An accurate semiautomated oscillometric blood pressure device for use in pregnancy (including pre-eclampsia) in a low-income and middle-income country population: the Microlife 3AS1-2
The DOI matches its Crossref record.
Abstract
OBJECTIVES: To assess the accuracy of the Microlife 3AS1-2 blood pressure device in pregnancy and pre-eclampsia in a low-resource setting. METHODS: Prospective validation according to the British Hypertension Society protocol. A total of 45 pregnant women were recruited from Kimberley Hospital (South Africa), of whom 15 had pre-eclampsia. RESULTS: The Microlife 3AS1-2 device achieved an overall B/A grade in pregnancy (including pre-eclampsia), passing the British Hypertension Society protocol requirements and achieving the International Organization for Standardization standard with a mean difference and SD of -3.8 ± 7.3 and -1.5 ± 6.2 mmHg for systolic and diastolic pressures, respectively. CONCLUSION: The Microlife 3AS1-2 device can be recommended for use in pregnancy, including pre-eclampsia. In addition, it fulfils the requirements stipulated by the WHO for an automated blood pressure device suitable for use in a low-resource setting. This makes it the ideal device for antenatal clinics and primary healthcare facilities in low-income and middle-income countries.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The device has additionally been validated as accurate for use in women with preeclampsia [12] and in those who develop low BP from sepsis or haemorrhage [13].”
Citations
Cited by
Miller, S., Abalos, E., Chamillard, M., Ciapponi, A., Colaci, D., Comandé, D., Diaz, V., Geller, S., Hanson, C., Langer, A., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31472-6 · Cited by 1,181
Preventing deaths due to the hypertensive disorders of pregnancy
von Dadelszen, P., & Magee, L. A. (2016) · Best Practice & Research Clinical Obstetrics & Gynaecology · doi:10.1016/j.bpobgyn.2016.05.005 · Cited by 192
Kitt, J., Fox, R., Tucker, K. L., & McManus, R. J. (2019) · Current Hypertension Reports · doi:10.1007/s11906-019-0949-4 · Cited by 181
Magee, L. A., Sharma, S., Nathan, H. L., Adetoro, O. O., Bellad, M. B., Goudar, S., Macuacua, S. E., Mallapur, A., Qureshi, R., Sevene, E., et al. (2019) · PLOS Medicine · doi:10.1371/journal.pmed.1002783 · Cited by 135
“To our knowledge, this is the first study of pregnancy hypertension incidence and type in less-developed countries that is both population-based in terms of participants and standardised in terms of BP measurement, by positioning and use of a pregnancy-validated BP device [23].”
MethodInfluentialSelf-citationEl Ayadi, A. M., Nathan, H. L., Seed, P. T., Butrick, E. A., Hezelgrave, N. L., Shennan, A. H., & Miller, S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0148729 · Cited by 121
Self-citationBello, N. A., Woolley, J. J., Cleary, K. L., Falzon, L., Alpert, B. S., Oparil, S., Cutter, G., Wapner, R., Muntner, P., Tita, A. T., et al. (2018) · Hypertension · doi:10.1161/hypertensionaha.117.10295 · Cited by 126
Pregnancy-Related Acute Kidney Injury in Preeclampsia: Risk Factors and Renal Outcomes
Conti-Ramsden, F. I., Nathan, H. L., de Greeff, A., Hall, D. R., Seed, P. T., Chappell, L. C., Shennan, A. H., & Bramham, K. (2019) · Hypertension · doi:10.1161/hypertensionaha.119.13089 · Cited by 122
“BP was measured using the Microlife CRADLE Vital Signs Alert—a device validated for use in pregnancy including preeclampsia.(13) The study was approved by the Ethics Committee at each hospital site: Stellenbosch University Ethics Committee (N14/06068), University of Cape Town Ethics Committees (410/2014), and the Univ…”
