Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations

Open access
Giagka, Vasiliki, Serdijn, Wouter A.·2018·Bioelectronic Medicine

Abstract

Patients suffering from conditions such as paralysis, diabetes or rheumatoid arthritis could in the future be treated in a personalised manner using bioelectronic medicines (BEms) (Nat Rev Drug Discov 13:399-400, 2013, Proc Natl Acad Sci USA 113:8284-9, 2016, J Intern Med 282:37-45, 2017). To deliver this personalised therapy based on electricity, BEms need to target various sites in the human body and operate in a closed-loop manner. The specific conditions and anatomy of the targeted sites pose unique challenges in the development of BEms. With a focus on BEms based on flexible substrates for accessing small peripheral nerves, this paper discusses several system-level technology considerations related to the development of such devices. The focus is mainly on miniaturisation and long-term operation. We present an overview of common substrate and electrode materials, related processing methods, and discuss assembly, miniaturisation and long-term stability issues.

AI snapshot

Written from the paper’s full text.

TL;DR

This review discusses system-level technology considerations for flexible bioelectronic medicines (BEms) that access the peripheral nervous system, focusing on miniaturisation and long-term operation. It surveys polymer substrates, electrode materials, interconnects, thin ASIC assembly, and protection, concluding that flexible hybrid implants are promising but require further characterisation for decade-long use.

Key findings

  • Flexible polymer substrates such as polyimide, parylene, PDMS, LCP, SU-8, and polyurethane are the main route for PNS interfaces, but they have limited working temperatures (< 350 °C), are permeable to gases and water vapour, and their processing can adversely affect biocompatibility and stability.
  • Electrode miniaturisation is limited by impedance: smaller electrodes lower recording signal-to-noise ratio and, during stimulation, can produce large electrode-tissue voltages that risk electrochemical reactions and complicate output-stage electronics. Coatings such as platinum black, nanostructured platinum grass, iridium oxide, titanium nitride, PEDOT, and carbon materials can reduce impedance or increase charge injection capacity, though long-term characterisation is often still needed.
  • The paper foresees first-generation flexible BEms with several up to tens of electrodes, recording compound neural activity rather than single-unit action potentials; in comparison, silicon-based in vivo systems commonly feature hundreds of electrodes, with recent improvements approaching one thousand, while flexible systems are usually in the order of tens for similar areas.
  • Hybrid flexible BEms require thinned ASICs (typical fabricated chips 500 μm–1 mm thick, active layers 5–10 μm, with functional structures thinner than 25 μm reported), reliable assembly on flexible substrates, and protection; hermetic packages with volumes less than 1 mm³ lose long-life predictability, so polymer and ceramic atomic layer deposition encapsulation approaches are being explored.

Methods

  • Review article; no primary experimental data, sample, dataset, or statistical analysis are reported.
  • Synthesises literature on materials, processing methods, assembly, miniaturisation, and long-term stability for flexible BEms accessing the peripheral nerves.
  • Draws on cited studies and reports device dimensions, material properties, and technology examples qualitatively.

Limitations

  • The authors state the review is not meant to be exhaustive and focuses on system-level technology considerations; power delivery, mapping organ functions onto specific nerves, biosensing for therapy regulation, and safety/security are out of scope and remain unsolved.
  • Evident from the text: no systematic search or selection criteria are described, and conclusions are qualitative rather than based on new experimental validation.
  • Evident from the text: many cited electrode, interconnect, and encapsulation technologies still require further characterisation and testing for the targeted decade-long lifetime.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1186/s42234-018-0010-y

Cites

103 of 103 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

0 of 64 cited sources checked · 0 supported · 0 flagged

  • Premisecitation

    State-of-the-art implants that target the PNS are mainly based on flexible materials.

    “State-of-the-art implants that target the PNS are, therefore, mainly based on flexible materials (Navarro et al. 2005).”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Navarro et al. 2005 source

      Not checked in the free run.

  • Interpretationreasoning

    The hardware implementation of BEms, including design, material selection and fabrication process, is largely dictated by the implantation site.

