A new induction‐based impact method for measuring intraocular pressure
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
: Purpose: An improved induction‐based impact method for measuring intraocular pressure is presented. Methods: A light probe containing a permanent magnet is launched towards the eye using a solenoid. The probe hits the eye and bounces back. The same solenoid, inside which the probe moves, is used to detect the movement and impact of the probe, because the moving magnet induces voltage in the solenoid. In the clinical study, the Goldmann applanation tonometer was used as a standard. Handheld and fixed tonometers were also compared by means of measurements in pressurized pig eyes. Results: At a low intraocular pressure (IOP), there is a longer impact time than at a high IOP. A handheld tonometer causes more variation in impact times, especially at low intraocular pressure (IOP), than a fixed tonometer. Conclusions: The new induction‐based impact method can be used to measure IOP. When a fixed tonometer is used, there is less variation in impact times.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Goldmann, H., & Schmidt, T. (1957) · Ophthalmologica · doi:10.1159/000303213 · Cited by 607
A new electromechanical method for measuring intraocular pressure
Kontiola, A. (1997) · Documenta Ophthalmologica · doi:10.1007/bf02569066 · Cited by 142
Self-citationClinical Evaluation of the Oculab Tono-Pen
Minckler, D. S., Baerveldt, G., Heuer, D. K., Quillen-Thomas, B., Walonker, A. F., & Weiner, J. M. (1987) · American Journal of Ophthalmology · doi:10.1016/0002-9394(87)90010-9 · Cited by 119
Cited by
Clinical Comparison of the Icare Tonometer and Goldmann Applanation Tonometry
Pakrou, N., Gray, T., Mills, R., Landers, J., & Craig, J. (2008) · Journal of Glaucoma · doi:10.1097/ijg.0b013e318133fb32 · Cited by 189
“IOP is calculated from the measurement of impact duration and/or maximum deceleration.(2) Use of the device is simple and can be rapidly taught.”
BackgroundMethodVidal-Sanz, M., Salinas-Navarro, M., Nadal-Nicolás, F. M., Alarcón-Martínez, L., Valiente-Soriano, F. J., Miralles de Imperial, J., Avilés-Trigueros, M., Agudo-Barriuso, M., & Villegas-Pérez, M. P. (2012) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2011.08.001 · Cited by 189
Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients
Brusini, P., Salvetat, M. L., Zeppieri, M., Tosoni, C., & Parisi, L. (2006) · Journal of Glaucoma · doi:10.1097/01.ijg.0000212208.87523.66 · Cited by 187
“It was also used in glaucoma studies involving measuring IOP in rodents(22) and was tested in other studies, with fair results.(21,23) The ICare tonometer is the first commercially available instrument based on the induction-based rebound method.”
MethodThe Influence of Corneal Properties on Rebound Tonometry
Chui, W.-S., Lam, A., Chen, D., & Chiu, R. (2008) · Ophthalmology · doi:10.1016/j.ophtha.2007.03.061 · Cited by 163
Noninvasive intraocular pressure monitoring: Current insights
De Smedt, S. (2015) · Clinical ophthalmology · doi:10.2147/opth.s53772 · Cited by 29
“It has the advantages of a noncontact unit, not requiring anesthetic drops.(55) Its use as a self-tonometer has been reported and its readings correlate well with GAT.”
Metadata
- Authors
- Kontiola, A. I.
- Venue
- Acta Ophthalmologica Scandinavica
- Publisher
- Wiley
- Published
- 2000-04
- Volume
- 78
- Issue
- 2
- Pages
- 142-145
- ISSN
- 1395-3907 (ISSN-L)
- 1600-0420
- OpenAlex
- W2070836631
- Cited by
- 214
- References
- 3
- FWCI
- 1.44
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A new induction‐based impact method for measuring intraocular pressure. https://app.cochat.ai/papers/10.1034/j.1600-0420.2000.078002142.x
- The paper
- Kontiola, A. I. (2000). A new induction‐based impact method for measuring intraocular pressure. Acta Ophthalmologica Scandinavica. https://doi.org/10.1034/j.1600-0420.2000.078002142.x
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.