MethodSelf-citationVousden, N., Lawley, E., Seed, P. T., Gidiri, M. F., Goudar, S. S., Sandall, J., Chappell, L. C., Shennan, A. H., & on behalf of the CRADLE Trial Collaborative Group (2019) · PLoS Medicine · doi:10.1371/journal.pmed.1002775 · Cited by 132
Self-citationHome blood‐pressure monitoring in a hypertensive pregnant population
Perry, H., Sheehan, E., Thilaganathan, B., & Khalil, A. A. (2018) · Ultrasound in Obstetrics and Gynecology · doi:10.1002/uog.19023 · Cited by 96
“Concerns over differences in the home and clinic readings of PE patients using an automated device and mercury sphygmomanometry have been reported,(20) and although validation studies for devices used in pregnancy often include PE patients, readings in this group may be less accurate.(11,21,22)”
MethodThe assessment of blood pressure in pregnant women: pitfalls and novel approaches
Hurrell, A., Webster, L., Chappell, L. C., & Shennan, A. H. (2021) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2020.10.026 · Cited by 79
Self-citationNathan, H. L., Seed, P. T., Hezelgrave, N. L., de Greeff, A., Lawley, E., Conti-Ramsden, F. I., Anthony, J., Steyn, W., Hall, D. R., Chappell, L. C., et al. (2018) · Journal of Global Health · doi:10.7189/jogh.08.020401 · Cited by 74
Self-citationNathan, H. L., Seed, P. T., Hezelgrave, N. L., de Greeff, A., Lawley, E., Anthony, J., Steyn, W., Hall, D. R., Chappell, L. C., & Shennan, A. H. (2019) · Acta Obstetricia Et Gynecologica Scandinavica · doi:10.1111/aogs.13626 · Cited by 76
Self-citationBellad, M. B., Goudar, S. S., Mallapur, A. A., Sharma, S., Bone, J. N., Charantimath, U., Katageri, G., Ramadurg, U. Y., Ansermino, J. M., Derman, R. J., et al. (2020) · Pregnancy Hypertension · doi:10.1016/j.preghy.2020.05.008 · Cited by 51
Self-citationVousden, N., Lawley, E., Nathan, H. L., Seed, P. T., Gidiri, M. F., Goudar, S., Sandall, J., Chappell, L. C., Shennan, A. H., Kachinjika, M., et al. (2019) · The Lancet Global Health · doi:10.1016/s2214-109x(18)30526-6 · Cited by 48
Self-citationthe CLIP Working Group, Nathan, H. L., Boene, H., Munguambe, K., Sevene, E., Akeju, D., Adetoro, O. O., Charanthimath, U., Bellad, M. B., de Greeff, A., et al. (2018) · Reproductive Health · doi:10.1186/s12978-017-0450-y · Cited by 41
“The device has been accuracy validated in pregnancy, including pre-eclampsia and hypotensive pregnancies, according to formal validation protocols [3, 4].”
MethodSelf-citationNathan, H. L., Seed, P. T., Hezelgrave, N. L., De Greeff, A., Lawley, E., Anthony, J., Hall, D. R., Steyn, W., Chappell, L. C., & Shennan, A. H. (2018) · Pregnancy Hypertension · doi:10.1016/j.preghy.2017.11.003 · Cited by 40
Self-citationQureshi, R. N., Sheikh, S. S., Hoodbhoy, Z. A., Sharma, S., Vidler, M., Payne, B. A., Ahmed, I., Ansermino, J. M., Bone, J. N., Dunsmuir, D. T., et al. (2020) · Pregnancy Hypertension · doi:10.1016/j.preghy.2020.07.011 · Cited by 41
Self-citationReliability of home blood pressure monitoring devices in pregnancy
Tremonti, C., Beddoe, J., & Brown, M. A. (2017) · Pregnancy Hypertension · doi:10.1016/j.preghy.2017.01.002 · Cited by 41
the PRECISE Network, von Dadelszen, P., Flint-O’Kane, M., Lucilla, P., Craik, R., Russell, D., Tribe, R. M., D’Alessandro, U., Roca, A., Jah, H., et al. (2020) · Reproductive Health · doi:10.1186/s12978-020-0872-9 · Cited by 37
Self-citationMcKinnon, B., Sall, M., Vandermorris, A., Traoré, M., Lamesse‐Diedhiou, F., McLaughlin, K., & Bassani, D. G. (2020) · Health Policy and Planning · doi:10.1093/heapol/czz178 · Cited by 39
