Systematic Evaluation of Phantom Fluids for Simultaneous PET/MR Hybrid Imaging
Open accessAbstract
UNLABELLED: With the recent advent of integrated PET/MR hybrid systems, the need for simultaneous PET and MR phantom measurements arises. Phantom fluids that are used in stand-alone MR systems, especially in larger phantoms and at a high magnetic field strength, are not necessarily applicable in PET imaging and vice versa. In this study, different approaches to fluid selection were considered and systematically evaluated with respect to their usability for simultaneous PET/MR phantom imaging. METHODS: Demineralized water, water with increased electrical conductivity, a water-oil emulsion, and monoethylene and triethylene glycol were investigated in MR and PET measurements using the most common PET tracer (18)F-FDG. As an alternative to (18)F-FDG, a modified PET tracer ((18)F-fluoride Kryptofix 222 complex) was investigated toward its ability to dissolve in pure oil, which provides good signal homogeneity in MR imaging. Measurements were performed on a 3.0 T integrated PET/MR whole-body system using a National Electrical Manufacturers Association quality-standard phantom. RESULTS: All tested fluids dissolved the radiotracer (18)F-FDG homogeneously. Regarding their suitability for MR at 3.0 T, all fluids significantly improved the homogeneity compared to pure water (increase of excitation flip angle within the tested phantom by a factor of 2.0). When the use of (18)F-FDG was preferred, triethylene glycol provided the best compromise (flip angle increase by a factor of 1.13). The potential alternative tracer, (18)F-fluoride Kryptofix 222 complex, dissolved in pure oil; however, it is not optimal in its tested composition because it accumulates at the bottom of the phantom during the time of measurement. CONCLUSION: This study provides a systematic approach toward phantom fluid selection for imaging a given quality-standard body phantom--and phantoms of comparable size--at 3.0 T. For simultaneous PET/MR scans using the standard tracer (18)F-FDG, an alternative fluid to water and oil is proposed that serves as a viable option for both imaging modalities. Nevertheless, when water is preferred, ways to improve MR image homogeneity are presented. The tested alternative PET tracer enables the use of pure oil in combined scans, but the tracer composition needs to be optimized for phantom measurement applications.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Cross-calibration of PET/MRI scanners is generally hampered by water-filled phantoms, which can produce substantial artifacts in the MR-AC sequence with corresponding SUV errors.
“It should also be noted that cross-calibration of PET/MRI scanners is generally hampered by water-filled phantoms, which are commonly used for SUV calibration in PET/CT and can produce substantial artifacts in the MR-AC sequence with corresponding SUV errors [63,64].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [63]
The MR Imaging section (page unavailable) reports that pure water produces artifacts in attenuation-correction images and assigned attenuation values. The Discussion (page unavailable) identifies homogeneous phantom imaging as a prerequisite for MR-based attenuation correction, but does not report corresponding SUV errors or establish that PET/MRI cross-calibration is generally hampered. (LLM, conf=0.84)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Hamacher, K., Coenen, H. H., & Stöcklin, G. (1986) · PubMed · Cited by 1,368 · see citing works
Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR Scanner
Delso, G., Fürst, S., Jakoby, B., Ladebeck, R., Ganter, C., Nekolla, S. G., Schwaiger, M., & Ziegler, S. I. (2011) · Journal of Nuclear Medicine · doi:10.2967/jnumed.111.092726 · Cited by 870 · see citing works
Self-citationNational electrical manufacturers association
Busemann-Sokole, E., & Cradduck, T. D. (1926) · Journal of the A.I.E.E. · doi:10.1109/jaiee.1926.6537344 · Cited by 866 · see citing works
International Electrotechnical Commission (IEC)
OECD (2016) · OECD eBooks · doi:10.1787/9789264244047-29-en · Cited by 802 · see citing works
de BAZELAIRE, C., Duhamel, G., Rofsky, N. M., & Alsop, D. C. (2004) · Radiology · doi:10.1148/radiol.2303021331 · Cited by 791 · see citing works
Martínez-Möller, A., Souvatzoglou, M., Delso, G., Bundschuh, R. A., Chefd’hotel, C., Ziegler, S. I., Navab, N., Schwaiger, M., & Nekolla, S. G. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.054726 · Cited by 712 · see citing works
Self-citationDesign and performance evaluation of a whole-body Ingenuity TF PET–MRI system
Zaidi, H., Ojha, N., Morich, M. A., Griesmer, J. J., Hu, Z., Maniawski, P. J., Ratib, O. M., Izquierdo‐Garcia, D., Fayad, Z. A., & Shao, L. (2011) · Physics in Medicine and Biology · doi:10.1088/0031-9155/56/10/013 · Cited by 400 · see citing works
Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI.
