Implantable RF Medical Devices: The Benefits of High-Speed Communication and Much Greater Communication Distances in Biomedical Applications
Abstract
In the early ages of implantable devices, radio frequency (RF) technologies were not commonplace due to the challenges stemming from the inherent nature of biological tissue boundaries. As technology improved and our understanding matured, the benefit of RF in biomedical applications surpassed the implementation challenges and is thus becoming more widespread. The fundamental challenge is due to the significant electromagnetic (EM) effects of the body at high frequencies. The EM absorption and impedance boundaries of biological tissue result in significant reduction of power and signal integrity for transcutaneous propagation of RF fields. Furthermore, the dielectric properties of the body tissue surrounding the implant must be accounted for in the design of its RF components, such as antennas and inductors, and the tissue is often heterogeneous and the properties are highly variable. Additional challenges for implantable applications include the need for miniaturization, power minimization, and often accounting for a conductive casing due to biocompatibility and hermeticity requirements [1]?[3]. Today, wireless technologies are essentially a must have in most electrical implants due to the need to communicate with the device and even transfer usable energy to the implant [4], [5]. Low-frequency wireless technologies face fewer challenges in this implantable setting than its higher frequency, or RF, counterpart, but are limited to much lower communication speeds and typically have a very limited operating distance. The benefits of high-speed communication and much greater communication distances in biomedical applications have spawned numerous wireless standards committees, and the U.S. Federal Communications Commission (FCC) has allocated numerous frequency bands for medical telemetry as well as those to specifically target implantable applications. The development of analytical models, advanced EM simulation software, and representative RF human phantom recipes has significantly facilitated design and optimization of RF components for implantable applications.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Radio Frequency (RF) technology employs far-field electromagnetic waves in the MHz and GHz range to wirelessly transmit energy and data over greater distances to brain implants differentiating it from the near-field magnetic fields utilized in inductive coupling and the intermediate-range operations of Midfield Transfer (MDF).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Chow et al., 2013
In the citing paper
“An external transmitter generates these waves, captured by an antenna connected to an internal receiver in the implant, where it is converted to electrical power (Chow et al., 2013; Singer and Robinson, 2021).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Chow et al., 2013
Citations
Cites
Yee, K. (1966) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.1966.1138693 · Cited by 15,008
Advanced engineering electromagnetics
Balanis, C. A. (1989) · Cited by 7,027
Gabriel, S., Lau, R.-W.-M., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/002 · Cited by 4,487
Gabriel, S., Lau, R. W., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/003 · Cited by 4,175
The dielectric properties of biological tissues: I. Literature survey
Gabriel, C., Gabriel, S., & Corthout, E. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/001 · Cited by 3,168
The Finite Difference Time Domain Method for Electromagnetics
Kunz, K. S., & Luebbers, R. J. (2018) · CRC Press eBooks · doi:10.1201/9780203736708 · Cited by 2,530
Implanted Antennas Inside a Human Body: Simulations, Designs, and Characterizations
Kim, J., & Rahmat‐Samii, Y. (2004) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2004.832018 · Cited by 758
Design of Implantable Microstrip Antenna for Communication With Medical Implants
Soontornpipit, P., Furse, C. M., & Chung, Y. C. (2004) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2004.831976 · Cited by 473
Bohorquez, J. L., Chandrakasan, A. P., & Dawson, J. L. (2009) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2009.2014728 · Cited by 246
Fully Wireless Implantable Cardiovascular Pressure Monitor Integrated with a Medical Stent
Chow, E. Y., Chlebowski, A. L., Chakraborty, S., Chappell, W. J., & Irazoqui, P. P. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2041058 · Cited by 223
Self-citationSimulated biological materials for electromagnetic radiation absorption studies
Hartsgrove, G. W., Kraszewski, A. W., & Surowiec, A. (1987) · Bioelectromagnetics · doi:10.1002/bem.2250080105 · Cited by 203
A brief history of cardiac pacing.
