The ReWalk Powered Exoskeleton to Restore Ambulatory Function to Individuals with Thoracic-Level Motor-Complete Spinal Cord Injury
The DOI matches its Crossref record; 13 of the 14 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The aim of this study was to assess the safety and performance of ReWalk in enabling people with paraplegia due to spinal cord injury to carry out routine ambulatory functions. DESIGN: This was an open, noncomparative, nonrandomized study of the safety and performance of the ReWalk powered exoskeleton. All 12 subjects have completed the active intervention; three remain in long-term follow-up. RESULTS: After training, all subjects were able to independently transfer and walk, without human assistance while using the ReWalk, for at least 50 to 100 m continuously, for a period of at least 5 to 10 mins continuously and with velocities ranging from 0.03 to 0.45 m/sec (mean, 0.25 m/sec). Excluding two subjects with considerably reduced walking abilities, average distances and velocities improved significantly. Some subjects reported improvements in pain, bowel and bladder function, and spasticity during the trial. All subjects had strong positive comments regarding the emotional/psychosocial benefits of the use of ReWalk. CONCLUSIONS: ReWalk holds considerable potential as a safe ambulatory powered orthosis for motor-complete thoracic-level spinal cord injury patients. Most subjects achieved a level of walking proficiency close to that needed for limited community ambulation. A high degree of performance variability was observed across individuals. Some of this variability was explained by level of injury, but other factors have not been completely identified. Further development and application of this rehabilitation tool to other diagnoses are expected in the future.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“With a built-in multisensor system, an exoskeleton can acquire wearer's motion intensions and accordingly assist the wearer's motion.”
From Recent developments and challenges of lower extremity exoskeletons (2016) · cites this paper as [15]
In the citing paper
“These results also indicate that the patients experienced a reduction in secondary complications resulting from life in a wheelchair such as depression and neuropathic pain.”
From Recent developments and challenges of lower extremity exoskeletons (2016) · cites this paper as [15]
In the citing paper
“Assistance from exoskeletons enable these patients to regain the ability to stand up, sit down, and walk, just as an able-bodied person [14], [15].”
From Recent developments and challenges of lower extremity exoskeletons (2016) · cites this paper as [15]
In the citing paper
“Based on clinical study results of the ReWalk exoskeleton [15], [34], paralyzed patients can practically stand upright and walk with increased independence and their life quality is greatly improved.”
From Recent developments and challenges of lower extremity exoskeletons (2016) · cites this paper as [15]
Citations
Cites
Schulz, R., & Decker, S. (1985) · Journal of Personality and Social Psychology · doi:10.1037//0022-3514.48.5.1162 · Cited by 541
Exercise Recommendations for Individuals with Spinal Cord Injury
Jacobs, P. L., & Nash, M. S. (2004) · Sports Medicine · doi:10.2165/00007256-200434110-00003 · Cited by 453 · cited 8×
“Exercise has been shown to be an effective contributor to overall health, bone density, a proper level of muscle tone, cardiovascular fitness, regular bowel and bladder function, reduced risk for obesity, heart disease, and reduction in type II diabetes for patients with SCI.(28,29) Therapeutic exercise for individual…”
BackgroundInfluentialCorrelates of life satisfaction among persons with spinal cord injury
Dijkers, M. P. J. M. (1999) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(99)90076-x · Cited by 289
Physical Activity Levels Are Low in Free‐Living Adults with Chronic Paraplegia
Buchholz, A. C., McGillivray, C. F., & Pencharz, P. B. (2003) · Obesity Research · doi:10.1038/oby.2003.79 · Cited by 220
Shoulder pain in persons with thoracic spinal cord injury: Prevalence and characteristics
Alm, M., Saraste, H., & Norrbrink, C. (2008) · Journal of Rehabilitation Medicine · doi:10.2340/16501977-0173 · Cited by 179 · cited 1×
“Upper body exercises including hand ergometry or weight lifting may increase the likelihood of overuse of already taxed upper limb joints, particularly for thoraciclevel injured individuals who rely on wheelchair propulsion for locomotion and their arms for transfers.(16,38,39) Having an SCI results in a 40% reduction…”
BackgroundPain and life quality within 2 years of spinal cord injury
Anke, A., Stenehjem, A. E., & Stanghelle, J. K. (1995) · Spinal Cord · doi:10.1038/sc.1995.120 · Cited by 138
Levins, S. M., Redenbach, D., & Dyck, I. (2004) · Physical Therapy · doi:10.1093/ptj/84.6.496 · Cited by 133 · cited 1×
“training during acute rehabilitation, decreased access to exercise facilities and adequate adaptive fitness equipment, limited time, and psychologic factors.(31) The physical deconditioning resulting from largely sedentary lives of individuals with thoracic or higher level SCI is well documented.”
BackgroundThrasher, T. A., Graham, G. M., & Popović, M. R. (2005) · Artificial Organs · doi:10.1111/j.1525-1594.2005.29076.x · Cited by 106
Wheelchair accessibility of public buildings: a review of the literature
Welage, N., & Liu, K. P. Y. (2010) · Disability and Rehabilitation Assistive Technology · doi:10.3109/17483107.2010.522680 · Cited by 63 · cited 1×
“However, because of significant architectural and environmental restrictions, there are limitations associated with wheelchair use.(11) Furthermore, the typical seated wheelchair posture does not load the legs in a normal manner(12) and can lead to pressure sores(13,14) and promote joint contractures.”
