Evaluation of an automated sphygmomanometer for use in the office setting
The DOI matches its Crossref record.
Abstract
OBJECTIVE: To determine whether the WatchBP Office sphygmomanometer can be used to obtain automated office blood pressure (AOBP) readings that are similar to the awake ambulatory BP. METHODS: One hundred patients referred for 24 h ambulatory BP monitoring had BP recorded three times using the WatchBP Office fully automated sphygmomanometer in accordance with standard AOBP measurement guidelines. The mean AOBP was compared with the mean awake ambulatory BP. RESULTS: The mean (± SD) AOBP (138.6 ± 13.7/79.7 ± 9.0 mmHg) was similar to the mean awake ambulatory BP (136.8 ± 12.4/79.0 ± 10.8 mmHg). The small difference in systolic BP (1.8 mmHg) was statistically significant (P=0.03), but was within the accepted range (5 mmHg) recommended by guidelines for equivalence between BP readings. There was a strong correlation (P<0.001) between the systolic/diastolic AOBP and awake ambulatory BP readings (r=0.819/0.801). CONCLUSION: The WatchBP Office produces BP readings that closely approximate the awake ambulatory BP, confirming that this automated sphygmomanometer is suitable for recording AOBP in clinical practice.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association
Muntner, P., Shimbo, D., Carey, R. M., Charleston, J. B., Gaillard, T., Misra, S., Myers, M. G., Ogedegbe, G., Schwartz, J. E., Townsend, R. R., et al. (2019) · Hypertension · doi:10.1161/hyp.0000000000000087 · Cited by 1,512
Self-citationRoerecke, M., Kaczorowski, J., & Myers, M. G. (2019) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2018.6551 · Cited by 216
Self-citationBlood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research
Muntner, P., Einhorn, P. T., Cushman, W. C., Whelton, P. K., Bello, N. A., Drawz, P. E., Green, B. B., Jones, D. W., Juraschek, S. P., Margolis, K. L., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.069 · Cited by 199
“These automated office blood pressure (AOBP) devices can measure BP with or without an observer present (17).”
BackgroundOscillometric Blood Pressure Estimation: Past, Present, and Future
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolic, M., Rajan, S., & Batkin, I. (2015) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2015.2434215 · Cited by 196
Barriers to accurate blood pressure measurement in the medical office
Hwang, K. O., Aigbe, A., Ju, H.-H., Jackson, V. C., & Sedlock, E. W. (2018) · Journal of Primary Care & Community Health · doi:10.1177/2150132718816929 · Cited by 53
“The average of multiple readings obtained by AOBP is comparable to mean awake ambulatory BP.(20,21) Clinics that use examination tables for BP measurement should be outfitted with chairs that support of the patient’s arms, back, and feet.”
Method
Metadata
- Authors
- Myers, M. G., & Valdivieso, M.
- Venue
- Blood Pressure Monitoring
- Published
- 2012-06
- Volume
- 17
- Issue
- 3
- Pages
- 116-119
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2314188828
- Cited by
- 36
- References
- 0
- FWCI
- 1.53
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of an automated sphygmomanometer for use in the office setting. https://app.cochat.ai/papers/10.1097/mbp.0b013e3283540785
- The paper
- Myers, M. G., & Valdivieso, M. (2012). Evaluation of an automated sphygmomanometer for use in the office setting. Blood Pressure Monitoring. https://doi.org/10.1097/mbp.0b013e3283540785
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.