Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
Abstract
OBJECTIVE: The measurement of blood pressure in atrial fibrillation is considered as difficult and uncertain, and current guidelines recommend the use of the auscultatory method. The accuracy of automated blood pressure monitors in atrial fibrillation remains controversial. METHOD: A systematic review and meta-analysis was performed of studies comparing automated (oscillometric or automated Korotkov) versus manual auscultatory blood pressure measurements (mercury or aneroid sphygmomanometer) in patients with sustained atrial fibrillation. RESULTS: Twelve validations were analyzed (566 patients; five home, three ambulatory and three office devices). Pooled correlation coefficients between automated and manual blood pressure measurements were stronger for SBP than DBP (r = 0.89 versus 0.76, P < 0.001). Automated measurements were higher than manual measurements [pooled average SBP difference 0.5 mmHg, 95% confidence interval (CI) -0.9, 1.9; DBP 2.5 mmHg, 95%CI -0.6, 5.7). The mean difference was within 5 mmHg in six and four (SBP and DBP, respectively) of six validations. The SD of mean difference was within 8 mmHg in two and three (SBP and DBP, respectively) of four validations. The proportion of absolute automated-manual differences within 5 mmHg was at least 65% in four and two (SBP and DBP, respectively) of eight validations. Three studies showed no impact of heart rate on the automated-manual blood pressure differences. CONCLUSION: There is limited evidence and significant heterogeneity in the studies that validated automated blood pressure monitors in atrial fibrillation. These monitors appear to be accurate in measuring SBP but not DBP. Given that atrial fibrillation is common in the elderly, in whom systolic hypertension is more common and important than diastolic hypertension, automated monitors appear to be appropriate for self-home but not for office measurement.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“the clinical application and the accuracy of HBPM may be limited by certain conditions related either to the individual [63–66], to the procedure itself [4,67], to the oscillometric technique for BP measurement or to cuff-related issues [4,65,68–74].”
In the citing paper
“Monitoring was repeated as necessary until ≥ 70% of the measured values obtained during both daytime and nighttime were satisfactory [20].”
From The effect of the physical activity on polymorphic premature ventricular complexes in chronic kidney disease (2017) · cites this paper as [20]
Citations
Cites
Bias in meta-analysis detected by a simple, graphical test
Egger, M., Smith, G. D., Schneider, M., & Minder, C. (1997) · BMJ · doi:10.1136/bmj.315.7109.629 · Cited by 56,722
Operating characteristics of a rank correlation test for publication bias
Begg, C. B., & Mazumdar, M. (1994) · Biometrics · doi:10.2307/2533446 · Cited by 16,955
Benjamin, E. J. (1994) · JAMA · doi:10.1001/jama.1994.03510350050036 · Cited by 2,613
Kannel, W. B., Wolf, P. A., Benjamin, E. J., & Levy, D. (1998) · The American Journal of Cardiology · doi:10.1016/s0002-9149(98)00583-9 · Cited by 2,033
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationDanaei, G., Finucane, M. M., Lin, J. K., Singh, G. M., Paciorek, C. J., Cowan, M. J., Farzadfar, F., Stevens, G. A., Lim, S. S., Riley, L. M., et al. (2011) · The Lancet · doi:10.1016/s0140-6736(10)62036-3 · Cited by 1,239
Brien, E. O., Petrie, J. C., Littler, W. A., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D. G., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 764
Bland, M., & Altman, D. G. (1990) · Computers in Biology and Medicine · doi:10.1016/0010-4825(90)90013-f · Cited by 689
