Precision and accuracy of blood glucose measurements using three different instruments
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Diabet. Med. 29, 260–265 (2012) Aims Assessment of insulin sensitivity by dynamic metabolic tests such as the hyperinsulinemic euglycemic clamp critically relies on the reproducible and fast measurement of blood glucose concentrations. Although various instruments have been developed over the last decades, little is known as to the accuracy and comparability. We therefore compared the best new instrument with the former gold standard instruments to measure glucose concentrations in metabolic tests. Methods Fasting blood samples of 15 diabetic and 10 healthy subjects were collected into sodium‐fluoride tubes, spiked with glucose (0, 2.8, 6.9 and 11.1 mmol/l) and measured either as whole blood (range 3.3–26.3 mmol/l) or following centrifugation as plasma (range 3.9–32.0 mmol/l). Plasma samples were analyzed in the YSI‐2300 STAT plus (YSI), EKF Biosen C‐Line (EKF) and the reference method, Beckman Glucose analyzer‐II (BMG), whole blood samples in EKF instruments with YSI as reference method. Results The average deviation of the EKF from the reference, BMG, was 3.0 ± 3.5% without any concentration‐dependent variability. Glucose measurements by YSI were in good agreement with that by BMG (plasma) and EKF (plasma and whole blood) up to concentrations of 13.13 mmol/l (0.5 ± 3.7%), but deviation increased to −6.2 ± 3.8% at higher concentrations. Precision (n = 6) was ±2.2% (YSI), ±3.9% (EKF) and ±5.2% (BMG). Conclusions The EKF instrument is comparable regarding accuracy and precision to the reference method BMG and can be used in metabolic tests, while the YSI showed a systematic shift at higher glucose concentrations. Based on these results we decided to replace BMG with EKF instrument in metabolic tests.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Validation of home blood glucose meters with respect to clinical and analytical approaches
Brunner, G. A., Ellmerer, M., Sendlhofer, G., Wutte, A., Trajanoski, Z., Schaupp, L., Quehenberger, F., Wach, P., Krejs, G. J., & Pieber, T. R. (1998) · Diabetes Care · doi:10.2337/diacare.21.4.585 · Cited by 119
Newer portable glucose meters—analytical improvement compared with previous generation devices?
Weitgasser, R., Gappmayer, B., & Pichler, M. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.10.1821 · Cited by 85
Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 74
Cited by
Measuring and estimating insulin resistance in clinical and research settings
Gastaldelli, A. (2022) · Obesity · doi:10.1002/oby.23503 · Cited by 244
“Glucose clamp protocols usually use a laboratoryquality glucose measurement device (Figure 4), such as the Yellow Springs Instrument or the Beckman glucose oxidase analyzer [3, 18].”
MethodAcute dietary fat intake initiates alterations in energy metabolism and insulin resistance
Hernández, E. Á., Kahl, S., Seelig, A., Begovatz, P., Irmler, M., Kupriyanova, Y., Nowotny, B., Nowotny, P., Herder, C., Barosa, C., et al. (2017) · Journal of Clinical Investigation · doi:10.1172/jci89444 · Cited by 194
Self-citationEffects of Intranasal Insulin on Hepatic Fat Accumulation and Energy Metabolism in Humans
Gancheva, S., Koliaki, C., Bierwagen, A., Nowotny, P., Heni, M., Fritsche, A., Häring, H.-U., Szendroedi, J., & Roden, M. (2015) · Diabetes · doi:10.2337/db14-0892 · Cited by 81
“Venous blood glucose concentrations were measured by a glucose oxidase method (EKF Biosen C-Line glucose analyzer; EKF Diagnostic GmbH, Barleben, Germany) (26).”
MethodSelf-citationFritsch, M., Koliaki, C., Livingstone, R., Phielix, E., Bierwagen, A., Meisinger, M., Jelenik, T., Strassburger, K., Zimmermann, S., Brockmann, K., et al. (2015) · The American Journal of Clinical Nutrition · doi:10.3945/ajcn.115.107599 · Cited by 33
Self-citation(2016) · doi:10.1177/1932296816634356 · Cited by 23
Metadata
- Authors
- Nowotny, B., Nowotny, P. J., Strassburger, K., & Roden, M.
- Venue
- Diabetic Medicine
- Publisher
- Wiley
- Published
- 2012-01-11
- Volume
- 29
- Issue
- 2
- Pages
- 260-265
- ISSN
- 0742-3071 (ISSN-L)
- 1464-5491
- OpenAlex
- W1526417179
- Cited by
- 43
- References
- 4
- FWCI
- 1.93
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Precision and accuracy of blood glucose measurements using three different instruments. https://app.cochat.ai/papers/10.1111/j.1464-5491.2011.03406.x
- The paper
- Nowotny, B., Nowotny, P. J., Strassburger, K., & Roden, M. (2012). Precision and accuracy of blood glucose measurements using three different instruments. Diabetic Medicine. https://doi.org/10.1111/j.1464-5491.2011.03406.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.