    “The hardware implementation of BEms, including the design, material selection and fabrication process, is largely dictated by the implantation site.”
    Quote in the paperSubstantial claim
  • Premisecitation

    Many flexible PNS implants have shown promising performance in long-term in vivo tests.

    “Many of them have shown promising performance in long-term in vivo tests (Sohal et al. 2016; Caravaca et al. 2017; Hara et al. 2016; Wurth et al. 2017; Pothhof et al. 2016).”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Sohal et al. 2016 source

      Not checked in the free run.

    • cites Caravaca et al. 2017 source

      Not checked in the free run.

    • cites Hara et al. 2016 source

      Not checked in the free run.

  • Premisecitation

    When higher selectivity is required, interfaces tend to be more invasive.

    “When higher selectivity is required, interfaces tend to be more invasive.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Tyler and Durand 1997 source

      Not checked in the free run.

    • cites Lawrence et al. 2003 source

      Not checked in the free run.

    • cites Boretius et al. 2010 source

      Not checked in the free run.

  • Premisenone given

    The majority of flexible PNS implants are passive devices, connected to remote electronics using long wires.

    “The majority of those implants are however passive devices, connected to remote electronics using long wires.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. 28
  • May combine two claims. 30
  • Probably not a substantial claim: may not pass S4 (based). May combine two claims.

    Interpretationnone given

    To deliver personalised therapy based on electricity, BEms need to target various sites in the human body and operate in a closed-loop manner.

    “To deliver this personalised therapy based on electricity, BEms need to target various sites in the human body and operate in a closed-loop manner.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based). May combine two claims.

    Interpretationnone given

    A flexible BEm for the PNS requires a compact unibody design including a three-dimensional neural interface and the electronics, wireless operation, and closed-loop stimulating, biomarker sensing and neural recording capabilities.

    “These BEms, belong to a new generation of implants: they have to be of a compact unibody design (including a three-dimensional neural interface and the electronics), operate wirelessly (either standalone or in a distributed system), and in a closed-loop fashion, having stimulating, biomarker sensing and neural recording possibilities.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationreasoning

    For a distributed wireless system of BEms to restore full control of the urinary tract, miniature (<1cm3) BEms placed close to their targeted nerves are required.

    “For such an implementation, miniature (<1cm3) BEms placed close to their targeted nerves are required.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given

    The specific conditions and anatomy of the targeted sites pose unique challenges in the development of BEms.

    “The specific conditions and anatomy of the targeted sites pose unique challenges in the development of BEms.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Peripheral nerve size and composition varies greatly; the major and minor diameters of the compound femoral nerve are about 10.5 mm and 2.3 mm, respectively, while pelvic splanchnic nerve bundles can be smaller than 0.05 mm or larger than 0.2 mm.

    “for example, the major and minor diameters of the compound femoral nerve are about 10.5 mm and 2.3 mm, respectively. In contrast, the diameter of pelvic splanchnic nerve bundles can be smaller than 0.05 mm, or larger than 0.2 mm”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites Gustafson et al. 2009 source

      Not checked in the free run.

    • cites Jang et al. 2015 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Extraneural interfaces leave the nerve intact but usually target several bundles of axons or fascicles within it, hence offer lower selectivity.

    “Extraneural interfaces leave the nerve intact but usually target several bundles of axons or fascicles within it, hence offer lower selectivity.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Brindley 1972 source

      Not checked in the free run.

    • cites Koller et al. 1992 source

      Not checked in the free run.

    • cites Loeb and Peck 1996 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Silicon-based implants are stiff and rigid, and micromotion of the device might cause tissue damage.

    “They are, however, stiff and rigid, and micromotion of the device might cause tissue damage”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Alexander Arts et al. 2003 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Softer materials match much better the mechanical properties of the surrounding tissue, and several have shown excellent performance during chronic implantation with only mild to minor foreign body reactions.

    “Softer materials match much better the mechanical properties of the surrounding tissue, and several of them have shown excellent performance during chronic implantation, with only mild to minor foreign body reactions (25).”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [25] Chuang CL, Liao QA, Li HT, Liao SJ, Huang GS. Increasing the bonding strength of chips on flex substrates using thermosonic flip-chip bonding process with nonconductive paste. Microelectron Eng. 2010;87:624–30. source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Titanium nitride coated electrodes have exhibited good electrochemical stability and charge injection limits similar to that of iridium oxide, but the biocompatibility of the coating still needs to be investigated further.

