Is DDD superior to VVI pacing in mixed carotid sinus syndrome? An acute and medium-term study
Abstract
The aim of this study was to evaluate fhe importance of atrial synchronism for pacing therapy of patients with mixed carotid sinus syndrome. In 23 patients (21 m, two f: mean age 69 ± 8 years) affected by symptomatic mixed carotid sinus syndrome we performed: (1) Research of ventriculo‐atrial conduction, orthostatic hypotension and pacemaker effect; and [2] Carotid sinus massage in the standing position during VVI and DVI temporary pacing. Next, ail patients received a permanent DDD pacemaker and entered a 2 month two period single‐blind, randomized, cross‐over study on DVI/DDD versus VVI mode. During the DVI/DDD period, no syncope occurred in any patients, minor symptoms persisted in 11 (48%) of them; during VVI period syncopes recurred in three patients, symptoms requiring the withdrawal of VVI pacing and premature DVI/DDD reprogramming in eight patients, minor symptoms in 17 (74%). A comparison between 14 patients, who preferred DVI/DDD period (Group A), and the remaining nine patients who noted no preference between DVI/DDD and VVI period (Group B) was performed on the basis of the preimplant evaluation. Group A patients had a greater pacemaker effect (‐34 ± 16 mmHg vs ‐16 ± 14 mmHg) and a higher prevalence of symptomatic pacemaker effect (50% vs 0%), of ventriculo‐atrial conduction (78% vs 44%) and of orthostatic hypotension (50% vs 11%), while the entity of the systolic pressure fall caused by carotid sinus massage was similar in the two groups either during VVI mode (Group A‐51 ± 16 mmHg vs Group B‐56 ± 27 mmHg) or DVI mode (Group A‐38 ± 17 mmHg vs Group 3–45 ± 17 mmHg). Thus, we conclude that vasodepressor reflex of carotid sinus syndrome is not prevented by DVI/DDD pacing even if the loss of atrial synchronism resulting from WI pacing, by adding an important pacemaker effect, causes a more severe hemodynamic consequence. DVI/DDD pacing is more effective than VVI in 61% of patients. When pacemaker effect, ventriculoatrial conduction and orthostatic hypotension are present, VVI failure is possible, therefore DVI/DDD stimulation is indicated; on the contrary the evaluation of the vasodepressor reflex during WJ or DVI temporary pacing is of little value in the choice of the mode of pacing.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In CSS, dual-chamber pacing is better than single-chamber ventricular mode in counteracting BP fall during CSM and in preventing symptom recurrences.
“In CSS, a few small controlled studies 304,305 and one registry 306 showed that dual-chamber pacing is better than the single chamber ventricular mode in counteracting BP fall during CSM and in preventing symptom recurrences.”
From 2018 ESC Guidelines for the diagnosis and management of syncope (2018) · cites this paper as [305]
Cites
Ausubel, K., & FURMAN, S. (1985) · Annals of Internal Medicine · doi:10.7326/0003-4819-103-3-420 · Cited by 211 · see citing works
MORLEY, C. A., Perrins, E. J., Grant, P., Chan, S. L., McBrien, D. J., & Sutton, R. (1982) · Heart · doi:10.1136/hrt.47.5.411 · Cited by 175 · see citing works
Carotid sinus hypersensitivity and syncope
Walter, P. F., Crawley, I. S., & Dorney, E. R. (1978) · The American Journal of Cardiology · doi:10.1016/0002-9149(78)90934-7 · Cited by 164 · see citing works
MORLEY, C. A., & Sutton, R. B. (1984) · International Journal of Cardiology · doi:10.1016/0167-5273(84)90188-8 · Cited by 102 · see citing works
Carotid sinus hypersensitivity: beneficial effects of dual-chamber pacing
Madigan, N. P., Flaker, G. C., Curtis, J. J., Reid, J., Mueller, K. J., & Murphy, T. J. (1984) · The American Journal of Cardiology · doi:10.1016/0002-9149(84)90632-5 · Cited by 95 · see citing works
Hypotension with ventricular pacing: An atrial vasodepressor reflex in human beings
Erlebacher, J. A., Danner, R. L., & Stelzer, P. E. (1984) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(84)80100-x · Cited by 62 · see citing works
Carotid sinus hypersensitivity in patients presenting with syncope.
