Precision of measurement of cerebral tissue oxygenation index using near-infrared spectroscopy in preterm neonates
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
The use of cerebral tissue oxygenation index (c-TOI) in a clinical setting is limited by doubts concerning the accuracy of the measurements. Since there is no gold standard, validation is difficult. Our modest aim was to quantify the precision of c-TOI doing repeated measurements by reapplying the optode several times presuming no regional differences in cerebral oxygenation. Thirty-seven premature infants were examined with several measurements of c-TOI using the NIRO 300 oximeter. Three to eight measurements were done on each infant over a period of 15 to 25 min. One-way analysis of variance was used to estimate within- and between-infant variation. The median gestational age was 27.6 weeks (23.9 to 33.0). Mean c-TOI (n = 253) was 74.6+/-8.5%. The within-infant variation was 5.2% when resiting the optode. For comparison, the between-infants variation was 6.9%, while the spontaneous 2-sec variation was 2.9%. The precision of a single measurement of c-TOI was not good. By measuring five times instead of one on each subject, the precision of the mean can be assumed to be comparable to pulse oximetry. This may be too cumbersome for clinical use, but may reduce sample size in research.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Positron emission tomography study of human brain functional development
Chugani, H. T., Phelps, M. E., & Mazziotta, J. C. (1987) · Annals of Neurology · doi:10.1002/ana.410220408 · Cited by 1,548
Tissue oxygenation monitor using NIR spatially resolved spectroscopy:
Suzuki, S., Takasaki, S., Ozaki, T., & Kobayashi, Y. (1999) · SPIE Proceedings · doi:10.1117/12.356862 · Cited by 484
Cited by
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438
“73 In order to isolate venous blood from arterial blood, NIRS was used together with venous occlusion by a pressure cuff 74,75 or by hand.”
Hyttel-Sorensen, S., Pellicer, A., Alderliesten, T., Austin, T., van Bel, F., Benders, M., Claris, O., Dempsey, E., Franz, A. R., Fumagalli, M., et al. (2015) · BMJ · doi:10.1136/bmj.g7635 · Cited by 309
Self-citationMonitoring Neonatal Regional Cerebral Oxygen Saturation in Clinical Practice: Value and Pitfalls
van Bel, F., Lemmers, P., & Naulaers, G. (2008) · Neonatology · doi:10.1159/000151642 · Cited by 258
Lemmers, P. M. A., Toet, M. C., & van Bel, F. (2008) · Pediatrics · doi:10.1542/peds.2007-0925 · Cited by 244
Alderliesten, T., Dix, L., Baerts, W., Caicedo, A., van Huffel, S., Naulaers, G., Groenendaal, F., van Bel, F., & Lemmers, P. (2016) · Pediatric Research · doi:10.1038/pr.2015.186 · Cited by 199
Metadata
- Authors
- Sorensen, L. C., & Greisen, G.
- Venue
- Journal of Biomedical Optics
- Publisher
- SPIE-Intl Soc Optical Eng
- Published
- 2006
- Volume
- 11
- Issue
- 5
- Pages
- 054005
- ISSN
- 1083-3668 (ISSN-L)
- OpenAlex
- W2080181287
- Cited by
- 112
- References
- 2
- FWCI
- 4.03
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.