CoChat

Error Analysis and Accuracy Calibration Method of U-Tube Coriolis Mass Flowmeter Under Pulsating Flow

Yang, Hai, Zhang, Lizao, Li, Li, Liang, Haibo, Zou, Jialing · 2021 · IEEE Transactions on Instrumentation and Measurement
Cited by 391

The DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.

Abstract

Coriolis mass flowmeter (CMF) is a kind of flow measurement instrument which can directly measure the high precision transient mass flow parameters. And the vibration characteristic of the U-shaped measuring tube inside is one of the important factors that determine the measuring accuracy. The pulsating flow through the measuring tube will lead to the motion component except the main vibration frequency, which will affect the phase difference calculation and reduce the measurement accuracy of the mass flowmeter. This paper presents a method to improve the accuracy of U-tube CMF based on variable step size Least Mean Square (LMS) filter and Hilbert transform with interval-shifting under pulsating flow. Experimental work was conducted on a dynamic experimental platform of pulsating flow. Experimental results show that the stability and accuracy of the proposed algorithm are better than the traditional CMF phase difference calibration method. The mean time difference error is 9.0525μs, and the mean time difference relative error is 5.806%. The calibration effect is more than 88.0934% better than other traditional algorithms. It is verified that it has good error calibration effect for pulsating flow at various frequencies.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/tim.2021.3126366

Cites

6 of 6 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 6 references

Research works printed but not linked: 25

  1. 10.1016/j.flowmeasinst.2019.101613 · doi:10.1016/j.flowmeasinst.2019.101613
  2. 10.1016/j.flowmeasinst.2006.12.006 · doi:10.1016/j.flowmeasinst.2006.12.006
  3. tu, 2008, Method for CMF signal processing based on the recursive DTFT algorithm with negative frequency contribution, IEEE Trans Instrum Meas, 57, 2647, 10.1109/TIM.2008.925006 · doi:10.1109/tim.2008.925006
  4. 10.1109/ICIT.2016.7474833 · doi:10.1109/icit.2016.7474833
  5. 10.1155/2020/7960546 · doi:10.1155/2020/7960546
  6. zhang, 2017, Applications of digital signal processing technology in Coriolis mass flowmeter, Chin J Sci Instrum, 38, 2087
  7. tu, 2015, Error analysis of DFT or DTFT based phase difference measurement considering the negative frequency contribution, J Vib Shock, 34, 85
  8. 10.1007/s12647-012-0017-x · doi:10.1007/s12647-012-0017-x
  9. 10.1109/JSEN.2020.3042078 · doi:10.1109/jsen.2020.3042078
  10. 10.1109/CISCE.2019.00070 · doi:10.1109/cisce.2019.00070
  11. qi, 2016, Principle and application of mass flow meter, Instrumentation, 23, 44
  12. 10.1016/j.flowmeasinst.2010.07.003 · DOI
  13. 10.1016/j.proeng.2014.11.535 · doi:10.1016/j.proeng.2014.11.535
  14. 10.1109/TIM.2006.877719 · doi:10.1109/tim.2006.877719
  15. yin, 2013, Computer aided design of Coriolis mass flow-meter based on fluid-structure coupling calculation, Mar Eng, 35, 140
  16. 10.1016/j.jsv.2015.05.014 · doi:10.1016/j.jsv.2015.05.014
  17. van, 2014, Quantification of the influence of external vibrations on the measurement error of a Coriolis mass-flow meter, Flow Meas Instrum, 40, 39, 10.1016/j.flowmeasinst.2014.08.005 · doi:10.1016/j.flowmeasinst.2014.08.005
  18. 10.1109/ICMTMA.2010.121 · doi:10.1109/icmtma.2010.121
  19. 10.1109/JSEN.2013.2260859 · doi:10.1109/jsen.2013.2260859
  20. binulal, 2014, Coriolis flow meter: A review from 1989 to 2014, Int J Sci Eng Res, 5, 718
  21. 10.1109/MEMSYS.2016.7421591 · doi:10.1109/memsys.2016.7421591
  22. 10.1016/j.sna.2019.06.020 · doi:10.1016/j.sna.2019.06.020
  23. 10.1109/I2MTC.2018.8409623 · doi:10.1109/i2mtc.2018.8409623
  24. ch, 2016, Calculation method of oil quantity in dynamic measurement of crude oil, Chemical Management, 32, 190
  25. 10.1109/ICEMI.2017.8265702 · doi:10.1109/icemi.2017.8265702

Cited by

Metadata

Authors
Yang, H., Zhang, L., Li, L., Liang, H., & Zou, J.
Venue
IEEE Transactions on Instrumentation and Measurement
Published
2021
Volume
70
Pages
1-13
ISSN
  • 0018-9456 (ISSN-L)
  • 1557-9662
OpenAlex
W3213777593
Cited by
391
References
6
FWCI
32.68
Type
journal article
Language
en
Open access
closed
Funders
  • National Natural Science Foundation of China (51775543)
  • Joint Funds of the National Natural Science Foundation of China (U1610111)
  • Sichuan Science and Technology Program (2019YFG0195)
  • Scientific Research Starting Project of (2018QHZ016)
  • Young Scholars Development Fund of SWPU (201799010002)
  • Science and Technology Cooperation Project of the CNPC-SWPU Innovation Alliance (2020CX040000)

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.