O'Brien, E. T., Atkins, N., Stergiou, G. S., Karpettas, N., Parati, G., Asmar, R. G., Imai, Y., Wang, J., Mengden, T., & Shennan, A. H. (2010) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3283360e98 · Cited by 609
Self-citationAccuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol
Brien, E. O., Mee, F. D., Atkins, N., & O??Malley, K. (1991) · Journal of Hypertension · doi:10.1097/00004872-199106000-00016 · Cited by 451
Cited by
Sorted by citations
2013 ESH/ESC Guidelines for the management of arterial hypertension
Authors Task Force Members, Mancia, G., Fagard, R., Narkiewicz, K., Redon, J., Zanchetti, A., Böhm, M., Christiaens, T., Cífková, R., Backer, G. D., et al. (2013) · European Heart Journal · doi:10.1093/eurheartj/eht151 · Cited by 13,739
Mancia, G., Fagard, R., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T., Cifkova, R., De Backer, G., Dominiczak, A., et al. (2013) · Journal of Hypertension · doi:10.1097/01.hjh.0000431740.32696.cc · Cited by 5,655
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association
Muntner, P., Shimbo, D., Carey, R. M., Charleston, J. B., Gaillard, T., Misra, S., Myers, M. G., Ogedegbe, G., Schwartz, J. E., Townsend, R. R., et al. (2019) · Hypertension · doi:10.1161/hyp.0000000000000087 · Cited by 1,512
Cheung, A. K., Chang, T. I., Cushman, W. C., Furth, S. L., Hou, F. F., Ix, J. H., Knoll, G. A., Muntner, P., Pecoits-Filho, R., Sarnak, M. J., et al. (2021) · Kidney International · doi:10.1016/j.kint.2020.11.003 · Cited by 1,452
European Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419
Self-citationDaskalopoulou, S. S., Rabi, D. M., Zarnke, K. B., Dasgupta, K., Nerenberg, K. A., Cloutier, L., Gelfer, M., Lamarre-Cliché, M., Milot, A., Bolli, P., et al. (2015) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2015.02.016 · Cited by 690
2013 ESH/ESC Guidelines for the management of arterial hypertension
Mancia, G., Fagard, R. H., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T. C., Cífková, R., Backer, G. D., Dominiczak, A. F., et al. (2013) · Blood Pressure · doi:10.3109/08037051.2013.812549 · Cited by 430
Hypertension and Atrial Fibrillation
Verdecchia, P., Angeli, F., & Reboldi, G. (2018) · Circulation Research · doi:10.1161/circresaha.117.311402 · Cited by 259
Mossello, E., Pieraccioli, M., Nesti, N., Bulgaresi, M., Lorenzi, C., Caleri, V., Tonon, E., Cavallini, M. C., Baroncini, C., Di Bari, M., et al. (2015) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2014.8164 · Cited by 247
Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research
Muntner, P., Einhorn, P. T., Cushman, W. C., Whelton, P. K., Bello, N. A., Drawz, P. E., Green, B. B., Jones, D. W., Juraschek, S. P., Margolis, K. L., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.069 · Cited by 199
Self-citationParati, G., Stergiou, G. S., Bilo, G., Kollias, A., Pengo, M., Ochoa, J. E., Agarwal, R., Asayama, K., Asmar, R., Burnier, M., et al. (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002922 · Cited by 178
Self-citationStergiou, G. S., Parati, G., Vlachopoulos, C. V., Achimastos, A. D., Andreadis, E., Asmar, R. G., Avolio, A. P., Bénétos, A., Bilo, G., Boubouchairopoulou, N., et al. (2016) · Journal of Hypertension · doi:10.1097/hjh.0000000000000969 · Cited by 158
Self-citationLakhal, K., Ehrmann, S., Martin, M., Faiz, S., Réminiac, F., Cinotti, R., Capdevila, X., Asehnoune, K., Blanloeil, Y., Rozec, B. R., et al. (2015) · British Journal of Anaesthesia · doi:10.1093/bja/aev304 · Cited by 143
Atrial fibrillation and arterial hypertension
Kallistratos, M. S., Poulimenos, L. E., & Manolis, A. J. (2017) · Pharmacological Research · doi:10.1016/j.phrs.2017.10.007 · Cited by 114
A New Algorithm for the Diagnosis of Hypertension in Canada
Cloutier, L., Daskalopoulou, S. S., Padwal, R. S., Lamarre-Cliché, M., Bolli, P., McLean, D., Milot, A., Tobe, S. W., Tremblay, G., McKay, D. W., et al. (2015) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2015.02.014 · Cited by 100