BackgroundNeuromusculoskeletal Syndromes in Wheelchair Athletes
Groah, S. L., & Lanig, I. S. (2000) · Seminars in Neurology · doi:10.1055/s-2000-9829 · Cited by 32
Cited by
Items 101–200 of 957
Recognition of walking environments and gait period by surface electromyography
Kyeong, S., Shin, W., Yang, M., Heo, U., Feng, J.-R., & Kim, J. (2019) · Frontiers of Information Technology & Electronic Engineering · doi:10.1631/fitee.1800601 · Cited by 68
Feasibility of integrating robotic exoskeleton gait training in inpatient rehabilitation
Swank, C., Sikka, S., Driver, S. J., Bennett, M., & Callender, L. (2019) · Disability and Rehabilitation Assistive Technology · doi:10.1080/17483107.2019.1587014 · Cited by 79
“Studies have indicated gait training with exoskeletons to be safe, well tolerated, with no significant complications (n1⁄4 6 & 8) [8,9] over distances of 40–100 m (n1⁄4 12) [10] for community dwelling people with chronic spinal cord injuries (SCI).”
BackgroundSequential time-dependent reliability analysis for the lower extremity exoskeleton under uncertainty
Yu, S., Wang, Z., & Zhang, K. (2017) · Reliability Engineering & System Safety · doi:10.1016/j.ress.2017.10.006 · Cited by 70
“Furthermore, several representatives such as BLEEX (Berkeley Lower Extremity Exoskeleton) [1, 2], HAL (Hybrid Assistive Leg) [3, 4], ReWalk [5], Rex [6], NTULEE (Nanyang Technology University Lower Extremity Exoskeleton) [7], Lokomat [8], LOPES (Lower-extremity Powered Exoskeleton) [9], EXPOS (Exoskeleton for patients…”
BackgroundModular Design and Decentralized Control of the Recupera Exoskeleton for Stroke Rehabilitation
Kumar, S., Wöhrle, H., Trampler, M., Simnofske, M., Peters, H., Mallwitz, M., Kirchner, E. A., & Kirchner, F. (2019) · Applied Sciences · doi:10.3390/app9040626 · Cited by 78
Kwon, S. H., Lee, B.-S., Lee, H.-J., Kim, E. J., Lee, J. A., Yang, S. P., Kim, T. Y., Pak, H. R., Kim, H. K., Kim, H. Y., et al. (2020) · Annals of Rehabilitation Medicine · doi:10.5535/arm.2020.44.2.131 · Cited by 71
“Gait-using exoskeleton gaitassistive robots that enable over-ground walking can be used in an actual gait environment, such as outdoors, indoors, and in an environment with obstacles, thereby motivating users to walk [10,11].”
BackgroundMethodInfluentialGagnon, D. H., Escalona, M. J., Vermette, M., Carvalho, L. P., Karelis, A. D., Duclos, C., & Aubertin‐Leheudre, M. (2018) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0354-2 · Cited by 74
Experience of Robotic Exoskeleton Use at Four Spinal Cord Injury Model Systems Centers
Heinemann, A. W., Jayaraman, A., Mummidisetty, C. K., Spraggins, J., Pinto, D., Charlifue, S. W., Tefertiller, C., Taylor, H. B., Chang, S., Stampas, A., et al. (2018) · Neurology Report · doi:10.1097/npt.0000000000000235 · Cited by 70
“Focus Group Participant Demographic Characteristics Demographic Characteristic Craig Hospital (n = 5) Shepherd Center (n = 10) Shirley Ryan AbilityLab (n = 4) TIRR Memorial Hermann (n = 11) Total (n = 30) Age (median years, range) 42 (33-48) 40 (28-53) 34 (30-40) 33 (29-46) 37 (28-53) Clinical experience (median years…”
Wu, X., Yuan, Y., Zhang, X., Wang, C., Xu, T., & Tao, D. (2022) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2021.3082067 · Cited by 67
“Though gait training with exoskeleton has been demonstrated to be safe and well-tolerated over distances of 40–100 m [9] with no significant complications [10], the accuracy of gait phase classification still has a big impact on the performance and safety of lower limb”
Robotics for Lower Limb Rehabilitation
Esquenazi, A., & Talaty, M. C. (2019) · Physical Medicine and Rehabilitation Clinics of North America · doi:10.1016/j.pmr.2018.12.012 · Cited by 83
Self-citationLi, Z., Li, Q., Huang, P., Xia, H., & Li, G. (2023) · IEEE Transactions on Cybernetics · doi:10.1109/tcyb.2023.3240231 · Cited by 67
Schrade, S. O., Dätwyler, K., Stücheli, M., Studer, K., Türk, D.-A., Meboldt, M., Gassert, R., & Lambercy, O. (2018) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0360-4 · Cited by 70
“Several studies reported that the regular use of an exoskeleton could have a positive impact on chronic neuropathic pain, emotional and psychological constitution [11], bowel and bladder function [11–13], and spasticity [9, 12–15].”
BackgroundSynergy-Based Control of Assistive Lower-Limb Exoskeletons by Skill Transfer
Wei, Q., Li, Z., Zhao, K., Kang, Y., & Su, C. (2019) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2019.2961567 · Cited by 68
Juszczak, M., Gallo, E., & Bushnik, T. L. (2018) · Topics in Spinal Cord Injury Rehabilitation · doi:10.1310/sci17-00055 · Cited by 65
“Consistent with Esquenazi et al’s findings of bowel/bladder function after exoskeleton use, eight of nine of those who reported changes in bowel function classified the change as positive, with fewer instances of neurogenic bowel dysfunction as well as decreased time and assistance required for bowel management.(12) O…”
BackgroundResultPerspectives of people with spinal cord injury learning to walk using a powered exoskeleton
Manns, P. J., Hurd, C., & Yang, J. F. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-019-0565-1 · Cited by 68
“The psychological benefits that others have alluded to anecdotally [4] were discussed at length by our participants.”
