Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC
Open accessThe DOI matches its Crossref record; 48 of the 48 references it cites were found in a registry or an index.
Abstract
For a first characterization, we used the two KETEK-PM3325-WB SiPMs each equipped with a 3 mm x 3 mm x 5 mm LYSO scintillation crystal provided with the PETsys TOFPET2 ASIC evaluation kit. We changed the lower of two discriminator thresholds (D_T1) in the timing branch from vth_t1 = 5-30. The overvoltage was varied in a range of 1.25-7.25 V. The ambient temperature was kept at 16 °C. For all measurements, we performed an energy calibration including a correction for saturation. We evaluated the energy resolution, the coincidence resolving time (CRT) and the coincidence rate. At an overvoltage of 6 V, we obtained an energy resolution of about 10% full width at half maximum (FWHM), a CRT of approximately 210 ps FWHM and 400 ps full width at tenth maximum, the coincidence rate showed only small variations of about 5%. To investigate the influence of the ambient temperature, it was varied between 12 °C and 20 °C. At 12 °C and an overvoltage of 6.5 V, a CRT of approx. 195 ps FWHM and an energy resolution of about 9.5% FWHM could be measured. Observed satellite peaks in the time difference spectra were investigated in more detail. We could show that the location of the satellite peaks is correlated with a programmable delay element in the trigger circuit.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures


![Fig. 3 Sketch of the TOFPET2 ASIC trigger scheme used in the timing branch. Adapted from [40].](https://figures.cochat.ai/paper-figures/02ae9b94dbc343e6ba137b0bc5e1ffea/initial-measurements-with-the-petsys-tofpet2-asic-evaluation-kit-and-a-figure-3.png)










Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Rinne, J. O., Brooks, D. J., Rossor, M. N., Fox, N. C., Bullock, R., Klunk, W. E., Mathis, C. A., Blennow, K., Barakos, J., Okello, A. A., et al. (2010) · The Lancet Neurology · doi:10.1016/s1474-4422(10)70043-0 · Cited by 701
Bar‐Shalom, R., Yefremov, N., Guralnik, L., Gaitini, D. E., Frenkel, A., Kuten, A., Altman, H., Keidar, Z., & Israel, O. (2003) · PubMed · Cited by 594
Ryan, C. G., Clayton, E., Griffin, W. L., Sie, S. H., & Cousens, D. R. (1988) · Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms · doi:10.1016/0168-583x(88)90063-8 · Cited by 510 · cited 1×
“The 511 keV and 1275 keV peaks in the energy value histogram (Figure 4a) were identified by calculating and subtracting the background [39]–[41] and performing a peak search using a Markov chain algorithm [42].”
MethodWeber, W. A., Petersen, V., Schmidt, B., Tyndale-Hines, L., Link, T., Peschel, C., & Schwaiger, M. (2003) · Journal of Clinical Oncology · doi:10.1200/jco.2003.12.004 · Cited by 493
Brain PET in the Diagnosis of Alzheimer’s Disease
Marcus, C., Mena, E., & Subramaniam, R. M. (2014) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000000547 · Cited by 382
Recent developments in time-of-flight PET
Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0138-3 · Cited by 368 · cited 1×
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundTowards picosecond resolution with single-photon avalanche diodes
Cova, S., Longoni, A., & Andreoni, A. (1981) · Review of Scientific Instruments · doi:10.1063/1.1136594 · Cited by 219 · cited 1×
“signal contains timing information in the order of a few tens of picoseconds [10], [12]–[15].”
BackgroundRecent advances and future advances in time-of-flight PET
Moses, W. W. (2007) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2007.06.038 · Cited by 205 · cited 3×
“Furthermore, SiPMs can be operated in a magnetic field why they are the photosensor of choice for applications that integrate a PET into an MRI [1]–[3].”
BackgroundMethodAdvances in time-of-flight PET
Surti, S., & Karp, J. S. (2016) · Physica Medica · doi:10.1016/j.ejmp.2015.12.007 · Cited by 173 · cited 12×
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundMethodInfluentialGundacker, S., Auffray, E., Pauwels, K., & Lecoq, P. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/7/2802 · Cited by 173 · cited 1×
“/MeV), short decay time (∼40 ns), and fast rise time (∼70 ps), which result in excellent CRT values [1], [10], [11].”
