Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC
Open accessThe DOI matches its Crossref record; 48 of the 48 references it cites were found in a registry or an index.
Abstract
For a first characterization, we used the two KETEK-PM3325-WB SiPMs each equipped with a 3 mm x 3 mm x 5 mm LYSO scintillation crystal provided with the PETsys TOFPET2 ASIC evaluation kit. We changed the lower of two discriminator thresholds (D_T1) in the timing branch from vth_t1 = 5-30. The overvoltage was varied in a range of 1.25-7.25 V. The ambient temperature was kept at 16 °C. For all measurements, we performed an energy calibration including a correction for saturation. We evaluated the energy resolution, the coincidence resolving time (CRT) and the coincidence rate. At an overvoltage of 6 V, we obtained an energy resolution of about 10% full width at half maximum (FWHM), a CRT of approximately 210 ps FWHM and 400 ps full width at tenth maximum, the coincidence rate showed only small variations of about 5%. To investigate the influence of the ambient temperature, it was varied between 12 °C and 20 °C. At 12 °C and an overvoltage of 6.5 V, a CRT of approx. 195 ps FWHM and an energy resolution of about 9.5% FWHM could be measured. Observed satellite peaks in the time difference spectra were investigated in more detail. We could show that the location of the satellite peaks is correlated with a programmable delay element in the trigger circuit.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures


![Fig. 3 Sketch of the TOFPET2 ASIC trigger scheme used in the timing branch. Adapted from [40].](https://figures.cochat.ai/paper-figures/02ae9b94dbc343e6ba137b0bc5e1ffea/initial-measurements-with-the-petsys-tofpet2-asic-evaluation-kit-and-a-figure-3.png)










Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Rinne, J. O., Brooks, D. J., Rossor, M. N., Fox, N. C., Bullock, R., Klunk, W. E., Mathis, C. A., Blennow, K., Barakos, J., Okello, A. A., et al. (2010) · The Lancet Neurology · doi:10.1016/s1474-4422(10)70043-0 · Cited by 701
Bar‐Shalom, R., Yefremov, N., Guralnik, L., Gaitini, D. E., Frenkel, A., Kuten, A., Altman, H., Keidar, Z., & Israel, O. (2003) · PubMed · Cited by 594
Ryan, C. G., Clayton, E., Griffin, W. L., Sie, S. H., & Cousens, D. R. (1988) · Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms · doi:10.1016/0168-583x(88)90063-8 · Cited by 510 · cited 1×
“The 511 keV and 1275 keV peaks in the energy value histogram (Figure 4a) were identified by calculating and subtracting the background [39]–[41] and performing a peak search using a Markov chain algorithm [42].”
MethodWeber, W. A., Petersen, V., Schmidt, B., Tyndale-Hines, L., Link, T., Peschel, C., & Schwaiger, M. (2003) · Journal of Clinical Oncology · doi:10.1200/jco.2003.12.004 · Cited by 493
Brain PET in the Diagnosis of Alzheimer’s Disease
Marcus, C., Mena, E., & Subramaniam, R. M. (2014) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000000547 · Cited by 382
Recent developments in time-of-flight PET
Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0138-3 · Cited by 368 · cited 1×
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundTowards picosecond resolution with single-photon avalanche diodes
Cova, S., Longoni, A., & Andreoni, A. (1981) · Review of Scientific Instruments · doi:10.1063/1.1136594 · Cited by 219 · cited 1×
“signal contains timing information in the order of a few tens of picoseconds [10], [12]–[15].”
BackgroundRecent advances and future advances in time-of-flight PET
Moses, W. W. (2007) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2007.06.038 · Cited by 205 · cited 3×
“Furthermore, SiPMs can be operated in a magnetic field why they are the photosensor of choice for applications that integrate a PET into an MRI [1]–[3].”
BackgroundMethodAdvances in time-of-flight PET
Surti, S., & Karp, J. S. (2016) · Physica Medica · doi:10.1016/j.ejmp.2015.12.007 · Cited by 173 · cited 12×
“[coincidence resolving time (CRT)] with a timing resolution in the order of a few hundred picoseconds [1], [2].”
