Abstract
This paper reviews metrology aspects of terahertz time-domain spectroscopy. Noise is discussed together with its effects on system performance, including frequency-dependent dynamic range, signal-to noise ratio, and the measurement bandwidth. Sources of uncertainties are addressed, as are measurement errors arising from system design and from interactions with the object under test. Finally, techniques of system calibration are described. Commonly encountered issues are addressed and simple practical solutions suggested. Existing literature on the subject is reviewed.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“produces artifacts that can degrade the overall measurement quality and yield misleading results [4-13].”
From Maximum-likelihood parameter estimation in terahertz time-domain spectroscopy (2021) · cites this paper as [6]
In the citing paper
“Such structured, signal-dependent, time-domain noise is common in THz-TDS waveform measurements”
From Maximum-likelihood parameter estimation in terahertz time-domain spectroscopy (2021) · cites this paper as [6]
In the citing paper
“leads to well-known ambiguities in defining the signal-to-noise ratio and dynamic range for them [5,6,8].”
From Maximum-likelihood parameter estimation in terahertz time-domain spectroscopy (2021) · cites this paper as [6]
In the citing paper
“The Fourier transform scrambles the THz-TDS measurement noise, which is more naturally represented in the time domain”
From Maximum-likelihood parameter estimation in terahertz time-domain spectroscopy (2021) · cites this paper as [6]
In the citing paper
“Errors in the positioning of the optical delay line lead to more complex distortions.”
From Correction of Optical Delay Line Errors in Terahertz Time-Domain Spectroscopy (2019) · cites this paper as [12]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Data Reduction and Error Analysis for the Physical Sciences
Bott, D. L. (1972) · Journal of Quality Technology · doi:10.1080/00224065.1972.11980528 · Cited by 658
Dynamic range in terahertz time-domain transmission and reflection spectroscopy
Jepsen, P. U., & Fischer, B. M. (2005) · Optics Letters · doi:10.1364/ol.30.000029 · Cited by 382 · cited 1×
“One of the most important papers dealing with the dynamic range of THz TDS systems and its implications on the bandwidth was published by Jepsen and Fischer [6].”
BackgroundHan, P. Y., Tani, M., Usami, M., Kono, S., Kersting, R., & Zhang, X.-C. (2001) · Journal of Applied Physics · doi:10.1063/1.1343522 · Cited by 259
Off-axis properties of silicon and quartz dielectric lens antennas
Filipovic, D. F., Gauthier, G. P., Raman, S., & Rebeiz, G. M. (1997) · IEEE Transactions on Antennas and Propagation · doi:10.1109/8.575618 · Cited by 238
Design and performance of a THz emission and detection setup based on a semi-insulating GaAs emitter
Zhao, G., Schouten, R. N., van der Valk, N., Wenckebach, W. T., & Planken, P. C. M. (2002) · Review of Scientific Instruments · doi:10.1063/1.1459095 · Cited by 233
Uncertainty in terahertz time-domain spectroscopy measurement
Withayachumnankul, W., Fischer, B. M., Lin, H., & Abbott, D. (2008) · Journal of the Optical Society of America B · doi:10.1364/josab.25.001059 · Cited by 181
Naftaly, M., & Dudley, R. (2009) · Optics Letters · doi:10.1364/ol.34.001213 · Cited by 174 · cited 2×
“The SNR and DR are defined (with respect to amplitude) as [4]”
BackgroundSelf-citationStrom, U., & Taylor, P. C. (1977) · Physical Review B · doi:10.1103/physrevb.16.5512 · Cited by 172
Influence of substrate-lens design in terahertz time-domain spectroscopy
Van Rudd, J., & Mittleman, D. M. (2002) · Journal of the Optical Society of America B · doi:10.1364/josab.19.000319 · Cited by 159
Influence of noise on the characterization of materials by terahertz time-domain spectroscopy
Duvillaret, L., Garet, F., & Coutaz, J.-L. (2000) · Journal of the Optical Society of America B · doi:10.1364/josab.17.000452 · Cited by 151
Research works printed but not linked: 2
- guenther, 1990, Modern Optics
- 0, The HITRAN Database of Gas Absorption Spectra
Cited by
Fundamentals of Measurement in Terahertz Time-Domain Spectroscopy
Withayachumnankul, W., & Naftaly, M. (2014) · Journal of Infrared, Millimeter, and Terahertz Waves · doi:10.1007/s10762-013-0042-z · Cited by 288
Self-citationThe progress and perspectives of terahertz technology for diagnosis of neoplasms: a review
Zaytsev, K. I., Dolganova, I. N., Chernomyrdin, N. V., Katyba, G. M., Gavdush, A. A., Cherkasova, O. P., Komandin, G. A., Shchedrina, M. A., Khodan, A. N., Ponomarev, D. S., et al. (2020) · Journal of Optics · doi:10.1088/2040-8986/ab4dc3 · Cited by 205
“The most common ones are signal apodization (windowing) and wavelet-domain denoising, which aim to distinguish between the useful part of the signal and random or systematic noises [205, 206].”
BackgroundTerahertz radiation and the skin: a review
Nikitkina, A. I., Bikmulina, P. Y., Gafarova, E. R., Kosheleva, N. V., Efremov, Y. M., Bezrukov, E. A., Butnaru, D. V., Dolganova, I. N., Chernomyrdin, N. V., Cherkasova, O. P., et al. (2021) · Journal of Biomedical Optics · doi:10.1117/1.jbo.26.4.043005 · Cited by 173
State-of-the-art in terahertz sensing for food and water security – A comprehensive review
Ren, A., Zahid, A., Fan, D., Yang, X., Imran, M. A., Alomainy, A., & Abbasi, Q. H. (2019) · Trends in Food Science & Technology · doi:10.1016/j.tifs.2019.01.019 · Cited by 165
High harmonic generation in undoped graphene: Interplay of inter- and intraband dynamics
Al-Naib, I. A. I., Sipe, J. E., & Dignam, M. M. (2014) · Physical Review B · doi:10.1103/physrevb.90.245423 · Cited by 154
Metadata
- Authors
- Naftaly, M.
- Affiliations
- National Physical Laboratory (GB)
- Corresponding author
- Naftaly, Mira
- ORCID
- Naftaly, Mira
- Venue
- IEEE Sensors Journal
- Published
- 2013-01
- Volume
- 13
- Issue
- 1
- Pages
- 8-17
- ISSN
- 1530-437X (ISSN-L)
- 1558-1748
- OpenAlex
- W1977261100
- Cited by
- 76
- References
- 34
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 3 · 2023: 4 · 2022: 9 · 2021: 6 · 2020: 3 · 2019: 9 · 2018: 6 · 2017: 7 · 2016: 9 · 2015: 5 · 2014: 8 · 2013: 3
- FWCI
- 3.27
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Sensors Journal (publishedVersion)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Terahertz technology and applications
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Microwave and Dielectric Measurement Techniques
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced Electrical Measurement Techniques
- Keywords
- terahertz time-domain spectroscopy, metrology, measurement noise, dynamic range, signal-to-noise ratio, measurement bandwidth, measurement uncertainty, measurement error, system calibration, system design
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Metrology Issues and Solutions in THz Time-Domain Spectroscopy: Noise, Errors, Calibration. https://app.cochat.ai/papers/10.1109/jsen.2012.2208624
- The paper
- Naftaly, M. (2013). Metrology Issues and Solutions in THz Time-Domain Spectroscopy: Noise, Errors, Calibration. IEEE Sensors Journal. https://doi.org/10.1109/jsen.2012.2208624
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.