Verberk, W., Omboni, S., Kollias, A., & Stergiou, G. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2015.10.182 · Cited by 89
Self-citationImpact of Atrial Fibrillation on the Accuracy of Oscillometric Blood Pressure Monitoring
Pagonas, N., Schmidt, S. C., Eysel, J., Compton, F., Hoffmann, C., Seibert, F. S., Hilpert, J. W., Tschöpe, C., Zidek, W., & Westhoff, T. H. (2013) · Hypertension · doi:10.1161/hypertensionaha.113.01426 · Cited by 82
Triage, Treatment, and Transfer
Middleton, S., Grimley, R., & Alexandrov, A. W. (2015) · Stroke · doi:10.1161/strokeaha.114.006139 · Cited by 72
Stergiou, G. S., Kyriakoulis, K. G., Stambolliu, E., Destounis, A., Karpettas, N., Kalogeropoulos, P., & Kollias, A. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002201 · Cited by 67
Self-citationClark, C. E., Steele, A. M., Taylor, R. S., Shore, A. C., Ukoumunne, O. C., & Campbell, J. L. (2014) · Diabetes Care · doi:10.2337/dc13-1576 · Cited by 67
Pinho‐Gomes, A., Azevedo, L. F., Copland, E., Canoy, D., Nazarzadeh, M., Ramakrishnan, R., Berge, E., Sundström, J., Kotecha, D., Woodward, M., et al. (2021) · PLoS Medicine · doi:10.1371/journal.pmed.1003599 · Cited by 62
Kiuchi, M. G., Chen, S., Hoye, N. A., & Pürerfellner, H. (2017) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-017-0302-2 · Cited by 54
Clark, C. E., McDonagh, S. T. J., & McManus, R. J. (2019) · Journal of Human Hypertension · doi:10.1038/s41371-018-0153-z · Cited by 49
Lin, H., Wang, T., Chen, M. Y., Hsu, C., Wang, K.-L., Huang, C., Hsieh, M., Chiu, Y.-W., Chiang, L.-T., Chuang, W.-P., et al. (2020) · PubMed Central · doi:10.6515/acs.202011_36(6).20201106a · Cited by 40
Sánchez, R. A., Boggia, J., Peñaherrera, E., Barroso, W. S., Barbosa, E., Villar, R., Cobos, L., Hernández Hernández, R., Lopez, J., Octavio, J. A., et al. (2020) · The Journal of Clinical Hypertension · doi:10.1111/jch.13816 · Cited by 38
Berkelmans, G. F. N., Kuipers, S., Westerhof, B. E., de Man, A., & Smulders, Y. M. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0044-9 · Cited by 35
Human Errors in Automated Office Blood Pressure Measurement: Still Room for Improvement
Elias, M. F., & Goodell, A. L. (2021) · Hypertension · doi:10.1161/hypertensionaha.120.16164 · Cited by 31
Cuff-based blood pressure measurement: challenges and solutions
Pilz, N., Picone, D., Patzak, A., Opatz, O., Lindner, T., Fesseler, L., Heinz, V., & Bothe, T. (2024) · Blood Pressure · doi:10.1080/08037051.2024.2402368 · Cited by 33
Myers, M. G. (2013) · Clinical and Experimental Pharmacology and Physiology · doi:10.1111/1440-1681.12149 · Cited by 33
Kiuchi, M. G., Chen, S., Silva, G. R. E., Paz, L. M. R., Kiuchi, T., Filho, A. G. D. P., & Souto, G. L. L. (2016) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-016-0186-6 · Cited by 31
Automated measurement of office, home and ambulatory blood pressure in atrial fibrillation
Κόλλιας, Α., & Stergiou, G. S. (2013) · Clinical and Experimental Pharmacology and Physiology · doi:10.1111/1440-1681.12103 · Cited by 29
Self-citationBlood pressure measurement in special populations and circumstances
Stergiou, G. S., Dolan, E., Kollias, A., Poulter, N. R., Shennan, A., Staessen, J. A., Zhang, Z., & Weber, M. A. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13296 · Cited by 29
Self-citationHalfon, M., Wuerzner, G., Marques-Vidal, P., Taffé, P., Vaucher, J., Waeber, B., Liaudet, L., Ltaief, Z., Popov, М., & Waeber, G. (2018) · Blood Pressure · doi:10.1080/08037051.2017.1383852 · Cited by 21
Methods of Blood Pressure Assessment Used in Milestone Hypertension Trials
Chen, Y., Lei, L., & Wang, J. (2018) · Pulse · doi:10.1159/000489855 · Cited by 26