BackgroundResultInfluentialPlatz, T., Gillner, A., Nicole, B., Kroll, S., & Roschka, S. (2016) · BioMed Research International · doi:10.1155/2016/8459018 · Cited by 68
“Results of a device-training for nonambulatory individuals with thoracic and lumbar spinal cord injury (SCI) using a powered exoskeleton for technically assisted mobility with regard to the achieved level of control of the system after training, user satisfaction, and effects on quality of life…”
BackgroundMethodAdaptive Ankle Exoskeleton Control: Validation Across Diverse Walking Conditions
Bishe, S. S. P. A., Nguyen, T. T., Fang, Y., & Lerner, Z. F. (2021) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2021.3091519 · Cited by 62
Active Human-Following Control of an Exoskeleton Robot With Body Weight Support
Li, G., Li, Z., Su, C., & Tian, X. (2023) · IEEE Transactions on Cybernetics · doi:10.1109/tcyb.2023.3253181 · Cited by 64
“In [11], [12], and [13], the effects of Lokomat, WAlKBOT, and ReWalk system were evaluated on stand, sitting, balance, and gait training by recording quantitative measures of locomotor improvements, which shows more effective performance of robot-assistive gait training than routine therapy.”
MethodGarate, V. R., Parri, A., Yan, T., Munih, M., Lova, R. M., Vitiello, N., & Ronsse, R. (2017) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2017.00015 · Cited by 67
“One possibility would be to trigger a change of locomotion intention with some dynamic movement from the torso or head, or with a similar system as the wrist-pad used in ReWalk (Esquenazi et al., 2012; Murtagh, 2015), or to augment the intention detection algorithm with inputs provided by kinetic signals.”
BackgroundMethodInfluentialEffects of a powered ankle-foot orthosis on perturbed standing balance
Emmens, A. R., van Asseldonk, E. H. F., & van der Kooij, H. (2018) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0393-8 · Cited by 65
“Except for the wearable exoskeleton “REX” (REX Bionics Ltd, New Zealand) [16], commercially available devices are used in combination with crutches for maintaining balance [3, 7, 17].”
Design and Voluntary Control of Variable Stiffness Exoskeleton Based on sEMG Driven Model
Zhu, Y., Wu, Q., Chen, B., & Zhao, Z. (2022) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2022.3160668 · Cited by 61
Wearable Power-Assist Locomotor (WPAL) for supporting upright walking in persons with paraplegia
Tanabe, S., Hirano, S., & Saitoh, E. (2013) · Neurorehabilitation · doi:10.3233/nre-130932 · Cited by 61
Gil-Agudo, A. M., Megía‐García, Á., Pons, J. L., Sinovas-Alonso, I., Comino‐Suárez, N., Lozano-Berrio, V., & del‐Ama, A. J. (2023) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-023-01158-z · Cited by 79
“In terms of the feasibility and adverse events, our results were similar to other studies [19, 27, 46, 47].”
ResultHuman Lower Limb Motion Intention Recognition for Exoskeletons: A Review
Li, L., Cao, G., Liang, H., Zhang, Y.-P., & Cui, F. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3328615 · Cited by 71
“Various representative LLEs have been launched in the market, such as Bleex [15], Rewalk [19], HAL [20], and Lokomat [21].”
MethodA Kinematic Model of a Humanoid Lower Limb Exoskeleton with Hydraulic Actuators
Głowiński, S., Krzyżyński, T., Bryndal, A., & Maciejewski, I. (2020) · Sensors · doi:10.3390/s20216116 · Cited by 59
“Typical examples of lower limb exoskeletons are ReWalk (Marlborough, MA, USA) and HAL (The Hybrid Assistive Limb, Gauken-Minami, Tsukuba, Ibaraki Prefecture, Japan) [10].”
BackgroundBayón, C., Keemink, A. Q. L., van Mierlo, M., Rampeltshammer, W. F., van der Kooij, H., & van Asseldonk, E. H. F. (2022) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-022-01000-y · Cited by 62
Wu, C.-H., Mao, H., Hu, J., Wang, T.-Y., Tsai, Y.-J., & Hsu, W. (2018) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0355-1 · Cited by 62
“ReWalk and EKSO are specially designed for SCI.”
Interactive Control of Lower Limb Exoskeleton Robots: A Review
Zhang, Y.-P., Cao, G.-Z., Li, L.-L., & Diao, D.-F. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3352005 · Cited by 58
“ReWalk, developed in Israel, is the first rehabilitation exoskeleton robot approved by the U.S. Food and Drug Administration (FDA); it can assist disabled individuals in walking normally and support rehabilitation training for patients with lower limb dysfunction [23], [24].”
