Single-chip CMUT-on-CMOS front-end system for real-time volumetric IVUS and ICE imaging
Open accessAbstract
Intravascular ultrasound (IVUS) and intracardiac echography (ICE) catheters with real-time volumetric ultrasound imaging capability can provide unique benefits to many interventional procedures used in the diagnosis and treatment of coronary and structural heart diseases. Integration of capacitive micromachined ultrasonic transducer (CMUT) arrays with front-end electronics in single-chip configuration allows for implementation of such catheter probes with reduced interconnect complexity, miniaturization, and high mechanical flexibility. We implemented a single-chip forward-looking (FL) ultrasound imaging system by fabricating a 1.4-mm-diameter dual-ring CMUT array using CMUT-on-CMOS technology on a front-end IC implemented in 0.35-μm CMOS process. The dual-ring array has 56 transmit elements and 48 receive elements on two separate concentric annular rings. The IC incorporates a 25-V pulser for each transmitter and a low-noise capacitive transimpedance amplifier (TIA) for each receiver, along with digital control and smart power management. The final shape of the silicon chip is a 1.5-mm-diameter donut with a 430-μm center hole for a guide wire. The overall front-end system requires only 13 external connections and provides 4 parallel RF outputs while consuming an average power of 20 mW. We measured RF A-scans from the integrated single- chip array which show full functionality at 20.1 MHz with 43% fractional bandwidth. We also tested and demonstrated the image quality of the system on a wire phantom and an ex vivo chicken heart sample. The measured axial and lateral point resolutions are 92 μm and 251 μm, respectively. We successfully acquired volumetric imaging data from the ex vivo chicken heart at 60 frames per second without any signal averaging. These demonstrative results indicate that single-chip CMUT-on-CMOS systems have the potential to produce realtime volumetric images with image quality and speed suitable for catheter-based clinical applications.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Efforts have been reported to integrate LEDs for optogenetic stimulation and receiving elements for wireless power transfer on chip for mm-scale devices.
“Other common off-chip components include light emitting diodes (LEDs) for optogenetic stimulation, and receiving elements for wireless power transfer, such as inductors, antennas, ultrasonic transducers. Efforts have been reported to integrate all of these on chip for mm-scale devices (Zargham and Glenn Gulak 2015; Gurun et al. 2014).”
From Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations (2018) · cites this paper as Gurun et al. 2014
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A low-power low-noise cmos for amplifier neural recording applications
Harrison, R. R., & Charles, C. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.811979 · Cited by 1,695 · see citing works
A Review of Tissue Substitutes for Ultrasound Imaging
Culjat, M. O., Goldenberg, D. M., Tewari, P., et al. (2010) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2010.02.012 · Cited by 666 · see citing works
Synthetic aperture imaging for small scale systems
Karaman, M., Li, P., & O’Donnell, M. (1995) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/58.384453 · Cited by 579 · see citing works
Self-citationIntegration of 2D CMUT arrays with front-end electronics for volumetric ultrasound imaging
Wygant, I. O., Zhuang, X., Yeh, D. T., et al. (2008) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2008.652 · Cited by 250 · see citing works
Volumetric ultrasound imaging using 2-D CMUT arrays
Oralkan, Ö., Ergun, A. S., Cheng, C., et al. (2003) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2003.1251142 · Cited by 188 · see citing works
Wygant, I. O., Jamal, N. S., Lee, H. J., et al. (2009) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2009.1297 · Cited by 147 · see citing works
Annular-ring CMUT arrays for forward-looking IVUS: transducer characterization and imaging
Degertekin, F. L., Guldiken, R., & Karaman, M. (2006) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2006.1593387 · Cited by 145 · see citing works
Self-citationIntracardiac Echocardiography During Interventional and Electrophysiological Cardiac Catheterization
Hijazi, Z. M., Shivkumar, K., & Sahn, D. J. (2009) · Circulation · doi:10.1161/circulationaha.107.753046 · Cited by 142 · see citing works
Yeh, D. T., Oralkan, Ö., Wygant, I. O., et al. (2006) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2006.1642519 · Cited by 139 · see citing works
Front-end receiver electronics for high-frequency monolithic CMUT-on-CMOS imaging arrays
