CoChat

Comparison of stethoscope bell and diaphragm, and of stethoscope tube length, for clinical blood pressure measurement

Open access
Liu, Chengyu, Griffiths, Clive, Murray, Alan, Zheng, Dingchang · 2016 · Blood Pressure Monitoring
Cited by 15

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: This study investigated the effect of stethoscope side and tube length on auscultatory blood pressure (BP) measurement. METHODS: Thirty-two healthy participants were studied. For each participant, four measurements with different combinations of stethoscope characteristics (bell or diaphragm side, standard or short tube length) were each recorded at two repeat sessions, and eight Korotkoff sound recordings were played twice on separate days to one experienced listener to determine the systolic and diastolic BPs (SBP and DBP). Analysis of variance was carried out to study the measurement repeatability between the two repeat sessions and between the two BP determinations on separate days, as well as the effects of stethoscope side and tube length. RESULTS: There was no significant paired difference between the repeat sessions and between the repeat determinations for both SBP and DBP (all P-values>0.10, except the repeat session for SBP using short tube and diaphragm). The key result was that there was a small but significantly higher DBP on using the bell in comparison with the diaphragm (0.66 mmHg, P=0.007), and a significantly higher SBP on using the short tube in comparison with the standard length (0.77 mmHg, P=0.008). CONCLUSION: This study shows that stethoscope characteristics have only a small, although statistically significant, influence on clinical BP measurement. Although this helps understand the measurement technique and resolves questions in the published literature, the influence is not clinically significant.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0000000000000175

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 15

Metadata

Authors
Liu, C., Griffiths, C., Murray, A., & Zheng, D.
Venue
Blood Pressure Monitoring
Published
2016-06
Volume
21
Issue
3
Pages
178-183
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2290662376
Cited by
15
References
1
FWCI
2.99
Type
journal article
Language
en
Open access
hybrid

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of stethoscope bell and diaphragm, and of stethoscope tube length, for clinical blood pressure measurement. https://app.cochat.ai/papers/10.1097/mbp.0000000000000175
The paper
Liu, C., Griffiths, C., Murray, A., & Zheng, D. (2016). Comparison of stethoscope bell and diaphragm, and of stethoscope tube length, for clinical blood pressure measurement. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0000000000000175

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.