Heating of the eye by a retinal prosthesis: modeling, cadaver and in vivo study
Abstract
In order to develop retinal implants with a large number of electrodes, it is necessary to ensure that they do not cause damage to the neural tissue by the heat that the electrical circuits generate. Knowledge of the amount of power that induces thermal damage will assist in development of power budgets for implants, which has a significant effect upon the design of the prostheses circuitry. In this study, temperatures were measured at multiple locations on the retina while the retina was heated in cadaver and in vivo preparations using a variety of prosthesis implantation sites. A finite element thermal model of the cat eye was also created and validated by the cadaver and in vivo tests, allowing for a much larger spectrum of thermal influences to be evaluated without additional animal experimentation. To ensure that retinal tissue temperatures are not increased by more than 2 °C, a 5 mm × 5 mm, suprachoroidally implanted heating element must not dissipate more than 135 mW (5.4 mW/mm (2)).
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Another advantage of epiretinal implants is the facilitation of heat dissipation via the vitreous.
“Moreover, another advantage of epiretinal implants is the facilitation of heat dissipation via the vitreous [42].”
From Advances in visual prostheses: engineering and biological challenges (2022) · cites this paper as [42]
Cites
Global data on visual impairment in the year 2002.
(2004) · Europe PMC (PubMed Central) · Cited by 4,157 · see citing works
Fundamentals of Temperature, Pressure, and Flow Measurements
Benedict, R. P. (1984) · Wiley eBooks · doi:10.1002/9780470172698 · Cited by 330 · see citing works
Anatomy and Histology of the Eye and Orbit in Domestic Animals.
Evans, J. W. (1961) · Archives of Neurology · doi:10.1001/archneur.1961.00450180115017 · Cited by 256 · see citing works
Visual Performance Using a Retinal Prosthesis in Three Subjects With Retinitis Pigmentosa
Yanai, D., Weiland, J. D., Mahadevappa, M., Greenberg, R. J., Fine, I., & Humayun, M. S. (2007) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2007.01.027 · Cited by 233 · see citing works
Anatomy and histology of the eye and orbit in domestic animals.
Prince, J., Diesem, C. D., Eglitis, I., & Ruskell, G. L. (1960) · Medical Entomology and Zoology · Cited by 233 · see citing works
Thermal Impact of an Active 3-D Microelectrode Array Implanted in the Brain
Kim, S., Tathireddy, P., Normann, R. A., & Solzbacher, F. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.908429 · Cited by 215 · see citing works
Thermal effects of bioimplants
Lazzi, G. (2005) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2005.1511503 · Cited by 193 · see citing works
Effect of Pulse Duration on Size and Character of the Lesion in Retinal Photocoagulation
Jain, A. (2008) · Archives of Ophthalmology · doi:10.1001/archophthalmol.2007.29 · Cited by 184 · see citing works
Facial Recognition Using Simulated Prosthetic Pixelized Vision
Thompson, R. W., Barnett, G. D., Humayun, M. S., & Dagnelie, G. (2003) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.03-0341 · Cited by 176 · see citing works
FEM simulation of the eye structure with bioheat analysis
Ng, E. Y. K., & Ooi, E. H. (2006) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2006.04.001 · Cited by 134 · see citing works
The computation of temperature rises in the human eye induced by infrared radiation
Scott, J. A. (1988) · Physics in Medicine and Biology · doi:10.1088/0031-9155/33/2/004 · Cited by 123 · see citing works
