In-silico pre-clinical trials for implantable cardioverter defibrillators
The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.
Abstract
Regulatory authorities require that the safety and efficacy of a new high-risk medical device be proven in a Clinical Trial (CT), in which the effects of the device on a group of patients are compared to the effects of the current standard of care. Phase III trials can run for several years, cost millions of dollars, and expose patients to an unproven device. In this paper, we demonstrate how to use a large group of synthetic patients based on computer modeling to improve the planning of a CT so as to increase the chances of a successful trial for implantable cardioverter defibrillators (ICDs). We developed a computer model of the electrical generation and propagation in the heart. This model was used to generate a large group of heart instances capable of producing episodes of 19 different arrhythmias. We also implemented two arrhythmia detection algorithms from the literature: Rhythm ID from Boston Scientific and PR Logic + Wavelet from Medtronic. Using this setup, we conducted multiple in-silico trials to compare the ability of the two algorithms to appropriately discriminate between potentially fatal Ventricular Tachy-arrhythmias (VT) and nonfatal Supra-Ventricular Tachy-arrhythmias (SVTs). The results of our in-silico trial indicate that Rhythm ID was less able to discriminate between SVT and VT and so may lead to more cases of inappropriate therapy. This corroborates the findings of the Rhythm ID Going Head to Head Trial (RIGHT), a clinical trial that compared the two algorithms in patients. We further demonstrated that the result continues to hold if we vary the distribution of arrhythmias in the synthetic population. We also used the same in-silico cohort to explore the sensitivity of the outcome to different parameter settings of the device algorithms, which is not feasible in a real clinical trial. In-silico trials can provide early insight into the factors which affect the outcome of a CT at a fraction of the cost and duration and without the ethical issues.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“The next example of ISCTs is the work of [46] which uses large numbers of automata-based heart models for evaluating performance of implantable cardioverter defibrillators (ICDs).”
From Credibility assessment of in silico clinical trials for medical devices (2024) · cites this paper as [46]
In the citing paper
“Unlike the example in Section 2.3.5 which involves a relatively small number (~20) of image-derived heart models, [46] develops a large number (> 11,000) of computationally inexpensive heart models.”
From Credibility assessment of in silico clinical trials for medical devices (2024) · cites this paper as [46]
In the citing paper
“The results matched a real clinical trial comparing the two ICDs, the Rhythm ID Going Head-to-Head Trial (RIGHT).”
From Credibility assessment of in silico clinical trials for medical devices (2024) · cites this paper as [46]
Citations
Cites
Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT II
Daubert, J. P., Zareba, W., Cannom, D. S., McNitt, S., Rosero, S. Z., Wang, P., Schuger, C., Steinberg, J. S., Higgins, S. L., Wilber, D. J., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.09.073 · Cited by 813 · cited 1×
“One of the main causes of VT/SVT mis-classifications is inappropriate parameter setting [11].”
Clinical Cardiac Pacing, Defibrillation, and Resynchronization Therapy
(2007) · Elsevier eBooks · doi:10.1016/b978-1-4160-2536-8.x5001-8 · Cited by 225 · cited 2×
“4 Device Testing: We implemented two VT/SVT discrimination algorithms: Rhythm ID from Boston Scientific [6], [7], and PR Logic + Wavelet (PRL+W) from Medtronic [8], [9], [7].”
Medical Device Development: From Prototype to Regulatory Approval
Kaplan, A. V., Baim, D. S., Smith, J. J., Feigal, D. A., Simons, M., Jefferys, D., Fogarty, T. J., Kuntz, R. E., & Leon, M. B. (2004) · Circulation · doi:10.1161/01.cir.0000134695.65733.64 · Cited by 194 · cited 1×
“For example, a 2002 trial for stents lasted 2 years, enrolled 800 patients and cost $10 to $12 million [3].”
Cyber–Physical Modeling of Implantable Cardiac Medical Devices
Jiang, Z., Pajic, M., & Mangharam, R. (2012) · Proceedings of the IEEE · doi:10.1109/jproc.2011.2161241 · Cited by 172 · cited 1×
“We developed an automata-based EP heart model to simulate the electrical activities of the heart (see [5] for heart modeling details).”
