RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Open accessAbstract
This study was designed to examine the RR interval signal qualities of a Holter device and a heart rate chest belt monitor at rest and during exercise. Ten healthy individuals completed five low- to high-intensity activities while simultaneously using the medilog® AR12plus Holter monitor and the Polar H10 heart rate monitor. The RR interval signal quality was based on the quantification of the missing RR intervals and RR interval detection errors. Therefore, both measurement systems were compared against visual inspection of the raw electrocardiography signal. The missing and wrong R-wave peak detections were counted manually for both measurement systems. RR interval signal quality was defined as the relative number of correctly detected RR intervals. Overall, RR interval signal qualities of 94.6% and 99.6% were demonstrated for the medilog® AR12plus and the Polar H10. During the high-intensity activities, the RR interval signal quality of the medilog® AR12plus dropped to 89.8%, whereas the Polar H10 maintained a signal quality of 99.4%. The correlation between both systems was high (r = 0.997, p > 0.001). The excellent RR interval signal quality during low- to moderate-intensity activities in the medilog® AR12plus and during low- to high-intensity activities in the Polar H10 demonstrates both measurement systems' validity for the detection of RR intervals throughout a wide range of activities. A simple chest strap such as the Polar H10 might be recommended as the gold standard for RR interval assessments if intense activities with strong body movements are investigated.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The Polar H10 has shown strong agreement with clinical ECG instrumentation in prior validation studies.
“The H10 has shown strong agreement with clinical ECG instrumentation in prior validation studies [36].”
From PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone Effects (2026) · cites this paper as [36]
In a previous report, the Polar H10 had substantially better signal quality during high-intensity activities compared to a 3-lead ECG Holter monitor.
“In this report, there was a substantially better signal quality during high-intensity activities for the Polar H10 in comparison to a 3-lead ECG Holter monitor”
From Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women (2022) · cites this paper as [13]
The Polar H10 chest strap device has already proven validity to assess RR intervals correctly during rest and physical exercise conditions.
“A frequently used chest strap device is the Polar H10 (Polar Electro Oy, Kempele, Finland) which has already proven validity to assess RR intervals correctly during rest and physical exercise conditions [13].”
From Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women (2022) · cites this paper as [13]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159 · see citing works
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917 · see citing works
Heart rate variability: Standards of measurement, physiological interpretation, and clinical use
Malik, M., Bigger, J. T., Camm, A. J., Kleiger, R. E., Malliani, A., Moss, A. J., & Schwartz, P. J. (1996) · European Heart Journal · doi:10.1093/oxfordjournals.eurheartj.a014868 · Cited by 15,289 · see citing works
Perceived exertion as an indicator of somatic stress
Borg, G. (1970) · Journal of Rehabilitation Medicine · doi:10.2340/1650197719702239298 · Cited by 4,723 · see citing works
Validity of the Polar S810 Heart Rate Monitor to Measure R-R Intervals at Rest
Gamelin, F.-X., Berthoin, S., & Bosquet, L. (2006) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000218135.79476.9c · Cited by 502 · see citing works
Power spectrum analysis of heart rate variability: a tool to explore neural regulatory mechanisms.
Malliani, A., Lombardi, F., & Pagani, M. (1994) · Heart · doi:10.1136/hrt.71.1.1 · Cited by 492 · see citing works
Vagal modulation of heart rate during exercise: effects of age and physical fitness
Tulppo, M. P., Mäkikallio, T. H., Seppänen, T., Laukkanen, R. M. T., & Huikuri, H. V. (1998) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1998.274.2.h424 · Cited by 413 · see citing works
Validity of the Polar V800 heart rate monitor to measure RR intervals at rest
Giles, D., Draper, N., & Neil, W. (2015) · European Journal of Applied Physiology · doi:10.1007/s00421-015-3303-9 · Cited by 376 · see citing works
Plews, D. J., Scott, B., Altini, M., Wood, M., Kilding, A. E., & Laursen, P. B. (2017) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2016-0668 · Cited by 323 · see citing works
Validity and Reliability of Short-Term Heart-Rate Variability from the Polar S810
NUNAN, D., DONOVAN, G., JAKOVLJEVIC, D. G., HODGES, L. D., SANDERCOCK, G. R. H., & BRODIE, D. A. (2009) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e318184a4b1 · Cited by 302 · see citing works
Weippert, M., Kumar, M., Kreuzfeld, S., Arndt, D., Rieger, A., & Stoll, R. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1415-9 · Cited by 302 · see citing works
Decrease in heart rate variability with overtraining: assessment by the Poincaré plot analysis
Mourot, L., Bouhaddi, M., Perrey, S., Cappelle, S., Henriet, M. T., Wolf, J., Rouillon, J. D., & Regnard, J. (2004) · Clinical Physiology and Functional Imaging · doi:10.1046/j.1475-0961.2003.00523.x · Cited by 257 · see citing works
