Image Quality and Reconstruction Intervals of Dual-Source CT Coronary Angiography: Recommendations for ECG-Pulsing Windowing
The DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
PURPOSE: To evaluate reconstruction intervals and image quality in dual-source computed tomography (DSCT) coronary angiography for optimal placement of the ECG-pulsing window. MATERIALS AND METHODS: DSCT coronary angiography was performed in 60 patients. Thirteen datasets were reconstructed in 5% increments from 20-80%. Two readers independently assessed image quality of each segment in each percentage-interval, using scores ranging from 1 (no motion artifacts) to 4 (nonevaluable). RESULTS: Mean heart rate (HR) was 69.0 +/- 18.9 beats per minute (bpm) (range, 35-117 bpm). Diagnostic image quality (scores 1-3) was found in 97.8% of all segments (763 of 780). The 70% RR-interval provided best image quality in all patients and all HRs. The narrowest reconstruction window providing diagnostic image quality was 60-70% for HR <60 bpm, 60-80% for 60-70 bpm, 55-80% for 70-80 bpm, and 30-80% for HR >80 bpm. CONCLUSIONS: DSCT coronary angiography provides best image quality for various HRs at 70%. The ECG-pulsing window can be adapted according to the HR while maintaining diagnostic image quality.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Measurement of Observer Agreement for Categorical Data
Landis, J. R., & Koch, G. G. (1977) · Biometrics · doi:10.2307/2529310 · Cited by 81,148 · cited 1×
“20 was considered as slight or poor agreement.(18) The Fisher exact test was used to evaluate categorical data (distribution of the sexes among HR groups).”
MethodAusten, W., Edwards, J., Frye, R., Gensini, G., Gott, V., Griffith, L., McGoon, D., Murphy, M., & Roe, B. (1975) · Circulation · doi:10.1161/01.cir.51.4.5 · Cited by 3,256 · cited 1×
“Coronary arteries were classified according to the guidelines of the American Heart Association (15 segments).(15) The right coronary artery was defined to include segments 1–4, the left main and left anterior descending artery to include segments 5–10, and the left circumflex artery to include segments 11–15.”
MethodDiagnostic Accuracy of Noninvasive Coronary Angiography Using 64-Slice Spiral Computed Tomography
Raff, G. L., Gallagher, M. J., O’Neill, W. W., & Goldstein, J. A. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.05.056 · Cited by 1,472
First performance evaluation of a dual-source CT (DSCT) system
Flohr, T. G., McCollough, C. H., Bruder, H., Petersilka, M., Gruber, K., Süβ, C., Grasruck, M., Stierstorfer, K., Krauss, B., Raupach, R., et al. (2006) · European Radiology · doi:10.1007/s00330-005-2919-2 · Cited by 1,413 · cited 3×
“With DSCT, a variable and individually tailored ECGpulsing approach for radiation dose reduction has become feasible.(11) The implementation of ECG-pulsing requires knowledge about the expected necessary reconstruction intervals in each individual patient before the scan, to correctly assign the scan period with peak…”
BackgroundMethodSelf-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl001 · Cited by 1,336
Quantification of Obstructive and Nonobstructive Coronary Lesions by 64-Slice Computed Tomography
Leber, A. W., Knez, A., von Ziegler, F., Becker, A., Nikolaou, K., Paul, S., Wintersperger, B., Reiser, M., Becker, C. R., Steinbeck, G., et al. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.03.071 · Cited by 1,109
Mollet, N. R., Cademartiri, F., van Mieghem, C. A. G., Runza, G., McFadden, E. P., Baks, T., Serruys, P. W., Krestin, G. P., & de Feyter, P. J. (2005) · Circulation · doi:10.1161/circulationaha.105.533471 · Cited by 1,018
Accuracy of MSCT coronary angiography with 64-slice technology: first experience
Leschka, S., Alkadhi, H., Plass, A., Desbiolles, L., Grünenfelder, J., Marincek, B., & Wildermuth, S. (2005) · European Heart Journal · doi:10.1093/eurheartj/ehi261 · Cited by 960
Self-citationNieman, K., Cademartiri, F., Lemos, P. A., Raaijmakers, R., Pattynama, P. M. T., & de Feyter, P. J. (2002) · Circulation · doi:10.1161/01.cir.0000037222.58317.3d · Cited by 956
Hausleiter, J., Meyer, T., Hadamitzky, M., Huber, E., Zankl, M., Martinoff, S., Kastrati, A., & Schömig, A. (2006) · Circulation · doi:10.1161/circulationaha.105.602490 · Cited by 705 · cited 1×
“4–11 mSv.(22) One reason that ECG pulsing has not been used might be the lack of clinical stability of conventional ECG-pulsing algorithms for patients with arrhythmia and ectopic beats.”
BackgroundJakobs, T. F., Becker, C. R., Ohnesorge, B., Flohr, T., Suess, C., Schoepf, U. J., & Reiser, M. F. (2002) · European Radiology · doi:10.1007/s00330-001-1278-x · Cited by 615 · cited 2×
“For dose reduction, it is proposed to apply the technique of ECGcontrolled tube current modulation technique that is characterized by peak tube output during mid- to end-diastole and a reduction of tube output to 20% during other parts of the cardiac cycle.(19) Nevertheless, this technique was not used in some 64-slic…”
MethodResultSelf-citationScheffel, H., Alkadhi, H., Plass, A., Vachenauer, R., Desbiolles, L., Gaemperli, O., Schepis, T., Frauenfelder, T., Schertler, T., Husmann, L., et al. (2006) · European Radiology · doi:10.1007/s00330-006-0474-0 · Cited by 420
Self-citationHong, C., Becker, C. R., Huber, A., Schoepf, U. J., Ohnesorge, B., Knez, A., Brüning, R., & Reiser, M. F. (2001) · Radiology · doi:10.1148/radiol.2203010055 · Cited by 255 · cited 7×
“Moreover, diagnostic image quality could be obtained for HR 65 bpm at a single reconstruction interval of 60%.23 Ferencik and colleagues24 have observed the optimal reconstruction window in mid-diastole from 60% to 70% of the RR-interval.”