Nathan, H. L., Vousden, N., Lawley, E., de Greeff, A., Hezelgrave, N. L., Sloan, N., Tanna, N., Goudar, S. S., Gidiri, M. F., Sandall, J., et al. (2018) · BMJ Innovations · doi:10.1136/bmjinnov-2017-000235 · Cited by 31
Self-citationon behalf of the CRADLE Trial Collaborative Group, Vousden, N., Lawley, E., Seed, P. T., Gidiri, M. F., Charantimath, U., Makonyola, G., Brown, A., Yadeta, L., Best, R., et al. (2019) · Implementation Science · doi:10.1186/s13012-019-0885-3 · Cited by 31
Self-citationSevene, E., Sharma, S., Munguambe, K., Sacoor, C. N., Valá, A., Macuácua, S., Boene, H. E., Ansermino, J. M., Augusto, O. J., Bique, C., et al. (2020) · Pregnancy Hypertension · doi:10.1016/j.preghy.2020.05.006 · Cited by 30
Self-citationon behalf of the CRADLE Trial Collaborative Group, Vousden, N., Lawley, E., Nathan, H. L., Seed, P. T., Brown, A., Muchengwa, T., Charantimath, U., Bellad, M., Gidiri, M. F., et al. (2018) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-018-1737-x · Cited by 30
Self-citationBone, J. N., Magee, L. A., Singer, J., Nathan, H. L., Qureshi, R. N., Sacoor, C. N., Sevene, E., Shennan, A. H., Bellad, M. B., Goudar, S. S., et al. (2021) · The Lancet Global Health · doi:10.1016/s2214-109x(21)00219-9 · Cited by 35
Self-citationInnovations in vital signs measurement for the detection of hypertension and shock in pregnancy
Vousden, N., Nathan, H. L., & Shennan, A. H. (2018) · Reproductive Health · doi:10.1186/s12978-018-0533-4 · Cited by 29
“as well as in pregnancy, including women with hypertension [45] or hypotension [46].”
BackgroundSelf-citationFeroz, A. S., Afzal, N., & Seto, E. (2022) · BMJ Open · doi:10.1136/bmjopen-2021-056130 · Cited by 48
Sharma, S., Adetoro, O. O., Vidler, M., Drebit, S. K., Payne, B. A., Akeju, D. O., Adepoju, A., Jaiyesimi, E., Sotunsa, J., Bhutta, Z. A., et al. (2017) · BMC Health Services Research · doi:10.1186/s12913-017-2124-4 · Cited by 31
“This second set of focus groups was used to assess the competency and experience of health workers with the Microlife VSA blood pressure device [20].”
First trimester screening for pre‐eclampsia in Chinese pregnancies: case–control study
Cheng, Y., Leung, T., Law, L. W., Ting, Y., Law, K., & Sahota, D. S. (2017) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/1471-0528.14970 · Cited by 23
and the Community Level Interventions for Pre-eclampsia (CLIP) India Feasibility Working Group, Charantimath, U., Vidler, M., Katageri, G., Ramadurg, U. Y., Karadiguddi, C., Kavi, A. J., Joshi, A. M., Mungarwadi, G., Bannale, S., et al. (2018) · Reproductive Health · doi:10.1186/s12978-018-0532-5 · Cited by 23
“This handheld device can be used effectively by unskilled personnel after minimal training [20, 21].”
BackgroundVousden, N., Holmes, E., Seed, P. T., Gidiri, M., Goudar, S. S., Sandall, J., Chinkoyo, S., Kumsa, L., Brown, A., Charantimath, U., et al. (2020) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/1471-0528.16309 · Cited by 21
Self-citationNathan, H. L., Duhig, K. E., Vousden, N., Lawley, E., Seed, P. T., Sandall, J., Bellad, M. B., Brown, A. C., Chappell, L. C., Goudar, S. S., et al. (2018) · Trials · doi:10.1186/s13063-018-2581-z · Cited by 20
“The CRADLE VSA device has been validated as accurate for use in pregnancy, including in women with preeclampsia [9] and is the first device to be validated as accurate in pregnant women with low blood pressure; from example, from haemorrhage or sepsis [10].”