Pichler, B. J., Judenhofer, M. S., Catana, C., Walton, J. H., Kneilling, M., Nutt, R. E., Siegel, S. B., Claussen, C. D., & Cherry, S. R. (2006) · PubMed · Cited by 396 · see citing works
Sources of intensity nonuniformity in spin echo images at 1.5 T
Simmons, A., Tofts, P. S., Barker, G. J., & Arridge, S. (1994) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910320117 · Cited by 277 · see citing works
Bernstein, M. A., Huston, J., & Ward, H. A. (2006) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20698 · Cited by 255 · see citing works
Central brightening due to constructive interference with, without, and despite dielectric resonance
Collins, C. M., Liu, W., Schreiber, W., Yang, Q., & Smith, M. B. H. (2005) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20245 · Cited by 245 · see citing works
Abdominal MRI at 3.0 T: The Basics Revisited
Merkle, E. M., & Dale, B. M. (2006) · American Journal of Roentgenology · doi:10.2214/ajr.05.0932 · Cited by 197 · see citing works
Sled, J. G., & Pike, G. B. (1998) · IEEE Transactions on Medical Imaging · doi:10.1109/42.730409 · Cited by 190 · see citing works
Ibrahim, T. S., Lee, R., Abduljalil, A., Baertlein, B. A., & Robitaille, P.-M. L. (2001) · Magnetic Resonance Imaging · doi:10.1016/s0730-725x(01)00300-9 · Cited by 163 · see citing works
Review of 18F-FDG synthesis and quality control
Yu, S. (2006) · Biomedical Imaging and Intervention Journal · doi:10.2349/biij.2.4.e57 · Cited by 155 · see citing works
Chemistry & chemical reactivity
Kotz, J. C., & Purcell, K. F. (1991) · Saunders College Pub. eBooks · Cited by 129 · see citing works
Simultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils
Paulus, D. H., Braun, H., Aklan, B., & Quick, H. H. (2012) · Medical Physics · doi:10.1118/1.4729716 · Cited by 114 · see citing works
Self-citationHigh-Field-Strength Magnetic Resonance
Ladd, M. E. (2007) · Topics in Magnetic Resonance Imaging · doi:10.1097/rmr.0b013e3180f612b3 · Cited by 114 · see citing works
PET Tracers and Radiochemistry
Schyler, D. J. (2004) · Annals of the Academy of Medicine Singapore · doi:10.47102/annals-acadmedsg.v33n2p146 · Cited by 76 · see citing works
Standing Waves in Uniform Water Phantoms
Tofts, P. S. (1994) · Journal of Magnetic Resonance Series B · doi:10.1006/jmrb.1994.1067 · Cited by 76 · see citing works
Saha, G. B. (2010) · Springer eBooks · doi:10.1007/978-1-4419-0805-6 · Cited by 73 · see citing works
Woishnis, W. A. (2006) · Elsevier eBooks · doi:10.1016/b978-081551522-7.50031-5 · Cited by 59 · see citing works
Biersack, H., & Freeman, L. M. (2007) · Springer eBooks · doi:10.1007/978-3-540-28026-2 · Cited by 34 · see citing works
Flagged first, then by citations
2 of the 25 references couldn’t be matched to a known paper.
Research works printed but not linked: 2
- Brown WH, Foote CS, Iverson BL, Anslyn EV. Organic Chemistry. Belmont, CA: Brooks/Cole; 2012:458-459.
- Martinez-Möller A, Souvatzoglou M, Delso G, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MR imaging. J Nucl Med. 2009;50:520-526.
Cited by
Sorted by citations
Vaquero, J. J., & Kinahan, P. E. (2015) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071114-040723 · Cited by 384 · see citing works
PET/MRI: Where might it replace PET/CT?