    “Titanium nitride (TiN) coated electrodes have exhibited good electrochemical stability and charge injection limits similar to that of iridium oxide, but the biocompatibility of the coating still needs to be investigated further (Brunton et al. 2015).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Brunton et al. 2015 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationreasoning

    Denser electrode integration is desirable as it offers higher spatial resolution, both for the stimulating and sensing operation mode.

    “Generally, denser electrode integration is desirable as it offers higher spatial resolution, both for the stimulating and sensing operation mode.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    PEDOT is a conductive polymer known to exhibit low electrochemical impedance due to its porous film surface, can be modified by the addition of bioactive molecules, exhibits large charge injection capacity, and has proven to be stable in vivo without causing any toxic effect.

    “Poly (3,4 – ethylene dioxytiophene) (PEDOT) is a conductive polymer known to exhibit low electrochemical impedance due to its porous film surface. PEDOT is interesting as it can be modified by the addition of bioactive molecules to e.g. promote cell adherence to the electrode surface. It exhibits large charge injection capacity and has proven to be stable in vivo without causing any toxic effect (”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Charkhar et al. 2015 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Roughening the surface of electrodes increases the electrochemical surface area while keeping the geometric surface area constant, reducing electrode impedance.

    “One popular technique is roughening the surface of electrodes, in order to increase the electrochemical surface area (Cogan 2008), while keeping the geometric surface area constant.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Cogan 2008 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S4 (based). May combine two claims.

    Premisenone given

    Off-the-shelf components could provide a rapid prototyping solution, but these are neither area nor power optimised.

    “Systems assembled from off-the-shelf components could provide a rapid prototyping solution, but these are neither area nor power optimised.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisenone given

    Existing technologies for realising flexible implants are mainly polymer-based.

    “Existing technologies for realising flexible implants are mainly polymer-based.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim. May combine two claims.

    Interpretationreasoning

    Bi-directional communication and programmability are necessary for BEms, while safe and secure chronic operation is of utmost importance.

    “To this end, bi-directional communication and programmability are necessary, while safe and secure chronic operation is of utmost importance.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Platinum black and nanostructured platinum grass coatings have both shown significant improvements in electrode impedance, due to an increased surface area.

    “More specifically, platinum black (Tand et al. 2014) and nanostructured platinum grass (Liu et al. 2012) coatings have both shown significant improvements in electrode impedance, due to an increased surface area.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Tand et al. 2014

      Not checked in the free run.

    • cites Liu et al. 2012 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationnone given

    Small electrode sizes come at the expense of higher impedances, an undesired feature both in the recording as well as in the stimulation domain.

    “One major bottleneck limiting the achievable electrode density is that small electrode sizes come at the expense of higher impedances, an undesired feature both in the recording as well as in the stimulation domain.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Flexible transistors can incorporate some on-site intelligence, but their level of maturity is not yet sufficient to meet the foreseen processing complexity or the strict requirements for clinical use.

    “Despite the fact that some on-site intelligence can be incorporated with the use of flexible transistors (Viventi et al. 2011), the level of maturity of such technologies is not yet sufficient to meet neither the foreseen processing complexity nor the strict requirements for clinical use.”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Viventi et al. 2011 source

      Not checked in the free run.

  • May combine two claims.

    Premisenone given

    Polymers have limited working temperatures (< 350 °C), which makes them incompatible with several important processing methods, such as chemical vapour deposition and baking, but also with conventional assembly methods, such as bonding and soldering.

    “Challenges of the use of polymers as substrate materials for BEms relate to their limited working temperatures (< 350 °C), which makes them incompatible with several important processing methods, such as chemical vapour deposition and baking, but also with conventional assembly methods, such as bonding and soldering.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationreasoning

    The development of active implants is, in many ways, more challenging compared to their passive counterparts.

    “The development of active implants is, in many ways, more challenging compared to their passive counterparts.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.