BackgroundMethodSelf-citationStrategies to reduce the global burden of direct maternal deaths
von Dadelszen, P., & Magee, L. A. (2017) · Obstetric Medicine · doi:10.1177/1753495x16686287 · Cited by 26
“An important advance in the move toward offering every pregnant woman accurate BP measurement and detection of pregnancy hypertension has been the development of the semi-automated Microlife BP 3AS1-2 sphygmomanometer , which has been validated in women with the hypertension of pre-eclampsia.(48) This device avoids th…”
MethodBone, J. N., Bellad, M. B., Goudar, S. S., Mallapur, A. A., Charantimath, U., Ramadurg, U. Y., Katageri, G., Lesperance, M., Kinshella, M. W., Suleman, R., et al. (2022) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-022-04714-y · Cited by 29
“Data were collected during POM visits using standardised methods and a semi-automated, pregnancy-validated digital device (Microlife BP 3AS1-2) [23].”
MethodGiblin, L., Vousden, N., Nathan, H. L., Gidiri, F., Goudar, S. S., Charantimath, U., Sandall, J., Seed, P. T., Chappell, L. C., & Shennan, A. H. (2021) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-021-03796-4 · Cited by 17
Self-citationAffordable and low‐maintenance obstetric devices
Ayres‐de‐Campos, D., Stones, W. R., Theron, G. B., & Commitee, T. F. S. M. A. N. H. (2019) · International Journal of Gynecology & Obstetrics · doi:10.1002/ijgo.12838 · Cited by 15
von Dadelszen, P., Magee, L. A., Payne, B. A., Dunsmuir, D. T., Drebit, S. K., Dumont, G. A., Miller, S., Norman, J. E., Pyne‐Mercier, L., Shennan, A. H., et al. (2015) · International Journal of Gynecology & Obstetrics · doi:10.1016/j.ijgo.2015.02.008 · Cited by 15
“The device has been validated for use in both non-pregnant and pregnant populations (including both women with pre-eclampsia and low blood pressure), according to the British Hypertension Society protocol [29,30].”
MethodSelf-citationAgreement of aneroid and oscillometric blood pressure devices used in pregnancy
Aldridge, E., Mollen, J., Verburg, P. E., Wittwer, M. R., Dekker, G., Roberts, C. T., & Arstall, M. A. (2019) · Pregnancy Hypertension · doi:10.1016/j.preghy.2019.05.005 · Cited by 11
“The Microlife VSA has been previously validated in preeclamptic populations [5].”
MethodToledo‐Jaldin, L., Bull, S. S., Contag, S. A., Escudero, C., Gutiérrez, S. P., Heath, A., Roberts, J. M., Scandlyn, J. N., Julian, C. G., & MOORE, L. G. (2019) · Pregnancy Hypertension · doi:10.1016/j.preghy.2019.03.008 · Cited by 13
“have been validated for use in pregnancy and preeclampsia in particular (28, 29).”
Preeclampsia in 2021—a Perioperative Medical Challenge for the Anesthesiologist
van Dyk, D., Dyer, R. A., & Fernandes, N. L. (2021) · Anesthesiology Clinics · doi:10.1016/j.anclin.2021.08.005 · Cited by 16
Magee, L. A., Bone, J. N., Owasil, S., Singer, J., Lee, T., Bellad, M. B., Goudar, S. S., Logan, A. G., Macuácua, S., Mallapur, A. A., et al. (2021) · Hypertension · doi:10.1161/hypertensionaha.120.16851 · Cited by 10
Self-citationShobo, O. G., Okoro, A., Okolo, M., Longtoe, P., Omale, I., Ofiemu, E., & Anyanti, J. (2020) · Implementation Science Communications · doi:10.1186/s43058-020-00076-8 · Cited by 14
“It has also fulfilled the standard required by the British Hypertension Society (BHS) grading criteria, Association for the Advancement of Medical Instrumentation (ANSI/AAMI/ISO), and World Health Organization (WHO) for low-resource settings [49, 50].”
MethodRidout, A. E., Moses, F. L., Herm-Singh, S., Turienzo, C. F., Seed, P. T., Goodhart, V., Vousden, N., Sam, B., Momoh, M., Kamara, D., et al. (2023) · Trials · doi:10.1186/s13063-023-07587-4 · Cited by 8
“The device has additionally been validated as accurate for use in women with preeclampsia [12] and in those who develop low BP from sepsis or haemorrhage [13].”