Nelson, G. (1993) · PubMed · Cited by 163
A Sub-100 $\mu$W MICS/ISM Band Transmitter Based on Injection-Locking and Frequency Multiplication
Pandey, J., & Otis, B. P. (2011) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2118030 · Cited by 161
Easy-to-Swallow Wireless Telemetry
Yuce, M. R., & Dissanayake, T. (2012) · IEEE Microwave Magazine · doi:10.1109/mmm.2012.2205833 · Cited by 144
A Miniature-Implantable RF-Wireless Active Glaucoma Intraocular Pressure Monitor
Chow, E. Y., Chlebowski, A. L., & Irazoqui, P. P. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2081364 · Cited by 138
Self-citationArtificial pacemaker for treatment of Adams-Stokes syndrome and slow heart rate
Elmgvist, R., Landegren, J., Pettersson, S. O., Senning, A., & William‐Olsson, G. (1963) · American Heart Journal · doi:10.1016/0002-8703(63)90239-4 · Cited by 105
Evaluation of Cardiovascular Stents as Antennas for Implantable Wireless Applications
Chow, E. Y., Ouyang, Y., Beier, B. L., Chappell, W. J., & Irazoqui, P. P. (2009) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2009.2029954 · Cited by 101
Self-citationFar-Field RF Powering of Implantable Devices: Safety Considerations
Bercich, R. A., Duffy, D. R., & Irazoqui, P. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2246787 · Cited by 95
Self-citationChow, E. Y., Yang, C., Ouyang, Y., Chlebowski, A. L., Irazoqui, P. P., & Chappell, W. J. (2011) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2011.2144551 · Cited by 67
Self-citationDETUNING STUDY OF IMPLANTABLE ANTENNAS INSIDE THE HUMAN BODY
Vidal, N., Curto, S., Lopez-Villegas, J. M., Sieiro, J. J., & Ramos, F. M. (2012) · Electromagnetic waves · doi:10.2528/pier11120515 · Cited by 64
Mixed-signal integrated circuits for self-contained sub-cubic millimeter biomedical implants
Chow, E. Y., Chakraborty, S., Chappell, W. J., & Irazoqui, P. P. (2010) · Digest of technical papers/Digest of technical papers - IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2010.5433933 · Cited by 63
Self-citationA 490uW fully MICS compatible FSK transceiver for implantable devices
Bae, J., Cho, N., & Yoo, H. (2009) · Symposium on VLSI Circuits · Cited by 53
Implantable Wireless Telemetry Boards for In Vivo Transocular Transmission
Chow, E. Y., Yang, C., Chlebowski, A. L., Moon, S.-W., Chappell, W. J., & Irazoqui, P. P. (2008) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2008.2007338 · Cited by 48
Self-citationPaulsen, K. D., Jia, X., & Sullivan, J. M. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.245615 · Cited by 46
Chow, E. Y., Beier, B. L., Francino, A., Chappell, W. J., & Irazoqui, P. P. (2009) · Journal of Interventional Cardiology · doi:10.1111/j.1540-8183.2009.00483.x · Cited by 38
Self-citationWave-propagation from medical implants-influence of body shape on radiation pattern
Johansson, A. J. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2002.1106454 · Cited by 27
Design of an Ultra Wideband System for In-Body Wireless Communications
Wang, J., & Su, D. (2006) · Asia-Pacific Conference on Environmental Electromagnetics · doi:10.1109/ceem.2006.258019 · Cited by 27
UWB for in-body medical implants: A viable option
Ghildiyal, A., Amara, K., Molin, R. D., Godara, B., Amara, A., & Shevgaonkar, R. K. (2010) · IEEE International Conference on Ultra-Wideband (ICUWB) · doi:10.1109/icuwb.2010.5615732 · Cited by 19
Chow, E. Y., Beier, B. L., Ouyang, Y., Chappell, W. J., & Irazoqui, P. P. (2009) · IEEE MTT-S International Microwave Symposium digest · doi:10.1109/mwsym.2009.5165947 · Cited by 18
Self-citationA fast and accurate automatic matching network designed for ultra low power medical applications
Chan, W. P. F., de Foucauld, E., Vincent, P., Hameau, F., Morche, D., Delaveaud, C., Molin, R. D., Pons, P., Pierquin, R., & Kerhervé, É. (2009) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2009.5117838 · Cited by 11
High Data-Rate 6.7 GHz Wireless ASIC Transmitter for Neural Prostheses
Chow, E. Y., Kahn, A. R., & Irazoqui, P. P. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353867 · Cited by 9
Self-citationMiniature antenna for RF telemetry through ocular tissue
Chow, E. Y., Yang, C., Chlebowski, A. L., Chappell, W. J., & Irazoqui, P. P. (2008) · IEEE MTT-S International Microwave Symposium digest · doi:10.1109/mwsym.2008.4633303 · Cited by 8
Self-citationChow, E. Y., Ha, D., Lin, T.-Y., de Vries, W. N., John, S. W. M., Chappell, W. J., & Irazoqui, P. P. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5627324 · Cited by 3
Self-citation
Flagged first, then by citations
3 of the 36 references couldn’t be matched to a known paper.
