Analysis and detection of motion artifact in photoplethysmographic data using higher order statistics
Abstract
In this work, we present novel methods for detecting the presence of motion artifact in photoplethysmographic (PPG) measurements based on higher order statistical information present in the data. We analyze both clean and corrupt PPG data in the time and frequency domains. In the time domain, skew and kurtosis measures of the signal are used as distinguishing metrics between clean and motion- corrupted data. In the frequency domain, the presence of random components due to motion artifact is analyzed using a frequency domain kurtosis measure. Additionally, bispectral analysis of PPG data indicates the presence of strong quadratic phase coupling (QPC) and more specifically self coupling in the case of clean PPG data. Though quadratic phase coupling is found in data corrupted by motion artifact, the self coupling feature is absent. A Neyman-Pearson (NP) detection rule is formulated for each of the measures. Additionally, treating each of the measures as observations from independent sensors, the Varshney-Chair rule is used to fuse individual decisions to form a global system decision.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Krishnan et al. discovered that skew and kurtosis measurements in the time domain were higher when there was corruption, and frequency-domain kurtosis was smaller when there was corruption.
“They discovered skew and kurtosis measurements in the time domain to be higher when there was corruption, and the frequency-domain kurtosis was smaller when there was corruption.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [173]
Quadratic phase was present in corrupted signals.
“The quadratic phase was present in corrupted signals.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [173]
Cites
Mendel, J. M. (1991) · Proceedings of the IEEE · doi:10.1109/5.75086 · Cited by 1,816 · see citing works
Bispectrum estimation: A digital signal processing framework
Nikias, C. L., & Raghuveer, M. R. (1987) · Proceedings of the IEEE · doi:10.1109/proc.1987.13824 · Cited by 1,380 · see citing works
Optimal Data Fusion in Multiple Sensor Detection Systems
Chair, Z., & Varshney, P. K. (1986) · IEEE Transactions on Aerospace and Electronic Systems · doi:10.1109/taes.1986.310699 · Cited by 1,140 · see citing works
Use of the kurtosis statistic in the frequency domain as an aid in detecting random signals
Dwyer, R. F. (1984) · IEEE Journal of Oceanic Engineering · doi:10.1109/joe.1984.1145602 · Cited by 116 · see citing works
Higher-order spectral analysis of burst patterns in EEG
Muthuswamy, J., Sherman, D. L., & Thakor, N. V. (1999) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.736762 · Cited by 107 · see citing works
Yao, J., & Warren, S. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1617256 · Cited by 83 · see citing works
Self-citationWhat should we say about the kurtosis?
Mansour, A., & Jutten, C. (1999) · IEEE Signal Processing Letters · doi:10.1109/97.803435 · Cited by 47 · see citing works
An Improved Pre-processing Approach for Photoplethysmographic Signal
Weng, J., Ye, Z., & Weng, J. (2005) · doi:10.1109/iembs.2005.1616337 · Cited by 29 · see citing works
Independent component analysis and high-order statistics for automatic artifact rejection
Mammone, N., & Morabito, F. C. (2006) · IEEE International Joint Conference on Neural Networks (IJCNN) · doi:10.1109/ijcnn.2005.1556286 · Cited by 17 · see citing works
Pilgram, B., Castiglioni, P., Parati, G., & Rienzo, M. D. (1997) · Biological Cybernetics · doi:10.1007/s004220050341 · Cited by 8 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Dynamic time warping and machine learning for signal quality assessment of pulsatile signals
Li, Q., & Clifford, G. D. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/9/1491 · Cited by 274 · see citing works
Signal quality measures for pulse oximetry through waveform morphology analysis
Sukor, J. A., Redmond, S. J., & Lovell, N. H. (2011) · Physiological Measurement · doi:10.1088/0967-3334/32/3/008 · Cited by 166 · see citing works
Atrial fibrillation detection from wrist photoplethysmography signals using smartwatches
Bashar, S. K., Han, D., Hajeb-Mohammadalipour, S., Ding, E., Whitcomb, C., McManus, D. D., & Chon, K. H. (2019) · Scientific Reports · doi:10.1038/s41598-019-49092-2 · Cited by 120 · see citing works
Couceiro, R., Carvalho, P., Paiva, R. P., Henriques, J., & Muehlsteff, J. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346497 · Cited by 33 · see citing works
- Authors
- Krishnan, R., Natarajan, B., & Warren, S.
- Affiliations
- Kansas State University (US)
- Venue
- 2008 IEEE International Conference on Acoustics, Speech and Signal Processing
- Publisher
- IEEE
- Published
- 2008-03
- Pages
- 613-616
- ISSN
- 1520-6149 (ISSN-L)
- OpenAlex
- W1992974489
- Cited by
- 45
- References
- 10
- Citations by year
- 2025: 4 · 2024: 5 · 2023: 2 · 2022: 4 · 2021: 2 · 2020: 4 · 2019: 4 · 2018: 2 · 2017: 2 · 2016: 2 · 2015: 1 · 2014: 2 · 2013: 1 · 2012: 3
- FWCI
- 5.07
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- IEEE International Conference on Acoustics Speech and Signal Processing (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- motion artifact detection, photoplethysmography, higher-order statistics, skewness, kurtosis, bispectral analysis, quadratic phase coupling, self-coupling, Neyman-Pearson detection, decision fusion, PPG signal quality
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review