MethodSelf-citationLihme, F., Basit, S., Persson, L. K. G., Larsen, M. O., Lauridsen, K. H., Lykke, J. A., Andersen, A. S., Halse, K., Thorsen‐Meyer, A., Melbye, M., et al. (2023) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/1471-0528.17667 · Cited by 8
Third‐Trimester Cardiovascular Function and Risk of Hypertensive Disorders of Pregnancy
Lihme, F., Basit, S., Persson, L. K. G., Larsen, M. O., Lauridsen, K. H., Lykke, J. A., Andersen, A. S., Thorsen‐Meyer, A., Pihl, K., Melbye, M., et al. (2024) · Journal of the American Heart Association · doi:10.1161/jaha.123.032673 · Cited by 7
Home blood pressure measurement in women with pregnancy-related hypertensive disorders
Lan, P. G., Hyett, J. A., & Gillin, A. G. (2017) · Pregnancy Hypertension · doi:10.1016/j.preghy.2017.09.005 · Cited by 7
Robbins, T., Shennan, A. H., Sandall, J., Guangul, T. E., Gebreyohannes, R. D., Abdella, A., Mayston, R., & Hanlon, C. (2023) · BMJ Open · doi:10.1136/bmjopen-2022-061500 · Cited by 14
Self-citationPopulation-level data on antenatal screening for proteinuria; India, Mozambique, Nigeria, Pakistan
Magee, L. A., Sharma, S., Sevene, E., Qureshi, R. N., Mallapur, A. A., Macuácua, S., Goudar, S. S., Bellad, M. B., Adetoro, O. O., Payne, B. A., et al. (2020) · Bulletin of the World Health Organization · doi:10.2471/blt.19.248898 · Cited by 5
Self-citationReverse innovation in maternal health
Firoz, T., Makanga, P. T., Nathan, H. L., Payne, B. A., & Magee, L. A. (2017) · Obstetric Medicine · doi:10.1177/1753495x17700515 · Cited by 7
Self-citationde Greeff, A., & Shennan, A. H. (2015) · Tropical Doctor · doi:10.1177/0049475515581542 · Cited by 5
Self-citationNgene, N. C., & Moodley, J. (2018) · Pregnancy Hypertension · doi:10.1016/j.preghy.2018.02.013 · Cited by 5
Persson, L. K. G., Lihme, F., Basit, S., Larsen, M. O., Scheller, N. M., Andersen, A. S., Halse, K., Thorsen‐Meyer, A., Sørensen, K. M., Wohlfahrt, J., et al. (2022) · Paediatric and Perinatal Epidemiology · doi:10.1111/ppe.12917 · Cited by 5
Lihme, F., Madsen, M. E., & Lykke, J. A. (2017) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000277 · Cited by 4
von Dadelszen, P., Bone, J. N., Sandhu, A., Ansermino, J. M., Qureshi, R. N., Sacoor, C. N., Sevene, E., Li, J., Vidler, M., Bellad, M. B., et al. (2023) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/1471-0528.17465 · Cited by 4
“8,9 These factors support the generalisability of our findings to the TABLE 2 Adjusted risk ratios (95% CI) comparing BP categories with normotension-1 (sBP 135 mmHg and dBP 85 mmHg) and their association with CLIP outcomes, according to BP outputs.”