Research works printed but not linked: 25
- knapp, 2011, ANSYS Inc. request for waiver of 47 C.F.R. sections 1.1307(b) (2) of commission rules; declaratory ruling concerning sections 1.1307(b)(2) of commission rules (DA 11-192), Federal Communications Commission Tech Rep 11-192
- 2011, § 95.1221-RF Exposure,” in “Code of Federal Regulations, Title 47—Telecommunication, Chapter 1—FCC, Subchapter D—Safety and Special Radio Services, Part 95—Personal Radio Services, Subpart I—Medical Device RadioCommunication Service, Tech Rep 95 1221
- means, 2001, Evaluating compliance with FCC guidelines for human exposure to radiofrequency electromagnetic fields, Technical report Federal Communications Commission Office of Engineering and Technology
- 2012, § 95.628 MedRadio transmitters in the 413-419 MHz, 426-432 MHz, 438-444, MHz, and 451-457 MHz and 2360-2400 MHz bands,” in “Code of Federal Regulations, Title 47—Telecommunication, Chapter 1—FCC, Subchapter D—Safety and Special Radio Services, Part 95—Personal Radio Services, Subpart I—Medical Device Radiocommunications Service, Tech Rep 95 628
- 2010, Z S Inc
- 2012, § 95.627 MedRadio transmitters in the 401-406 MHz band,” in “Code of Federal Regulations, Title 47—Telecommunication, Chapter1—FCC, Subchapter D—Safety and Special Radio Services, Part 95—Personal Radio Services, Subpart E—Technical Regulations, Tech Rep 95 627
- 0, Boston Scientific Corp
- 2011, Axis Research Mind, Global cardiac rhythm management (CRM) devices-market growth analysis 2009-2015
- greatbatch, 1962, Medical cardiac pacemaker, U S Patent 3 057 356
- 2002, FCC 02-48, ET Docket 98-153, Tech Rep 98-153
- 1999, Amendment of Parts 2 and 95 of the Commission's Rules to Establish a Medical Implant Communications Service in the 402-405 MHz Band, Technical Report 99–9
- 1999, Subpart I—medical implant communications,” in “Code of Federal Regulations, Title 47—Telecommunication, Chapter 1—FCC, Subchapter D-Safety and Special Radio Services, Part 95—Personal Radio Services, Federal Communications Commission
- 2013, Biotronik SE and Co KG
- 2013, Medtronic Inc
- 2013, St Jude Medical Inc
- 2009, ET Docket No. 06-135, ” in “FCC Adopted Rules for New Advanced Medical Technologies, Tech Rep 06-123
- 2009, Subpart I— medical device radiocommunication service,” in “Code of Federal Regulations, Title 47—Telecommunication, Chapter 1—FCC, Subchapter D—Safety and Special Radio Services, Part 95—Personal Radio Services, Tech Rep
- 2011, FCC 11-176, ET Docket No. 09-36,” in “Amendment of Parts 2 and 95 of the Commission‘s Rules to Provide Additional Spectrum for the Medical Device Radiocommunication Service in the 413-457 MHz Band, Tech Rep 09-36 Washington D C
- 0, Documents of the international radio conference—Doc. no. 1–100, Proc Int Radio Conf, 18e
- 2012, § 95.628 Medradio transmitters in the 413-419 MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz and 2360-2400 MHz bands, in “Code of Federal Regulations Title 47—Telecommunication Chapter 1—FCC Subchapter D-Safety and Special Radio Services Part 95—Personal Radio Services Subpart E—Technical Regulations ” Tech Rep 95 628