Park, S., Bergmark, B. A., Shi, M., Lanz, H., Chung, N., Ruff, C. T., Antman, E. M., D, E. B. M., & Giugliano, R. P. (2019) · Hypertension · doi:10.1161/hypertensionaha.119.13138 · Cited by 25
Nolde, J. M., Kiuchi, M. G., Lugo‐Gavidia, L. M., Ho, J. K., Chan, J., Matthews, V. B., Herat, L. Y., Carnagarin, R., Azzam, O., & Schlaich, M. P. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002620 · Cited by 23
Fukuda, K., Matsuo, R., Kamouchi, M., Kiyuna, F., Sato, N., Nakamura, K., Hata, J., Wakisaka, Y., Ago, T., Imaizumi, T., et al. (2021) · Stroke · doi:10.1161/strokeaha.120.033751 · Cited by 22
Xie, F., Xu, J., Xia, L.-L., Luo, X., Jiang, Z., Wu, Y., & Su, H. (2020) · Hypertension Research · doi:10.1038/s41440-019-0386-4 · Cited by 21
Systematic Review: The Association between Late Life Depression and Hypotension
Briggs, R., Kenny, R. A., & Kennelly, S. (2016) · Journal of the American Medical Directors Association · doi:10.1016/j.jamda.2016.06.027 · Cited by 21
Phillips, R., Smith, B. E., & Madigan, V. (2017) · Current Anesthesiology Reports · doi:10.1007/s40140-017-0235-4 · Cited by 21
Topouchian, J. A., Hakobyan, Z., Asmar, J. T., Sv, G., Zelveian, P., & Asmar, R. G. (2018) · Vascular Health and Risk Management · doi:10.2147/vhrm.s165524 · Cited by 21
Chazova, I. E., Михайловна, Ч. Н., Blinova, N. V., Владимировна, А. А., Alekseeva, T. A., Ambatiello, L. G., Balanova, Y. А., Bragina, A. E., Danilov, N. M., Drapkina, O. M., et al. (2024) · Systemic Hypertension · doi:10.38109/2075-082x-2024-4-5-109 · Cited by 23
Šelmytė–Besusparė, A., Barysienė, J., Petrikonytė, D., Aidietis, A., Marinskis, G., & Laucevičius, A. (2017) · BMC Cardiovascular Disorders · doi:10.1186/s12872-017-0521-6 · Cited by 18
Kiuchi, M. G., Chen, S., Silva, G. R. E., Paz, L. M. R., Kiuchi, T., Filho, A. G. D. P., & Souto, G. L. L. (2016) · Kidney Research and Clinical Practice · doi:10.1016/j.krcp.2016.08.005 · Cited by 18
Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia
Lakhal, K., Martin, M., Ehrmann, S., Faiz, S., Rozec, B. R., & Boulain, T. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0067-2 · Cited by 18
Does Atrial Fibrillation Affect the Automated Oscillometric Blood Pressure Measurement?
Stergiou, G. S., Κόλλιας, Α., & Karpettas, N. (2013) · Hypertension · doi:10.1161/hypertensionaha.113.02211 · Cited by 17
Self-citationMaselli, M., Giantin, V., Corrado, D., Franchin, A., Attanasio, F., Pengo, V., Tramontano, A., Toni, P. D., Perissinotto, E., & Manzato, E. (2015) · Journal of Clinical Hypertension · doi:10.1111/jch.12547 · Cited by 16
Accuracy of oscillometric blood pressure measurement in atrial fibrillation
Feenstra, R. K., Allaart, C. P., Berkelmans, G. F. N., Westerhof, B. E., & Smulders, Y. M. (2017) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000305 · Cited by 16
Chang, Y., Kim, J., Kim, M. H., Kim, Y., & Song, T. (2018) · Journal of Clinical Neurology · doi:10.3988/jcn.2018.14.4.555 · Cited by 16
Nolde, J. M., Frost, S. M., Kannenkeril, D., Lugo‐Gavidia, L. M., Chan, J., Joyson, A., Azzam, O., Carnagarin, R., Kiuchi, M. G., Vignarajan, J., et al. (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002863 · Cited by 16
Blood pressure variability in hypertensive patients with atrial fibrillation in the VALUE trial
Mehlum, M. H., Liestøl, K., Wyller, T. B., Hua, T. A., Rostrup, M., & Berge, E. (2019) · Blood Pressure · doi:10.1080/08037051.2018.1524707 · Cited by 17
Hemodynamic monitoring devices: Putting it all together
Naik, B. I., & Durieux, M. E. (2014) · Best practice & research. Clinical anaesthesiology · doi:10.1016/j.bpa.2014.09.004 · Cited by 15
Κόλλιας, Α., Kyriakoulis, K. G., Stambolliu, E., & Stergiou, G. S. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002244 · Cited by 15
Self-citationPark, S.-H., & Choi, Y.-K. (2020) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000414 · Cited by 14