    Premisenone given

    Air can be trapped in the polymer during fabrication and manifest itself as air bubbles, which can gradually compromise adhesion between layers and promote delamination.

    “Air can be trapped in the polymer during fabrication and manifest itself as air bubbles, which can gradually compromise adhesion between layers and promote delamination.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Iridium oxide coating is corrosion resistant and exhibits, in general, good biocompatibility and higher charge delivery capacity, but its exact properties depend greatly on the fabrication method, which affects the stability of the coating.

    “Iridium oxide (IrO2) coating is corrosion resistant and exhibits, in general, good biocompatibility and higher charge delivery capacity (Wang et al. 2009). Nevertheless, its exact properties depend greatly on the fabrication method – activated, sputtered and electrodeposited variants have been reported – which, in turn, affects the stability of the coating.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Wang et al. 2009 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Deposition processes for platinum black have reportedly caused cytotoxic reactions due to lead traces found in the electroplating electrolyte, whereas this was not observed for nanostructured platinum grass.

    “The deposition processes for platinum black have reportedly caused cytotoxic reactions due to lead traces found in the electroplating electrolyte (Schuettler et al. 2005b). This was not observed for nanostructured platinum grass (Boehler et al. 2015).”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Schuettler et al. 2005b

      Not checked in the free run.

    • cites Boehler et al. 2015 source

      Not checked in the free run.

  • May combine two claims.

    Interpretationreasoning

    Although dissimilar metals are generally desired to be avoided because galvanic corrosion could occur if exposed to body liquid, failures for the gold-platinum intermetallic combination are unlikely to occur because their galvanic potentials are close to each other.

    “It is generally desired that dissimilar metals are avoided, as galvanic corrosion could occur if exposed to body liquid. However, as the galvanic potentials of gold and platinum are close to each other, such failures for this intermetallic combination are unlikely to occur.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationnone given

    Thinner devices are preferred as these usually can withstand larger bending radii.

    “Although thicker substrates can be fabricated, thinner devices are preferred as these usually can withstand larger bending radii.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationreasoning

    Thinned ASICs will add to the flexibility of the BEm, but may have to be bent and conform to the features or structures of the implantation site, and care should be taken that the resulting mechanical strength of the thin die is adequate for the application.

    “Thinned ASICs will add to the flexibility of the BEm, but may have to be bent and conform to the features or structures of the implantation site. Care should therefore be taken that the resulting mechanical strength of the thin die is adequate for the application.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    It is foreseen that the first generation of flexible BEms will feature several, up to tens of electrodes, rather than hundreds, with relatively larger sizes, recording compound neural activity as closed-loop control signals, rather than single-unit action potentials.

    “Based on the above, it is foreseen that the first generation of flexible BEms will feature several, up to tens of electrodes, rather than hundreds, with relatively larger sizes, recording compound neural activity as closed-loop control signals, rather than single-unit action potentials – which are anyway difficult to acquire chronically.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Interpretationreasoning

    Due to the softer nature of the polymeric substrates, deformation due to motion from nearby organs is likely to occur, and the mechanical properties of interconnects and their substrates should allow the implant to withstand the expected deformation over millions or even billions of cycles without failing.

    “Due to the softer nature of the polymeric substrates, in which these interconnects need to be embedded, however, deformation due to motion from nearby organs is likely to occur. The degree of this deformation will of course vary depending on the implantation site. Therefore, the mechanical properties of these interconnects (and their substrates) should be such that they allow the implant to withst”
    Quote in the paperSubstantial claim
    • Rests on an assumption: The implantation site will subject the implant to deformation over millions or even billions of cycles.
  • Probably not a substantial claim: may not pass S2 (checkable).

    Interpretationreasoning

    Due to the limited available space around the peripheral nerves it is desirable that the majority of components required for the operation of the system are integrated on chip.

    “Due to the limited available space around the peripheral nerves it is desirable that the majority of components required for the operation of the system are integrated on chip, as in Fig. 1.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Premisecitation

    Current state-of-the-art silicon-based in vivo systems do not commonly feature more than hundreds of electrodes, with recent improvements bringing this number closer to one thousand, even on a single probe.

