Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare®) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare® was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare® rebound tonometer with applanation tonometry (AT) ( p < 0.05). The mean difference between the two tonometers was 1.34 ± 2.03 mmHg (mean ± S.D.) and the 95% limits of agreement were ±3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare® and the AT. In the present population the ICare® overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare® tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of ±3.00 mmHg in more than 80% of the population.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Comparing methods of measurement: why plotting difference against standard method is misleading
Bland, J. M., & Altman, D. G. (1995) · The Lancet · doi:10.1016/s0140-6736(95)91748-9 · Cited by 2,343 · cited 1×
“The 95% limits of agreement (mean of the difference ± 1.96 · S.D. of the differences) were also calculated as recommended by Bland and Altman (1986, 1995).”
MethodInfluentialSources of error with use of Goldmann-type tonometers
Whitacre, M. M., & Stein, R. (1993) · Survey of Ophthalmology · doi:10.1016/0039-6257(93)90053-a · Cited by 747 · cited 2×
“Measurements of IOP are subject to many sources of error, and the results cannot always be taken as exact figures (Whitacre and Stein, 1993).”
BackgroundTwenty-four-hour pattern of intraocular pressure in the aging population.
Liu, J. H., Kripke, D. F., Twa, M. D., Hoffman, R. E., Mansberger, S. L., Rex, K. M., Girkin, C. A., & Weinreb, R. N. (1999) · PubMed · Cited by 293
A new induction‐based impact method for measuring intraocular pressure
Kontiola, A. I. (2000) · Acta Ophthalmologica Scandinavica · doi:10.1034/j.1600-0420.2000.078002142.x · Cited by 214 · cited 3×
“…The College of Optometrists436 improved more than 30 years ago and recently developed further by Kontiola, and it is based on launching a magnetized probe against the eye using a solenoid (Dekking and Coster, 1967; Kontiola, 2000; Kontiola et al., 2001; Goldblum et al., 2002; Danias et al., 2003).”
BackgroundMethodMethod for the Noninvasive Measurement of Intraocular Pressure in Mice
Danias, J., Kontiola, A., Filippopoulos, T., & Mittag, T. W. (2003) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.02-0553 · Cited by 148 · cited 2×
“…The College of Optometrists436 improved more than 30 years ago and recently developed further by Kontiola, and it is based on launching a magnetized probe against the eye using a solenoid (Dekking and Coster, 1967; Kontiola, 2000; Kontiola et al., 2001; Goldblum et al., 2002; Danias et al., 2003).”
BackgroundAccuracy of the New ICare Rebound Tonometer vs. Other Portable Tonometers in Healthy Eyes
García‐Resúa, C., González‐Méijome, J. M., Gilino, J., & Yebra‐Pimentel, E. (2006) · Optometry and Vision Science · doi:10.1097/01.opx.0000200673.96758.7b · Cited by 105
Kontiola, A., & Puska, P. M. (2003) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-003-0671-3 · Cited by 100 · cited 2×
“The software discarded the highest and lowest IOP readings and calculated the average IOP value from the rest (Kontiola and Puska, 2004).”
BackgroundMethodGoldblum, D., Kontiola, A., Mittag, T. W., Chen, B., & Danias, J. (2002) · Graefe s Archive for Clinical and Experimental Ophthalmology · doi:10.1007/s00417-002-0571-y · Cited by 90 · cited 1×
“…The College of Optometrists436 improved more than 30 years ago and recently developed further by Kontiola, and it is based on launching a magnetized probe against the eye using a solenoid (Dekking and Coster, 1967; Kontiola, 2000; Kontiola et al., 2001; Goldblum et al., 2002; Danias et al., 2003).”
Background
Cited by
Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry
Pakrou, N., Gray, T., Mills, R., Landers, J., & Craig, J. (2008) · Journal of Glaucoma · doi:10.1097/ijg.0b013e318133fb32 · Cited by 189
“Another small study of subjects with normal IOPs concluded that the Icare overestimates the GAT, although with an absolute bias of <3mm Hg in approximately 80%.(16) A study of 103 subjects compared the Icare with the GAT and found no significant difference between the 2 modalities.”
BackgroundSurveillance for ocular hypertension: an evidence synthesis and economic evaluation.
Burr, J., Botello-Pinzon, P., Takwoingi, Y., Hernández, R., Vazquez-Montes, M., Elders, A., Asaoka, R., Banister, K., van der Schoot, J., Fraser, C., et al. (2012) · Health Technology Assessment · doi:10.3310/hta16290 · Cited by 180
The Influence of Corneal Properties on Rebound Tonometry
Chui, W.-S., Lam, A., Chen, D., & Chiu, R. (2008) · Ophthalmology · doi:10.1016/j.ophtha.2007.03.061 · Cited by 163
Effect of Corneal Thickness on Dynamic Contour, Rebound, and Goldmann Tonometry
Martinez-de-la-Casa, J. M., Garcia-Feijoo, J., Vico, E., Fernandez-Vidal, A., Benitez del Castillo, J. M., Wasfi, M., & Garcia-Sanchez, J. (2006) · Ophthalmology · doi:10.1016/j.ophtha.2006.06.016 · Cited by 161
Metadata
- Authors
- Fernandes, P., Díaz‐Rey, J. A., Queirós, A., Gonzalez‐Meijome, J. M., & Jorge, J.
- Venue
- Ophthalmic and Physiological Optics
- Published
- 2005-08-10
- Volume
- 25
- Issue
- 5
- Pages
- 436-440
- ISSN
- 0275-5408 (ISSN-L)
- 1475-1313
- OpenAlex
- W2060608359
- Cited by
- 167
- References
- 25
- FWCI
- 10.18
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.