Multi‐Center Clinical Experience with a Lumenless, Catheter‐Delivered, Bipolar, Permanent Pacemaker Lead: Implant Safety and Electrical Performance

GAMMAGE, MICHAEL D., LIEBERMAN, RANDY A., YEE, RAYMOND, MANOLIS, ANTONIS S., COMPTON, STEVEN J., KHAZEN, CÉSAR, SCHAAF, KATIE, OLESON, KIMBERLY A., CROSSLEY, GEORGE H., for the Worldwide SelectSecure Clinical Investigators · 2006 · Pacing and Clinical Electrophysiology
Cited by 86

The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.

Abstract

Purpose: Reduced lead diameter and reliability can be designed into transvenous permanent pacing leads through use of redundant insulation and removal of the stylet lumen. The model 3830 lead (Medtronic Inc., Minneapolis, MN, USA) is a bipolar, fixed‐screw, steroid‐eluting, lumenless, 4.1‐Fr pacing lead. Implantation can be performed in a variety of right heart sites using a deflectable catheter (Model 10600, Medtronic). Lead performance and safety were studied. Methods: Two prospective trials of 338 implanted subjects from 56 global sites were conducted. Electrical and safety data were obtained at implant, pre‐discharge, and up to 18 months post‐implant. Leads were implanted at traditional and alternate right heart sites. Results: The study enrolled 338 subjects (204 males, 70.6 ± 11.6 years) followed‐up for a mean of 10.2 months (range, 0–21.6). Mean P‐wave amplitudes ranged from 3.2 mV at 3 months to 2.9 mV at 18 months, while mean atrial pulse width thresholds at 2.5 V ranged from 0.07 ms at 3 months to 0.09 ms at 18 months. Mean R‐wave amplitudes ranged from 11.3 mV to 11.1 mV and mean ventricular pulse width thresholds at 2.5 V ranged from 0.10 ms to 0.14 ms. There were 22 ventricular and 12 atrial lead complications within 3 months post‐implant. Survival from lead‐related complications improved to a clinically acceptable rate in the cohort of patients when revised implant techniques were employed. Conclusions: With the use of recommended implant techniques, the study results support the electrical efficacy and safety of a catheter‐delivered, lumenless lead in traditional or alternate right atrium or right ventricle sites through 18 months post‐implant.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1111/j.1540-8159.2006.00452.x

Cites

3 of 3 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 3 references

Cited by

See all 86

Metadata

Authors
GAMMAGE, M. D., LIEBERMAN, R. A., YEE, R., MANOLIS, A. S., COMPTON, S. J., KHAZEN, C., SCHAAF, K., OLESON, K. A., CROSSLEY, G. H., & for the Worldwide SelectSecure Clinical Investigators
Venue
Pacing and Clinical Electrophysiology
Publisher
Wiley
Published
2006-08
Volume
29
Issue
8
Pages
858-865
ISSN
  • 0147-8389 (ISSN-L)
  • 1540-8159
OpenAlex
W2007126234
Cited by
86
References
3
FWCI
3.72
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Multi‐Center Clinical Experience with a Lumenless, Catheter‐Delivered, Bipolar, Permanent Pacemaker Lead: Implant Safety and Electrical Performance. https://app.cochat.ai/papers/10.1111/j.1540-8159.2006.00452.x
The paper
GAMMAGE, M. D., LIEBERMAN, R. A., YEE, R., MANOLIS, A. S., COMPTON, S. J., KHAZEN, C., SCHAAF, K., OLESON, K. A., CROSSLEY, G. H., & for the Worldwide SelectSecure Clinical Investigators. (2006). Multi‐Center Clinical Experience with a Lumenless, Catheter‐Delivered, Bipolar, Permanent Pacemaker Lead: Implant Safety and Electrical Performance. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/j.1540-8159.2006.00452.x

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.