An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country population: the Microlife 3AS1-2
The DOI matches its Crossref record.
Abstract
OBJECTIVES: To assess the accuracy of the Microlife 3AS1-2 blood pressure device in pregnancy and pre-eclampsia in a low-resource setting. METHODS: Prospective validation according to the British Hypertension Society protocol. A total of 45 pregnant women were recruited from Kimberley Hospital (South Africa), of whom 15 had pre-eclampsia. RESULTS: The Microlife 3AS1-2 device achieved an overall B/A grade in pregnancy (including pre-eclampsia), passing the British Hypertension Society protocol requirements and achieving the International Organization for Standardization standard with a mean difference and SD of -3.8 ± 7.3 and -1.5 ± 6.2 mmHg for systolic and diastolic pressures, respectively. CONCLUSION: The Microlife 3AS1-2 device can be recommended for use in pregnancy, including pre-eclampsia. In addition, it fulfils the requirements stipulated by the WHO for an automated blood pressure device suitable for use in a low-resource setting. This makes it the ideal device for antenatal clinics and primary healthcare facilities in low-income and middle-income countries.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The device has additionally been validated as accurate for use in women with preeclampsia [12] and in those who develop low BP from sepsis or haemorrhage [13].”
Citations
Cited by
Miller, S., Abalos, E., Chamillard, M., Ciapponi, A., Colaci, D., Comandé, D., Diaz, V., Geller, S., Hanson, C., Langer, A., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31472-6 · Cited by 1,181
Preventing deaths due to the hypertensive disorders of pregnancy
von Dadelszen, P., & Magee, L. A. (2016) · Best Practice & Research Clinical Obstetrics & Gynaecology · doi:10.1016/j.bpobgyn.2016.05.005 · Cited by 192
Kitt, J., Fox, R., Tucker, K. L., & McManus, R. J. (2019) · Current Hypertension Reports · doi:10.1007/s11906-019-0949-4 · Cited by 177
Magee, L. A., Sharma, S., Nathan, H. L., Adetoro, O. O., Bellad, M. B., Goudar, S., Macuacua, S. E., Mallapur, A., Qureshi, R., Sevene, E., et al. (2019) · PLOS Medicine · doi:10.1371/journal.pmed.1002783 · Cited by 135
“To our knowledge, this is the first study of pregnancy hypertension incidence and type in less-developed countries that is both population-based in terms of participants and standardised in terms of BP measurement, by positioning and use of a pregnancy-validated BP device [23].”
MethodInfluentialSelf-citationEl Ayadi, A. M., Nathan, H. L., Seed, P. T., Butrick, E. A., Hezelgrave, N. L., Shennan, A. H., & Miller, S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0148729 · Cited by 121
Self-citation
Metadata
- Authors
- Nathan, H. L., de Greeff, A., Hezelgrave, N. L., Chappell, L. C., & Shennan, A. H.
- Affiliations
- King's College London (GB)
- Corresponding author
- Nathan, Hannah L.
- Venue
- Blood Pressure Monitoring
- Published
- 2015-02
- Volume
- 20
- Issue
- 1
- Pages
- 52-55
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2323067445
- Cited by
- 66
- References
- 0
- Citations by year
- 2026: 2 · 2024: 1 · 2023: 6 · 2022: 5 · 2021: 6 · 2020: 8 · 2019: 11 · 2018: 12 · 2017: 9 · 2016: 4 · 2015: 2
- FWCI
- 5.32
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Blood Pressure Monitoring (publishedVersion)
- PubMed (publishedVersion)
- Research Portal (King's College London) (submittedVersion)
- Funders
- Bill and Melinda Gates Foundation
- National Institute for Health and Care Research (ACF-2009-17-019, DRF-2013-06-171, RP-2014-05-019)
- Universiteit van die Vrystaat
- Medical Research Council (MR/N006240/1)
- Topics
- Health Sciences › Medicine › Obstetrics and Gynecology › Pregnancy and preeclampsia studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- oscillometric blood pressure measurement, pregnancy, preeclampsia, automated blood pressure devices, resource-limited settings, prenatal care, hypertension in pregnancy
- MeSH terms
- Adult, Blood Pressure, Blood Pressure Determination, Developing Countries, Female, Humans, Oscillometry, Poverty, Pre-Eclampsia, Pregnancy, South Africa, Global Health, World Health Organization
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country population: the Microlife 3AS1-2. https://app.cochat.ai/papers/10.1097/mbp.0000000000000086
- The paper
- Nathan, H. L., de Greeff, A., Hezelgrave, N. L., Chappell, L. C., & Shennan, A. H. (2015). An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country population: the Microlife 3AS1-2. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0000000000000086
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.