BackgroundMethodInfluentialGundacker, S., Auffray, E., Pauwels, K., & Lecoq, P. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/7/2802 · Cited by 173 · cited 1×
“/MeV), short decay time (∼40 ns), and fast rise time (∼70 ps), which result in excellent CRT values [1], [10], [11].”
BackgroundBackground elimination methods for multidimensional coincidence γ-ray spectra
Morháč, M., Kliman, J., Matoušek, V., Veselský, M., & Turzo, I. (1997) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/s0168-9002(97)01023-1 · Cited by 162
Francesco, A. D., Bugalho, R., Oliveira, L. B., Pacher, L., Rivetti, A., Rolo, M. D. D. R., Silva, J. C., Silva, R., & Varela, J. (2016) · Journal of Instrumentation · doi:10.1088/1748-0221/11/03/c03042 · Cited by 142 · cited 1×
“The digital part of the ASIC is clocked with a frequency of 200 MHz [37], [38].”
MethodSub-100 ps coincidence time resolution for positron emission tomography with LSO:Ce codoped with Ca
Nemallapudi, M. V., Gundacker, S., Lecoq, P. R., Auffray, E., Ferri, A., Gola, A. G., & Piemonte, C. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/12/4635 · Cited by 138 · cited 1×
“There are benchtop experiments which can achieve coincidence time resolutions below 100 ps [17], [18].”
Background20-ps timing resolution with single-photon avalanche diodes
Cova, S., Lacaita, A. L., Ghioni, M., Ripamonti, G., & Louis, T. A. (1989) · Review of Scientific Instruments · doi:10.1063/1.1140324 · Cited by 136
TOFPET ASIC for PET applications
Rolo, M. D. D. R., Bugalho, R., Gonçalves, F. M., Mazza, G., Rivetti, A., Silva, J. C., Silva, R., & Varela, J. (2013) · Journal of Instrumentation · doi:10.1088/1748-0221/8/02/c02050 · Cited by 126 · cited 1×
“can be switched between a ToT mode and an integration mode (called “qdc” mode by PETsys) [19], [39].”
BackgroundA Digital Preclinical PET/MRI Insert and Initial Results
Weissler, B., Gebhardt, P., Dueppenbecker, P. M., Wehner, J., Schug, D., Lerche, C. W., Goldschmidt, B., Salomon, A., Verel, I., Heijman, E., et al. (2015) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2015.2427993 · Cited by 106 · cited 12×
“Various different single-channel SiPMs as well as SiPM arrays of different vendors will be tested in the future using a similiar evaluation protocol as presented in this work.”
MethodSelf-citation35 ps time resolution at room temperature with large area single photon avalanche diodes
Gulinatti, A., Maccagnani, P., Rech, I., Ghioni, M., & Cova, S. (2005) · Electronics Letters · doi:10.1049/el:20047445 · Cited by 96
Myocardial perfusion imaging with PET
Nakazato, R., Berman, D. S., Alexánderson, E., & Slomka, P. J. (2013) · Imaging in Medicine · doi:10.2217/iim.13.1 · Cited by 92
Identification of peaks in multidimensional coincidence γ-ray spectra
Morháč, M., Kliman, J., Matoušek, V., Veselsky, M., & Turzo, I. (2000) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/s0168-9002(99)01005-0 · Cited by 84 · cited 1×
“4(a)] were identified by calculating and subtracting the background [44]–[46] and performing a peak search using a Markov chain algorithm [47].”
MethodInitial PET performance evaluation of a preclinical insert for PET/MRI with digital SiPM technology
Schug, D., Lerche, C., Weissler, B., Gebhardt, P., Goldschmidt, B., Wehner, J., Dueppenbecker, P. M., Salomon, A., Hallen, P., Kiessling, F. M., et al. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/7/2851 · Cited by 68 · cited 1×
“defined range of the respective spectrum [(details of the fitting methods are described in [48], see Fig.”