Ehman, E. C., Johnson, G. B., Villanueva-Meyer, J. E., Cha, S., Leynes, A. P., Larson, P. E. Z., & Hope, T. A. (2017) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.25711 · Cited by 276 · see citing works
Grant, A. M., Deller, T. W., Khalighi, M. M., Maramraju, S. H., Delso, G., & Levin, C. S. (2016) · Medical Physics · doi:10.1118/1.4945416 · Cited by 246 · see citing works
Quick, H. H. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24523 · Cited by 137 · see citing works
Self-citationSimultaneous PET/MR imaging: MR-based attenuation correction of local radiofrequency surface coils
Paulus, D. H., Braun, H., Aklan, B., & Quick, H. H. (2012) · Medical Physics · doi:10.1118/1.4729716 · Cited by 114 · see citing works
Self-citationQuantitative comparison of PET performance—Siemens Biograph mCT and mMR
Karlberg, A. M., Sæther, O., Eikenes, L., & Goa, P. E. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0142-7 · Cited by 79 · see citing works
Integrated PET/MR imaging: Automatic attenuation correction of flexible RF coils
Kartmann, R., Paulus, D. H., Braun, H., Aklan, B., Ziegler, S., Navalpakkam, B. K., Lentschig, M. G., & Quick, H. H. (2013) · Medical Physics · doi:10.1118/1.4812685 · Cited by 71 · see citing works
Self-citationBeyer, T., Lassen, M. L., Boellaard, R., Delso, G., Yaqub, M., Sattler, B., & Quick, H. H. (2016) · Magnetic Resonance Materials in Physics, Biology and Medicine · doi:10.1007/s10334-015-0505-4 · Cited by 66 · see citing works
Self-citationBailey, D. L., Antoch, G., Bartenstein, P., Barthel, H., Beer, A. J., Bisdas, S., Bluemke, D. A., Boellaard, R., Claussen, C. D., Franzius, C., et al. (2015) · Molecular Imaging and Biology · doi:10.1007/s11307-014-0818-0 · Cited by 64 · see citing works
Self-citationImage quality and detectability in Siemens Biograph PET/MRI and PET/CT systems—a phantom study
Øen, S. K., Aasheim, L. B., Eikenes, L., & Karlberg, A. M. (2019) · EJNMMI Physics · doi:10.1186/s40658-019-0251-1 · Cited by 64 · see citing works
Boellaard, R., Rausch, I., Beyer, T., Delso, G., Yaqub, M., Quick, H. H., & Sattler, B. (2015) · Medical Physics · doi:10.1118/1.4930962 · Cited by 62 · see citing works
Self-citationZiegler, S., Jakoby, B., Braun, H., Paulus, D. H., & Quick, H. H. (2015) · EJNMMI Physics · doi:10.1186/s40658-015-0122-3 · Cited by 60 · see citing works
Self-citationTowards improved hardware component attenuation correction in PET/MR hybrid imaging
Paulus, D. H., Tellmann, L., & Quick, H. H. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/22/8021 · Cited by 59 · see citing works
Self-citationOehmigen, M., Ziegler, S., Jakoby, B., Georgi, J.-C., Paulus, D. H., & Quick, H. H. (2014) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.139147 · Cited by 57 · see citing works
Self-citationSoderlund, A. T., Chaal, J., Tjio, G., Totman, J. J., Conti, M., & Townsend, D. W. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.156711 · Cited by 53 · see citing works
Influences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 50 · see citing works
“It should also be noted that cross-calibration of PET/MRI scanners is generally hampered by waterfilled phantoms, which are commonly used for SUV calibration in PET/CT and can produce substantial artifacts in the MR-AC sequence with corresponding SUV errors [63,64].”
BackgroundA review of harmonization strategies for quantitative PET
Akamatsu, G., Tsutsui, Y., Daisaki, H., Mitsumoto, K., Baba, S., & Sasaki, M. (2023) · Annals of Nuclear Medicine · doi:10.1007/s12149-022-01820-x · Cited by 49 · see citing works
Lindemann, M. E., Oehmigen, M., Blumhagen, J. O., Gratz, M., & Quick, H. H. (2017) · Medical Physics · doi:10.1002/mp.12449 · Cited by 41 · see citing works
Self-citationBailey, D. L., Barthel, H., Beuthien‐Baumann, B., Beyer, T., Bisdas, S., Boellaard, R., Czernin, J. G., Drzezga, A. E., Ernemann, U., Franzius, C., et al. (2014) · Molecular Imaging and Biology · doi:10.1007/s11307-014-0725-4 · Cited by 39 · see citing works