Gurun, G., Hasler, P. E., & Degertekin, F. L. (2011) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2011.1993 · Cited by 135 · see citing works
Self-citationReal-time volume imaging using a crossed electrode array
Demore, C. E. M., Joyce, A. W., Wall, K., et al. (2009) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2009.1167 · Cited by 105 · see citing works
Forward-viewing CMUT arrays for medical imaging
Demirci, U., Ergun, A. S., Oralkan, Ö., et al. (2004) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2004.1320749 · Cited by 93 · see citing works
Monolithic CMUT-on-CMOS Integration for Intravascular Ultrasound Applications
Zahorian, J., Hochman, M., Xu, T., et al. (2011) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2011.2128 · Cited by 91 · see citing works
Self-citationKimura, B. J., Bhargava, V., & DeMaria, A. N. (1995) · American Heart Journal · doi:10.1016/0002-8703(95)90457-3 · Cited by 82 · see citing works
Wang, Y., Stephens, D. N., & O’Donnell, M. P. (2002) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2002.1159845 · Cited by 72 · see citing works
Coherent-array imaging using phased subarrays. Part I: basic principles
Johnson, J. A., Karaman, M., & Khuri-Yakub, B. T. (2005) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2005.1397349 · Cited by 71 · see citing works
Self-citationCoded excitation for synthetic aperture ultrasound imaging
O'Donnell, M., & Wang, Y. (2005) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2005.1406544 · Cited by 65 · see citing works
Microfabricated ultrasonic transducers monolithically integrated with high voltage electronics
Daft, C. M. W., Calmes, S., da Graca, D., et al. (2005) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.2004.1417770 · Cited by 61 · see citing works
Pipelined Sampled-Delay Focusing in Ultrasound Imaging Systems
Kim, J. H., Song, T. K., & Park, S. B. (1987) · Ultrasonic Imaging · doi:10.1177/016173468700900201 · Cited by 56 · see citing works
Guldiken, R., Zahorian, J., Yamaner, F. Y., et al. (2009) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2009.1169 · Cited by 55 · see citing works
Self-citationCMUTS with dual electrode structure for improved transmit and receive performance
Guldiken, R., McLean, J., & Degertekin, F. L. (2006) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2006.1593388 · Cited by 49 · see citing works
Self-citationLow-temperature micromachined cMUTs with fully-integrated analogue front-end electronics
Noble, R. A., Davies, R. R., King, D. O., et al. (2003) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.2002.1192474 · Cited by 45 · see citing works
Annular array imaging with full-aperture resolution
Norton, S. J. (1992) · The Journal of the Acoustical Society of America · doi:10.1121/1.404169 · Cited by 42 · see citing works
Volumetric real-time imaging using a CMUT ring array
Choe, J. W., Oralkan, Ö., Nikoozadeh, A., et al. (2012) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2012.2310 · Cited by 42 · see citing works
CMOS chip for invasive ultrasound imaging
Black, W. C., & Stephens, D. N. (1994) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.328640 · Cited by 41 · see citing works
New Fabrication Techniques for Ring-Array Transducers for Real-Time 3D Intravascular Ultrasound
Light, E. D., Lieu, V., & Smith, S. W. (2009) · Ultrasonic Imaging · doi:10.1177/016173460903100403 · Cited by 38 · see citing works
GPU-Based Real-Time Volumetric Ultrasound Image Reconstruction for a Ring Array
Choe, J. W., Nikoozadeh, A., Oralkan, Ö., et al. (2013) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2013.2253117 · Cited by 31 · see citing works
Stephens, D. N., Truong, U. T., Nikoozadeh, A., et al. (2012) · Journal of Ultrasound in Medicine · doi:10.7863/jum.2012.31.2.247 · Cited by 30 · see citing works
Optimizing circular ring arrays for forward- looking IVUS imaging
Tekeş, C., Karaman, M., & Degertekin, F. L. (2011) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2011.2123 · Cited by 27 · see citing works
Self-citationAttenuation and Backscattering of Ultrasound in Freshly Excised Animal Tissues
Kadaba, M. P., Bhagat, P. K., & Wu, V. C. (1980) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1980.326710 · Cited by 27 · see citing works
Modular & scalable ultrasound platform with GPU processing
Lewandowski, M., Walczak, M., Witek, B., et al. (2012) · IEEE International Ultrasonics Symposium · doi:10.1109/ultsym.2012.0518 · Cited by 26 · see citing works
Transcatheter valve repair/implantation
Siegel, R. J., Luo, H., & Biner, S. (2011) · International journal of cardiac imaging · doi:10.1007/s10554-011-9833-2 · Cited by 26 · see citing works
Thermal-mechanical-noise-based CMUT characterization and sensing
