The DOI matches its Crossref record.
150-μm Spatial Resolution Using Photon-Counting Detector Computed Tomography Technology: Technical Performance and First Patient Images
Open accessAbstract
Objective: The aims of this study were to quantitatively assess two new scan modes on a photon-counting detector computed tomography system, each designed to maximize spatial resolution, and to qualitatively demonstrate potential clinical impact using patient data. Materials and Methods: This Health Insurance Portability Act–compliant study was approved by our institutional review board. Two high–spatial-resolution scan modes (Sharp and UHR) were evaluated using phantoms to quantify spatial resolution and image noise, and results were compared with the standard mode (Macro). Patients were scanned using a conventional energy-integrating detector scanner and the photon-counting detector scanner using the same radiation dose. In first patient images, anatomic details were qualitatively evaluated to demonstrate potential clinical impact. Results: Sharp and UHR modes had a 69% and 87% improvement in in-plane spatial resolution, respectively, compared with Macro mode (10% modulation-translation-function values of 16.05, 17.69, and 9.48 lp/cm, respectively). The cutoff spatial frequency of the UHR mode (32.4 lp/cm) corresponded to a limiting spatial resolution of 150 μm. The full-width-at-half-maximum values of the section sensitivity profiles were 0.41, 0.44, and 0.67 mm for the thinnest image thickness for each mode (0.25, 0.25, and 0.5 mm, respectively). At the same in-plane spatial resolution, Sharp and UHR images had up to 15% lower noise than Macro images. Patient images acquired in Sharp mode demonstrated better delineation of fine anatomic structures compared with Macro mode images. Conclusions: Phantom studies demonstrated superior resolution and noise properties for the Sharp and UHR modes relative to the standard Macro mode and patient images demonstrated the potential benefit of these scan modes for clinical practice.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Flohr, T., Petersilka, M., Henning, A., Ulzheimer, S., Ferda, J., & Schmidt, B. (2020) · Physica Medica · doi:10.1016/j.ejmp.2020.10.030 · Cited by 581
Self-citationPhoton-counting Detector CT: System Design and Clinical Applications of an Emerging Technology
Leng, S., Bruesewitz, M., Tao, S., Rajendran, K., Halaweish, A. F., Campeau, N. G., Fletcher, J. G., & McCollough, C. H. (2019) · RadioGraphics · doi:10.1148/rg.2019180115 · Cited by 545
Self-citationLell, M. M., & Kachelrieß, M. (2020) · Investigative Radiology · doi:10.1097/rli.0000000000000601 · Cited by 254
Euler, A., Nowak, T., Bucher, B., Eberhard, M., Schmidt, B., Flohr, T. G., Frey, D., Distler, O., & Alkadhi, H. (2021) · Investigative Radiology · doi:10.1097/rli.0000000000000781 · Cited by 19
Metadata
- Authors
- Leng, S., Rajendran, K., Gong, H., Zhou, W., Halaweish, A. F., Henning, A., Kappler, S., Baer, M., Fletcher, J. G., & McCollough, C. H.
- Venue
- Investigative Radiology
- Published
- 2018-11
- Volume
- 53
- Issue
- 11
- Pages
- 655-662
- ISSN
- 0020-9996 (ISSN-L)
- 1536-0210
- OpenAlex
- W2806792627
- Cited by
- 208
- References
- 0
- FWCI
- 5.66
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: 150-μm Spatial Resolution Using Photon-Counting Detector Computed Tomography Technology: Technical Performance and First Patient Images. https://app.cochat.ai/papers/10.1097/rli.0000000000000488
- The paper
- Leng, S., Rajendran, K., Gong, H., Zhou, W., Halaweish, A. F., Henning, A., Kappler, S., Baer, M., Fletcher, J. G., & McCollough, C. H. (2018). 150-μm Spatial Resolution Using Photon-Counting Detector Computed Tomography Technology: Technical Performance and First Patient Images. Investigative Radiology. https://doi.org/10.1097/rli.0000000000000488
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.