Oscillometric estimation of central blood pressure: validation of the Mobil-O-Graph in comparison with the SphygmoCor device

Weiss, Wolfgang, Gohlisch, Christopher, Harsch-Gladisch, Christl, Tölle, Markus, Zidek, Walter, van der Giet, Markus · 2012 · Blood Pressure Monitoring
Cited by 208

The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.

Abstract

BACKGROUND: Hypertension is a major risk factor for a wide range of cardiovascular diseases and is typically identified by measuring blood pressure (BP) at the brachial artery. Although such a measurement may accurately determine diastolic BP, systolic BP is not reflected accurately. Current noninvasive techniques for assessing central aortic BP require additional recording of an arterial pressure wave using a high-fidelity applanation tonometer. Within one measurement cycle, the Mobil-O-Graph BP device uses brachial oscillometric BP waves for a noninvasive estimation of central BP. We therefore validated the Mobil-O-Graph against the SphygmoCor device, which is widely known as the commonly used approach for a noninvasive estimation of central BP. METHODS: For each individual, we compared three readings of the central BP values obtained by the Mobil-O-Graph and SphygmoCor device consecutively. One hundred individuals (mean age 56.1 ± 15.4 years) were recruited for measurement.Differences between the central BP values of the test device and the SphygmoCor device were calculated for each measurement. RESULTS: The mean difference (95% confidence interval) for the estimated central systolic BP between both devices was -0.6 ± 3.7 mmHg. Comparison of the central BP values measured by the two devices showed a statistically significant linear correlation (R=0.91, P<0.0001). The mean between-method difference was 0.50 mmHg for central systolic BP estimation. The intrarater reproducibility between both the devices was also comparable. Bland and Altman analyses showed that the mean differences (95% confidence interval) between repeated measurements were 1.89 (0.42-3.36) mmHg and 1.36 (-0.16 to 2.83) mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively. Thus, neither of these differences was statistically significantly different from 0. The limits of agreement were -16.34 to 19.73 and -15.23 to 17.17 mmHg for the SphygmoCor and the Mobil-O-Graph device, respectively. CONCLUSION: Oscillometric noninvasive estimation of central BP with the Mobil-O-Graph BP device is as effective as using the well-established SphygmoCor applanation tonometry device. In comparison, the Mobil-O-Graph combines the widespread benefits of brachial BP measurement and also provides central BP within one measurement.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/mbp.0b013e328353ff63

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 12 references

Cited by

See all 208

Metadata

Authors
Weiss, W., Gohlisch, C., Harsch-Gladisch, C., Tölle, M., Zidek, W., & van der Giet, M.
Venue
Blood Pressure Monitoring
Published
2012-06
Volume
17
Issue
3
Pages
128-131
ISSN
1359-5237 (ISSN-L)
OpenAlex
W2331662939
Cited by
208
References
12
FWCI
4.80
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Oscillometric estimation of central blood pressure: validation of the Mobil-O-Graph in comparison with the SphygmoCor device. https://app.cochat.ai/papers/10.1097/mbp.0b013e328353ff63
The paper
Weiss, W., Gohlisch, C., Harsch-Gladisch, C., Tölle, M., Zidek, W., & van der Giet, M. (2012). Oscillometric estimation of central blood pressure: validation of the Mobil-O-Graph in comparison with the SphygmoCor device. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e328353ff63

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.