BackgroundSelf-citationWehner, J., Weissler, B., Dueppenbecker, P. M., Gebhardt, P., Goldschmidt, B., Schug, D., Kiessling, F. M., & Schulz, V. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/6/2231 · Cited by 65
Self-citationPET/MRI insert using digital SiPMs: Investigation of MR-compatibility
Wehner, J., Weissler, B., Dueppenbecker, P. M., Gebhardt, P., Schug, D., Ruetten, W., Kiessling, F. M., & Schulz, V. (2013) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2013.08.077 · Cited by 57
Self-citationPET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner
Omidvari, N., Cabello, J., Topping, G. J., Schneider, F. R., Paul, S. M., Schwaiger, M., & Ziegler, S. I. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa910d · Cited by 56 · cited 1×
“scanner (MADPET4) using the TOFPET v1 ASIC [35], a tri-modal PET/MR/EEG scanner (TRIMAGE) using the Triroc ASIC [29], [30], [36] and a multimodal setup combining TOF-”
MethodTriroc: A Multi-Channel SiPM Read-Out ASIC for PET/PET-ToF Application
Ahmad, S., Fleury, J., Taille, C. D. L., Seguin-Moreau, N., Dulucq, F., Martin-Chassard, G., Callier, S., Thienpont, D., & Raux, L. (2015) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2015.2397973 · Cited by 54 · cited 3×
“PETA series [24], [27], [31]–[34] as well as the Weeroc series are using an integration method for the energy measurement [26], [29], [30].”
MethodFast Self Triggered Multi Channel Readout ASIC for Time- and Energy Measurement
Fischer, P., Peric, I., Ritzert, M., & Koniczek, M. (2009) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2008.2008807 · Cited by 52 · cited 2×
“PETA series [24], [27], [31]–[34] as well as the Weeroc series are using an integration method for the energy measurement [26], [29], [30].”
MethodBackground estimation for gamma-ray spectrometry
Burgess, D. D., & Tervo, R. J. (1983) · Nuclear Instruments and Methods in Physics Research · doi:10.1016/0167-5087(83)90612-9 · Cited by 49 · cited 1×
“The 511 keV and 1275 keV peaks in the energy value histogram (Figure 4a) were identified by calculating and subtracting the background [39]–[41] and performing a peak search using a Markov chain algorithm [42].”
MethodASIC development for SiPM readout
Corsi, F., Foresta, M., Marzocca, C., Matarrese, G., & Guerra, A. D. (2009) · Journal of Instrumentation · doi:10.1088/1748-0221/4/03/p03004 · Cited by 47 · cited 2×
“The recently developed TOF-PET ASICs are designed in standard CMOS technology with a varying number of input channels ranging from eight channels [23]”
BackgroundMethodExperimental results with TOFPET2 ASIC for time-of-flight applications
Bugalho, R., Francesco, A. D., Ferramacho, L., Leong, C., Niknejad, T., Oliveira, L. B., Pacher, L., Rolo, M. D. D. R., Rivetti, A., Silveira, M. I. L., et al. (2017) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2017.11.034 · Cited by 47 · cited 2×
“300 ps or even as low as 200 ps [19]–[21].”
BackgroundResultClinical Utility of Positron Emission Mammography
Glass, S. B., & Shah, Z. A. (2013) · Baylor University Medical Center Proceedings · doi:10.1080/08998280.2013.11928996 · Cited by 42 · cited 1×
“, in measuring cardiac perfusion and the assessment of Alzheimer’s disease [4]–[9].”
BackgroundPetiroc, a new front-end ASIC for time of flight application
Fleury, J., Callier, S., Taille, C. D. L., Seguin-Moreau, N., Thienpont, D., Dulucq, F., Ahmad, S., & Martin-Chassard, G. (2013) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2013.6829018 · Cited by 34 · cited 1×
“K LYSO 3:0mm 3:0mm 5mm 20 190 12.0 [27], [32] PETA3 FBK LYSO 3:0mm 3:0mm 15mm 20 250 n.a. [32] PETA5 FBK RGB-HD LYSO 2:5mm 2:5mm 10mm 20 205 n.a. [33] Petiroc KETEK LYSO 3:0mm 3:0mm 3mm n.a. n.a. 9.5 [53] Triroc ADVANSID NUV LYSO 3:0mm 3:0mm 8mm 23 420 10.7 [36] Triroc ADVANSID NUV LYSO 3:0mm 3:0mm 10mm n.a. 433 11.0…”
InfluentialPETA4: a multi-channel TDC/ADC ASIC for SiPM readout
Sacco, I., Fischer, P., & Ritzert, M. (2013) · Journal of Instrumentation · doi:10.1088/1748-0221/8/12/c12013 · Cited by 33 · cited 2×
“PETA series [24], [27], [31]–[34] as well as the Weeroc series are using an integration method for the energy measurement [26], [29], [30].”