BackgroundJohns, J. S., Krogh, K., Rodriguez, G. M., Eng, J. J., Haller, E., Heinen, M., Laredo, R., Longo, W. E., Montero-Colon, W., Wilson, C. S., et al. (2021) · Topics in Spinal Cord Injury Rehabilitation · doi:10.46292/sci2702-75 · Cited by 69
Human motion intent learning based motion assistance control for a wearable exoskeleton
Long, Y., Du, Z., Wang, W., & Dong, W. (2017) · Robotics and Computer-Integrated Manufacturing · doi:10.1016/j.rcim.2017.08.007 · Cited by 57
McIntosh, K., Charbonneau, R., Bensaada, Y., Bhatiya, U., & Ho, C. (2019) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2019.09.005 · Cited by 61
Tamburella, F., Lorusso, M., Tramontano, M., Fadlun, S., Masciullo, M., & Scivoletto, G. (2022) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-022-01003-9 · Cited by 68
“Spasticity studies employed quite different outcome measures: Modified Ashworth Scale (MAS) [85] was employed in 3 studies (Ekso N = 2 [61, 62], Indego N = 1 [46]), 2 studies reported Ashworth Scale (AS) (ReWalk [67, 69]), or Numeric Rating Scale for spasticity (NRS_sp) (N = 1 Ekso [62]; N = 1 Indego [46]), and in one…”
BackgroundMethodResultInfluentialGait Trajectory and Event Prediction from State Estimation for Exoskeletons During Gait
Tanghe, K., Groote, F. D., Lefeber, D., Schutter, J. D., & Aertbeliën, E. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2950309 · Cited by 64
“Other exoskeletons allow the subject to move freely, and can be used for rehabilitation or permanent assistance [3]–[5].”
BackgroundGagnon, D. H., Vermette, M., Duclos, C., Aubertin‐Leheudre, M., Ahmed, S., & Kairy, D. (2017) · Disability and Rehabilitation Assistive Technology · doi:10.1080/17483107.2017.1413145 · Cited by 57
“It also aligns with another study, which found that three of 11 participants reported improved lower extremity spasticity, whereas five of 11 participants reported improved bowel movement upon completion of a locomotor training program with a robotic exoskeleton [25].”
ResultAlibeji, N. A., Molazadeh, V., Moore-Clingenpeel, F., & Sharma, N. (2018) · IEEE Control Systems · doi:10.1109/mcs.2018.2866603 · Cited by 60
Human Locomotion in Hypogravity: From Basic Research to Clinical Applications
Lacquaniti, F., Ivanenko, Y. P., Labini, F. S., Scaleia, V. L., Scaleia, B. L., Willems, P. A., & Zago, M. (2017) · Frontiers in Physiology · doi:10.3389/fphys.2017.00893 · Cited by 52
“Subsequently, exoskeletons have been used in rehabilitation, for instance the Indego, EKSO Bionics, ReWalk, Mindwalker (Figure 3E) (del-Ama et al., 2012; Esquenazi et al., 2012; Wang et al., 2015).”
MethodDesign and Control of a Polycentric Knee Exoskeleton Using an Electro-Hydraulic Actuator
Lee, T., Lee, D., Song, B., & Baek, Y. S. (2019) · Sensors · doi:10.3390/s20010211 · Cited by 52
Tamburella, F., Tagliamonte, N. L., Pisotta, I., Masciullo, M., Arquilla, M., van Asseldonk, E. H. F., van der Kooij, H., Wu, A. R., Dzeladini, F., Ijspeert, A. J., et al. (2020) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2020.2984790 · Cited by 50
“For people with iSCI, the focus is generally on rehabilitation [7], [8] and gait restoration [9], [10].”
BackgroundStepping Forward with Exoskeletons: Team IHMC?s Design and Approach in the 2016 Cybathlon
Griffin, R., Cobb, T., Craig, T., Daniel, M., van Dijk, N., Gines, J., Kramer, K., Shah, S., Siebinga, O., Smith, J., et al. (2017) · IEEE Robotics & Automation Magazine · doi:10.1109/mra.2017.2754284 · Cited by 59
“To address this, commercial exoskeletons such as ReWalk [5], Ekso [6], and Indego [7] have made significant strides forward since the first prototypes.”
BackgroundClinician-Focused Overview of Bionic Exoskeleton Use After Spinal Cord Injury
Palermo, A. E., MAHER, J. L., Baunsgaard, C. B., & Nash, M. S. (2017) · Topics in Spinal Cord Injury Rehabilitation · doi:10.1310/sci2303-234 · Cited by 54
User-Centered Design and Development of the Modular TWIN Lower Limb Exoskeleton
Laffranchi, M., D'Angella, S., Vassallo, C., Piezzo, C., Canepa, M., Giuseppe, S. D., Salvo, M. D., Succi, A., Cappa, S., Cerruti, G., et al. (2021) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2021.709731 · Cited by 54
Nazari, F., Mohajer, N., Nahavandi, D., Khosravi, A., & Nahavandi, S. (2023) · IEEE Transactions on Cognitive and Developmental Systems · doi:10.1109/tcds.2023.3241632 · Cited by 62
“ATLAS [104], HAL [105], and ReWalk [106] are examples of this implementation.”
BackgroundOptimization-Based Motion Planning in Joint Space for Walking Assistance With Wearable Robot
Kagawa, T., Ishikawa, H., Kato, T., Sung, C., & Uno, Y. (2015) · IEEE Transactions on Robotics · doi:10.1109/tro.2015.2409434 · Cited by 54
“sults from the user’s voluntary motion [11], [12].”
BackgroundMethoddel‐Ama, A. J., Gil-Agudo, A. M., Pons, J. L., & Moreno, J. C. (2014) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2014.00298 · Cited by 50
Delisle-Rodríguez, D., Cardoso, V. F., Gurve, D., Loterio, F. A., Romero-Laiseca, M. A., Krishnan, S., & Bastos-Filho, T. (2019) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab08c8 · Cited by 52
“1741-2552/19/056005+16$33.00 D Delisle-Rodriguez et al systems for gait rehabilitation, such as the hybrid assistive limb (HAL) [7–10], ReWalk [11], Lokomat [12, 13], and Lopes [14].”