Background
Research works printed but not linked: 9
- 37. PETsys TOF ASIC Evaluation Kit (Flyer), V21 Edition, PETsys Electron. SA, Porto Salvo, Portugal, Nov. 2017. [Online]. Available: http://www.petsys.com/
- 28. TOFPET ASIC V1-Short Data Sheet (Revision 1.2), PETsys Electron. SA, Porto Salvo, Portugal, 2014.
- 54. G. F. Knoll, Radiation Detection and Measurement. Hoboken, NJ, USA: Wiley, 2010.
- 40. TOFPET 2 SiPM Readout ASIC, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 12. Product Data Sheet SiPM-Silicon Photomultiplier PM3315- WB/PM3325-WB, KETEK GmbH, Munich, Germany, Nov. 2016. [Online]. Available: https://www.ketek.net/
- 49. TOFPET2 ASIC-Evaluation Report, PETsys Electron. SA, Porto Salvo, Portugal, 2017. [Online]. Available: http://www.petsys.com/
- 42. TOFPET2 ASIC Evaluation Kit-Software User Guide V2018.04, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 41. TOFPET2 ASIC Evaluation Kit-Hardware User Guide (V1.2), PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 43. Personal Communication, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
Cited by
Liang, S., Wu, H., Zhen, L., Hua, Q., Garg, S., Kaddoum, G., Hassan, M. M., & Yu, K. (2022) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2022.3158253 · Cited by 367
“At present, devices that can implement edge computing include SoC, FPGA, Application Specific Integrated Circuit(ASIC) [36], Central Processing Units (CPU) and GPU.”
MethodSND@LHC: the scattering and neutrino detector at the LHC
Acampora, G., Ahdida, C., Albanese, R., Albrecht, C., Alexandrov, A., Andreini, M., Anokhina, A., Asada, T., Auberson, N., Baldanza, C., et al. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/05/p05067 · Cited by 105
Sensors for Positron Emission Tomography Applications
Jiang, W., Chalich, Y., & Deen, M. J. (2019) · Sensors · doi:10.3390/s19225019 · Cited by 112
Park, H., Yi, M., & Lee, J. S. (2022) · Biomedical Engineering Letters · doi:10.1007/s13534-022-00234-y · Cited by 70
Evaluation of the PETsys TOFPET2 ASIC in multi-channel coincidence experiments
Nadig, V., Schug, D., Weissler, B., & Schulz, V. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00370-x · Cited by 65
“These single events were further processed using an analysis software developed at our institute [52].”
BackgroundMethodResultInfluentialSelf-citationLamprou, E., González, A. J., Sánchez, F., & Benlloch, J. (2020) · Physica Medica · doi:10.1016/j.ejmp.2019.12.004 · Cited by 53
“Detailed analysis of this effect can be found in [28].”
BackgroundMueller, F., Naunheim, S., Kuhl, Y., Schug, D., Solf, T., & Schulz, V. (2022) · Medical Physics · doi:10.1002/mp.16015 · Cited by 55
Self-citationHRFlexToT: A High Dynamic Range ASIC for Time-of-Flight Positron Emission Tomography
Sánchez, D., Fernandez, S. G., Mauricio, J., Coromina, L. F., Sanuy, A., Guixe, G., Huertas, A. L., Escalero, R. M., Marin, J., Perez, J. M., et al. (2021) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2021.3066426 · Cited by 49
“Several ASICs have been developed over the years to provide precise time and energy measurements [21], [22].”
BackgroundMethodInfluentialA Comprehensive Study on the Timing Limits of the TOFPET2 ASIC and on Approaches for Improvements
Nadig, V., Yusopova, M., Radermacher, H., Schug, D., Weissler, B., Schulz, V., & Gundacker, S. (2022) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2022.3158704 · Cited by 40
“This might also flag the appearance of the satellite peaks characterized in [14] due to a longer signal rise time.”
