Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus
Open accessThe DOI matches its Crossref record; 28 of the 28 references it cites were found in a registry or an index.
Abstract
Objective: To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods. Research design and methods: 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed with StatStrip as semigold standard. The glucose response after a standardized oral glucose load was measured by FSL-CGM and capillary samples analyzed by perchloric acid hexokinase (PCA-HK) method, StatStrip and FSL test strip (FSLC), and a commonly used CGM system (iPro2). Results: FSL-CGM arm sensor readings showed 85.5% of paired readings falling within Clarke Error Grid (ISO 15197:2013) zone A when compared with StatStrip. For FSL-CGM abdomen and FSLC, these percentages were 64% and 98%, respectively. The overall correlation of FSL-CGM in the arm and the StatStrip indicates a performance with lower results with the FSL-CGM in the arm than expected based on the StatStrip in the lower glucose ranges, and higher results than expected in the higher ranges. Following a standardized glucose load, a slower rise in glucose level was observed for FSL-CGM arm as compared with PCA-HK, StatStrip, FSLC, and iPro2 during the first 45–60 min after glucose load ingestion. Conclusions: Certain matters need attention while using the FSL-CGM in daily life including the observed lower values in the lower ranges, and the underestimation of the effect of a meal on glucose response. These effects of such deviations can partly be overcome by optimizing the available user instructions. Trial registration number: TC5348; results.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · cited 1×
“Outcomes from FSL-CGM and capillary blood glucose readings were superimposed on the error grids as described by Clarke et al.(17) The point accuracy of sensor-based glucose values versus finger stick blood glucose was determined as per cent within consensus error grid zone A.”
MethodContinuous Glucose Monitoring: A Review of Successes, Challenges, and Opportunities
Rodbard, D. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0417 · Cited by 737
Klonoff, D. C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1231 · Cited by 724
The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
Bailey, T., Bode, B. W., Christiansen, M. P., Klaff, L. J., & Alva, S. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0378 · Cited by 685 · cited 1×
“Independent accuracy assessments of the FSL-CGM are scarce; previous assessments were performed comparing the FSL-CGM to capillary blood evaluated with a point of care (POC) measurement.(10) Whether such a comparison can be seen as sufficient accurate remains to be seen, however.”
MethodPrice, M. J., Ades, A., Soldan, K., Welton, N. J., Macleod, J., Simms, I., DeAngelis, D., Turner, K. M., & Horner, P. J. (2016) · Health Technology Assessment · doi:10.3310/hta20220 · Cited by 494
Reliability of point-of-care testing for glucose measurement in critically ill adults*
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000189939.10881.60 · Cited by 395
Continuous glucose monitoring systems for type 1 diabetes mellitus
Langendam, M., Luijf, Y. M., Hooft, L., DeVries, J. H., Mudde, A. H., & Scholten, R. J. (2012) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd008101.pub2 · Cited by 262 · cited 1×
“While improved glycemic control has been demonstrated with the use of CGM systems, 3 – 5 CGM accuracy also remains a challenge; most of the available systems need calibrating at least twice daily to allow a suf fi ciently reliable correlation between interstitial and capillary glucose results.”
BackgroundDelays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300528 · Cited by 198 · cited 1×
“Further evaluation is needed to identify the proper target population most likely to bene fi t from the FSL-CGM.”
BackgroundPoolsup, N., Suksomboon, N., & Kyaw, A. M. (2013) · Diabetology & Metabolic Syndrome · doi:10.1186/1758-5996-5-39 · Cited by 179
Accuracy of bedside glucose measurement from three glucometers in critically ill patients*
Hoedemaekers, C. W. E., Klein Gunnewiek, J. M. T., Prinsen, M. A., Willems, J. L., & Van der Hoeven, J. G. (2008) · Critical Care Medicine · doi:10.1097/ccm.0b013e318186ffe6 · Cited by 165
Cited by
Charleer, S., De Block, C., Van Huffel, L., Broos, B., Fieuws, S., Nobels, F., Mathieu, C., & Gillard, P. (2020) · Diabetes Care · doi:10.2337/dc19-1610 · Cited by 261
Porumb, M., Stranges, S., Pescapè, A., & Pecchia, L. (2020) · Scientific Reports · doi:10.1038/s41598-019-56927-5 · Cited by 196
7. Diabetes Technology: Standards of Medical Care in Diabetes—2019
American Diabetes Association (2019) · Diabetes Care · doi:10.2337/dc19-s007 · Cited by 179
Flash forward: a review of flash glucose monitoring
Leelarathna, L., & Wilmot, E. G. (2018) · Diabetic Medicine · doi:10.1111/dme.13584 · Cited by 167
“[32] has compared the accuracy of the FreeStyle Libre monitor during 14 days of home use and during an oral glucose tolerance test.”
MethodAlva, S., Bailey, T., Brazg, R., Budiman, E. S., Castorino, K., Christiansen, M. P., Forlenza, G., Kipnes, M., Liljenquist, D. R., & Liu, H. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820958754 · Cited by 164
Fokkert, M., van Dijk, P., Edens, M., Barents, E., Mollema, J., Slingerland, R., Gans, R., & Bilo, H. (2019) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-000809 · Cited by 163
Self-citationFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Self-citationCastellana, M., Parisi, C., Di Molfetta, S., Di Gioia, L., Natalicchio, A., Perrini, S., Cignarelli, A., Laviola, L., & Giorgino, F. (2020) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2019-001092 · Cited by 84
“37 38 Thus, while both FGM and RT-CGM can be regarded as effective options for CGM, some patient’s characteristics (eg, hypoglycemia unawareness) and the glycemic goals may lead to consider one over the other device.”
Lin, Y. K., Fisher, S. J., & Pop‐Busui, R. (2020) · Journal of Diabetes Investigation · doi:10.1111/jdi.13290 · Cited by 82
A Century of Diabetes Technology: Signals, Models, and Artificial Pancreas Control
Kovatchev, B. (2019) · Trends in Endocrinology & Metabolism · doi:10.1016/j.tem.2019.04.008 · Cited by 80
Varma, N., Cygankiewicz, I., Turakhia, M. P., Heidbuchel, H., Hu, Y.-F., Chen, L. Y., Couderc, J.-P., Cronin, E. M., Estep, J. D., Grieten, L., et al. (2021) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.120.009204 · Cited by 74
Evaluation of flash glucose monitoring after long-term use: a pediatric survey
Vergier, J., Samper, M., Dalla-Vale, F., Ventura, V., Baucher, F., Joubert, F., Pons, C., De Oliveira, M., Salet, R., Faure-Galon, N., et al. (2019) · Primary Care Diabetes · doi:10.1016/j.pcd.2018.08.004 · Cited by 32
Metadata
- Authors
- Fokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G.
- Venue
- BMJ Open Diabetes Research & Care
- Publisher
- BMJ
- Published
- 2017-02
- Volume
- 5
- Issue
- 1
- Pages
- e000320
- ISSN
- 2052-4897 (ISSN-L)
- OpenAlex
- W2587847154
- Cited by
- 209
- References
- 28
- FWCI
- 19.11
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.