BackgroundInfluentialXue, X., Yang, X., Deng, Z., Tu, H., Kong, D., Li, N., & Xu, F. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2021.806723 · Cited by 55
Research of intent recognition in rehabilitation robots: a systematic review
Luo, S., Meng, Q., Li, S., & Yu, H. (2023) · Disability and Rehabilitation Assistive Technology · doi:10.1080/17483107.2023.2170477 · Cited by 64
“ReWalk [48,49] is a lower limb exoskeleton with the motor as the power system.”
BackgroundExoskeletons for Personal Use After Spinal Cord Injury
Kandilakis, C., & Sasso-Lance, E. (2019) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2019.05.028 · Cited by 52
Zahradka, N., Verma, K., Behboodi, A., Bodt, B., Wright, H., & Lee, S. C. K. (2020) · Sensors · doi:10.3390/s20185272 · Cited by 58
A Muscle Synergy-Inspired Adaptive Control Scheme for a Hybrid Walking Neuroprosthesis
Alibeji, N. A., Kirsch, N. A., & Sharma, N. (2015) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2015.00203 · Cited by 50
“Functional electrical stimulation (FES) and powered orthoses are two viable technologies that have the potential to restore the walking function in persons with SCI (Kralj and Bajd, 1989; Kobetic et al., 1997; Farris et al., 2011; Neuhaus et al., 2011; Esquenazi et al., 2012; del-Ama et al., 2014a; Ha et al., 2015).”
BackgroundMechatronic Wearable Exoskeletons for Bionic Bipedal Standing and Walking: A New Synthetic Approach
Onose, G., Cârdei, V., Crăciunoiu, Ş. T., Avramescu, V., Opriș, I. C., Lebedev, M., & Constantinescu, M. V. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00343 · Cited by 47
“…wearable exoskeletons can be used by patients with post-stroke hemiparesis (Bortole et al., 2015), as well as nonstroke neurological pathologies24 (Esquenazi et al., 2012), such as cerebral palsy, myelome-ningocele, traumatic brain injury and Guillain Barré syndrome (Zeilig et al., 2012).”
BackgroundDesign and evaluation of a modular lower limb exoskeleton for rehabilitation
dos Santos, W. M., Nogueira, S. L., de Oliveira, G. C., Peña, G. G., & Siqueira, A. A. G. (2017) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2017.8009288 · Cited by 51
“com provides powered hip and knee motion to assist individuals with SCI in standing upright and walking, as well as ascending and descending stairs [3].”
BackgroundWalkON Suit: A Medalist in the Powered Exoskeleton Race of Cybathlon 2016
Choi, J., Na, B., Jung, P.-G., Rha, D., & Kong, K. (2017) · IEEE Robotics & Automation Magazine · doi:10.1109/mra.2017.2752285 · Cited by 47
A Semi-active Exoskeleton Based on EMGs Reduces Muscle Fatigue When Squatting
Wang, Z., Wu, X., Zhang, Y., Chen, C., Liu, S., Liu, Y., Peng, A., & Ma, Y. (2021) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2021.625479 · Cited by 47
“15:625479. doi: 10.3389/fnbot.2021.625479 A Semi-active Exoskeleton Based on EMGs Reduces Muscle Fatigue When Squatting Zhuo Wang 1,2,3†, Xinyu Wu 1,2,4†, Yu Zhang 1,2,3, Chunjie Chen 1,2,4*, Shoubin Liu 3, Yida Liu 1,2,4, Ansi Peng 1,2,4 and Yue Ma 1,2,4 1Chinese Academy of Sciences Key Laboratory of Human-Machine-In…”
BackgroundInfluentialMatsuda, M., Mataki, Y., Mutsuzaki, H., Yoshikawa, K., Takahashi, K., Enomoto, K., Sano, K., Mizukami, M., Tomita, K., Ohguro, H., et al. (2018) · Journal of Physical Therapy Science · doi:10.1589/jpts.30.207 · Cited by 47
Adaptive Oscillator-Based Control for Active Lower-Limb Exoskeleton and its Metabolic Impact
Seo, K., Kim, K., Park, Y. J., Cho, J.-K., Lee, J.-W., Choi, B., Lim, B., Lee, Y., & Shim, Y. (2018) · Proceedings - IEEE International Conference on Robotics and Automation/Proceedings · doi:10.1109/icra.2018.8460841 · Cited by 46
“Some exoskeletons like ATLAS [4], Ekso [5], ReWalk [6] were developed for paraplegic or quadriplegic people and cover full lower limb for assistance.”
MethodLiu, X., Zhou, Z., Mai, J., & Wang, Q. (2019) · Robotics and Autonomous Systems · doi:10.1016/j.robot.2019.06.008 · Cited by 51
Xiang, X.-N., Ding, M., Zong, H.-Y., Liu, Y., Cheng, H., He, C., & He, H. (2020) · Spinal Cord · doi:10.1038/s41393-020-0423-9 · Cited by 50
“[25] measured walking speeds that ranged from 0.”