BackgroundResultSelf-citationLaBella, A., Cao, X., Zeng, X., Zhao, W., & Goldan, A. H. (2020) · Medical Physics · doi:10.1002/mp.14654 · Cited by 34
Technical opportunities and challenges in developing total-body PET scanners for mice and rats
Du, J., & Jones, T. (2023) · EJNMMI Physics · doi:10.1186/s40658-022-00523-6 · Cited by 38
“Companies like PETsys Electronics and Hamamatsu also provide complete data acquisition system solutions based on their ASICs for PET scanners [148, 149].”
BackgroundConvolutional Neural Network for Crystal Identification and Gamma Ray Localization in PET
LaBella, A., Vaska, P., Zhao, W., & Goldan, A. H. (2020) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2020.2980985 · Cited by 31
“PETsys TOFPET2 ASICs and an FEB/D board (PETsys Electronics, Oeiras, Portugal) were used for data acquisition and readout [42].”
MethodCao, X., LaBella, A., Zeng, X., Zhao, W., & Goldan, A. H. (2021) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2021.3110902 · Cited by 31
Investigation of the Power Consumption of the PETsys TOFPET2 ASIC
Nadig, V., Weissler, B., Radermacher, H., Schulz, V., & Schug, D. (2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2019.2955032 · Cited by 28
“A detailed description of the operation of the trigger circuit and the appearance of satellite peaks can be found in [49].”
BackgroundMethodSelf-citationToward Sub-100 ps TOF-PET Systems Employing the FastIC ASIC With Analog SiPMs
Mariscal-Castilla, A., Fernandez, S. G., Escalero, R. M., Arribas, J. M. F.-T., Mauricio, J., Kratochwil, N., Alozy, J., Piller, M., Portero, S., Sanuy, A., et al. (2024) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2024.3414578 · Cited by 28
“…proposed for the readout of photo-detectors that can be used in PET applications, such as the NINO ASIC [29], the Weeroc family (PETIROC2A and more recently RADIOROC) [30], [31], the TOFPET2 [32], [33] ASIC developed by PETsys electronics [34], and the FlexToT family [14], [35], [36], [37], [38].”
MethodDevelopment of Multichannel High Time Resolution Data Acquisition System for TOT-ASIC
Sato, S., Uenomachi, M., & Shimazoe, K. (2021) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2021.3084144 · Cited by 24
“PETsys [12] consists of TOF-PET2 ASIC with 64 channels for signal processing and digitization.”
MethodYoshida, E., Obata, F., Kamada, K., & Yamaya, T. (2020) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2020.3032466 · Cited by 22
Binder, T., Kang, H. G., Nitta, M., Schneider, F., Yamaya, T., Parodi, K., Wiest, F., & Thirolf, P. G. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abfbf3 · Cited by 21
Nadig, V., Hornisch, M., Oehm, J., Herweg, K., Schulz, V., & Gundacker, S. (2023) · EJNMMI Physics · doi:10.1186/s40658-023-00594-z · Cited by 20
“The energy spectra were calibrated using the positions of the peaks at 511 keV and 1275 keV in the 22 Na spectrum and a saturation correction model [21] where e is the raw energy value in arbitrary units, E is the saturation-corrected energy value, c is a calibration factor, and s is the maximum…”
MethodSelf-citationD’Ascenzo, N., Antonecchia, E., Gao, M., Zhang, X., Baumgartner, G., Brensing, A., Li, Z., Liu, Q., Rose, G., Shi, X., et al. (2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2019.2937681 · Cited by 20
“The design of fast readout electronics, capable of handling this complex configuration, is thus an important issue in the development of compact brain PET scanners [20], [21].”
