Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid–liquid extraction method for the ABSCIEX API-5000 mass spectrometry system
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
Aims: Accurate serum aldosterone determination is critical to the screening and diagnosis of primary aldosteronism, the localisation of aldosterone producing tumours, and the investigation of other disorders of the renin-angiotensin system. Mass spectrometry offers a means to overcome problems with method-dependent bias between competitive immunoassays for aldosterone. The authors have developed a simple, sensitive and precise liquid–liquid extraction aldosterone method for the ABSCIEX API-5000 liquid chromatography and tandem mass spectrometry (LC-MS/MS) system. Methods: Using d7-aldosterone internal standard, 500 μl of sample is extracted with 2500 μl of methyl tertbutyl ether followed by dry-down, reconstitution and LC-MS/MS analysis in ESI negative mode. Method validation was undertaken using standard approaches and comparison made against a commercial radioimmunoassay. Accuracy was assessed using EQA material with assigned aldosterone concentrations. Results: The assay was linear up to 3420 pmol/l (LOQ=50 pmol/l, LOD<22 pmol/l). Total CVs were ≤5% for concentrations ≥120 pmol/l and 10% at the LOQ. Mean accuracy was 98.5% against GCMS assigned material. Conclusion: The authors present a precise, sensitive and simple aldosterone method suitable for routine clinical use that requires no solid phase extraction or specialised ion sources.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Funder, J. W., Carey, R. M., Fardella, C., Gomez-Sanchez, C. E., Mantero, F., Stowasser, M., Young, W. F., & Montori, V. M. (2008) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2008-0104 · Cited by 1,554 · cited 2×
“As a matter of practicality, we have defined our lower reportable limit at LOQ (S:N1⁄410:1) of 50 pmol/l (CVz10%), a concentration well below any diagnostic threshold used in routine clinical practice.(1) Our LOD (S:N1⁄43:1) is <22 pmol/l, which is slightly better than the 28 pmol/l value observed using APPI.”
MethodPassing, H., & Bablok, W. (1983) · cclm · doi:10.1515/cclm.1983.21.11.709 · Cited by 1,513
Approach to the patient with an adrenal incidentaloma
Nieman, L. K. (2010) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-0457 · Cited by 290
Toward Excellence in Testosterone Testing: A Consensus Statement
Rosner, W., & Vesper, H. (2010) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-1314 · Cited by 248
Cited by
Liquid chromatography–tandem mass spectrometry for clinical diagnostics
Thomas, S. N., French, D., Jannetto, P. J., Rappold, B. A., & Clarke, W. A. (2022) · Nature Reviews Methods Primers · doi:10.1038/s43586-022-00175-x · Cited by 229
Eisenhofer, G., Peitzsch, M., Kaden, D., Langton, K., Pamporaki, C., Masjkur, J., Tsatsaronis, G., Mangelis, A., Williams, T. A., Reincke, M., et al. (2017) · Clinica Chimica Acta · doi:10.1016/j.cca.2017.05.002 · Cited by 171
Eisenhofer, G., Dekkers, T., Peitzsch, M., Dietz, A. S., Bidlingmaier, M., Treitl, M., Williams, T. A., Bornstein, S. R., Haase, M., Rump, L. C., et al. (2016) · Clinical Chemistry · doi:10.1373/clinchem.2015.251199 · Cited by 158
LC–MS/MS analysis of steroids in the clinical laboratory
Keevil, B. G. (2016) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2016.04.009 · Cited by 149
Laboratory challenges in primary aldosteronism screening and diagnosis
Rehan, M., Raizman, J. E., Cavalier, E., Don-Wauchope, A. C., & Holmes, D. T. (2015) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2015.01.003 · Cited by 89
Self-citation
Metadata
- Authors
- Van Der Gugten, J. G., Dubland, J., Liu, H.-F., Wang, A., Joseph, C., & Holmes, D. T.
- Venue
- Journal of Clinical Pathology
- Publisher
- BMJ
- Published
- 2012-03-12
- Volume
- 65
- Issue
- 5
- Pages
- 457-462
- ISSN
- 0021-9746 (ISSN-L)
- 1472-4146
- OpenAlex
- W2116813588
- Cited by
- 55
- References
- 4
- FWCI
- 2.99
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid–liquid extraction method for the ABSCIEX API-5000 mass spectrometry system. https://app.cochat.ai/papers/10.1136/jclinpath-2011-200564
- The paper
- Van Der Gugten, J. G., Dubland, J., Liu, H.-F., Wang, A., Joseph, C., & Holmes, D. T. (2012). Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid–liquid extraction method for the ABSCIEX API-5000 mass spectrometry system. Journal of Clinical Pathology. https://doi.org/10.1136/jclinpath-2011-200564
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.