Gurun, G., Hochman, M., Hasler, P. E., et al. (2012) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2012.2317 · Cited by 22 · see citing works
Self-citationGurun, G., Hasler, P. E., & Degertekin, F. L. (2011) · IEEE International Ultrasonics Symposium · doi:10.1109/ultsym.2011.0115 · Cited by 19 · see citing works
Self-citationCMUTs with dual electrode structure for improved transmit and receive performance
McLean, J., Guldiken, R., & Degertekin, F. L. (2005) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.2004.1417772 · Cited by 19 · see citing works
Self-citationVolumetric imaging using single chip integrated CMUT-on-CMOS IVUS array
Tekeş, C., Zahorian, J., Gurun, G., et al. (2012) · doi:10.1109/embc.2012.6346644 · Cited by 15 · see citing works
Self-citationHigh-voltage devices for 0.5-μm standard CMOS technology
Bassin, C., Ballan, H., & Declercq, M. J. (2000) · IEEE Electron Device Letters · doi:10.1109/55.817446 · Cited by 14 · see citing works
P0-18 Forward-Looking IVUS Imaging Using a Dual-Annular Ring CMUT Array: Experimental Results
Guldiken, R., Zahorian, J., Gurun, G., et al. (2007) · Proceedings/Proceedings - IEEE Ultrasonics Symposium · doi:10.1109/ultsym.2007.313 · Cited by 14 · see citing works
Self-citationNew Fabrication Techniques for Ring-Array Transducers for Real-Time 3D Intravascular Ultrasound
Light, E. D., Lieu, V., & Smith, S. W. (2009) · Ultrasonic Imaging · doi:10.1177/016173460903100405 · Cited by 11 · see citing works
Transmit optimization of CMUTs in non-collapse mode using a transient array model
Satir, S., Zahorian, J., & Degertekin, F. L. (2012) · IEEE International Ultrasonics Symposium · doi:10.1109/ultsym.2012.0021 · Cited by 8 · see citing works
Self-citationIntegrated electronics design for high-frequency intravascular ultrasound imaging
Gurun, G. (2011) · SMARTech Repository (Georgia Institute of Technology) · Cited by 5 · see citing works
Self-citationMeasurements of Ultrasonic Attenuation Properties of Midgestational Fetal Pig Hearts
Gibson, A. A., Singh, G. K., Hoffman, J. J., et al. (2008) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2008.08.014 · Cited by 4 · see citing works
Experimental study of dual-ring CMUT array optimization for forward-looking IVUS
Tekeş, C., Zahorian, J., Gurun, G., et al. (2011) · IEEE International Ultrasonics Symposium · doi:10.1109/ultsym.2011.0277 · Cited by 3 · see citing works
Self-citationNikoozadeh, A., Wygant, I. O., Lin, D.-S., et al. (2011) · Acoustical imaging · doi:10.1007/978-90-481-3255-3_24 · Cited by 2 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- [28]Salvia J Lajevardi P Hekmat M Murmann B A 56M Omega CMOS TIA for MEMS Applications Proceedings of the IEEE 2009 Custom Integrated Circuits Conference 2009 199 202
Cited by
Sorted by citations
Giagka, V., & Serdijn, W. A. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0010-y · Cited by 58 · see citing works
- Authors
- Gurun, G., Tekes, C., Zahorian, J., Xu, T., Satir, S., Karaman, M., Hasler, J., & Degertekin, F. L.
- Affiliations
- Georgia Institute of Technology (US)
- Istanbul Technical University (TR)
- Venue
- IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Published
- 2014-02
- Volume
- 61
- Issue
- 2
- Pages
- 239-250
- ISSN
- 0885-3010 (ISSN-L)
- PMC
- PMC4070885
- OpenAlex
- W2161817031
- Cited by
- 176
- References
- 44
- Citations by year
- 2026: 1 · 2025: 8 · 2024: 16 · 2023: 9 · 2022: 14 · 2021: 23 · 2020: 15 · 2019: 15 · 2018: 23 · 2017: 20 · 2016: 14 · 2015: 9 · 2014: 9
- FWCI
- 17.96
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Istanbul Technical University Academic Open Archive (Istanbul Technical University) (submittedVersion)
- Funders
- Georgia Institute of Technology
- National Institute of Biomedical Imaging and Bioengineering (R01EB010070)
- Topics
- Physical Sciences › Engineering › Mechanics of Materials › Ultrasonics and Acoustic Wave Propagation
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Ultrasound Imaging and Elastography
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Radio Frequency Integrated Circuit Design
- Keywords
- capacitive micromachined ultrasonic transducers, intravascular ultrasound, intracardiac echocardiography, volumetric ultrasound imaging, capacitive transimpedance amplifier, pulser, power management, catheter-based imaging, real-time imaging, point resolution
- MeSH terms
- Amplifiers, Electronic, Animals, Chickens, Computer Systems, Echocardiography, Equipment Design, Sensitivity and Specificity, Signal Processing, Computer-Assisted, Transducers, Reproducibility of Results, Ultrasonography, Interventional, Phantoms, Imaging, Equipment Failure Analysis, Imaging, Three-Dimensional, Electric Capacitance
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review