    “Current state-of-the-art silicon-based in vivo systems do not commonly feature more than hundreds of electrodes (Campbell et al. 1991), with recent improvements bringing this number closer to one thousand (Shobe et al. 2015; Scholvin et al. 2016; Buszáki et al. 2015), even on a single probe (Jun et al. 2017).”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Campbell et al. 1991 source

      Not checked in the free run.

    • cites Shobe et al. 2015

      Not checked in the free run.

    • cites Jun et al. 2017 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Fail-safe blocking capacitor-free stimulator designs have been reported and could be employed instead of blocking capacitors.

    “Blocking capacitors are still useful to block direct currents (DCs), but fail-safe blocking capacitor-free stimulator designs have been reported (Liu et al. 2008; Sit and Sarpeshkar 2007), and could be employed instead.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Liu et al. 2008 source

      Not checked in the free run.

    • cites Sit and Sarpeshkar 2007 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Denser electrode integration can be achieved using multiple metal layers, but flexible systems are rarely multi-layered, due to the more complex fabrication process and the higher chance of failure.

    “Denser electrode integration can be achieved using multiple metal layers (Suaning et al. 2007), but flexible systems are rarely multi-layered, due to the more complex fabrication process this implies and the higher chance of failure.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Suaning et al. 2007

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Blocking capacitors do not offer the protection the community thought they did; they introduce a voltage offset across the electrode-tissue interface, which is often neglected and needs to be properly characterised to ensure safe operation.

    “Complete elimination of blocking capacitors could be an option, as they anyway do not offer the protection the community thought they did–they introduce a voltage offset across the electrode-tissue interface, which is often neglected and needs to be properly characterised to ensure safe operation (van Dongen and Serdijn 2016).”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites van Dongen and Serdijn 2016 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    When large capacitors are still needed, for example for energy storage purposes, silicon integration of such structures could be another viable alternative.

    “When large capacitors are still needed, for example for energy storage purposes, silicon integration of such structures could be another viable alternative (Lallemand and Voiron 2013).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Lallemand and Voiron 2013

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    Even after thinning and die separation, millimeter-small, micrometer-thin individual dice have to be handled, mounted, aligned and assembled on the flexible substrates, and related challenges are far from trivial as they are radically different from wafer level processes.

    “However, even after thinning and die separation, millimeter-small, micrometer-thin individual dice have to be handled, mounted, aligned and assembled on the flexible substrates, and related challenges are far from trivial as they are radically different from wafer level processes (Bosman et al. 2011).”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites Bosman et al. 2011 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.

    Premisereasoning

    Platinum is preferred as stimulating electrode material due to its higher charge storage capacity, but its low conductivity (more than four times lower compared to gold) renders it undesirable for narrow, thin and long interconnects.

    “Generally, although platinum is preferred as stimulating electrode material due to its higher charge storage capacity, its low conductivity (more than four times lower compared to gold) renders it undesirable for narrow, thin and long interconnects.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Interpretationreasoning

    The hybrid nature of flexible BEms implies that the custom designed ASICs required for the performance of the system will have to be assembled on the flexible substrates in a heterogeneous process.

    “The hybrid nature of flexible BEms implies that the custom designed ASICs required for the performance of the system will have to be assembled on the flexible substrates in a heterogeneous process.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    In flip-chip bonding with polymeric adhesives, the adhesive acts as the interconnection medium between the bumped chip and the substrate.

    “The adhesive acts as the interconnection medium between the bumped chip and the substrate, while it at the same time protects the contacts and provides mechanical support.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Lai and Liu 1999

      Not checked in the free run.

    • cites Chen et al. 2000 source

      Not checked in the free run.

    • cites Lu and Chen 2010 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Polymeric adhesives are very popular for flip-chip bonding on flexible substrates.

    “Polymeric adhesives are also very popular for flip-chip bonding on flexible substrates (Lai and Liu 1999; Chen et al. 2000; Lu and Chen 2010; Chiang et al. 2006; Chuang et al. 2010) as they can eliminate the need for underfill.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Lai and Liu 1999

      Not checked in the free run.

    • cites Chen et al. 2000 source

      Not checked in the free run.