Bilateral, Misalignment-Compensating, Full-DOF Hip Exoskeleton: Design and Kinematic Validation
Junius, K., Degelaen, M., Lefeber, N., Swinnen, E., Vanderborght, B., & Lefeber, D. (2017) · Applied Bionics and Biomechanics · doi:10.1155/2017/5813154 · Cited by 47
Narayan, J., & Dwivedy, S. K. (2021) · Applied Bionics and Biomechanics · doi:10.1155/2021/5573041 · Cited by 43
“The researchers, in the literature, have regarded the predefined gait tracking control as the basis of every control scheme for exoskeleton systems, where the joint movements of the lower limb could be estimated using gait analysis experiments [15, 16, 17].”
BackgroundRetractedConcept of a soft exosuit for the support of leg function in rehabilitation
Bartenbach, V., Schmidt, K., Naef, M., Wyss, D., & Riener, R. (2015) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2015.7281187 · Cited by 56
“Examples for such systems are ekso [1], ReWalk [2] and REX [3].”
BackgroundStability of Mina v2 for Robot-Assisted Balance and Locomotion
Mummolo, C., Peng, W. Z., Agarwal, S., Griffin, R. J., Neuhaus, P. D., & Kim, J. H. (2018) · Frontiers in Neurorobotics · doi:10.3389/fnbot.2018.00062 · Cited by 48
“In several cases, the operator relies on crutches or walkers in order to maintain balance (Acosta-Marquez and Bradley, 2005; Strausser and Kazerooni, 2011; Esquenazi et al., 2012; Farris et al., 2014; Stücheli et al., 2017).”
BackgroundDriving Robotic Exoskeletons Using Cable-Based Transmissions: A Qualitative Analysis and Overview
Grosu, S., Rijcke, L. D., Grosu, V., Geeroms, J., Vanderboght, B., Lefeber, D., & Rodriguez‐Guerrero, C. (2018) · Applied Mechanics Reviews · doi:10.1115/1.4042399 · Cited by 45
Bodenhagen, L., Suvei, S.-D., Juel, W. K., Brander, E., & Krüger, N. (2019) · Health and Technology · doi:10.1007/s12553-019-00302-x · Cited by 55
“[6] assess the effect of ReWalk [120], a wearable robotic exoskeleton that was tested on 12 subjects with paraplegia due to spinal cord injury.”
BackgroundGuanziroli, E., Cazzaniga, M., Colombo, L., Basilico, S., Legnani, G., & Molteni, F. (2019) · European Journal of Physical and Rehabilitation Medicine · doi:10.23736/s1973-9087.18.05308-x · Cited by 50
User-driven walking assistance: first experimental results using the MyoSuit
Haufe, F. L., Kober, A. M., Schmidt, K., Sancho-Puchades, A., Duarte, J. E., Wolf, P., & Riener, R. (2019) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2019.8779375 · Cited by 49
“A variety of wearable lower-limb robots have been developed as training tools for rehabilitation following stroke [1, 2], spinal cord injury [3, 4], or as assistive devices for users with spinal cord injury [5-9].”
BackgroundTan, X., Zhang, B., Liu, G., Zhao, X., & Zhao, Y. (2021) · IEEE Transactions on Automation Science and Engineering · doi:10.1109/tase.2021.3066403 · Cited by 42
“Lower extremity exoskeletons (LEEs), a type of robotic device designed for lower extremity augmentation, can apply torques to biological joints to assist healthy individuals [2]–[4] or patients with motor dysfunction [5]–[7].”
BackgroundA muscle-driven approach to restore stepping with an exoskeleton for individuals with paraplegia
Chang, S. R., Nandor, M. J., Li, L., Kobetic, R., Foglyano, K. M., Schnellenberger, J. R., Audu, M. L., Pinault, G., Quinn, R. D., & Triolo, R. J. (2017) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-017-0258-6 · Cited by 49
“The IMU subsystem could be used as an alternative signal or in combination with the other sensor signals for step initiation or balance control, similar to commercial exoskeleton systems [14].”
Bio-Inspired Balance Control Assistance Can Reduce Metabolic Energy Consumption in Human Walking
Zhao, G., Sharbafi, M. A., Vlutters, M., van Asseldonk, E. H. F., & Seyfarth, A. (2019) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2019.2929544 · Cited by 40
“[16], ReWalk [17], and eLEGS [18] for assisting pathological gaits.”
BackgroundFont-Llagunes, J. M., Lugrís, U., Costa, D. C., Sánchez, F. J. A., & Cuadrado, J. (2019) · Journal of Mechanisms and Robotics · doi:10.1115/1.4045510 · Cited by 40
Overground Robot-Assisted Gait Training for Pediatric Cerebral Palsy
Kim, S. K., Park, D., Yoo, B., Shim, D., Choi, J.-O., Choi, T. Y., & Park, E. S. (2021) · Sensors · doi:10.3390/s21062087 · Cited by 47
Qian, Y., Yu, H., & Fu, C. (2022) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2022.3213810 · Cited by 43
“For instance, exoskeletons, such as the ReWalk [2] and eLEGS [3], have achieved noticeable success in providing walking assistance to patients with spinal cord injure, in which the exoskeletons guide the patients to walk following a normative gait trajectory with litter or no active participation from the patients.”
BackgroundSun, J., Wang, J., Yang, P., & Guo, S. (2022) · IEEE Transactions on Industrial Informatics · doi:10.1109/tii.2022.3215197 · Cited by 39
“The ReWalk powered exoskeleton is designed to help spinal cord injury patients to restore ambulatory function [2].”
Background$\rm PI^{{\text{2}}}$-Based Adaptive Impedance Control for Gait Adaption of Lower Limb Exoskeleton
Wang, X., Zhang, R., Miao, Y., An, M., Wang, S., & Zhang, Y. (2024) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2024.3370954 · Cited by 39
“Thus, there has been noteworthy growth in the research works on lower limb exoskeletons, such as ReWalk [4], EKSO [5], and HAL [6].”
