Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population

Open access
Fernandes, P., Díaz‐Rey, J. A., Queirós, A., Gonzalez‐Meijome, J. M., Jorge, J. · 2005 · Ophthalmic and Physiological Optics
Cited by 167

The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.

Abstract

The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare®) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare® was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare® rebound tonometer with applanation tonometry (AT) ( p < 0.05). The mean difference between the two tonometers was 1.34 ± 2.03 mmHg (mean ± S.D.) and the 95% limits of agreement were ±3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare® and the AT. In the present population the ICare® overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare® tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of ±3.00 mmHg in more than 80% of the population.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1475-1313.2005.00327.x

Cites

2 of 2 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 167

Metadata

Authors
Fernandes, P., Díaz‐Rey, J. A., Queirós, A., Gonzalez‐Meijome, J. M., & Jorge, J.
Venue
Ophthalmic and Physiological Optics
Published
2005-08-10
Volume
25
Issue
5
Pages
436-440
ISSN
  • 0275-5408 (ISSN-L)
  • 1475-1313
OpenAlex
W2060608359
Cited by
167
References
2
FWCI
10.18
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population. https://app.cochat.ai/papers/10.1111/j.1475-1313.2005.00327.x
The paper
Fernandes, P., Díaz‐Rey, J. A., Queirós, A., Gonzalez‐Meijome, J. M., & Jorge, J. (2005). Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic and Physiological Optics. https://doi.org/10.1111/j.1475-1313.2005.00327.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.