Fallon, J. B., Irvine, D. R. F., & Shepherd, R. K. (2008) · The Journal of Comparative Neurology · doi:10.1002/cne.21886 · Cited by 112 · see citing works
Thermal Elevation in the Human Eye and Head Due to the Operation of a Retinal Prosthesis
Gosalia, K. C., Weiland, J. D., Humayun, M. S., & Lazzi, G. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.827548 · Cited by 95 · see citing works
Shivdasani, M. N., Luu, C. D., Cicione, R., Fallon, J. B., Allen, P. J., Leuenberger, J., Suaning, G. J., Lovell, N. H., Shepherd, R. K., & Williams, C. E. (2010) · Journal of Neural Engineering · doi:10.1088/1741-2560/7/3/036008 · Cited by 86 · see citing works
Flyckt, V. M. M., Raaymakers, B. W., & LAGENDIJK, J. J. W. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/19/018 · Cited by 85 · see citing works
Ocular surface temperature: A 3D FEM prediction using bioheat equation
Ng, E. Y. K., & Ooi, E. H. (2006) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2006.08.023 · Cited by 56 · see citing works
Piyathaisere, D. V., Margalit, E., Chen, S., Shyu, J., D’Anna, S. A., Weiland, J. D., Grebe, R. R., Grebe, L., Fujii, G. Y., Kim, S. Y., et al. (2003) · Ophthalmic surgery, lasers & imaging retina · doi:10.3928/1542-8877-20030301-07 · Cited by 43 · see citing works
Sailer, H., Shinoda, K., Blatsios, G., Köhler, K., Bondzio, L., Zrenner, E., & Gekeler, F. (2007) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-006-0513-1 · Cited by 31 · see citing works
Improved Heat Transfer Modeling of the Eye for Electromagnetic Wave Exposures
Hirata, A. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.893492 · Cited by 22 · see citing works
Thermal heating of a retinal prosthesis: Thermal model and in-vitro study
Opie, N. L., Burkitt, A. N., Hamish Meffin, A., & Grayden, D. B. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5626670 · Cited by 12 · see citing works
Thermal safety of a retinal prosthesis
Opie, N. L. (2011) · Minerva Access (University of Melbourne) · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Implantable devices: issues and challenges
Bazaka, K., & Jacob, M. (2012) · Electronics · doi:10.3390/electronics2010001 · Cited by 358 · see citing works
Retinal stimulation strategies to restore vision: fundamentals and systems
Yue, L., Weiland, J. D., Roska, B., & Humayun, M. S. (2016) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2016.05.002 · Cited by 272 · see citing works
Stochastic circuits for real-time image-processing applications
Alaghi, A., Li, C., & Hayes, J. P. (2013) · ACM/IEEE Design Automation Conference (DAC) · doi:10.1145/2463209.2488901 · Cited by 262 · see citing works
Printable Ultrathin Metal Oxide Semiconductor-Based Conformal Biosensors
Rim, Y. S., Bae, S., Chen, H., Yang, J. L., Kim, J., Andrews, A. M., Weiss, P. S., Yang, Y., & Tseng, H. (2015) · ACS Nano · doi:10.1021/acsnano.5b05325 · Cited by 151 · see citing works
The development of neural stimulators: a review of preclinical safety and efficacy studies
Shepherd, R. K., Villalobos, J., Burns, O., & Nayagam, D. A. X. (2018) · Journal of Neural Engineering · doi:10.1088/1741-2552/aac43c · Cited by 81 · see citing works
Advances in visual prostheses: engineering and biological challenges
Borda, E., & Ghezzi, D. (2022) · Progress in Biomedical Engineering · doi:10.1088/2516-1091/ac812c · Cited by 44 · see citing works
Progress in artificial vision through suprachoroidal retinal implants
Bareket, L., Barriga‐Rivera, A., Zapf, M. P., Lovell, N. H., & Suaning, G. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6cbb · Cited by 34 · see citing works
Opie, N. L., Greferath, U., Vessey, K. A., Burkitt, A. N., Hamish Meffin, A., Grayden, D. B., & Fletcher, E. L. (2012) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.12-10600 · Cited by 33 · see citing works
