicine · doi:10.1016/0010-4825(90)90013-f · Cited by 689
O'Brien, E. T., Atkins, N., Stergiou, G. S., Karpettas, N., Parati, G., Asmar, R. G., Imai, Y., Wang, J., Mengden, T., & Shennan, A. H. (2010) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283360e98 · Cited by 609
Self-citationAccuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol
Brien, E. O., Mee, F. D., Atkins, N., & O??Malley, K. (1991) · Journal of Hypertension · doi:10.1097/00004872-199106000-00016 · Cited by 451
Cited by
Sorted by citations
Items 101–120 of 120
Home blood pressure monitoring in hypertension
Lambie, J. (2014) · British Journal of Cardiac Nursing · doi:10.12968/bjca.2014.9.1.13 · Cited by 1
Mean Arterial Pressure in Acute Ischemic Stroke Study
Grove, M. (2020) · doi:10.21007/etd.cghs.2020.0511 · Cited by 1
Revisão das orientações terapêuticas da hipertensão arterial em ambulatório
Relva, D. E. B. (2014) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Clinical Case 6: Patient with Essential Hypertension and Atrial Fibrillation
Virdis, A. (2016) · Practical case studies in hypertension management · doi:10.1007/978-3-319-39164-9_6
Kiuchi, M. G., Chen, S., & Pürerfellner, H. (2016) · IJC Metabolic & Endocrine · doi:10.1016/j.ijcme.2016.09.006
Robust Estimation of Mean Arterial Pressure in Atrial Fibrillation Using Oscillometry
Tannous, M. (2014) · uO Research (University of Ottawa) · doi:10.20381/ruor-5036
I. Accurate Measurement of Blood Pressure
Gelfer, M., Cloutier, L., Tobe, S. W., Lamarre-Cliché, M., Bolli, P., Tremblay, G., McLean, D., McKay, D. W., Padwal, R. S., Rabi, D. M., et al. (2016)
Pulmonary vein reconnection in patients with and without CKD
Kiuchi, M. G., & Chen, S. (2017) · IJC Metabolic & Endocrine · doi:10.1016/j.ijcme.2017.03.002
Kiuchi, M. G., Chen, S., Paz, L. M. R., & Pürerfellner, H. (2017) · Hipertensión y Riesgo Vascular · doi:10.1016/j.hipert.2017.07.002
Devices for Home Blood Pressure Monitoring
Asmar, R. G., Κόλλιας, Α., Palatini, P., Parati, G., Shennan, A. H., Stergiou, G. S., Topouchian, J. A., Wang, J., White, W. B., & O'Brien, E. T. (2019) · Updates in hypertension and cardiovascular protection · doi:10.1007/978-3-030-23065-4_1
Self-citationArmitage, L. C., Lawson, B. K., Román, C., Thompson, B., Biggs, C. A., Rutter, H., Lewis-Jones, M., Ede, J., Tarassenko, L., Farmer, A. A., et al. (2022) · Wellcome Open Research · doi:10.12688/wellcomeopenres.17537.2
Methods of Blood Pressure Assessment Used in Milestone Hypertension Trials
Chen, Y., Lei, L., & Wang, J. (2018) · Springer eBooks · doi:10.1007/978-3-319-92946-0_14
Mesures de la pression artérielle dans des situations particulières
Theiler, K., Kirsch, M. E. W., & Wuerzner, G. (2022) · JMIG · doi:10.53738/revmed.2022.18.795.1698
Ovchinnikov, Y. V., Борисович, П. М., Анатольевич, А. А., & Pavlova, E. M. (2022) · Clinical Medicine (Russian Journal) · doi:10.30629/0023-2149-2022-100-6-253-260
Ranting in 2026: Waiting for 2027
Hirsch, I. B. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261416448
How Accurate Is Inpatient Blood Pressure Measurement?
Gallagher, A., Smyth, B. M., Alexeev, S., Mangos, G., Schutte, A. E., Sharman, J. E., Jardine, M., Brown, M. A., Burrell, L., Levin, A., et al. (2026) · Hypertension · doi:10.1161/hypertensionaha.125.26355
Korneev, A. B., Обрезан, А. Г., & Shubik, Y. V. (2026) · Vrach · doi:10.29296/25877305-2026-05-09
Problems and Solutions of Accurate Blood Pressure Measurement in Atrial Fibrillation
Рогоза, А. Н., Тихоненко, В. М., Shubik, Y. V., Frolova, N. L., Борисович, К. А., Пивоваров, В. В., Иосифович, Х. Г., & Zaitsev, G. К. (2026) · Systemic Hypertension · doi:10.38109/2075-082x-2026-2-27-34
Farzad, M. S., Bouadjenek, M. R., Hobbs, M., Zohaid, M., Zheng, D., & Shah, S. M. T. (2026) · SSRN Electronic Journal · doi:10.2139/ssrn.7395422
Metadata
- Authors
- Stergiou, G. S., Kollias, A., Destounis, A., & Tzamouranis, D.
- Affiliations
- Sotiria General Hospital (GR)
- Corresponding author
- Stergiou, George S.
- Venue
- Journal of Hypertension
- Published
- 2012-11
- Volume
- 30
- Issue
- 11
- Pages
- 2074-2082
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W1967096675
- Cited by
- 120
- References
- 31
- Citations by year
- 2026: 6 · 2024: 4 · 2023: 4 · 2022: 9 · 2021: 9 · 2020: 10 · 2019: 14 · 2018: 9 · 2017: 15 · 2016: 13 · 2015: 8 · 2014: 6 · 2013: 11 · 2012: 2
- FWCI
- 5.56
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Hypertension (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Atrial Fibrillation Management and Outcomes
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- atrial fibrillation, automated blood pressure measurement, oscillometric blood pressure monitor, auscultatory blood pressure measurement, systolic blood pressure, diastolic blood pressure, measurement accuracy, home blood pressure monitoring, ambulatory blood pressure monitoring, office blood pressure measurement
- MeSH terms
- Analysis of Variance, Atrial Fibrillation, Automation, Blood Pressure Determination, Heart Rate, Humans, Hypertension, Systole, Reproducibility of Results, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review