Assessment of Bone Mineral Density From a Computed Tomography Topogram of Photon-Counting Detector Computed Tomography—Effect of Phantom Size and Tube Voltage

Euler, André, Nowak, Tristan, Bucher, Brian, Eberhard, Matthias, Schmidt, Bernhard, Flohr, Thomas G., Frey, Diana, Distler, Oliver +1 more·2021·Investigative Radiology

Abstract

Purpose: The aim of this study was to assess the accuracy and impact of different sizes and tube voltages on bone mineral density (BMD) assessment using a computed tomography (CT) topogram acquired with photon-counting detector CT in an osteopenic ex vivo animal spine. Materials and Methods: The lumbar back of a piglet was used to simulate osteopenia of the lumbar spine. Five fat layers (each with a thickness of 3 cm) were consecutively placed on top of the excised spine to emulate a total of 5 different sizes. Each size was repeatedly imaged on (A) a conventional dual-energy x-ray absorptiometry scanner as the reference standard, (B) a prototype photon-counting detector CT system at 120 kVp with energy thresholds at 20 and 70 keV, and (C) the same prototype system at 140 kVp with thresholds at 20 and 75 keV. Material-specific data were reconstructed from spectral topograms for B and C. Bone mineral density was measured for 3 lumbar vertebrae (L2–L4). A linear mixed-effects model was used to estimate the impact of vertebra, imaging setup, size, and their interaction term on BMD. Results: The BMD of the lumbar spine corresponded to a T score in humans between −4.2 and −4.8, which is seen in osteoporosis. Averaged across the 3 vertebrae and 5 sizes, mean BMD was 0.56 ± 0.03, 0.55 ± 0.02, and 0.55 ± 0.02 g/cm 2 for setup A, B, and C, respectively. There was no significant influence of imaging setup (P = 0.7), simulated size (P = 0.67), and their interaction term (both P > 0.2) on BMD. Bone mineral density decreased significantly from L2 to L4 for all 3 setups (all P < 0.0001). Bone mineral density was 0.59 ± 0.01, 0.57 ± 0.01, and 0.52 ± 0.02 g/cm 2 for L2, L3, and L4, respectively, for setup A; 0.57 ± 0.02, 0.55 ± 0.01, and 0.53 ± 0.01 g/cm 2 for setup B; and 0.57 ± 0.01, 0.55 ± 0.01, and 0.53 ± 0.01 g/cm 2 for setup C. Conclusion: A single CT topogram acquired on photon-counting detector CT with 2 energy thresholds enabled BMD quantification with similar accuracy compared with dual-energy x-ray absorptiometry over a range of simulated sizes and tube voltages in an osteopenic ex vivo animal spine.

Integrity signals

Registry

Crossref · 10.1097/rli.0000000000000781

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Two prior studies on a prototype PCD-CT scanner in a phantom and in an ex vivo animal experiment have already indicated accurate aBMD measurements with spectral localizer radiographs.

    “Two prior studies on a prototype PCD-CT scanner in a phantom and in an ex vivo animal experiment have already indicated accurate aBMD measurements with spectral localizer radiographs.9,11”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Euler, A., Nowak, T., Bucher, B., Eberhard, M., Schmidt, B., Flohr, T. G., Frey, D., Distler, O., & Alkadhi, H.
Affiliations
  • University of Zurich (CH)
  • Siemens Healthineers (Germany) (DE)
  • Independent Researcher, Los Angeles, California
Corresponding author
  • Euler, André
  • Eberhard, Matthias
  • Alkadhi, Hatem
Venue
Investigative Radiology
Published
2021-04-08
Volume
56
Issue
10
Pages
614-620
ISSN
  • 0020-9996 (ISSN-L)
  • 1536-0210
OpenAlex
W3150426754
Cited by
19
References
2
Citations by year
2026: 1 · 2025: 5 · 2024: 5 · 2023: 4 · 2022: 3 · 2021: 1
FWCI
0.99
Type
journal article
Language
en
Open access
closed
Locations
  • Investigative Radiology (publishedVersion)
  • PubMed (publishedVersion)
  • Zurich Open Repository and Archive (University of Zurich) (submittedVersion)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
  • Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
  • Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
Keywords
bone mineral density, photon-counting CT, tube voltage, phantom size, lumbar spine, osteopenia, osteoporosis, dual-energy X-ray absorptiometry, measurement accuracy, energy threshold
MeSH terms
Animals, Humans, Lumbar Vertebrae, Osteoporosis, Swine, Tomography, X-Ray Computed, Absorptiometry, Photon, Bone Density

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review