Mechatronic design of a sit-to-stance exoskeleton
Junius, K., Brackx, B., Grosu, V., Cuypers, H., Geeroms, J., Moltedo, M., Vanderborght, B., & Lefeber, D. (2014) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2014.6913902 · Cited by 41
“The ReWalk [9], Ekso [10] and HAL [11] exoskeletons are all full lower limb devices.”
BackgroundKim, J., Gu, G. M., & Heo, P. (2015) · Biosystems & biorobotics · doi:10.1007/978-3-319-21813-7_21 · Cited by 43
Exoskeletal Assisted Rehabilitation After Spinal Cord Injury
Gorgey, A. S., Sumrell, R. M., & Goetz, L. L. (2019) · Elsevier eBooks · doi:10.1016/b978-0-323-48323-0.00044-5 · Cited by 40
Walking With a Robotic Exoskeleton Does Not Mimic Natural Gait: A Within-Subjects Study
Swank, C., Wang-Price, S. S., Gao, F., & Almutairi, S. M. (2018) · JMIR Rehabilitation and Assistive Technologies · doi:10.2196/11023 · Cited by 45
“Although gait training with exoskeletons has been shown to be safe and well tolerated, with no significant complications [14] over distances of 40-100 m [15], it is unclear how closely the gait of a person wearing a robotic exoskeleton approximates normal gait.”
BackgroundChun, A. K., Asselin, P. K., Knezevic, S., Kornfeld, S. D., Bauman, W. A., Korsten, M. A., Harel, N. Y., Huang, V. S., & Spungen, A. M. (2019) · Spinal Cord · doi:10.1038/s41393-019-0392-z · Cited by 41
del‐Ama, A. J., Gil-Agudo, A. M., Bravo-Esteban, E., Pérez‐Nombela, S., Pons, J. L., & Moreno, J. C. (2014) · Robotics and Autonomous Systems · doi:10.1016/j.robot.2014.10.014 · Cited by 39
Towards Variable Assistance for Lower Body Exoskeletons
Gurriet, T., Tucker, M., Duburcq, A., Boeris, G., & Ames, A. D. (2019) · IEEE Robotics and Automation Letters · doi:10.1109/lra.2019.2955946 · Cited by 39
Exoskeleton-assisted Gait Training in Persons With Multiple Sclerosis: A Single-Group Pilot Study
Afzal, T., Tseng, S.-C., Lincoln, J. A., Kern, M., Francisco, G. E., & Chang, S. (2019) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/j.apmr.2019.10.192 · Cited by 40
A biarticular passive exosuit to support balance control can reduce metabolic cost of walking
Barazesh, H., & Sharbafi, M. A. (2020) · Bioinspiration & Biomimetics · doi:10.1088/1748-3190/ab70ed · Cited by 42
“Lower limb exoskeletons are mainly used for three types of applications [5]: (i) rehabilitation for patients with mobility disabilities [6–8], (ii) supporting patients with lost mobility in the lower extremity [9, 10], and (iii) improving the abilities of able-bodied humans by reducing energy consumption and increasin…”
BackgroundFramework for Assessment of the Usability of Lower-Extremity Robotic Exoskeletal Orthoses
Bryce, T. N., Dijkers, M. P. J. M., & Kozlowski, A. J. (2015) · American Journal of Physical Medicine & Rehabilitation · doi:10.1097/phm.0000000000000321 · Cited by 40
Kim, H., Lee, J.-W., Jang, J., Park, S., & Han, C.-S. (2015) · International Journal of Control Automation and Systems · doi:10.1007/s12555-013-0386-0 · Cited by 37
Duddy, D., Doherty, R., Connolly, J., McNally, S. J., Loughrey, J., & Faulkner, M. (2021) · Sensors · doi:10.3390/s21093207 · Cited by 40
“Powered exoskeleton devices have been shown to be safe to use with patients who suffer from neurological impairments, with no falls, no serious adverse events [52], no fractures or major skin effects [59] or incidences of autonomic dysreflexia [60,61] reported by previous studies.”
BackgroundGait Deviation Correction Method for Gait Rehabilitation With a Lower Limb Exoskeleton Robot
Zhang, S., Guan, X., Ye, J. B., Chen, G., Zhang, Z., & Leng, Y. (2022) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2022.3194360 · Cited by 40
Knudsen, G.-K., Jørgensen, V., & Normann, B. (2019) · Spinal Cord Series and Cases · doi:10.1038/s41394-019-0243-3 · Cited by 35
“[17] noted that all subjects in a study of the safety and performance of the ReWalk powered exoskeleton had strongly positive comments regarding the emotional and psychosocial benefits of the use.”
BackgroundGao, S., Wang, Y., Fang, C., & Xu, L. (2020) · Applied Sciences · doi:10.3390/app10082638 · Cited by 41
Hussain, S., & Ficuciello, F. (2024) · IEEE Transactions on Medical Robotics and Bionics · doi:10.1109/tmrb.2024.3407374 · Cited by 43
“Other assistive robots like HAL [8], Vanderbilt [9], REX [10], HES Hand [11], ReWalk [12], HandeXos-Beta [13], HandSOME [14], and HexoSYS [15] utilize motors and tendons to aid activities of daily life (ADLs) such as climbing stair, balancing, grasping, pouring liquids, and buttoning.”