Retinal Prosthetic Approaches to Enhance Visual Perception for Blind Patients
Shim, S., Eom, K., Jeong, J., & Kim, S. J. (2020) · Micromachines · doi:10.3390/mi11050535 · Cited by 34 · see citing works
Wang, J., Xie, H., Chung, T., Chan, L. L. H., & Pang, S. W. (2017) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2016.2634584 · Cited by 30 · see citing works
Lee, S., Jung, S. W., Ahn, J., Yoo, H. J., Oh, S. J., & Cho, D. (2014) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/24/7/075018 · Cited by 23 · see citing works
Oh, S., Ahn, J.-H., Lee, S., Ko, H., Seo, J. M., Goo, Y., & Cho, D. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2336799 · Cited by 23 · see citing works
Feasibility of a chronic, minimally invasive endovascular neural interface
Opie, N. L., Rind, G. S., John, S. E., Ronayne, S. M., Grayden, D. B., Burkitt, A. N., May, C. N., O'Brien, T. J., & Oxley, T. J. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591716 · Cited by 21 · see citing works
Laser Driven Miniature Diamond Implant for Wireless Retinal Prostheses
Ahnood, A., Cheriton, R., Bruneau, A., Belcourt, J. A., Ndabakuranye, J. P., Lemaire, W., Hilkes, R., Fontaine, R., Cook, J. P. D., Hinzer, K., et al. (2020) · Advanced Biosystems · doi:10.1002/adbi.202000055 · Cited by 18 · see citing works
Feasibility of Intraocular Projection for Treatment of Intractable Corneal Opacity
Shim, S. Y., Gong, S., Rosenblatt, M. I., Palanker, D., Al-Qahtani, A. F., Sun, M. G., Zhou, Q., Kanu, L. N., Chau, F. Y., & Yu, C. Q. (2019) · Cornea · doi:10.1097/ico.0000000000001852 · Cited by 17 · see citing works
Opie, N. L., & O'Brien, T. J. (2021) · Artificial Organs · doi:10.1111/aor.14007 · Cited by 12 · see citing works
Adaptive Thermal Management of Implantable Device
Chai, R., & Zhang, Y. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2879427 · Cited by 12 · see citing works
Development of a Magnetic Attachment Method for Bionic Eye Applications
Fox, K. E., Hamish Meffin, A., Burns, O., Abbott, C. J., Allen, P. J., Opie, N. L., McGowan, C. C., Yeoh, J., Ahnood, A., Luu, C. D., et al. (2015) · Artificial Organs · doi:10.1111/aor.12582 · Cited by 12 · see citing works
The mechanism of human color vision and potential implanted devices for artificial color vision
Zhang, B., Zhang, R., Zhao, J., Yang, J., & Xu, S. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2024.1408087 · Cited by 11 · see citing works
Opie, N. L., Ayton, L. N., Apollo, N. V., Ganesan, K., Guymer, R. H., & Luu, C. D. (2014) · Artificial Organs · doi:10.1111/aor.12287 · Cited by 9 · see citing works
Neuromodulation using electroosmosis
Kare, S. S., Rountree, C. M., Troy, J. B., Finan, J. D., & Saggere, L. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac00d3 · Cited by 9 · see citing works
Mechanical suitability of an endovascular braincomputer interface
Opie, N. L., Ronayne, S. M., Rind, G. S., Yoo, P. E., & Oxley, T. J. (2020) · International Winter Conference on Brain-Computer Interface (BCI) · doi:10.1109/bci48061.2020.9061643 · Cited by 7 · see citing works
Developments in the design and microfabrication of photovoltaic retinal implants
Kusumanchi, P., Keller, S. S., & Davidsen, R. S. (2024) · Micro and Nano Engineering · doi:10.1016/j.mne.2024.100289 · Cited by 6 · see citing works
Development and Experimental Basis for the Future of Prosthetic Vision
Allen, P. J., & Ayton, L. N. (2020) · Springer eBooks · doi:10.1007/978-981-15-7644-7_33 · Cited by 6 · see citing works
Chen, Y., & Tang, K. (2015) · International Symposium on VLSI Design, Automation and Test · doi:10.1109/vlsi-dat.2015.7114516 · Cited by 6 · see citing works
