A real-time microprocessor QRS detector system with a 1-ms timing accuracy for the measurement of ambulatory HRV
Abstract
The design, test methods and results of an ambulatory QRS detector are presented. The device is intended for the accurate measurement of heart rate variability (HRV) and reliable QRS detection in both ambulatory and clinical use. The aim of the design work was to achieve high QRS detection performance in terms of timing accuracy and reliability, without compromising the size and power consumption of the device. The complete monitor system consists of a host computer and the detector unit. The detector device is constructed of a commonly available digital signal processing (DSP) microprocessor and other components. The QRS detection algorithm uses optimized prefiltering in conjunction with a matched filter and dual edge threshold detection. The purpose of the prefiltering is to attenuate various noise components in order to achieve improved detection reliability. The matched filter further improves signal-to-noise ratio (SNR) and symmetries the QRS complex for the threshold detection, which is essential in order to achieve the desired performance. The decision for detection is made in real-time and no search-back method is employed. The host computer is used to configure the detector unit, which includes the setting of the matched filter impulse response, and in the retrieval and postprocessing of the measurement results. The QRS detection timing accuracy and detection reliability of the detector system was tested with an artificially generated electrocardiogram (ECG) signal corrupted with various noise types and a timing standard deviation of less than 1 ms was achieved with most noise types and levels similar to those encountered in real measurements. A QRS detection error rate (ER) of 0.1 and 2.2% was achieved with records 103 and 105 from the MIT-BIH Arrhythmia database, respectively.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855
Quantitative Investigation of QRS Detection Rules Using the MIT/BIH Arrhythmia Database
Hamilton, P. S., & Tompkins, W. J. (1986) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1986.325695 · Cited by 1,142
A comparison of the noise sensitivity of nine QRS detection algorithms
Friesen, G. M., Jannett, T. C., Jadallah, M. A., Yates, S. L., Quint, S. R., & Nagle, H. T. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.43620 · Cited by 1,043
Applications of adaptive filtering to ECG analysis: noise cancellation and arrhythmia detection
Thakor, N. V., & Zhu, Y.-S. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83591 · Cited by 885
Estimation of QRS Complex Power Spectra for Design of a QRS Filter
Thakor, N. V., Webster, J. G., & Tompkins, W. J. (1984) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1984.325393 · Cited by 427
Neural-network-based adaptive matched filtering for QRS detection
Xue, Q., Hu, Y. H., & Tompkins, W. J. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.126604 · Cited by 423
Autonomic control of heart rate during exercise studied by heart rate variability spectral analysis
Yamamoto, Y., Hughson, R. L., & Peterson, J. C. (1991) · Journal of Applied Physiology · doi:10.1152/jappl.1991.71.3.1136 · Cited by 383
Information Transmission, Modulation, and Noise
Barker, R. H. (1960) · Physics Bulletin · doi:10.1088/0031-9112/11/2/012 · Cited by 275
Malík, M., Cripps, T., Farrell, T., & Camm, A. J. (1989) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441642 · Cited by 90
Uijen, G. J. H., de Weerd, J. P. C., & Vendrik, A. J. H. (1979) · Medical & Biological Engineering & Computing · doi:10.1007/bf02447064 · Cited by 41
Cited by
The principles of software QRS detection
Kohler, B.-U., Hennig, C., & Orglmeister, R. (2002) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.993193 · Cited by 1,242
Software for advanced HRV analysis
Niskanen, J.-P., Tarvainen, M. P., Ranta-aho, P. O., & Karjalainen, P. A. (2004) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2004.03.004 · Cited by 743
Validity of the Polar S810 Heart Rate Monitor to Measure R-R Intervals at Rest
GAMELIN, F. X., BERTHOIN, S., & BOSQUET, L. (2006) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000218135.79476.9c · Cited by 509
The use of the Hilbert transform in ECG signal analysis
Benitez, D., Gaydecki, P. A., Zaidi, A., & Fitzpatrick, A. P. (2001) · Computers in Biology and Medicine · doi:10.1016/s0010-4825(01)00009-9 · Cited by 489
RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438
Metadata
- Authors
- Ruha, A., Sallinen, S., & Nissila, S.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 1997-03
- Volume
- 44
- Issue
- 3
- Pages
- 159-167
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2112020801
- Cited by
- 255
- References
- 10
- FWCI
- 3.02
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A real-time microprocessor QRS detector system with a 1-ms timing accuracy for the measurement of ambulatory HRV. https://app.cochat.ai/papers/10.1109/10.554762
- The paper
- Ruha, A., Sallinen, S., & Nissila, S. (1997). A real-time microprocessor QRS detector system with a 1-ms timing accuracy for the measurement of ambulatory HRV. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/10.554762
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.