Scatter correction for cone‐beam CT in radiation therapy
Open accessAbstract
Cone‐beam CT (CBCT) is being increasingly used in modern radiation therapy for patient setup and adaptive replanning. However, due to the large volume of x‐ray illumination, scatter becomes a rather serious problem and is considered as one of the fundamental limitations of CBCT image quality. Many scatter correction algorithms have been proposed in literature, while a standard practical solution still remains elusive. In radiation therapy, the same patient is scanned repetitively during a course of treatment, a natural question to ask is whether one can obtain the scatter distribution on the first day of treatment and then use the data for scatter correction in the subsequent scans on different days. To realize this scatter removal scheme, two technical pieces must be in place: (i) A strategy to obtain the scatter distribution in on‐board CBCT imaging and (ii) a method to spatially match a prior scatter distribution with the on‐treatment CBCT projection data for scatter subtraction. In this work, simple solutions to the two problems are provided. A partially blocked CBCT is used to extract the scatter distribution. The x‐ray beam blocker has a strip pattern, such that partial volume can still be accurately reconstructed and the whole‐field scatter distribution can be estimated from the detected signals in the shadow regions using interpolation/extrapolation. In the subsequent scans, the patient transformation is determined using a rigid registration of the conventional CBCT and the prior partial CBCT. From the derived patient transformation, the measured scatter is then modified to adapt the new on‐treatment patient geometry for scatter correction. The proposed method is evaluated using physical experiments on a clinical CBCT system. On the Catphan©600 phantom, the errors in Hounsfield unit (HU) in the selected regions of interest are reduced from about 350 to below ; on an anthropomorphic phantom, the error is reduced from 15.7% to 5.4%. The proposed method is attractive in applications where a high CBCT image quality is critical, for example, dose calculation in adaptive radiation therapy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Paired cycle‐GAN‐based image correction for quantitative cone‐beam computed tomography
Harms, J., Lei, Y., Wang, T., Zhang, R., Zhou, J., Tang, X., Curran, W. J., Liu, T., & Yang, X. (2019) · Medical Physics · doi:10.1002/mp.13656 · Cited by 257
Sisniega, A., Zbijewski, W., Badal, A., Kyprianou, I. S., Stayman, J. W., Vaquero, J. J., & Siewerdsen, J. H. (2013) · Medical Physics · doi:10.1118/1.4801895 · Cited by 117
Metadata
- Authors
- Zhu, L., Xie, Y., Wang, J., & Xing, L.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2009-05-18
- Volume
- 36
- Issue
- 6Part1
- Pages
- 2258-2268
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2072021306
- Cited by
- 202
- References
- 0
- FWCI
- 13.73
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Scatter correction for cone‐beam CT in radiation therapy. https://app.cochat.ai/papers/10.1118/1.3130047
- The paper
- Zhu, L., Xie, Y., Wang, J., & Xing, L. (2009). Scatter correction for cone‐beam CT in radiation therapy. Medical Physics. https://doi.org/10.1118/1.3130047
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.