Thermal finite element analysis of localized hypothermia treatment of the human eye
Gongal, D., Thakur, S., Panse, A., Shankarrao, P., Stark, J. A., Hetling, J. R., Özgen, B., & Foster, C. D. (2022) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2022.103928 · Cited by 5 · see citing works
Heat exchange in the human eye: a review
Zadorozhnyy, O., Korol, A. R., Naumenko, V., Пасєчнікова, Н. В., & Butenko, L. L. (2022) · Oftalmologicheskii Zhurnal · doi:10.31288/oftalmolzh202265058 · Cited by 5 · see citing works
Characterization of an Electronic Corneal Prosthesis System
Shim, S. Y., Gong, S., Fan, V. H., Rosenblatt, M. I., Al-Qahtani, A. F., Sun, M. G., Zhou, Q., Kanu, L. N., Vieira, I. V., & Yu, C. Q. (2019) · Current Eye Research · doi:10.1080/02713683.2019.1708957 · Cited by 5 · see citing works
Introduction to biomaterials and implantable device design
Ivanova, E. P., Bazaka, K., & Crawford, R. J. (2014) · Elsevier eBooks · doi:10.1533/9781782422662.1 · Cited by 4 · see citing works
Optimized single pulse stimulation strategy for retinal implants
Savage, C. O., Grayden, D. B., Hamish Meffin, A., & Burkitt, A. N. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/1/016003 · Cited by 4 · see citing works
Retinal Prostheses: A Brief History
Yue, L., Weiland, J. D., & Humayun, M. S. (2017) · Essentials in ophthalmology/Essentials ophthalmology · doi:10.1007/978-3-319-67260-1_1 · Cited by 4 · see citing works
Thermal analysis of intraocular electronic display projector visual prosthesis
Gongal, D., Thakur, S., Panse, A., Pawar, R., Yu, C. Q., & Foster, C. D. (2020) · Numerical Heat Transfer Part A Applications · doi:10.1080/10407782.2020.1805230 · Cited by 4 · see citing works
Liu, C.-Y., Yang, F. F., Chung, C.-H., Yang, Z.-H., Chen, T., Yang, C., Yang, C., & Fan, L.-S. (2014) · Journal of Micro/Nanopatterning Materials and Metrology · doi:10.1117/1.jmm.13.1.013006 · Cited by 4 · see citing works
Axisymmetric Thermal Finite Element Analysis of Effects of Intraocular Projector in the Human Eye
Stark, J. A., Foster, C. D., & Yu, C. Q. (2021) · American Journal of Undergraduate Research · doi:10.33697/ajur.2020.035 · Cited by 4 · see citing works
Temperature Elevation in the Human Eye Due To Intraocular Projection Prosthesis Device
Gongal, D., Thakur, S., Panse, A., Stark, J. A., Yu, C. Q., & Foster, C. D. (2021) · Journal of Thermal Science and Engineering Applications · doi:10.1115/1.4050237 · Cited by 3 · see citing works
Liang, Y., Du, B., Nakamura, K., Wang, S., Inoue, B., Aruga, Y., Kino, H., Fukushima, T., Kiyoyama, K., & Tanaka, T. (2022) · IEICE Electronics Express · doi:10.1587/elex.19.20220363 · Cited by 3 · see citing works
Yousefi, T., & Kassiri, H. (2024) · IEEE Access · doi:10.1109/access.2024.3399711 · Cited by 3 · see citing works
Ermis, A., Lai, Y.-P., & Zhang, Y. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3020753 · Cited by 3 · see citing works
Yoon, Y. H., Yue, L., & Humayun, M. S. (2020) · Springer eBooks · doi:10.1007/978-981-15-7644-7_34 · Cited by 2 · see citing works
Asghar, S. A., & Mahadevappa, M. (2023) · IEEE Transactions on NanoBioscience · doi:10.1109/tnb.2023.3319084 · Cited by 2 · see citing works
Yousefi, T., Zoidl, G., & Kassiri, H. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3488713 · Cited by 2 · see citing works
Biswas, S., Das, S., & Mahadevappa, M. (2017) · IET Science Measurement & Technology · doi:10.1049/iet-smt.2016.0315 · Cited by 2 · see citing works
A Survey of Wireless Power Transfer Methods for Artificial Retina
Rosalinekavitha, B., & Malarkodi, B. (2019) · National Power Electronics Conference (NPEC) · doi:10.1109/npec47332.2019.9034753 · Cited by 2 · see citing works