MethodDevelopment of analog solid-state photo-detectors for Positron Emission Tomography
Bisogni, M. G., & Morrocchi, M. (2015) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2015.09.114 · Cited by 31 · cited 3×
“rate a PET into an MRI [1]–[3]. An SiPM consists of several thousands of Geigermode cells, so-called single-photon avalanche diodes (SPAD), each generating a very similar analog signal upon breakdown [10]. This breakdown is ideally caused by an impinging optical photon but can also be thermally induced in the semicond…”
BackgroundMethodInfluentialSchug, D., Wehner, J., Dueppenbecker, P. M., Weissler, B., Gebhardt, P., Goldschmidt, B., Salomon, A., Kiessling, F. M., & Schulz, V. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/18/7045 · Cited by 30
Self-citationA 64-channel ASIC for TOFPET applications
Rolo, M. D. D. R., Bugalho, R., Gonçalves, F. M., Rivetti, A., Mazza, G., Silva, J., Silva, R., & Varela, J. (2012) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2012.6551353 · Cited by 30 · cited 3×
“can be switched between a ToT mode and an integration mode (called “qdc” mode by PETsys) [19], [39].”
BackgroundThe TRIMAGE PET Data Acquisition System: Initial Results
Sportelli, G., Ahmad, S., Belcari, N., Bisogni, M. G., Camarlinghi, N., Pasquale, A. D., Dussoni, S., Fleury, J., Morrocchi, M., Zaccaro, E., et al. (2016) · IEEE Transactions on Radiation and Plasma Medical Sciences · doi:10.1109/tns.2016.2633237 · Cited by 27 · cited 1×
“scanner (MADPET4) using the TOFPET v1 ASIC [35], a tri-modal PET/MR/EEG scanner (TRIMAGE) using the Triroc ASIC [29], [30], [36] and a multimodal setup combining TOF-”
MethodSTIC3 – Silicon Photomultiplier Timing Chip with picosecond resolution
Stankova, V., Shen, W., Briggl, K., Chen, H., Fischer, P., Gil, A., Harion, T., Kiworra, V., Munwes, Y., Ritzert, M., et al. (2014) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2014.12.073 · Cited by 26 · cited 1×
“[20], [52] ToT ASIC S13361-3050AE-08 MPPC GFAG 2.”
A comparative study of the time performance between NINO and FlexToT ASICs
Sarasola, I., Nemallapudi, M. V., Gundacker, S., Sánchez, D., Gascon, D., Mendes, P. R., Marin, J., & Auffray, E. (2017) · Journal of Instrumentation · doi:10.1088/1748-0221/12/04/p04016 · Cited by 26
The current mode Time-over-Threshold ASIC for a MPPC module in a TOF-PET system
Orita, T., Koyama, A., Yoshino, M., Kamada, K., Yoshikawa, A., Shimazoe, K., & Sugawara, H. (2017) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2017.11.097 · Cited by 26 · cited 1×
“the time-over-threshold (ToT) value can be deducted which contains information about the integral of signal [20], [22].”
BackgroundChen, H., Briggl, K., Fischer, P., Gil, A., Harion, T., Munwes, Y., Ritzert, M., Schimansky, D., Schultz-Coulon, H.-C., Shen, W., et al. (2014) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2014.7431045 · Cited by 23 · cited 5×
“[17], [47] ToT ASIC S13361-3050AE-08 MPPC GFAG 2.”
BackgroundMethodInfluentialA compact, high-density gamma-detection module for Time-of-Flight measurements in PET applications
Sacco, I., Dohle, R., Fischer, P., Piemonte, C., & Ritzert, M. (2015) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2015.11.004 · Cited by 22 · cited 1×
“Of these previously mentioned TOF-PET ASICs, the PETA series is the only one using a integration method for the energy measurement [21], [23], [26]–[28].”