Effects of atrial fibrillation on the arterial fluid dynamics: a modelling perspective
Scarsoglio, S., Gallo, C., & Ridolfi, L. (2018) · Meccanica · doi:10.1007/s11012-018-0867-6 · Cited by 14
Atrial Fibrillation and Hypertension: “Quo Vadis”
Kwon, Y., Walker, M., Patel, P., Kwon, O., Koene, R. J., & Duprez, D. A. (2022) · Current Hypertension Reviews · doi:10.2174/1573402118666220112122403 · Cited by 15
Standardized home blood pressure monitoring: Rationale behind the 722 protocol
Lin, H.-J., Pan, H.-Y., Chen, C.-H., Cheng, H.-M., Chia, Y., Sogunuru, G., Tay, J., Turana, Y., Verma, N., Kario, K., et al. (2022) · The Journal of Clinical Hypertension · doi:10.1111/jch.14549 · Cited by 12
Kumar, S., Yadav, S., & Kumar, A. (2023) · Journal of Human Hypertension · doi:10.1038/s41371-023-00856-4 · Cited by 13
Kiuchi, M. G., Chen, S., & Hoye, N. A. (2017) · Kidney Research and Clinical Practice · doi:10.23876/j.krcp.2017.36.3.264 · Cited by 13
Chang, Y., Kim, J., Kim, Y., & Song, T. (2019) · Scientific Reports · doi:10.1038/s41598-019-49294-8 · Cited by 12
Should Oscillometric Blood Pressure Monitors Be Used in Patients With Atrial Fibrillation?
Myers, M. G., & Stergiou, G. S. (2015) · Journal of Clinical Hypertension · doi:10.1111/jch.12545 · Cited by 11
Self-citationShubik, V. V., Пивоваров, В. В., Zaitsev, G. К., Борисович, К. А., Тихоненко, В. М., Kormilitsyn, A. Y., Гордеева, М. В., Berman, M. V., Lobov, G. I., Bondarev, S., et al. (2021) · Journal of Arrhythmology · doi:10.35336/va-2021-1-23-32 · Cited by 11
Nolde, J., Kiuchi, M., Carnagarin, R., Frost, S., Kannenkeril, D., Lugo-Gavidia, L. M., Chan, J., Joyson, A., Matthews, V., Herat, L., et al. (2020) · The Journal of Clinical Hypertension · doi:10.1111/jch.14114 · Cited by 9
Accuracy of non-invasive blood pressure measurement in patients with atrial fibrillation
Su, H., & Guo, Z. (2021) · Journal of Human Hypertension · doi:10.1038/s41371-021-00596-3 · Cited by 10
Stergiou, G. S., Kyriakoulis, K. G., Bountzona, I., Menti, A., Destounis, A., Kalogeropoulos, P. G., & Κόλλιας, Α. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002684 · Cited by 10
Self-citationObtaining Accurate Noninvasive Blood Pressure Measurements in Adults
(2016) · Critical Care Nurse · doi:10.4037/ccn2016590 · Cited by 10
Wang, X., Shuai, W., Hong, K., Xu, J., Li, J., Li, P., Cheng, X., & Su, H. (2016) · Hypertension Research · doi:10.1038/hr.2016.31 · Cited by 10
Almroth, H., Karlsson, L., Carlhäll, C., & Charitakis, E. (2023) · European Heart Journal Open · doi:10.1093/ehjopen/oead112 · Cited by 9
Maselli, M., Giantin, V., Franchin, A., Attanasio, F., Tramontano, A., Toni, P. D., Pengo, V., Corrado, D., & Manzato, E. (2016) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000211 · Cited by 9
Incidence of ventricular arrhythmic events in CKD patients with ICD
Kiuchi, M. G., Chen, S., & Pürerfellner, H. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2016.11.094 · Cited by 9
Atrial fibrillation does not affect ankle–brachial index measured using the Doppler method
Dąbrowski, M., Lewandowski, J., Abramczyk, P., Łoń, I., Gaciong, Z. A., & Siński, M. (2017) · Hypertension Research · doi:10.1038/hr.2017.89 · Cited by 9
Automated blood pressure measurement in atrial fibrillation
Stergiou, G. S., Κόλλιας, Α., Destounis, A., & Tzamouranis, D. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32835ae9b3 · Cited by 9
Self-citationΚωνσταντίνου, Κ., Apostolos, A., Tsiachris, D., Dimitriadis, K. S., Papakonstantinou, P. Ε., Pappelis, K., Panoulas, V. F., & Tsioufis, K. P. (2024) · Journal of Human Hypertension · doi:10.1038/s41371-024-00936-z · Cited by 8
Xie, F., Xu, J., Liu, H., Li, X., Wu, Y., & Su, H. (2021) · Journal of Human Hypertension · doi:10.1038/s41371-021-00512-9 · Cited by 8
Home blood pressure monitoring: what does the evidence say?