Simulation-Informed Power Budget Estimate of a Fully-Implantable Brain–Computer Interface
Serrano-Amenos, C., Hu, F., Wang, P. T., Heydari, P., Do, A. H., & Nenadić, Z. (2024) · Annals of Biomedical Engineering · doi:10.1007/s10439-024-03528-7 · Cited by 2 · see citing works
Online Thermal Effect Modeling and Prediction of Implantable Devices
Chai, R., & Zhang, Y. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.3025874 · Cited by 1 · see citing works
Uluaydın, N. K., & Seker, S. S. (2022) · Sakarya University Journal of Science · doi:10.16984/saufenbilder.972989 · Cited by 1 · see citing works
Nonlinear Model Predictive Control for Thermal Management of Bio-Implants
Ermis, A., Lai, Y.-P., & Zhang, Y. (2022) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.23919/acc53348.2022.9867198 · Cited by 1 · see citing works
Neurostimulators for high-resolution artificial retina: ASIC design challenges and solutions
Jeong, H., Kim, J., Seo, J.-M., & Neviani, A. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/aca262 · Cited by 1 · see citing works
The Development of Visual Prosthetic Devices to Restore Vision to the Blind
Weiland, J. D., Humayun, M. S., & Calle, A. G. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00101-7 · Cited by 1 · see citing works
Thermal Performance of Class- FF Converter Used for Wireless Power Transfer in Retinal Implants
Mashhadi, I. A., Mazaheri‐Tehrani, E., Poorali, B., Pahlevani, M., & Pahlevani, H. (2019) · IEEE Wireless Power Transfer Conference · doi:10.1109/wptc45513.2019.9055588 · Cited by 1 · see citing works
Lee, P. J.-H., Law, M. K., & Bermak, A. (2017) · Analog Integrated Circuits and Signal Processing · doi:10.1007/s10470-017-0960-0
Current Advances in the Design of Retinal and Cortical Visual Prostheses
Bareket, L., Barriga‐Rivera, A., Rosenfeld, J. V., Lovell, N. H., & Suaning, G. J. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-812539-7.00005-2
Ermis, A., Zhou, M., Lai, Y.-P., & Zhang, Y. (2020) · IEEE Conference on Control Technology and Applications (CCTA) · doi:10.1109/ccta41146.2020.9206293
Chaudhury, A. R. (2018) · Deep Blue (University of Michigan)
Hypothermia treatment for ocular ischemia: A multiphase porous thermal model of human eye
Gongal, D., Hetling, J. R., & Foster, C. D. (2025) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2025.104139
Adaptive Thermal Management of Implantable Medical Devices
Ermis, A., Lai, Y.-P., & Zhang, Y. (2025) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2025.3635695
Báčová, T., Báča, V., & Horák, Z. (2026) · Bratislavské lekárske listy/Bratislava medical journal · doi:10.1007/s44411-026-00802-z
- Authors
- Opie, N. L., Burkitt, A. N., Meffin, H., & Grayden, D. B.
- Affiliations
- The University of Melbourne (AU)
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2012-02
- Volume
- 59
- Issue
- 2
- Pages
- 339-345
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- OpenAlex
- W1997184856
- Cited by
- 59
- References
- 21
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 5 · 2023: 1 · 2022: 7 · 2021: 5 · 2020: 9 · 2019: 5 · 2018: 5 · 2017: 5 · 2016: 2 · 2015: 3 · 2014: 5 · 2013: 1 · 2012: 3
- FWCI
- 1.19
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Retinal Development and Disorders
- Keywords
- retinal prosthesis, thermal damage, retinal temperature, power budget, finite element thermal model, cat eye, cadaveric study, in vivo evaluation
- MeSH terms
- Animals, Cadaver, Cats, Electrodes, Implanted, Hot Temperature, Models, Biological, Prosthesis Design, Retina, Finite Element Analysis, Visual Prosthesis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review