Davies, A. B., Stephens, M., & Davies, A. (1979) · Heart · doi:10.1136/hrt.42.5.583 · Cited by 50 · see citing works
Permanent demand pacing for hypersensitive carotid sinus syndrome.
Peretz, D. I., Gerein, A. N., & Miyagishima, R. T. (1973) · PubMed · Cited by 46 · see citing works
Stryjer, D., FRIEDENSOHN, A., & Schlesinger, Z. (1986) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1986.tb05419.x · Cited by 22 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
2018 ESC Guidelines for the diagnosis and management of syncope
Brignole, M., Moya, A., de Lange, F. J., Deharo, J.-C., Elliott, P. M., Fanciulli, A., Fedorowski, A., Furlan, R., Kenny, R. A., Martín, A., et al. (2018) · European Heart Journal · doi:10.1093/eurheartj/ehy037 · Cited by 1,951 · see citing works
Self-citation2017 ACC/AHA/HRS Guideline for the Evaluation and Management of Patients With Syncope
Shen, W.-K., Sheldon, R. S., Benditt, D. G., Cohen, M. I., Forman, D. E., Goldberger, Z. D., Grubb, B. P., Hamdan, M. H., Krahn, A. D., Link, M. S., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.03.004 · Cited by 484 · see citing works
2017 ACC/AHA/HRS Guideline for the Evaluation and Management of Patients With Syncope
Shen, W.-K., Sheldon, R. S., Benditt, D. G., Cohen, M. I., Forman, D. E., Goldberger, Z. D., Grubb, B. P., Hamdan, M. H., Krahn, A. D., Link, M. S., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.03.003 · Cited by 327 · see citing works
Shen, W.-K., Sheldon, R. S., Benditt, D. G., Cohen, M. I., Forman, D. E., Goldberger, Z. D., Grubb, B. P., Hamdan, M. H., Krahn, A. D., Link, M. S., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.03.002 · Cited by 155 · see citing works
Claesson, J.-E., Kristensson, B.-E., Edvardsson, N., & Wahrborg, P. (2007) · Europace · doi:10.1093/europace/eum180 · Cited by 89 · see citing works
Dretzke, J., Toff, W. D., Lip, G. Y., Raftery, J., Fry-Smith, A., & Taylor, R. S. (2004) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd003710.pub2 · Cited by 89 · see citing works
The natural history of carotid sinus syncope and the effect of cardiac pacing
Brignole, M., & Menozzi, C. (2011) · Europace · doi:10.1093/europace/euq516 · Cited by 49 · see citing works
Self-citation
- Authors
- BRIGNOLE, M., SARTORE, B., BARRA, M., MENOZZI, C., & LOLLI, G.
- Affiliations
- Presidio Ospedaliero (IT)
- Corresponding author
- BRIGNOLE, MICHELE
- ORCID
- BRIGNOLE, MICHELE
- Venue
- Pacing and Clinical Electrophysiology
- Publisher
- Wiley
- Published
- 1988-11
- Volume
- 11
- Issue
- 11
- Pages
- 1902-1910
- ISSN
- 0147-8389 (ISSN-L)
- 1540-8159
- OpenAlex
- W2099773135
- Cited by
- 59
- References
- 9
- Citations by year
- 2023: 1 · 2018: 4 · 2017: 6 · 2016: 1 · 2015: 1 · 2014: 6 · 2013: 4 · 2012: 2
- FWCI
- 1.71
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Pacing and Clinical Electrophysiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Cardiovascular Syncope and Autonomic Disorders
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Keywords
- DDD pacing, VVI pacing, orthostatic hypotension, carotid sinus massage, syncope, carotid sinus syndrome
- MeSH terms
- Aged, Cardiac Pacing, Artificial, Carotid Sinus, Clinical Trials as Topic, Female, Follow-Up Studies, Humans, Hypotension, Orthostatic, Male, Pacemaker, Artificial, Random Allocation, Reflex, Abnormal, Syncope, Syndrome
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review