Khong, T. K., & Fok, H. (2021) · Drug and Therapeutics Bulletin · doi:10.1136/dtb.2021.000025 · Cited by 8
The accuracy of SpaceLabs 90207 in blood pressure monitoring in patients with atrial fibrillation
Miszkowska-Nagórna, E., Neubauer‐Geryk, J., Wolf, J., Wielicka, M., Raczak, G., Narkiewicz, K., & Bieniaszewski, L. (2017) · Blood Pressure · doi:10.1080/08037051.2017.1356200 · Cited by 7
Norioka, N., Iwata, S., Ito, A., Tamura, S., Kawai, Y., Nonin, S., Ishikawa, S., Doi, A., Hanatani, A., & Yoshiyama, M. (2018) · Hypertension Research · doi:10.1038/s41440-018-0060-2 · Cited by 7
Xu, J., Wu, Y., Su, H., Hu, W., Li, J., Wang, W., Liu, X., & Cheng, X. (2014) · PLoS ONE · doi:10.1371/journal.pone.0100287 · Cited by 7
Ambulatory Blood Pressure Measurement in the Elderly
O'Brien, E. T., & Dolan, E. (2019) · Hypertension · doi:10.1161/hypertensionaha.118.11977 · Cited by 7
Omboni, S., Ballatore, T., Rizzi, F., Tomassini, F., Campolo, L., Panzeri, E., Lundwall, K., & Kahan, T. (2022) · Hypertension Research · doi:10.1038/s41440-022-00932-1 · Cited by 6
Shuai, W., Wang, X., Hong, K., Peng, Q., Li, J., Li, P., Cheng, X., & Su, H. (2016) · International Journal of Cardiology · doi:10.1016/j.ijcard.2016.08.091 · Cited by 6
Artificial Intelligence in Cardiovascular Medicine: A Giant Step in Personalized Medicine?
Jankauskas, S. S., Varzideh, F., Kansakar, U., & Santulli, G. (2026) · Journal of Personalized Medicine · doi:10.3390/jpm16040192 · Cited by 6
Ambulatory hypertensive patients treated by cardiologists in France
Blacher, J., Peroz, J., Huberman, J.-P., Ferrini, M., Jullien, G., Guenoun, M., & Guedj-Meynier, D. (2013) · Archives of cardiovascular diseases · doi:10.1016/j.acvd.2012.11.002 · Cited by 5
Tasić, D., Topouchian, J. A., Dragisic, D., Tasić, N., Hakobyan, Z., Vatinyan, S., Zelveian, P., & Asmar, R. G. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002106 · Cited by 5
Minhas, J. S., Coles, B., Mistri, A. K., Eveson, D., Hussain, S. T., Potter, J. F., Khunti, K. K., & Robinson, T. G. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002487 · Cited by 5
Kiuchi, M. G., & Chen, S. (2017) · Kidney Research and Clinical Practice · doi:10.23876/j.krcp.2017.36.2.167 · Cited by 4
Armitage, L. C., Lawson, B. K., Román, C., Thompson, B., Biggs, C. A., Rutter, H., Lewis-Jones, M., Ede, J., Tarassenko, L., Farmer, A. A., et al. (2022) · Wellcome Open Research · doi:10.12688/wellcomeopenres.17537.3 · Cited by 4
Should the measurement of blood pressure in the office be redefined?