BackgroundRashid, T., Ahmad, S., Cizel, J.-B., Taille, C. D. L., Morenas, M., Nadig, V., Perez, F., Schulz, V., Gundacker, S., & Fleury, J. (2023) · Journal of Instrumentation · doi:10.1088/1748-0221/18/10/p10005 · Cited by 19
Self-citationTang, Y., Townsend, T., Deng, H., Liu, Y., Zhang, R., & Chen, J. (2021) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2021.3096162 · Cited by 15
Application-specific nuclear medical in vivo imaging devices
Chaudhari, A. J., & Badawi, R. D. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abf275 · Cited by 16
A finely segmented semi‐monolithic detector tailored for high‐resolution PET
Kuhl, Y., Mueller, F., Naunheim, S., Bovelett, M., Lambertus, J., Schug, D., Weissler, B., Gegenmantel, E., Gebhardt, P., & Schulz, V. (2024) · Medical Physics · doi:10.1002/mp.16928 · Cited by 14
Self-citationTOFHIR2: the readout ASIC of the CMS barrel MIP Timing Detector
Albuquerque, E., Araújo, M., Benaglia, A., Boletti, A., Bugalho, R., Coutinho, T., Guio, F. D., Faccioli, P., Ferramacho, L., Firlej, M., et al. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/05/p05048 · Cited by 14
First imaging demonstration of a crosshair light-sharing PET detector
Yoshida, E., Akamatsu, G., Tashima, H., Kamada, K., Yoshikawa, A., & Yamaya, T. (2021) · Physics in Medicine and Biology · doi:10.1088/1361-6560/abe839 · Cited by 14
A Comprehensive Survey of Readout Strategies for SiPMs Used in Nuclear Imaging Systems
Hesari, S. H., Haque, M. A., & McFarlane, N. M. (2021) · Photonics · doi:10.3390/photonics8070266 · Cited by 14
ASICs in PET: what we have and what we need
Nadig, V., Gundacker, S., Herweg, K., Naunheim, S., Schug, D., Weissler, B., & Schulz, V. (2025) · EJNMMI Physics · doi:10.1186/s40658-025-00717-8 · Cited by 14
Self-citationInterleaved signal multiplexing readout in depth encoding Prism‐PET detectors
Li, Y., LaBella, A., Zeng, X., Wang, Z., Petersen, E. W., Cao, X., Zhao, W., & Goldan, A. H. (2023) · Medical Physics · doi:10.1002/mp.16456 · Cited by 13
Design of a high-accuracy time-to-digital converter based on dual-edge signals
Xiao, Y., Liang, H., Jin, L., Wang, Y., Shao, Z., Wang, Y., & Lu, Y. (2023) · Journal of Instrumentation · doi:10.1088/1748-0221/18/07/p07023 · Cited by 12
“HIGH-performance time-to-digital converter (TDC) is the core component for time interval measurements in scientific instruments and industrial applications, such as quantum imaging and communications [1, 2], automatic test equipment (ATE) [3], high-energy particle physics [4], satellite measurement and control [5], Li…”
BackgroundHerweg, K., Rutstrom, D. J., Nadig, V., Stand, L., Melcher, C. L., Zhuravleva, M., Schulz, V., & Gundacker, S. (2024) · EJNMMI Physics · doi:10.1186/s40658-024-00663-x · Cited by 9
“A saturation curve was fit to the 511-keV photopeak and the 1275-keV Compton edge (848 keV) according to the model described by Schug et al [33] and showed an almost linear behaviour (see Fig.”
Self-citationBaek, M. K., Choi, H. J., Kim, K. B., Ahn, J. J., Min, C. H., & Chung, Y. H. (2022) · Radiation Physics and Chemistry · doi:10.1016/j.radphyschem.2022.110162 · Cited by 8
Gapless implementation of crosshair light-sharing PET detector
Yoshida, E., Obata, F., Kamada, K., Yoshikawa, A., & Yamaya, T. (2021) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2021.165922 · Cited by 8
Dual time-over-threshold: estimation of decay time and pulse height for scintillation detectors
Ota, R., Nakajima, K., Ogawa, I., & Tamagawa, Y. (2019) · Journal of Instrumentation · doi:10.1088/1748-0221/14/11/p11012 · Cited by 7
“2 0 1 9 J I N S T 1 4 P 1 1 0 1 2 than those of similar detectors [20, 21].”