    • cites Lu and Chen 2010 source

      Not checked in the free run.

  • May combine two claims.

    Premisecitation

    In flip-chip bonding with polymeric adhesives, the adhesive protects the contacts and provides mechanical support.

    “The adhesive acts as the interconnection medium between the bumped chip and the substrate, while it at the same time protects the contacts and provides mechanical support.”
    Quote in the paperNot a substantial claim0 of 3 sources checked
    • cites Lai and Liu 1999

      Not checked in the free run.

    • cites Chen et al. 2000 source

      Not checked in the free run.

    • cites Lu and Chen 2010 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Flip-chip bonding has been extensively used on flexible substrates in several variations, including reflow soldering and thermocompression processes for thin dice (50 μm).

    “Flip-chip bonding has been extensively used on flexible substrates in several variations, including reflow soldering and thermocompression processes for thin dice (50 μm) (Banda et al. 2008; Zhang et al. 2009), but dispensing the underfill without allowing the material to flow on top of the very thin die is challenging.”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites Banda et al. 2008 source

      Not checked in the free run.

    • cites Zhang et al. 2009 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Soldering would probably not be suitable for very fine pitches.

    “When thick film rather than thin film is used options could include soldering, which provides reliable results, although it would probably not be suitable for very fine pitches (Schuettler et al. 2008), parallel-gap welding (Schuettler et al. 2008) and laser welding, which can be used to weld dissimilar materials with very different thicknesses (ratio 50:1) (Rischall and Shackleton 1964).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Schuettler et al. 2008

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    The microflex assembly technique is based on biocompatible materials.

    “Another interesting approach is the microflex assembly technique (Stieglitz et al. 2000), which is based on biocompatible materials and can be used both with thick but also thin films.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Stieglitz et al. 2000

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    The microflex assembly technique can be used both with thick and thin films.

    “Another interesting approach is the microflex assembly technique (Stieglitz et al. 2000), which is based on biocompatible materials and can be used both with thick but also thin films.”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Stieglitz et al. 2000

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Laser welding can be used to weld dissimilar materials with very different thicknesses (ratio 50:1).

    “When thick film rather than thin film is used options could include soldering, which provides reliable results, although it would probably not be suitable for very fine pitches (Schuettler et al. 2008), parallel-gap welding (Schuettler et al. 2008) and laser welding, which can be used to weld dissimilar materials with very different thicknesses (ratio 50:1) (Rischall and Shackleton 1964).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Rischall and Shackleton 1964 source

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Soldering provides reliable results for thick film assembly.

    “When thick film rather than thin film is used options could include soldering, which provides reliable results, although it would probably not be suitable for very fine pitches (Schuettler et al. 2008), parallel-gap welding (Schuettler et al. 2008) and laser welding, which can be used to weld dissimilar materials with very different thicknesses (ratio 50:1) (Rischall and Shackleton 1964).”
    Quote in the paperNot a substantial claim0 of 1 sources checked
    • cites Schuettler et al. 2008

      Not checked in the free run.

  • Probably not a substantial claim: may not pass S4 (based).

    Interpretationnone given

    Substrate processing steps might adversely affect the overall biocompatibility of the implant.

    “Nevertheless, substrate processing steps need careful consideration, as they might adversely affect both the reliability, but also the overall biocompatibility of the implant.”
    Quote in the paperNot a substantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given

    Some polymer candidates for flexible implants stand out because of their proven biocompatibility.

    “Technologies for developing flexible implants are clearly focusing on polymer processing, with some clear prominent candidates standing out because of their proven biocompatibility.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Interpretationnone given

    Substrate processing steps might adversely affect the reliability of the implant.

    “Nevertheless, substrate processing steps need careful consideration, as they might adversely affect both the reliability, but also the overall biocompatibility of the implant.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given

    Technologies for developing flexible implants are focusing on polymer processing.

    “Technologies for developing flexible implants are clearly focusing on polymer processing, with some clear prominent candidates standing out because of their proven biocompatibility.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Interpretationnone given

    Reduced impedance and/or increased electrochemical surface of electrode coatings allow for denser electrode integration.