Ruha, A., Sallinen, S., & Nissilä, S. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.554762 · Cited by 257 · see citing works
Kingsley, M., Lewis, M. J., & Marson, R. E. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-817878 · Cited by 230 · see citing works
Singh, N., Moneghetti, K. J., Christle, J. W., Hadley, D., Plews, D. J., & Froelicher, V. F. (2018) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2018.27.2 · Cited by 140 · see citing works
de Rezende Barbosa, M. P. D. C., Silva, N. T. D., de Azevedo, F. M., Pastre, C. M., & Vanderlei, L. C. M. (2016) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12203 · Cited by 129 · see citing works
Nunan, D., Jakovljević, D. G., Donovan, G., Hodges, L. D., SANDERCOCK, G. R. H., & Brodie, D. A. (2008) · European Journal of Applied Physiology · doi:10.1007/s00421-008-0742-6 · Cited by 123 · see citing works
Cardiac autonomic imbalance in an overtrained athlete
Hedelin, R., Wiklund, U., Bjerle, P., & Henriksson-Larsén, K. (2000) · Medicine & Science in Sports & Exercise · doi:10.1097/00005768-200009000-00001 · Cited by 120 · see citing works
Caminal, P., Sola, F., Gomis, P., Guasch, E., Perera-Lluna, A., Soriano, N. P., & Mont, L. (2018) · European Journal of Applied Physiology · doi:10.1007/s00421-018-3808-0 · Cited by 108 · see citing works
Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during Exercise
HORTON, J. F., STERGIOU, P., FUNG, T. S., & KATZ, L. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001388 · Cited by 79 · see citing works
Comparison of Omega Wave System and Polar S810i to Detect R-R Intervals at Rest
Parrado, E., García, M. Á., Ramos, J. L., et al. (2010) · International Journal of Sports Medicine · doi:10.1055/s-0030-1248319 · Cited by 45 · see citing works
Cheatham, S. W., Kolber, M. J., & Ernst, M. P. (2014) · Journal of Sport Rehabilitation · doi:10.1123/jsr.2013-0145 · Cited by 43 · see citing works
Accuracy of the Garmin 920 XT HRM to perform HRV analysis
Cassirame, J., Vanhaesebrouck, R., Chevrolat, S., & Mourot, L. (2017) · Australasian Physical & Engineering Sciences in Medicine · doi:10.1007/s13246-017-0593-8 · Cited by 20 · see citing works
Heart rate variability in relation to psychological factors
Rompelman, O., KAMPEN, W. H. A. V., Backer, E., et al. (1980) · Ergonomics · doi:10.1080/00140138008924817 · Cited by 14 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Mountjoy, M., Ackerman, K. E., Bailey, D. M., Burke, L. M., Constantini, N., Hackney, A. C., Heikura, I. A., Melin, A., Pensgaard, A. M., Stellingwerff, T., et al. (2023) · British Journal of Sports Medicine · doi:10.1136/bjsports-2023-106994 · Cited by 724 · see citing works
Flexible Wearable Sensors for Cardiovascular Health Monitoring
Chen, S., Qi, J., Fan, S., Qiao, Z., Yeo, J. C., & Lim, C. T. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100116 · Cited by 453 · see citing works
Schaffarczyk, M., Rogers, B., Reer, R., & Gronwald, T. (2022) · Sensors · doi:10.3390/s22176536 · Cited by 349 · see citing works
Lee, J., Kim, M., Park, H.-K., & Kim, I. Y. (2020) · Sensors · doi:10.3390/s20051493 · Cited by 153 · see citing works
Moya-Ramon, M., Mateo-March, M., Peña-González, I., Zabala, M., & Javaloyes, A. (2022) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2022.106696 · Cited by 122 · see citing works
Rogers, B., Giles, D., Draper, N., Mourot, L., & Gronwald, T. (2021) · Sensors · doi:10.3390/s21030821 · Cited by 54 · see citing works
Rogers, B., Schaffarczyk, M., Clauß, M., Mourot, L., & Gronwald, T. (2022) · Sensors · doi:10.3390/s22052032 · Cited by 41 · see citing works
Bozzini, B. N., McFadden, B. A., Scruggs, S. K., & Arent, S. M. (2021) · Journal of Strength & Conditioning Research · doi:10.1519/jsc.0000000000004051 · Cited by 11 · see citing works
Physiological synchrony during partnered sexual activity
Freihart, B. K., & Meston, C. M. (2024) · Journal of Sex & Marital Therapy · doi:10.1080/0092623x.2024.2321127 · Cited by 5 · see citing works
Kostrna, J., Oparina, E., Palacios, C., Rodriguez, A. J., Pei, J., Ajmal, A., & Ramella-Roman, J. C. (2026) · Sensors · doi:10.3390/s26102922 · Cited by 2 · see citing works
- Authors
- Gilgen-Ammann, R., Schweizer, T., & Wyss, T.
- Affiliations
- Swiss Federal Institute of Sport Magglingen SFISM (CH)
- Corresponding author
- Gilgen-Ammann, Rahel
- Venue
- European Journal of Applied Physiology
- Published
- 2019-04-19
- Volume
- 119
- Issue
- 7
- Pages
- 1525-1532
- ISSN
- 0020-9376 (ISSN-L)
- 1439-6319
- 1439-6327
- OpenAlex
- W2938169034
- Cited by
- 438
- References
- 26
- Citations by year
- 2026: 75 · 2025: 75 · 2024: 77 · 2023: 68 · 2022: 66 · 2021: 51 · 2020: 23 · 2019: 2
- FWCI
- 25.64
- Type
- journal article
- Language
- en
- Open access
- green
- doi.org
- Licence
- www.springer.com/tdm (tdm)
- Locations
- European Journal of Applied Physiology (publishedVersion)
- PubMed (publishedVersion)
- ARBOR - Bern University of Applied Sciences Repository (submittedVersion)
- BFH: ARBOR
- Funders
- Polar Electro Oy
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- heart rate monitoring, Holter monitor, exercise intensity, measurement validity, body movement artifacts, heart rate response to exercise
- MeSH terms
- Adult, Female, Heart Rate, Humans, Male, Reference Standards, Reproducibility of Results, Exercise, Electrocardiography, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review