Assessment of Bone Mineral Density From a Computed Tomography Topogram of Photon-Counting Detector Computed Tomography—Effect of Phantom Size and Tube Voltage
Abstract
Purpose: The aim of this study was to assess the accuracy and impact of different sizes and tube voltages on bone mineral density (BMD) assessment using a computed tomography (CT) topogram acquired with photon-counting detector CT in an osteopenic ex vivo animal spine. Materials and Methods: The lumbar back of a piglet was used to simulate osteopenia of the lumbar spine. Five fat layers (each with a thickness of 3 cm) were consecutively placed on top of the excised spine to emulate a total of 5 different sizes. Each size was repeatedly imaged on (A) a conventional dual-energy x-ray absorptiometry scanner as the reference standard, (B) a prototype photon-counting detector CT system at 120 kVp with energy thresholds at 20 and 70 keV, and (C) the same prototype system at 140 kVp with thresholds at 20 and 75 keV. Material-specific data were reconstructed from spectral topograms for B and C. Bone mineral density was measured for 3 lumbar vertebrae (L2–L4). A linear mixed-effects model was used to estimate the impact of vertebra, imaging setup, size, and their interaction term on BMD. Results: The BMD of the lumbar spine corresponded to a T score in humans between −4.2 and −4.8, which is seen in osteoporosis. Averaged across the 3 vertebrae and 5 sizes, mean BMD was 0.56 ± 0.03, 0.55 ± 0.02, and 0.55 ± 0.02 g/cm 2 for setup A, B, and C, respectively. There was no significant influence of imaging setup (P = 0.7), simulated size (P = 0.67), and their interaction term (both P > 0.2) on BMD. Bone mineral density decreased significantly from L2 to L4 for all 3 setups (all P < 0.0001). Bone mineral density was 0.59 ± 0.01, 0.57 ± 0.01, and 0.52 ± 0.02 g/cm 2 for L2, L3, and L4, respectively, for setup A; 0.57 ± 0.02, 0.55 ± 0.01, and 0.53 ± 0.01 g/cm 2 for setup B; and 0.57 ± 0.01, 0.55 ± 0.01, and 0.53 ± 0.01 g/cm 2 for setup C. Conclusion: A single CT topogram acquired on photon-counting detector CT with 2 energy thresholds enabled BMD quantification with similar accuracy compared with dual-energy x-ray absorptiometry over a range of simulated sizes and tube voltages in an osteopenic ex vivo animal spine.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Two prior studies on a prototype PCD-CT scanner in a phantom and in an ex vivo animal experiment have already indicated accurate aBMD measurements with spectral localizer radiographs.
“Two prior studies on a prototype PCD-CT scanner in a phantom and in an ex vivo animal experiment have already indicated accurate aBMD measurements with spectral localizer radiographs.9,11”
From In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT (2025) · cites this paper as [11]
Cites
Li, X., Li, X., Li, J., Jiao, X., Jia, X., Zhang, X., Fan, G., Yang, J., & Guo, J. (2020) · Clinical Radiology · doi:10.1016/j.crad.2019.11.008 · Cited by 41 · see citing works
DXA-equivalent quantification of bone mineral density using dual-layer spectral CT scout scans
Laugerette, A., Schwaiger, B. J., Brown, K., Frerking, L. C., Kopp, F. K., Mei, K., Sellerer, T., Kirschke, J., Baum, T., Gersing, A. S., et al. (2019) · European Radiology · doi:10.1007/s00330-019-6005-6 · Cited by 29 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Moser, L. J., Klambauer, K., Diaz Machicado, M. C., Frey, D., Mergen, V., Eberhard, M., Nowak, T., Schmidt, B., Flohr, T., Distler, O., et al. (2025) · Investigative Radiology · doi:10.1097/rli.0000000000001159 · Cited by 8 · see citing works
Self-citation
- Authors
- Euler, A., Nowak, T., Bucher, B., Eberhard, M., Schmidt, B., Flohr, T. G., Frey, D., Distler, O., & Alkadhi, H.
- Affiliations
- University of Zurich (CH)
- Siemens Healthineers (Germany) (DE)
- Independent Researcher, Los Angeles, California
- Corresponding author
- Euler, André
- Eberhard, Matthias
- Alkadhi, Hatem
- ORCID
- Venue
- Investigative Radiology
- Published
- 2021-04-08
- Volume
- 56
- Issue
- 10
- Pages
- 614-620
- ISSN
- 0020-9996 (ISSN-L)
- 1536-0210
- OpenAlex
- W3150426754
- Cited by
- 19
- References
- 2
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 5 · 2023: 4 · 2022: 3 · 2021: 1
- FWCI
- 0.99
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Investigative Radiology (publishedVersion)
- PubMed (publishedVersion)
- Zurich Open Repository and Archive (University of Zurich) (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Keywords
- bone mineral density, photon-counting CT, tube voltage, phantom size, lumbar spine, osteopenia, osteoporosis, dual-energy X-ray absorptiometry, measurement accuracy, energy threshold
- MeSH terms
- Animals, Humans, Lumbar Vertebrae, Osteoporosis, Swine, Tomography, X-Ray Computed, Absorptiometry, Photon, Bone Density
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review