Self-citationHybrid Positron Emission Tomography/Magnetic Resonance Imaging
Paulus, D. H., & Quick, H. H. (2016) · Investigative Radiology · doi:10.1097/rli.0000000000000289 · Cited by 36 · see citing works
Self-citationAttenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging
Eldib, M., Bini, J., Faul, D. D., Oesingmann, N., Tsoumpas, C., & Fayad, Z. A. (2015) · PET Clinics · doi:10.1016/j.cpet.2015.10.004 · Cited by 36 · see citing works
Size‐adaptable 13‐channel receive array for brain MRI in human neonates at 3 T
Rios, N. L., Foias, A., Lodygensky, G. A., Dehaes, M., & Cohen‐Adad, J. (2018) · NMR in Biomedicine · doi:10.1002/nbm.3944 · Cited by 33 · see citing works
MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging
Heußer, T., Rank, C. M., Berker, Y., Freitag, M. T., & Kachelrieß, M. (2017) · EJNMMI Physics · doi:10.1186/s40658-017-0177-4 · Cited by 31 · see citing works
Oehmigen, M., Lindemann, M. E., Lanz, T. E., Kinner, S., & Quick, H. H. (2016) · Medical Physics · doi:10.1118/1.4959546 · Cited by 29 · see citing works
Self-citationMultimodal phantoms for clinical PET/MRI
Lennie, E., Tsoumpas, C., & Sourbron, S. P. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00408-0 · Cited by 26 · see citing works
Pitfalls and Limitations in Simultaneous PET/MRI
Delso, G., ter Voert, E. E. G. W., Barbosa, F. D. G., & Veit‐Haibach, P. (2015) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2015.04.002 · Cited by 22 · see citing works
Technical Note: Human tissue‐equivalent MRI phantom preparation for 3 and 7 Tesla
Woletz, M., Roat, S., Hummer, A., Tik, M., & Windischberger, C. (2021) · Medical Physics · doi:10.1002/mp.14986 · Cited by 15 · see citing works
Keller, S. H., Jakoby, B., Svalling, S., Kjær, A., Højgaard, L., & Klausen, T. L. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0146-3 · Cited by 15 · see citing works
Optimization of yttrium-90 PET for simultaneous PET/MR imaging: A phantom study
Eldib, M., Oesingmann, N., Faul, D. D., Kostakoğlu, L., Knešaurek, K., & Fayad, Z. A. (2016) · Medical Physics · doi:10.1118/1.4958958 · Cited by 13 · see citing works
Teuho, J., Johansson, J. J., Lindén, J., Hansen, A. E., Holm, S., Keller, S. H., Delso, G., Veit‐Haibach, P., Magota, K., Saunavaara, V., et al. (2016) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.166165 · Cited by 12 · see citing works
Rausch, I., Valladares, A., Sundar, L. K. S., Beyer, T., Hacker, M., Meyerspeer, M., & Unger, E. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00364-9 · Cited by 11 · see citing works
The Role of PET/MR Imaging in Precision Medicine
Huo, E. J., Wilson, D. M., Eisenmenger, L. B., & Hope, T. A. (2017) · PET Clinics · doi:10.1016/j.cpet.2017.05.006 · Cited by 9 · see citing works
Catana, C., Laforest, R. A., An, H., Boada, F. E., Cao, T., Faul, D. D., Jakoby, B., Jansen, F. P., Kemp, B. J., Kinahan, P. E., et al. (2021) · Journal of Nuclear Medicine · doi:10.2967/jnumed.120.261881 · Cited by 8 · see citing works
Keller, S. H., Jakoby, B., Hansen, A. E., Svalling, S., & Klausen, T. L. (2015) · EJNMMI Physics · doi:10.1186/2197-7364-2-s1-a28 · Cited by 8 · see citing works
Harmonization of PET image reconstruction parameters in simultaneous PET/MRI
Laforest, R. A., Khalighi, M. M., Natsuaki, Y., Rajagopal, A., Chandramohan, D., Byrd, D. W., An, H., Larson, P. E. Z., James, S. S., Sunderland, J. J., et al. (2021) · EJNMMI Physics · doi:10.1186/s40658-021-00416-0 · Cited by 7 · see citing works
Min, S.-H., Kwon, O., Sattorov, M. A., Kim, S., Baek, I., Park, S., Barik, R. K., Bera, A., Hong, D., Kim, S., et al. (2021) · Defence Technology · doi:10.1016/j.dt.2021.10.014 · Cited by 6 · see citing works
PET/MRI: Attenuation Correction
Delso, G., & Nuyts, J. L. (2018) · Springer eBooks · doi:10.1007/978-3-319-68517-5_4 · Cited by 6 · see citing works
Park, C., Kang, S.-H., & Lee, Y. J. (2021) · Diagnostics · doi:10.3390/diagnostics11020319 · Cited by 6 · see citing works
