Validation of the mobil-O-Graph: 24 h-blood pressure measurement device

Wei, Wolfgang, Tölle, Markus, Zidek, Walter, van der Giet, Markus · 2010 · Blood Pressure Monitoring

Abstract

OBJECTIVE: Twenty-four-hour blood pressure measurement is of importance not only in the detection of hypertension but also in the detection of blood pressure changes in hypertensive and nonhypertensives over the day to identify, for example, nondipper hypertensives. This study describes the validation of the mobil-O-Graph according to the criteria of the British Hypertension Society (BHS). METHODS: For each patient three readings obtained by the mobil-O-Graph were compared with auscultatory sphygmomanometric readings obtained by two trained clinicians. The sphygmomanometric reference measurements were alternated with the readings obtained by the device. Eighty-five patients (mean age 53.4+/-18.4 years) were recruited for the BHS protocol. Differences between blood pressure values of the test device and the mercury reading were calculated for each measurement. RESULTS: In the BHS validation procedure the mean differences of the observer readings and the test device were -2.2+/-6.7 (systolic) and -0.6+/-5.6 mmHg (diastolic) for observer 1 and -2.2+/-7.3 mmHg (systolic) and-0.4+/-6.1 mmHg (diastolic) for observer 2. The device achieved grade A for systolic and diastolic blood pressure for both the observers 1 and 2 leading to a final grade A/A. According to the BHS protocol the measurements of the device have to be considered 'very accurate and with no error of clinical relevance'. CONCLUSION: The device met the accuracy requirements of the BHS standard and can be recommended for clinical use.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0b013e328338892f

Cites

6 of 6 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 6 references

Cited by

See all 260

Metadata

Authors
Wei, W., Tölle, M., Zidek, W., & van der Giet, M.
Venue
Blood Pressure Monitoring
Published
2010-08
Volume
15
Issue
4
Pages
225-228
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2033328508
Cited by
260
References
6
FWCI
3.47
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Validation of the mobil-O-Graph: 24 h-blood pressure measurement device. https://app.cochat.ai/papers/10.1097/mbp.0b013e328338892f
The paper
Wei, W., Tölle, M., Zidek, W., & van der Giet, M. (2010). Validation of the mobil-O-Graph: 24 h-blood pressure measurement device. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e328338892f

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.