Reliability of point-of-care testing for glucose measurement in critically ill adults*

Kanji, Salmaan, Buffie, Jennifer, Hutton, Brian, Bunting, Peter S., Singh, Avinder, McDonald, Kevin, Fergusson, Dean, McIntyre, Lauralyn A., Hebert, Paul C.·2005·Critical Care Medicine

Abstract

BACKGROUND: Glycemic control is increasingly being recognized as a priority in the treatment of critically ill patients. Titration and monitoring of insulin infusions involve frequent blood glucose measurement to achieve target glucose ranges and prevent adverse events related to hypoglycemia. Therefore, it is imperative that bedside glucose testing methods be safe and accurate. OBJECTIVE: To determine the accuracy and clinical impact of three common methods of bedside point-of-care testing for glucose measurements in critically ill patients receiving insulin infusions. DESIGN: Prospective observational study. SETTING: A 21-bed mixed medical/surgical intensive care unit of a tertiary care teaching hospital. PATIENTS: Thirty consecutive critically ill patients who were vasopressor-dependent (n = 10), had significant peripheral edema (n = 10), or were admitted following major surgery (n = 10). MEASUREMENTS: Findings from three different methods of glucose measurement were compared with central laboratory measurements: (1) glucose meter analysis of capillary blood (fingerstick); (2) glucose meter analysis of arterial blood; and (3) blood gas/chemistry analysis of arterial blood. Patients were enrolled for a maximum of 3 days and had a maximum of nine sets of measurements determined during this time. RESULTS: Clinical agreement with the central laboratory was significantly better with arterial blood analysis (69.9% and 76.5% for glucose meter and blood gas/chemistry analysis, respectively) than with capillary blood analysis (56.8%; p = .039 and .001, respectively). During hypoglycemia, clinical agreement was only 26.3% with capillary blood analysis and 55.6% and 64.9% for glucose meter and blood gas/chemistry analysis of arterial blood (p = .010 and <.001, respectively). Glucose meter analysis of both arterial and capillary blood tended to provide higher glucose values, whereas blood gas/chemistry analysis of arterial blood tended to yield lower glucose values. CONCLUSIONS: The magnitude of the differences in the glucose values offered by the four different methods of glucose measurement led to frequent clinical disagreements regarding insulin dose titration in the context of an insulin infusion protocol for aggressive glucose control.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/01.ccm.0000189939.10881.60

Cites

13 of 13 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 13 references

Cited by

See all 395

Metadata

Authors
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C.
Venue
Critical Care Medicine
Published
2005-12
Volume
33
Issue
12
Pages
2778-2785
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2008504588
Cited by
395
References
13
FWCI
13.50
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Reliability of point-of-care testing for glucose measurement in critically ill adults*. https://app.cochat.ai/papers/10.1097/01.ccm.0000189939.10881.60
The paper
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C. (2005). Reliability of point-of-care testing for glucose measurement in critically ill adults*. Critical Care Medicine. https://doi.org/10.1097/01.ccm.0000189939.10881.60

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.