Self-citationClinical Cardiac Electrophysiology
Rankin, A. C., Quinn, F. R., & Rae, A. P. (2012) · Cardiac Arrhythmias and Mapping Techniques · doi:10.1007/978-0-85729-877-5_2 · Cited by 162
SWERDLOW, C. D., BROWN, M. L., LURIE, K., ZHANG, J., WOOD, N. M., OLSON, W. H., & GILLBERG, J. M. (2002) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2002.00432.x · Cited by 107
Clinical Cardiac Electrophysiology
Longmore, D. B. (2012) · SpringerReference · doi:10.1007/springerreference_188094 · Cited by 84
Gold, M. R., Ahmad, S., Browne, K. F., Berg, K. C., Thackeray, L., & Berger, R. D. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2011.10.004 · Cited by 66 · cited 9×
“Our findings are consistent with the observations of the RIGHT trial itself [2], and are purely model-based.”
InfluentialClinical Cardiac Pacing, Defibrillation and Resynchronization Therapy
Ellenbogen, K. A. (2011) · Elsevier eBooks · doi:10.1016/c2009-0-44191-6 · Cited by 66
Interventional Electrophysiology
Janeira, L. F., & Evansville, I. N. (1994) · Critical Care Medicine · doi:10.1097/00003246-199401000-00044 · Cited by 41
Research works printed but not linked: 3
- abbas, 2016, Model-based Clinical Trials for Implantable Cardiac Defibrillators, Technical Report
- 2007, The Compass - Technical Guide to Boston Scientific Cardiac Rhythm Management Products, Development Documentation
- 0, Ask The ICD
Cited by
Credibility assessment of in silico clinical trials for medical devices
Pathmanathan, P., Aycock, K., Badal, A., Bighamian, R., Bodner, J., Craven, B. A., & Niederer, S. (2024) · PLOS Computational Biology · doi:10.1371/journal.pcbi.1012289 · Cited by 62
Decision support for personalized therapy in implantable medical devices: A digital twin approach
Yang, H., & Jiang, Z. (2023) · Expert Systems with Applications · doi:10.1016/j.eswa.2023.122883 · Cited by 24
Self-citationReal-Time Decision Policies With Predictable Performance
Abbas, H., Rajeev Alur, A., Mamouras, K., Mangharam, R., & Rodionova, A. (2018) · Proceedings of the IEEE · doi:10.1109/jproc.2018.2853608 · Cited by 14
“The beat timing in the EGMs (in other words, the Boolean stream s) was generated by the heart model of [37] and [38].”
MethodSelf-citationQuantitative Regular Expressions for Arrhythmia Detection
Abbas, H., Rodionova, A., Mamouras, K., Bartocci, E., Smolka, S. A., & Grosu, R. (2018) · IEEE Transactions on Computational Biology and Bioinformatics · doi:10.1109/tcbb.2018.2885274 · Cited by 14
“The input stream was generated by the heart model of [3], [22], whose outputs had been validated for realism by cardiologists [22].”
MethodSelf-citationDigital-twin-based Online Parameter Personalization for Implantable Cardiac Defibrillators
Lai, M., Yang, H., Gu, J., Chen, X., & Jiang, Z. (2022) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc48229.2022.9871142 · Cited by 12
“The implementation of the algorithm was obtained using open information from clinical literature and was analyzed in [15].”
BackgroundMethodSelf-citation
Metadata
- Authors
- Jiang, Z., Abbas, H., Jang, K. J., Beccani, M., Liang, J., Dixit, S., & Mangharam, R.
- Affiliations
- University of Pennsylvania (US)
- ORCID
- Venue
- 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Publisher
- IEEE
- Published
- 2016-08
- Volume
- 2016
- Pages
- 169-172
- ISSN
- 2375-7477 (ISSN-L)
- OpenAlex
- W2537390084
- Cited by
- 16
- References
- 10
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 1 · 2023: 2 · 2022: 3 · 2021: 1 · 2019: 2 · 2018: 3 · 2017: 1
- FWCI
- 5.62
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Arrhythmias and Treatments
- Keywords
- implantable cardioverter-defibrillator, in silico clinical trials, arrhythmia detection algorithms, ventricular tachyarrhythmia, supraventricular tachyarrhythmias, clinical trial design
- MeSH terms
- Algorithms, Arrhythmias, Cardiac, Clinical Trials as Topic, Computer Simulation, Heart, Humans, Models, Cardiovascular, Tachycardia, Supraventricular, Defibrillators, Implantable, Tachycardia, Ventricular
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.