Stergiou, G. S., & Parati, G. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e328358a221 · Cited by 5
Self-citationHome Monitoring of Blood Pressure
Stergiou, G. S., & Κόλλιας, Α. (2017) · Elsevier eBooks · doi:10.1016/b978-0-323-42973-3.00010-x · Cited by 3
Self-citationMohammad, J., Khan, K. A., Hafeez, I., Rather, H., & Lone, A. A. (2022) · Indian Heart Journal · doi:10.1016/j.ihj.2022.10.194 · Cited by 4
O’Connell, D. G., Jumper, J., Hicks, B., Klostermann, B., Carl, C., & Bell, C. (2023) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/cpt.0000000000000223 · Cited by 3
Xu, Y., Zai, W., Yang, M., Zhu, L., Zhang, Y., Fu, X. Y., & Dai, T. (2024) · Medicine · doi:10.1097/md.0000000000038179 · Cited by 3
Lin, T., Wang, C., Liao, M., & Lai, C. (2018) · The Journal of Cardiovascular Nursing · doi:10.1097/jcn.0000000000000486 · Cited by 3
Eysenck, W., Ammar, A. S., Kanthasamy, V., Freemantle, N., Veasey, R. A., Patel, N., FURNISS, S. S., & Sulke, N. (2019) · International Journal of Clinical Practice · doi:10.1111/ijcp.13410 · Cited by 3
Dabbagh, A. (2013) · Springer eBooks · doi:10.1007/978-3-642-40418-4_4 · Cited by 3
Mai, S., Zhu, H., Li, M., Zeng, Y., Zhang, Y., Huo, Y., Pan, X., & Huang, Y. (2022) · Therapeutic Advances in Chronic Disease · doi:10.1177/20406223221137040 · Cited by 2
Elias, M. F., Goodell, A. L., & Davey, A. (2020) · American Journal of Hypertension · doi:10.1093/ajh/hpaa216 · Cited by 2
Ambulatory blood pressure in atrial fibrillation: an irregular conundrum of rate and rhythm
Kotecha, D., & Lip, G. Y. H. (2013) · Hypertension Research · doi:10.1038/hr.2013.59 · Cited by 2
Eckert, S., Ott, C., Buchner, N. J., Dechend, R., Lüders, S., Mengden, T., & Sanner, B. M. (2019) · DMW - Deutsche Medizinische Wochenschrift · doi:10.1055/a-0986-7103 · Cited by 1
Abdel-Galeel, A., Fawzy, N. N., Ali, W. A., & Fouad, D. A. (2023) · BMC Cardiovascular Disorders · doi:10.1186/s12872-023-03292-x · Cited by 1
Metadata
- Authors
- Stergiou, G. S., Kollias, A., Destounis, A., & Tzamouranis, D.
- Affiliations
- Sotiria General Hospital (GR)
- Corresponding author
- Stergiou, George S.
- Venue
- Journal of Hypertension
- Published
- 2012-11
- Volume
- 30
- Issue
- 11
- Pages
- 2074-2082
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W1967096675
- Cited by
- 120
- References
- 31
- Citations by year
- 2026: 6 · 2024: 4 · 2023: 4 · 2022: 9 · 2021: 9 · 2020: 10 · 2019: 14 · 2018: 9 · 2017: 15 · 2016: 13 · 2015: 8 · 2014: 6 · 2013: 11 · 2012: 2
- FWCI
- 5.56
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Hypertension (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Atrial Fibrillation Management and Outcomes
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- atrial fibrillation, automated blood pressure measurement, oscillometric blood pressure monitor, auscultatory blood pressure measurement, systolic blood pressure, diastolic blood pressure, measurement accuracy, home blood pressure monitoring, ambulatory blood pressure monitoring, office blood pressure measurement
- MeSH terms
- Analysis of Variance, Atrial Fibrillation, Automation, Blood Pressure Determination, Heart Rate, Humans, Hypertension, Systole, Reproducibility of Results, Blood Pressure Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review