BackgroundStructure design of active power-assist lower limb exoskeleton APAL robot
Deng, J., Wang, P., Li, M., Guo, W., Zha, F., & Wang, X. (2017) · Advances in Mechanical Engineering · doi:10.1177/1687814017735791 · Cited by 35
Active Compliance Control Reduces Upper Body Effort in Exoskeleton-Supported Walking
Uğurlu, B., Oshima, H., Sarıyıldız, E., Narikiyo, T., & Babič, J. (2020) · IEEE Transactions on Human-Machine Systems · doi:10.1109/thms.2019.2961969 · Cited by 35
“A single walking experiment is as follows: 1) the participant receives walking support when the exoskeleton is position controlled via the conventional method [6] (1 min); 2) a rest period of 15 min; and 3) the participant receives walking support when”
MethodSlope Gradient Adaptive Gait Planning for Walking Assistance Lower Limb Exoskeletons
Zou, C., Huang, R., Qiu, J., Chen, Q., & Cheng, H. (2020) · IEEE Transactions on Automation Science and Engineering · doi:10.1109/tase.2020.3037973 · Cited by 35
“Digital Object Identifier 10.1109/TASE.2020.3037973 Most of these wearable robots are designed with a two-legged structure to help paraplegic patients walk overground, such as ReWalk [1], EKSO [2], Indego [3], and HAL [4].”
BackgroundInfluentialFernández, A. R., Lobo-Prat, J., Tarragó, R., Chaverri, D., Iglesias, X., Guirao-Cano, L., & Font-Llagunes, J. M. (2022) · Scientific Reports · doi:10.1038/s41598-022-23556-4 · Cited by 42
Template model inspired leg force feedback based control can assist human walking
Zhao, G., Sharbafi, M. A., Vlutters, M., van Asseldonk, E. H. F., & Andre Seyfarth, A. (2017) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2017.8009293 · Cited by 34
Shackleton, C., Evans, R., Shamley, D. R., West, S. J., & Albertus, Y. (2019) · Journal of Rehabilitation Medicine · doi:10.2340/16501977-2601 · Cited by 40
“(2012) (18) T0–T1 = spasticity reduced* T0–T1 = 3/11 reported reduced spasticity; none reported increased spasticity* Kressler et al.”
BackgroundMethodInfluentialNistor-Cseppento, C. D., Gherle, A., Negruţ, N., BUNGAU, S. G., SABĂU, A. M., Radu, A.-F., Bungău, A. F., Ţiţ, D. M., Uivaraseanu, B., Ghitea, T. C., et al. (2022) · Medicina · doi:10.3390/medicina58101447 · Cited by 41
“Patients with SCI can benefit from Rex®, EKSOTM and ReWalk® to improve their mobility (walking, climbing stairs) [82,83].”
BackgroundAssistance strategy for stair ascent with a robotic hip exoskeleton
Jang, J., Kim, K., Lee, J., Lim, B., & Shim, Y. (2016) · IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) · doi:10.1109/iros.2016.7759832 · Cited by 34
“Recently many studies for walking assistance activity have been performed in the form of robotic lower limb exoskeletons [1]–[5].”
BackgroundThe Efficacy of State of the Art Overground Gait Rehabilitation Robotics: A Bird's Eye View
Alias, N. A., Huq, M. S., Ibrahim, B. S. K. K., & Omar, R. (2017) · Procedia Computer Science · doi:10.1016/j.procs.2017.01.235 · Cited by 34
“All other DOF of the BWS framework are sans let, either by a latent system or basically by the associating straps [16].”
BackgroundNarayan, J., Abbas, M., & Dwivedy, S. K. (2022) · Automatika · doi:10.1080/00051144.2022.2119498 · Cited by 36
Lower Limb Exoskeleton Sensors: State-of-the-Art
Neťuková, S., Bejtic, M., Malá, C., Horáková, L., Kutílek, P., Kauler, J., & Krupička, R. (2022) · Sensors · doi:10.3390/s22239091 · Cited by 44
Metadata
- Authors
- Esquenazi, A., Talaty, M., Packel, A., & Saulino, M.
- Affiliations
- Moss Rehabilitation Hospital (US)
- Corresponding author
- Esquenazi, Alberto
- Venue
- American Journal of Physical Medicine & Rehabilitation
- Published
- 2012-11
- Volume
- 91
- Issue
- 11
- Pages
- 911-921
- ISSN
- 0894-9115 (ISSN-L)
- OpenAlex
- W2070825536
- Cited by
- 958
- References
- 14
- Citations by year
- 2026: 56 · 2025: 66 · 2024: 89 · 2023: 55 · 2022: 96 · 2021: 100 · 2020: 87 · 2019: 102 · 2018: 79 · 2017: 73 · 2016: 54 · 2015: 55 · 2014: 30 · 2013: 14 · 2012: 1
- FWCI
- 18.22
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- American Journal of Physical Medicine & Rehabilitation (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Prosthetics and Rehabilitation Robotics
- Health Sciences › Medicine › Clinical Neurology › Spinal Cord Injury Research
- Health Sciences › Medicine › Rehabilitation › Stroke Rehabilitation and Recovery
- Keywords
- powered exoskeleton, thoracic spinal cord injury, motor complete spinal cord injury, paraplegia, ambulatory function, gait training, walking performance, safety, psychosocial benefits, bowel and bladder function, spasticity, injury level
- MeSH terms
- Adolescent, Adult, Equipment Design, Female, Humans, Locomotion, Male, Middle Aged, Orthotic Devices, Paraplegia, Spinal Cord Injuries, Thoracic Vertebrae, Walking
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.