CoChat

The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control

Tholl, Ulrich, Lüders, Stephan, Bramlage, Peter, Dechend, Ralf, Eckert, Siegfried, Mengden, Thomas, Nürnberger, Jens, Sanner, Bernd, Anlauf, Manfred · 2016 · Blood Pressure Monitoring
Cited by 33

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: The German Hypertension League (Deutsche Hochdruckliga) established a program to assess the accuracy and reliability of blood pressure (BP)-measuring devices in 1999 (Quality Seal Protocol). Here, we report on the results of a testing series of 105 devices designed for BP self-measurement. METHODS: The test protocol for the validation of upper-arm, wrist, and finger devices was developed to compare device to conventional Riva-Rocci measurements based on five criteria: mean systolic and mean diastolic differences, their standard deviations, and a point score representing the correlation of systolic and diastolic errors of individual comparisons. The results of this testing are summarized. RESULTS: From 1999 to 2014, a total of 105 BP devices for self-measurement were tested according to the Quality Seal Protocol. Of these, 47.6% fulfilled all five validation criteria, 55.7% of the upper-arm devices (39 of 71) and 32.4% (11 of 34) of the wrist devices. Finger devices were not offered for testing. Forty-four devices (41.9%) failed multiple test criteria of the validation procedure. A subanalysis with 51 devices tested showed that a stricter definition of the passing point score with a limit of at least 55% would slightly increase the consistency with the conventional criteria in comparison with a point score criterion of at least 50%. It was therefore introduced in 2007. CONCLUSION: The results indicate the importance of a rigorous testing of a BP-measuring device used for home BP measurement to prevent patients from making erroneous treatment decisions.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0000000000000186

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 2 references

Cited by

See all 33

Metadata

Authors
Tholl, U., Lüders, S., Bramlage, P., Dechend, R., Eckert, S., Mengden, T., Nürnberger, J., Sanner, B., & Anlauf, M.
Venue
Blood Pressure Monitoring
Published
2016-08
Volume
21
Issue
4
Pages
197-205
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2328769839
Cited by
33
References
2
FWCI
2.74
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control. https://app.cochat.ai/papers/10.1097/mbp.0000000000000186
The paper
Tholl, U., Lüders, S., Bramlage, P., Dechend, R., Eckert, S., Mengden, T., Nürnberger, J., Sanner, B., & Anlauf, M. (2016). The German Hypertension League (Deutsche Hochdruckliga) Quality Seal Protocol for blood pressure-measuring devices: 15-year experience and results from 105 devices for home blood pressure control. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0000000000000186

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.