MethodInfluentialWintersperger, B. J., Nikolaou, K., von Ziegler, F., Johnson, T. R. C., Rist, C., Leber, A. W., Flohr, T. G., Knez, A., Reiser, M. F., & Becker, C. R. (2006) · Investigative Radiology · doi:10.1097/01.rli.0000202639.99949.c6 · Cited by 194 · cited 1×
“This approach improves temporal resolution at a specific HR; however, its use is limited by artifacts occurring with variable HRs, with interheartbeat variability of coronary artery position, and by the necessity of lower pitch factors prolonging data acquisition time and increasing radiation dose.(9,10) Dual-source C…”
BackgroundSelf-citationGhostine, S., Caussin, C., Daoud, B., Habis, M., Perrier, E., Pesenti-Rossi, D., Sigal-Cinqualbre, A. B., Angel, C.-Y., Lancelin, B., Capderou, A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.04.103 · Cited by 127 · cited 1×
“Nevertheless, this technique was not used in some 64-slice scanner studies, but was used in others.(20,21) This has resulted in estimated effective radiation doses for 64-slice CT coronary angiography studies up to 21.”
MethodOptimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT
Leschka, S., Husmann, L., Desbiolles, L. M., Gaemperli, O., Schepis, T., Koepfli, P., Boehm, T., Marinček, B., Kaufmann, P. A., & Alkadhi, H. (2006) · European Radiology · doi:10.1007/s00330-006-0262-x · Cited by 117 · cited 1×
“Moreover, diagnostic image quality could be obtained for HR 65 bpm at a single reconstruction interval of 60%.(23) Ferencik and colleagues(24) have observed the optimal reconstruction window in mid-diastole from 60% to 70% of the RR-interval.”
MethodSelf-citationAn information theoretic measure for the evaluation of ordinal scale data
Tastle, W. J., & Wierman, M. J. (2006) · Behavior Research Methods · doi:10.3758/bf03192803 · Cited by 62 · cited 1×
“A Likert scale allows to determine degrees of agreement when analyzing ordinal-ranked categories.(16) Scores 1–3 were considered diagnostic.”
MethodDewey, M., Müller, M., Teige, F., Schnapauff, D., Schink, T., Hamm, B., & Lembcke, A. E. (2006) · Investigative Radiology · doi:10.1097/01.rli.0000201233.42994.9b · Cited by 42 · cited 1×
“Another method to compensate for motion artifacts at higher HR represents the use of two- or multi-segment reconstruction algorithms that merge data from 2 or more consecutive heart beats for image reconstruction.(8) This approach improves temporal resolution at a specific HR; however, its use is limited by artifacts…”
Method
Cited by
Schroeder, S., Achenbach, S., Bengel, F., Burgstahler, C., Cademartiri, F., de Feyter, P., George, R., Kaufmann, P., Kopp, A. F., Knuuti, J., et al. (2008) · European Heart Journal · doi:10.1093/eurheartj/ehm544 · Cited by 554
Self-citationSCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT
Halliburton, S. S., Abbara, S., Chen, M. Y., Gentry, R., Mahesh, M., Raff, G. L., Shaw, L. J., & Hausleiter, J. (2011) · Journal of Cardiovascular Computed Tomography · doi:10.1016/j.jcct.2011.06.001 · Cited by 491
Contemporary carotid imaging: from degree of stenosis to plaque vulnerability
Brinjikji, W., Huston, J., Rabinstein, A. A., Kim, G.-M., Lerman, A., & Lanzino, G. (2016) · Journal of Neurosurgery · doi:10.3171/2015.1.jns142452 · Cited by 401
Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance
Scheffel, H., Alkadhi, H., Leschka, S., Plass, A., Desbiolles, L., Guber, I., Krauss, T., Gruenenfelder, J., Genoni, M., Luescher, T. F., et al. (2008) · Heart · doi:10.1136/hrt.2008.149971 · Cited by 291
Self-citationLow kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose
Leschka, S., Stolzmann, P., Schmid, F. T., Scheffel, H., Stinn, B., Marincek, B., Alkadhi, H., & Wildermuth, S. (2008) · European Radiology · doi:10.1007/s00330-008-0966-1 · Cited by 290
Self-citation
Metadata
- Authors
- Leschka, S., Scheffel, H., Desbiolles, L., Plass, A., Gaemperli, O., Valenta, I., Husmann, L., Flohr, T. G., Genoni, M., Marincek, B., Kaufmann, P. A., & Alkadhi, H.
- Venue
- Investigative Radiology
- Published
- 2007-08
- Volume
- 42
- Issue
- 8
- Pages
- 543-549
- ISSN
- 0020-9996 (ISSN-L)
- OpenAlex
- W1974239818
- Cited by
- 172
- References
- 18
- FWCI
- 23.03
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.