BackgroundMethodMalawi: Findings from a national census of health facilities
Stones, W. R. (2018) · Malawi Medical Journal · doi:10.4314/mmj.v30i3.2 · Cited by 3
Blood pressure measurement in pregnancy
Beardmore‐Gray, A., Dyer, R. A., & Shennan, A. H. (2018) · International Journal of Obstetric Anesthesia · doi:10.1016/j.ijoa.2018.11.006 · Cited by 2
Self-citationKuhrt, K., Seed, P. T., & Shennan, A. H. (2023) · Tropical Doctor · doi:10.1177/00494755221134975 · Cited by 1
Self-citationShe, J., Guan, X., Liu, Y., & Xiang, H. (2016) · Journal of Clinical Hypertension · doi:10.1111/jch.12859 · Cited by 1
The challenge of measuring blood pressure in low-resource settings
Roberts, J. M., & Assibey‐Mensah, V. (2019) · The Lancet Global Health · doi:10.1016/s2214-109x(19)30019-1 · Cited by 1
Eley, V. A., Christensen, R., Kumar, S., & Callaway, L. K. (2018) · International Journal of Obstetric Anesthesia · doi:10.1016/j.ijoa.2018.11.003
Pibernat, A. D., Millán, R. G., & Pibernat, E. D. (2021) · International Nursing Review · doi:10.1111/inr.12664
Hedetoft, K., Lihme, F., & Lykke, J. A. (2022) · Danish Journal of Obstetrics and Gynaecology · doi:10.56182/djog.v1i1.12
“Blood pressure measurements were conducted using the semiautomatic, validated device “Microlife Vital Sign alert (VSA)” which is based on an internal algorithm identical to the “blood pressure 3AS1-2”; this device has been validated for use in pregnancy including hypotensive states and hypertensive disorders of pregna…”
MethodRidout, A. E., Moses, F., Singh, S. H., Turienzo, C. F., Seed, P. T., Goodhart, V., Vousden, N., Sam, B., Momoh, M., Kamara, D., et al. (2023) · Research Square · doi:10.21203/rs.3.rs-3019173/v1
Self-citationBone, J. N., Bellad, M., Goudar, S., Mallapur, A. A., Charantimath, U., Ramadurg, U., Katageri, G., Lesperance, M., Woo Kinshella, M.-L., Suleman, R., et al. (2022) · Figshare · doi:10.6084/m9.figshare.19766456
Femi‐Lawal, V. O., Mzuma, A., Alameen, A. M., Adeleke, O. A., Akanbi, O. P., Arinade, J. A., Abib, O. M., Mohamed, D. Y., Inalegwu, A. F., Nwuta, C., et al. (2026) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-026-09352-2
“The study recruited 45 pregnant women at Kimberly Hospital in South Africa to evaluate the accuracy of the device in preeclampsia and pregnant women in community healthcare centres and antenatal clinics in LMICs [33].”
von Dadelszen, P., Sandhu, A., Bone, J. N., Qureshi, R. N., Dada, O. A., Mallapur, A. A., Temmerman, M., Volvert, M., Waiswa, J. A., Mukhanya, M. W., et al. (2026) · JMIR mhealth and uhealth · doi:10.2196/73307 · Cited by 1
“…access to BP and other vital signs measurement in low-income and middle-income countries, the semiautomated, low-cost Microlife BP 3AS1-2 and CRADLE Vital Signs Alert (VSA) devices were developed and have been validated in pregnant women with normal BP and pregnant women with hypertension [4,5].”
Metadata
- Authors
- Nathan, H. L., de Greeff, A., Hezelgrave, N. L., Chappell, L. C., & Shennan, A. H.
- Affiliations
- King's College London (GB)
- Corresponding author
- Nathan, Hannah L.
- Venue
- Blood Pressure Monitoring
- Published
- 2015-02
- Volume
- 20
- Issue
- 1
- Pages
- 52-55
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2323067445
- Cited by
- 66
- References
- 0
- Citations by year
- 2026: 2 · 2024: 1 · 2023: 6 · 2022: 5 · 2021: 6 · 2020: 8 · 2019: 11 · 2018: 12 · 2017: 9 · 2016: 4 · 2015: 2
- FWCI
- 5.32
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Blood Pressure Monitoring (publishedVersion)
- PubMed (publishedVersion)
- Research Portal (King's College London) (submittedVersion)
- Funders
- Bill and Melinda Gates Foundation
- National Institute for Health and Care Research (ACF-2009-17-019, DRF-2013-06-171, RP-2014-05-019)
- Universiteit van die Vrystaat
- Medical Research Council (MR/N006240/1)
- Topics
- Health Sciences › Medicine › Obstetrics and Gynecology › Pregnancy and preeclampsia studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- oscillometric blood pressure measurement, pregnancy, preeclampsia, automated blood pressure devices, resource-limited settings, prenatal care, hypertension in pregnancy
- MeSH terms
- Adult, Blood Pressure, Blood Pressure Determination, Developing Countries, Female, Humans, Oscillometry, Poverty, Pre-Eclampsia, Pregnancy, South Africa, Global Health, World Health Organization
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.