MethodPerformance of FBK SiPMs coupled to PETA3 read-out ASIC for PET application
Piemonte, C., Gola, A. G., Tarolli, A., Fisher, P., Ritzert, M., Schulz, V., & Solf, T. (2012) · Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment · doi:10.1016/j.nima.2012.10.012 · Cited by 21 · cited 1×
“.5 [51] STiC v3 S12643-050CN(X) MPPC LYSO 3:1mm 3:1mm 15mm 22 214 n.a. [20], [52] ToT ASIC S13361-3050AE-08 MPPC GFAG 2:0mm 2:0mm 2mm 25 200 13.5 [22] PETA3 FBK LYSO 3:0mm 3:0mm 5mm 20 190 12.0 [27], [32] PETA3 FBK LYSO 3:0mm 3:0mm 15mm 20 250 n.a. [32] PETA5 FBK RGB-HD LYSO 2:5mm 2:5mm 10mm 20 205 n.a. [33] Petiroc K…”
Self-citationSTiC - A mixed mode chip for SiPM ToF applications
Shen, W., Briggl, K., Chen, H., Fischer, P., Gil, A., Harion, T., Ritzert, M., & Schultz-Coulon, H.-C. (2012) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2012.6551231 · Cited by 18 · cited 2×
“over 16 channels [24], [25], 32 channels [26], 36 channels [27], and 64 channels [19], [20], [28]–[30].”
MethodTriroc, a versatile 64-channel SiPM readout ASIC for time-of-flight PET
Ahmad, S., Taille, C. D. L., Fleury, J., Seguin-Moreau, N., Raux, L., Callier, S., Martin-Chassard, G., Dulucq, F., & Thienpont, D. (2016) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2016.8069882 · Cited by 17 · cited 3×
“PETA series [24], [27], [31]–[34] as well as the Weeroc series are using an integration method for the energy measurement [26], [29], [30].”
MethodMulti-Channel Readout ASIC for ToF-PET
Fischer, P., Peric, I., Ritzert, M., & Solf, T. (2006) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2006.354423 · Cited by 16 · cited 1×
“PETA series [24], [27], [31]–[34] as well as the Weeroc series are using an integration method for the energy measurement [26], [29], [30].”
MethodCharacterization of TOF-PET Detectors Based on Monolithic Blocks and ASIC-ReadoutPreprint
Lamprou, E., González-Montoro, A., Cañizares, G., Ilisie, V., Barrio, J., Sánchez, F., González, A. J., & Benlloch, J. (2018) · arXiv (Cornell University) · doi:10.48550/arxiv.1806.08715 · Cited by 11 · cited 1×
“Other groups using the same setup report results in the same order of magnitude [50].”
Validation of a highly integrated SiPM readout system with a TOF-PET demonstrator
Niknejad, T., Setayeshi, S., Tavernier, S., Bugalho, R., Ferramacho, L., Francesco, A. D., Leong, C., Rolo, M. D. D. R., Shamshirsaz, M., Silva, J. C., et al. (2016) · Journal of Instrumentation · doi:10.1088/1748-0221/11/12/p12003 · Cited by 8 · cited 1×
“5 [51] STiC v3 S12643-050CN(X) MPPC LYSO 3.”
Measurements with a PET Coincidence Setup Based on the PETA5 ASIC and FBK RGB-HD SiPMs
Schug, D., Gebhardt, P., Weissler, B., Groß‐Weege, N., Dey, T., & Schulz, V. (2017) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2017.8532761 · Cited by 8 · cited 1×
“PETA series [24], [27], [31]–[34] as well as the Weeroc series are using an integration method for the energy measurement [26], [29], [30].”
MethodSelf-citationApplication specific integrated circuits for ionizing-radiation detectors (review, part 1)
Basiladze, S. G. (2016) · Instruments and Experimental Techniques · doi:10.1134/s0020441216010280 · Cited by 5 · cited 1×
“In general, there’s a wide variety of ASICs designed for various applications in medical imaging or nuclear physics [15].”