Evaluation of a modular all-in-one high-resolution PET detector and readout electronics setup
Schmidt, F. P., Krämer, J. C., Parl, C., Schmand, M. J., & Pichler, B. J. (2023) · Physics in Medicine and Biology · doi:10.1088/1361-6560/acd432 · Cited by 7
Decision Tree-Based Demultiplexing for Prism-PET
Li, Y., Zeng, X., & Goldan, A. H. (2023) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2023.3282831 · Cited by 6
“The authors gratefully appreciate the helpful contribution of colleagues from PETsys Electronics for scientific discussion.”
MethodPulse shape measurements for neutron/gamma discrimination using the TOFPET2 ASIC
Weinfurther, K., Marleau, P. A., & Sweany, M. (2023) · Journal of Instrumentation · doi:10.1088/1748-0221/18/07/p07015 · Cited by 5
“The TOFPET2 ASIC provides 64 channels of charge integration and pulse timing measurements and is optimized for 8 × 8 arrays of 3 mm × 3 mm SiPMs from Hamamatsu or KeTeK coupled to inorganic scintillators.”
MethodInfluentialEnríquez-Mier-y-Terán, F. E., Pineda-Valencia, W., Judenhofer, M. S., Meikle, S. R., & Kyme, A. Z. (2024) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2024.3505135 · Cited by 5
“The LSB of the discriminator D_T1 was set to 57, corresponding to about 5 mV per DAC unit; D_T2 and D_E were kept at their nominal values of 48 and 40, respectively, corresponding to about 15 and 20 mV per DAC unit [46].”
Li, V. A., Sutanto, F., Classen, T., Dazeley, S., Jovanovic, I., & Wu, T. (2022) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2022.167624 · Cited by 4
“O($10)/channel with a time resolution of a few-hundred picoseconds [17, 18].”
BackgroundCoincidence time resolution measurements for dual-ended readout PET detectors
Sun, M. D., Zhang, C., & Zhao, B. Q. (2023) · Journal of Instrumentation · doi:10.1088/1748-0221/18/07/p07003 · Cited by 4
Preliminary results of a spent nuclear fuel internal tomography (SFIT) system
Chung, Y. S., Choi, H.-J., Min, C. H., & Chung, Y. H. (2024) · Nuclear Engineering and Technology · doi:10.1016/j.net.2024.103415 · Cited by 3
On-Chip Fast Signal Generation and Low-Power SAR ADC for SiPM Readout
Tang, Y., Deng, H., Fan, Q., Feng, Y., Liu, Y., Renshaw, A., Zhang, R., Sun, X., & Chen, J. (2023) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2023.3241223 · Cited by 3
New probe for the improvement of the Spatial Resolution in total-body PET (PROScRiPT)
Ros, A., Barrientos, L., Borja-Lloret, M., Casaña, J. V., Muñoz, E., Röser, J., Udías, J. M., Viegas, R., & Llosa, G. (2021) · EPJ Web of Conferences · doi:10.1051/epjconf/202125309004 · Cited by 3
Performance investigation of LabPET II detector technology in an MRI-like environment
Moghadam, N., Arpin, L., Espagnet, R., Bouchard, J., Viscogliosi, N., Lecomte, R., & Fontaine, R. (2019) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ab57e0 · Cited by 3
Wen, X., & Hayward, J. P. (2025) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2025.3591060 · Cited by 2
“The evaluation board has been explored by many researchers to evaluate the timing performance of various fast scintillators specialized for medical imaging and high-energy physics [27], [29], [30].”
Kang, H. G., Yoshida, E., Akamatsu, G., & Yamaya, T. (2025) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ae1789 · Cited by 3
Wang, M., Wang, T., Wang, Y., Li, D., Cai, J., Li, X., Zhang, X., Li, X., Zhao, Y., Wei, L., et al. (2025) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2025.3568962 · Cited by 2
“Also, TOFPET2 [15], [16], STiC3 [17], [18], PETA4 [19], [20], and other chips use the same technical route to achieve multichannel time measurement (32–64 channels).”