    “Their reduced impedance and/or increased electrochemical surface allow for denser electrode integration, hence higher specificity and further miniaturisation.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationnone given

    Electrode materials have moved from noble metals to advanced coatings.

    “Electrode materials have moved from noble metals to advanced coatings, constructed by either different materials, or different geometries.”
    Quote in the paperNot a substantial claim
52 of 52
  • The paper relies on this without support.

    Peripheral nerves are often small and fragile.

    “Interfacing with the peripheral nerves is challenging due to their often small size and fragile nature.”
  • The paper relies on this without support.

    The implantation site will subject the implant to deformation over millions or even billions of cycles.

    “the implant to withstand the expected deformation over millions or even billions of cycles without failing”
  • The paper relies on this without support.

    BEms are developed to live decades in the human body.

    “BEms are developed to live decades in the human body and it is essential that they are reliably protected from the surrounding environment.”
  • The paper relies on this without support.

    The thin silicon chip and the flexible substrate have very different mechanical properties.

    “The two materials will have very different mechanical properties and high temperatures like the ones used for thermocompression bonding are often above the substrate’s melting point.”
  • The paper relies on this without support.

    High temperatures used for thermocompression bonding are often above the flexible substrate's melting point.

    “The two materials will have very different mechanical properties and high temperatures like the ones used for thermocompression bonding are often above the substrate’s melting point.”
  • The paper relies on this without support.

    Polymers are permeable to water vapour.

    “Candidates include polymers, despite the fact that they are permeable to water vapour.”
  • The paper relies on this without support.

    The conformal encapsulant layer protects components from the formation of liquid water, thereby preventing corrosion.

    “In encapsulation, the encapsulant attaches to the surface of all components to be protected, forming a conformal layer which protects them from the formation of liquid water, thereby preventing corrosion.”
7 of 7

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

  • Same reference is assigned two publication years

    Minor

    The references list the same Birmingham article, with the same journal, volume, pages, DOI and PMID, once as 2013 and once as 2014.

1 of 1

Cites

Flagged first, then by citations

1 of the 104 references couldn’t be matched to a known paper.

Research works printed but not linked: 3

  1. Birmingham K, et al. Bioelectronic medicines: a research roadmap. Nat Rev Drug Discov. 2014;13:399–400. · DOI · Paper page
  2. Lallemand F, Voiron F. Silicon interposers with integrated passive devices, an excellent alternative to discrete components in Proc. Europ. Microelectr. Packag. Conf. (EMPC) 2013, Grenoble, France. 2013.
  3. Schuettler M, et al.. Cytotoxicity of platinum black. in: 10th Annual Conference of the International Functional Electrical Stimulation Society. Montreal, Canada. 2005b.

Cited by

Sorted by citations

Authors
Giagka, V., & Serdijn, W. A.
Affiliations
  • Fraunhofer Institute for Reliability and Microintegration (DE)
  • Delft University of Technology (NL)
Corresponding author
Giagka, Vasiliki
Venue
Bioelectronic Medicine
Published
2018-06-26
Volume
4
Issue
1
Pages
8-8
ISSN
2332-8886 (ISSN-L)
OpenAlex
W2810368887
Cited by
58
References
104
Citations by year
2026: 4 · 2025: 4 · 2024: 7 · 2023: 5 · 2022: 8 · 2021: 11 · 2020: 10 · 2019: 9
FWCI
0.97
Type
journal article
Language
en
Open access
Locations
  • Bioelectronic Medicine (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Publikationsdatenbank der Fraunhofer-Gesellschaft (Fraunhofer-Gesellschaft) (submittedVersion, other-oa)
  • Fraunhofer-Publica (Fraunhofer-Gesellschaft) (submittedVersion)
  • Research Repository (Delft University of Technology) (submittedVersion, cc-by)
  • Data Archiving and Networked Services (DANS) (submittedVersion, other-oa)
Funders
Technische Universiteit Delft
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
  • Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
Keywords
bioelectronic medicine, peripheral nerves, flexible substrates, flexible electrodes, miniaturization, long-term device stability, closed-loop therapy, personalized medicine, electrode materials, device assembly

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review