- 2000, § 95.630 WMTS transmitter frequencies, Code of Federal Regulations Title 47—Telecommunication Chapter 1—FCC Subchapter D—Safety and Special Radio Services Part 95—Personal Radio Services Subpart E —Technical Regulations ” Tech Rep 95 630
- 2011, Human torso simulator and testing technique, publication number: 617965, Federal Communications Commission
- 2012, Cuming Microwave
- 2010, Investigation of the spectrum requirements for advanced medical technologies, amendment of parts 2 and 95 of the commission's rules to establish the medical device radiocommunication service at 401-402 and 405-206 MHz, Tech Rep FCC 10-128
- 1999, § 95.639 maximum transmitter power,” in “Code of Federal Regulations, Title 47—Telecommunication, Chapter 1—FCC, Subchapter D—Safety and Special Radio Services, Part 95—Personal Radio Services, Subpart E—Technical Regulations, Tech Rep 95 639
Cited by
Sorted by citations
Soft Microfluidic Assemblies of Sensors, Circuits, and Radios for the Skin
Xu, S., Zhang, Y., Jia, L., Mathewson, K. E., Jang, K.-I., Kim, J., Fu, H., Huang, X., Chava, P., Wang, R., et al. (2014) · Science · doi:10.1126/science.1250169 · Cited by 1,199
Wireless power transfer techniques for implantable medical devices: a review
Khan, S. R., Pavuluri, S. K., Cummins, G., & Desmulliez, M. P. Y. (2020) · Sensors · doi:10.3390/s20123487 · Cited by 355
“Radio frequency (RF) modeling, different frequency bands and a brief study on the RF implantable device testing inside phantoms, ex vivo and in vivo were presented in [47].”
BackgroundA Review of In-Body Biotelemetry Devices: Implantables, Ingestibles, and Injectables
Kiourti, A., & Nikita, K. S. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2668612 · Cited by 276
Compact EBG-Backed Planar Monopole for BAN Wearable Applications
Abbasi, M. A. B., Nikolaou, S., Antoniades, M. A., Stevanović, M. N., & Vryonides, P. (2016) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2016.2635588 · Cited by 184
Das, R., & Yoo, H. (2017) · IEEE transactions on microwave theory and techniques · doi:10.1109/tmtt.2017.2647945 · Cited by 142
Metadata
- Authors
- Chow, E., Morris, M., & Irazoqui, P.
- Affiliations
- Purdue University West Lafayette (US)
- ORCID
- Irazoqui, Pedro
- Venue
- IEEE Microwave Magazine
- Published
- 2013-06
- Volume
- 14
- Issue
- 4
- Pages
- 64-73
- ISSN
- 1527-3342 (ISSN-L)
- OpenAlex
- W1970635841
- Cited by
- 111
- References
- 36
- Citations by year
- 2026: 4 · 2025: 6 · 2024: 14 · 2023: 5 · 2022: 4 · 2021: 4 · 2020: 7 · 2019: 9 · 2018: 13 · 2017: 12 · 2016: 7 · 2015: 12 · 2014: 9 · 2013: 5
- FWCI
- 6.07
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Microwave Magazine (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Physical Sciences › Computer Science › Computer Networks and Communications › RFID technology advancements
- Keywords
- high-speed wireless communication, biotelemetry, tissue dielectric properties, electromagnetic wave absorption, impedance matching, RF power transmission
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review