Quality control for quantitative multicenter whole-body PET/MR studies: A NEMA image quality phantom study with three current PET/MR systems

Open access
Boellaard, Ronald, Rausch, Ivo, Beyer, Thomas, Delso, Gaspar, Yaqub, Maqsood, Quick, Harald H., Sattler, Bernhard·2015·Medical Physics

Abstract

Purpose: Integrated positron emission tomography/magnetic resonance (PET/MR) systems derive the PET attenuation correction (AC) from dedicated MR sequences. While MR‐AC performs reasonably well in clinical patient imaging, it may fail for phantom‐based quality control (QC). The authors assess the applicability of different protocols for PET QC in multicenter PET/MR imaging. Methods: The National Electrical Manufacturers Association NU 2 2007 image quality phantom was imaged on three combined PET/MR systems: a Philips Ingenuity TF PET/MR, a Siemens Biograph mMR, and a GE SIGNA PET/MR (prototype) system. The phantom was filled according to the EANM FDG‐PET/CT guideline 1.0 and scanned for 5 min over 1 bed. Two MR‐AC imaging protocols were tested: standard clinical procedures and a dedicated protocol for phantom tests. Depending on the system, the dedicated phantom protocol employs a two‐class (water and air) segmentation of the MR data or a CT‐based template. Differences in attenuation‐ and SUV recovery coefficients (RC) are reported. PET/CT‐based simulations were performed to simulate the various artifacts seen in the AC maps (μ‐map) and their impact on the accuracy of phantom‐based QC. Results: Clinical MR‐AC protocols caused substantial errors and artifacts in the AC maps, resulting in underestimations of the reconstructed PET activity of up to 27%, depending on the PET/MR system. Using dedicated phantom MR‐AC protocols, PET bias was reduced to −8%. Mean and max SUV RC met EARL multicenter PET performance specifications for most contrast objects, but only when using the dedicated phantom protocol. Simulations confirmed the bias in experimental data to be caused by incorrect AC maps resulting from the use of clinical MR‐AC protocols. Conclusions: Phantom‐based quality control of PET/MR systems in a multicenter, multivendor setting may be performed with sufficient accuracy, but only when dedicated phantom acquisition and processing protocols are used for attenuation correction.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.4930962

Cites

41 of 41 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • MR-based attenuation correction (MR-AC) in PET/MRI relies on vendor-dependent MR sequences to automatically segment the body into three or four compartments (air, bone, soft tissue, fat).

    “In contrast, MR-based attenuation correction (MR-AC) in PET/MRI relies on different, vendor-dependent MR sequences that are used to automatically segment the body into three or four compartments (air, bone, soft tissue, and fat) [58].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [58]

    Section 2.B (Clinical protocol) describes vendor-dependent MR sequences that segment images into three or four tissue classes. The listed classes are air, lungs, soft tissue, and—for two systems—fat, not bone; the Introduction also says the implemented approaches do not include bones or substitute them with soft tissue. (LLM, conf=0.90)

    Experimental — accuracy not yet measured
  • Inter-vendor differences in MR-AC maps can lead to SUV differences in phantom spheres of 10-20%.

    “In such phantoms, inter-vendor differences in MR-AC maps [61] can lead to SUV differences in phantom spheres of 10–20% [58].”

    From Influences on PET Quantification and Interpretation (2022) · cites this paper as [58]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Boellaard, R., Rausch, I., Beyer, T., Delso, G., Yaqub, M., Quick, H. H., & Sattler, B.
Affiliations
  • European Association of Nuclear Medicine (AT)
  • Medical University of Vienna (AT)
  • University Hospital Zurich (CH)
  • Amsterdam UMC Location Vrije Universiteit Amsterdam (NL)
  • Friedrich-Alexander-Universität Erlangen-Nürnberg (DE)
  • University Hospital Leipzig (DE)
Corresponding author
Boellaard, Ronald
Venue
Medical Physics
Publisher
Wiley
Published
2015-09-22
Volume
42
Issue
10
Pages
5961-5969
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W1940721824
Cited by
62
References
41
Citations by year
2026: 1 · 2025: 3 · 2024: 3 · 2023: 1 · 2022: 9 · 2021: 12 · 2020: 7 · 2019: 6 · 2018: 5 · 2017: 7 · 2016: 7 · 2015: 1
FWCI
5.39
Type
journal article
Language
en
Open access
Locations
  • Medical Physics (publishedVersion)
  • PubMed (publishedVersion)
  • Amsterdam UMC (VU Amsterdam) - Institutional Repository (publishedVersion)
  • Pure Amsterdam UMC (submittedVersion, other-oa)
  • Data Archiving and Networked Services (DANS) (submittedVersion)
Funders
EANM Research Limited
Topics
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
  • Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
Keywords
PET/MR, quality control, MR-based attenuation correction, standardized uptake value, FDG-PET
MeSH terms
Humans, Image Processing, Computer-Assisted, Magnetic Resonance Imaging, Quality Control, Phantoms, Imaging, Positron-Emission Tomography, Whole Body Imaging, Multimodal Imaging

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review