Background
Research works printed but not linked: 9
- 37. PETsys TOF ASIC Evaluation Kit (Flyer), V21 Edition, PETsys Electron. SA, Porto Salvo, Portugal, Nov. 2017. [Online]. Available: http://www.petsys.com/
- 28. TOFPET ASIC V1-Short Data Sheet (Revision 1.2), PETsys Electron. SA, Porto Salvo, Portugal, 2014.
- 54. G. F. Knoll, Radiation Detection and Measurement. Hoboken, NJ, USA: Wiley, 2010.
- 40. TOFPET 2 SiPM Readout ASIC, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 12. Product Data Sheet SiPM-Silicon Photomultiplier PM3315- WB/PM3325-WB, KETEK GmbH, Munich, Germany, Nov. 2016. [Online]. Available: https://www.ketek.net/
- 49. TOFPET2 ASIC-Evaluation Report, PETsys Electron. SA, Porto Salvo, Portugal, 2017. [Online]. Available: http://www.petsys.com/
- 42. TOFPET2 ASIC Evaluation Kit-Software User Guide V2018.04, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 41. TOFPET2 ASIC Evaluation Kit-Hardware User Guide (V1.2), PETsys Electron. SA, Porto Salvo, Portugal, 2018.
- 43. Personal Communication, PETsys Electron. SA, Porto Salvo, Portugal, 2018.
Cited by
Liang, S., Wu, H., Zhen, L., Hua, Q., Garg, S., Kaddoum, G., Hassan, M. M., & Yu, K. (2022) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2022.3158253 · Cited by 367
“At present, devices that can implement edge computing include SoC, FPGA, Application Specific Integrated Circuit(ASIC) [36], Central Processing Units (CPU) and GPU.”
MethodSND@LHC: the scattering and neutrino detector at the LHC
Acampora, G., Ahdida, C., Albanese, R., Albrecht, C., Alexandrov, A., Andreini, M., Anokhina, A., Asada, T., Auberson, N., Baldanza, C., et al. (2024) · Journal of Instrumentation · doi:10.1088/1748-0221/19/05/p05067 · Cited by 105
Sensors for Positron Emission Tomography Applications
Jiang, W., Chalich, Y., & Deen, M. J. (2019) · Sensors · doi:10.3390/s19225019 · Cited by 112
Park, H., Yi, M., & Lee, J. S. (2022) · Biomedical Engineering Letters · doi:10.1007/s13534-022-00234-y · Cited by 70
Evaluation of the PETsys TOFPET2 ASIC in multi-channel coincidence experiments
Nadig, V., Schug, D., Weissler, B., & Schulz, V. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00370-x · Cited by 65
“These single events were further processed using an analysis software developed at our institute [52].”
BackgroundMethodResultInfluentialSelf-citation
Metadata
- Authors
- Schug, D., Nadig, V., Weissler, B., Gebhardt, P., & Schulz, V.
- Affiliations
- RWTH Aachen University (DE)
- Venue
- IEEE Transactions on Radiation and Plasma Medical Sciences
- Published
- 2019-07
- Volume
- 3
- Issue
- 4
- Pages
- 444-453
- ISSN
- 2469-7303 (ISSN-L)
- 2469-7311
- OpenAlex
- W2885357483
- Cited by
- 56
- References
- 48
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 9 · 2023: 11 · 2022: 7 · 2021: 13 · 2020: 4 · 2019: 6
- FWCI
- 5.27
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- IEEE Transactions on Radiation and Plasma Medical Sciences (publishedVersion, cc-by)
- arXiv (Cornell University) (submittedVersion)
- RWTH Publications (RWTH Aachen) (submittedVersion, other-oa)
- arXiv (Cornell University)
- Funders
- Deutsche Forschungsgemeinschaft (SCHU 2973/2-1)
- Horizon 2020 Framework Programme (667211)
- Topics
- Physical Sciences › Physics and Astronomy › Radiation › Radiation Detection and Scintillator Technologies
- Physical Sciences › Physics and Astronomy › Nuclear and High Energy Physics › Particle Detector Development and Performance
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Keywords
- time-of-flight PET, LYSO scintillator, coincidence time resolution, energy resolution, coincidence frequency, overvoltage, energy calibration, ambient temperature dependence, satellite peaks, trigger circuit
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.