Talalwa, L., Natour, G., Bauer, A., Drzezga, A. E., Gordji-Nejad, A., & Beer, S. (2020) · Materials Research Express · doi:10.1088/2053-1591/abc76f · Cited by 6 · see citing works
Teuho, J., Johansson, J. J., Lindén, J., Saunavaara, V., Tolvanen, T., & Teräs, M. (2013) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nssmic.2013.6829260 · Cited by 5 · see citing works
Evaluation of 3D printable rubber-elastomeric polymer as phantom material for Hybrid PET/MRI
Talalwa, L., Gordji-Nejad, A., Natour, G., Drzezga, A. E., Bauer, A., & Beer, S. (2019) · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · doi:10.1109/nss/mic42101.2019.9059949 · Cited by 4 · see citing works
Lindemann, M. E., Gratz, M., Blumhagen, J. O., Jakoby, B., & Quick, H. H. (2022) · Medical Physics · doi:10.1002/mp.15446 · Cited by 4 · see citing works
Self-citationAdvanced quantitative evaluation of PET systems using the ACR phantom and NiftyPET software
Markiewicz, P. J., da Costa-Luis, C. O., Dickson, J. C., Barnes, A., Krokos, G., Mackewn, J. E., Clark, T., Wimberley, C., Macnaught, G., Yaqub, M. M., et al. (2022) · Medical Physics · doi:10.1002/mp.15596 · Cited by 4 · see citing works
Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm
Calatayud-Jordán, J., Vela, N. C., Chimeno-Hernández, J., Carles-Fariña, M., Arroyo, C. O., Bello-Arqués, P., Pérez-Enguix, D., Martí‐Bonmatí, L., & Torres-Espallardo, I. (2024) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-024-01452-7 · Cited by 3 · see citing works
Integrated PET/MRI in preclinical studies State of the art
Brunotte, F., Hein, H.-W., Collin, B., Oudot, A., Bricq, S., Lalande, A., Tizon, X., Vrigneaud, J., & WALKER, P. M. (2013) · HAL · Cited by 2 · see citing works
Rovelli, A., Brodie, J., Rashid, B., Tay, W. J., & Pini, R. (2024) · Transport in Porous Media · doi:10.1007/s11242-024-02146-0 · Cited by 2 · see citing works
Jeljeli, S., Cook, G., Hammers, A., Chiribiri, A., Rogers, H. J., Kinsella, S., Krokos, G., & Neji, R. (2025) · Springer eBooks · doi:10.1007/978-3-031-86228-1_6 · Cited by 1 · see citing works
Rothe, M., Riedel, S., Slawig, A., Deistung, A., Bohndorf, K., Brill, R., Wohlgemuth, W. A., & Gussew, A. (2025) · Magnetic Resonance in Medicine · doi:10.1002/mrm.70099 · Cited by 1 · see citing works
Recent progress on imaging technology and performance testing of PET/MR
Liu, J., & Geng, J. (2023) · Radiation Detection Technology and Methods · doi:10.1007/s41605-022-00376-8 · Cited by 1 · see citing works
Evaluation of MR-based PET Images with a Homogeneous μmap by Using Integrated PET/MR
Park, C., Kim, J. S., Baek, B., Kim, S. H., & Yang, H.-J. (2018) · Journal of the Korean Physical Society · doi:10.3938/jkps.73.671
- Authors
- Ziegler, S., Braun, H., Ritt, P., Hocke, C., Kuwert, T., & Quick, H. H.
- Affiliations
- Friedrich-Alexander-Universität Erlangen-Nürnberg (DE)
- Universitätsklinikum Erlangen (DE)
- Venue
- Journal of Nuclear Medicine
- Publisher
- Society of Nuclear Medicine
- Published
- 2013-06-21
- Volume
- 54
- Issue
- 8
- Pages
- 1464-1471
- ISSN
- 0161-5505 (ISSN-L)
- 2159-662X
- OpenAlex
- W2163030916
- Cited by
- 50
- References
- 25
- Citations by year
- 2025: 2 · 2024: 2 · 2023: 2 · 2022: 3 · 2021: 7 · 2020: 1 · 2019: 2 · 2018: 3 · 2017: 4 · 2016: 8 · 2015: 8 · 2014: 2 · 2013: 5 · 2012: 1
- FWCI
- 5.92
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Nuclear Medicine (publishedVersion)
- PubMed (publishedVersion)
- Universitätsbibliographie, Universität Duisburg-Essen (submittedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Advanced MRI Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiopharmaceutical Chemistry and Applications
- Keywords
- PET/MR imaging, phantom fluid, 18F-FDG, triethylene glycol, water-oil emulsions
- MeSH terms
- Feasibility Studies, Fluorides, Image Processing, Computer-Assisted, Magnetic Resonance Imaging, Time Factors, Phantoms, Imaging, Fluorodeoxyglucose F18, Positron-Emission Tomography
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review