Performance of a non-contact infrared thermometer in healthy newborns
Open accessThe DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
Objective: To evaluate the performance of a non-contact infrared thermometer (NCIT) in comparison with digital axillary thermometer (DAT) and infrared tympanic thermometers (ITT) in a population of healthy at term and preterm newborns nursed in incubators. Setting: 1 level III maternity hospital, and its intensive neonatal care unit. Participants: 119 healthy at term newborns and 70 preterm newborns nursed in incubators were consecutively enrolled. Exclusion criteria were unstable/critical conditions, polymalformative congenital syndromes and severe congenital syndromes. Interventions: Body temperature readings were prospectively collected. Each participant underwent bilateral axillary temperature measurement with DAT, bilateral tympanic measurement with ITT and mid-forehead temperature measurements using NCIT. Primary outcome measures: Degree of agreement between methods was evaluated by the Bland and Altman method. Results: 714 measurements in 119 healthy at term newborns and 420 measurements in 70 preterm newborns nursed in incubators were performed. Clinical reproducibility of NCIT was 0.0455°C for infants in incubators and 0.0861°C for infants outside an incubator. Bias was 0.029°C for infants in incubators and <0.0001°C for infants outside an incubator. Zero outliers were recorded. The mean difference between methods was good both for newborns at term (0.12°C for NCIT vs DAT and 0.02°C for NCIT vs ITT) and preterm newborns in incubators (0.10°C for NCIT vs DAT and 0.14°C for NCIT vs ITT). Limits of agreement were 0.99 to −0.75 and 0.78 to −0.75 in at term newborns and were particularly satisfactory in preterm newborns in incubators (95% CI: 0.48 to −0.27 and 0.68 to −0.40). Conclusions: Our results with Bland and Altman analysis demonstrate that NCIT is a very promising tool, especially in preterm newborns nursed in incubators. Trial registration: The study was approved by the Careggi University Hospital Ethics Committee (07/2011).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Understanding bland altman analysis
Giavarina, D. (2015) · Biochemia Medica · doi:10.11613/bm.2015.015 · Cited by 4,706
Dodd, S. R., Lancaster, G. A., Craig, J. V., Smyth, R. L., & Williamson, P. R. (2006) · Journal of Clinical Epidemiology · doi:10.1016/j.jclinepi.2005.10.004 · Cited by 90
Cited by
Liu, Y., Liu, C., Gao, M., Wang, Y., Bai, Y., Xu, R., & Gong, R. (2020) · BMJ Open · doi:10.1136/bmjopen-2020-039474 · Cited by 35
Hayward, G., Verbakel, J. Y., Ismail, F. A., Edwards, G., Wang, K., Fleming, S., Holtman, G. A., Glogowska, M., Morris, E., Curtis, K., et al. (2020) · British Journal of General Practice · doi:10.3399/bjgp20x708845 · Cited by 26
Digital Axillary and Non-Contact Infrared Thermometers for Children
Franconi, I., La Cerra, C., Marucci, A. R., Petrucci, C., & Lancia, L. (2018) · Clinical Nursing Research · doi:10.1177/1054773816676538 · Cited by 13
Dante, A., Franconi, I., Marucci, A. R., Alfes, C. M., & Lancia, L. (2020) · Journal of Pediatric Nursing · doi:10.1016/j.pedn.2019.11.014 · Cited by 13
Metadata
- Authors
- Sollai, S., Dani, C., Berti, E., Fancelli, C., Galli, L., de Martino, M., & Chiappini, E.
- Venue
- BMJ Open
- Publisher
- BMJ
- Published
- 2016-03
- Volume
- 6
- Issue
- 3
- Pages
- e008695
- ISSN
- 2044-6055 (ISSN-L)
- OpenAlex
- W2324199789
- Cited by
- 49
- References
- 3
- FWCI
- 3.42
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Performance of a non-contact infrared thermometer in healthy newborns. https://app.cochat.ai/papers/10.1136/bmjopen-2015-008695
- The paper
- Sollai, S., Dani, C., Berti, E., Fancelli, C., Galli, L., de Martino, M., & Chiappini, E. (2016). Performance of a non-contact infrared thermometer in healthy newborns. BMJ Open. https://doi.org/10.1136/bmjopen-2015-008695
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.