The CuboDAQ data acquisition system
Zaffaroni, E., & Haefeli, G. (2025) · Journal of Instrumentation · doi:10.1088/1748-0221/20/01/t01003 · Cited by 2
Wen, X., & Hayward, J. P. (2024) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2024.169667 · Cited by 1
A SiPM Based Sensor For Nuclear Detection Applications
Hesari, S. H., & McFarlane, N. M. (2021) · IEEE International System-on-Chip Conference (SOCC) · doi:10.1109/socc52499.2021.9739509 · Cited by 1
Study of ToF Capabilities in PET Detector Array Modules Designed for High Spatial Resolution Systems
Pinto, A. S. X. M., Borges, J. F., Correia, P. M. M., Encarnação, P. M. C. C., Veloso, J. F. C. A., & Silva, A. L. M. (2023) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmicrtsd49126.2023.10338207 · Cited by 1
An FPGA Based energy correction method for one-to-one coupled PET detector: model and evaluation
Ma, C., Zhao, X., Gao, S., Zhang, F., Wu, G., Li, X., Yu, L. Z., Lu, L., Ye, H., & Qian, H. (2022) · Journal of Instrumentation · doi:10.1088/1748-0221/17/01/t01003 · Cited by 1
Lamprou, E. (2021) · RiuNet (Universitat Politècnica de València) · doi:10.4995/thesis/10251/162991 · Cited by 1
2019 Index IEEE Transactions on Radiation and Plasma Medical Sciences Vol. 3
(2019) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/trpms.2019.2949183
Novgorodova, O., Glatte, A., Hentges, R., Kögler, T., Lutz, B., Romer, K. E., Teichmann, T., Weinberger, D., & Straessner, A. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/02/p02030
“For the PETsys, it was already demonstrated that the time resolution can be optimized by varying the thresholds [20] with the tendency for the first threshold 𝑉 th_T1 to be set to around 10 DAC units, which is recommended by PETsys and was used as a default value.”
BackgroundMethodInfluentialReadout Strategies and Asynchronous Architectures
Dandin, M. P., McFarlane, N. M., Sajal, M., Dehghandehnavi, F., & Nouri, B. (2024) · Springer eBooks · doi:10.1007/978-3-031-64334-7_7
An optically coupled hybrid semi-monolithic PET detector for simplified DOI and position decoding
Zhao, Y., Yang, H., Huang, W., Zhang, X., Jiang, X., Niu, M., Zeng, X., He, W. J., Wei, Q., & Gu, Z. (2026) · Physics in Medicine and Biology · doi:10.1088/1361-6560/ae97d3
“The raw summed QDC value e was converted to the corrected energy E using a simple saturation model (Schug et al 2019): where c is the energy conversion factor and s is the saturation parameter.”
Metadata
- Authors
- Schug, D., Nadig, V., Weissler, B., Gebhardt, P., & Schulz, V.
- Affiliations
- RWTH Aachen University (DE)
- Venue
- IEEE Transactions on Radiation and Plasma Medical Sciences
- Published
- 2019-07
- Volume
- 3
- Issue
- 4
- Pages
- 444-453
- ISSN
- 2469-7303 (ISSN-L)
- 2469-7311
- OpenAlex
- W2885357483
- Cited by
- 56
- References
- 48
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 9 · 2023: 11 · 2022: 7 · 2021: 13 · 2020: 4 · 2019: 6
- FWCI
- 5.27
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Radiation and Plasma Medical Sciences (publishedVersion, cc-by)
- arXiv (Cornell University) (submittedVersion)
- RWTH Publications (RWTH Aachen) (submittedVersion, other-oa)
- arXiv (Cornell University)
- Funders
- Deutsche Forschungsgemeinschaft (SCHU 2973/2-1)
- Horizon 2020 Framework Programme (667211)
- Topics
- Physical Sciences › Physics and Astronomy › Radiation › Radiation Detection and Scintillator Technologies
- Physical Sciences › Physics and Astronomy › Nuclear and High Energy Physics › Particle Detector Development and Performance
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- time-of-flight PET, LYSO scintillator, coincidence time resolution, energy resolution, coincidence frequency, overvoltage, energy calibration, ambient temperature dependence, satellite peaks, trigger circuit
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.