Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis

Open access
Yokoo, Takeshi, Serai, Suraj D., Pirasteh, Ali, Bashir, Mustafa R., Hamilton, Gavin, Hernando, Diego, Hu, Houchun H., Hetterich, Holger, Kühn, Jens-Peter, Kukuk, Guido M., Loomba, Rohit, Middleton, Michael S., Obuchowski, Nancy A., Song, Ji Soo, Tang, An, Wu, Xinhuai, Reeder, Scott B., Sirlin, Claude B., For the RSNA-QIBA PDFF Biomarker Committee·2018·Radiology

Abstract

Purpose To determine the linearity, bias, and precision of hepatic proton density fat fraction (PDFF) measurements by using magnetic resonance (MR) imaging across different field strengths, imager manufacturers, and reconstruction methods. Materials and Methods This meta-analysis was performed in accordance with Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A systematic literature search identified studies that evaluated the linearity and/or bias of hepatic PDFF measurements by using MR imaging (hereafter, MR imaging–PDFF) against PDFF measurements by using colocalized MR spectroscopy (hereafter, MR spectroscopy-PDFF) or the precision of MR imaging–PDFF. The quality of each study was evaluated by using the Quality Assessment of Studies of Diagnostic Accuracy 2 tool. De-identified original data sets from the selected studies were pooled. Linearity was evaluated by using linear regression between MR imaging–PDFF and MR spectroscopy-PDFF measurements. Bias, defined as the mean difference between MR imaging–PDFF and MR spectroscopy-PDFF measurements, was evaluated by using Bland-Altman analysis. Precision, defined as the agreement between repeated MR imaging–PDFF measurements, was evaluated by using a linear mixed-effects model, with field strength, imager manufacturer, reconstruction method, and region of interest as random effects. Results Twenty-three studies (1679 participants) were selected for linearity and bias analyses and 11 studies (425 participants) were selected for precision analyses. MR imaging–PDFF was linear with MR spectroscopy-PDFF (R2 = 0.96). Regression slope (0.97; P < .001) and mean Bland-Altman bias (−0.13%; 95% limits of agreement: −3.95%, 3.40%) indicated minimal underestimation by using MR imaging–PDFF. MR imaging–PDFF was precise at the region-of-interest level, with repeatability and reproducibility coefficients of 2.99% and 4.12%, respectively. Field strength, imager manufacturer, and reconstruction method each had minimal effects on reproducibility. Conclusion MR imaging–PDFF has excellent linearity, bias, and precision across different field strengths, imager manufacturers, and reconstruction methods. © RSNA, 2017 Online supplemental material is available for this article. An earlier incorrect version of this article appeared online. This article was corrected on October 2, 2017.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1148/radiol.2017170550

Cites

87 of 87 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 87 references

Cited by

See all 338

Metadata

Authors
Yokoo, T., Serai, S. D., Pirasteh, A., Bashir, M. R., Hamilton, G., Hernando, D., Hu, H. H., Hetterich, H., Kühn, J.-P., Kukuk, G. M., Loomba, R., Middleton, M. S., Obuchowski, N. A., Song, J. S., Tang, A., Wu, X., Reeder, S. B., Sirlin, C. B., & For the RSNA-QIBA PDFF Biomarker Committee
Venue
Radiology
Published
2018-02
Volume
286
Issue
2
Pages
486-498
ISSN
  • 0033-8419 (ISSN-L)
  • 1527-1315
OpenAlex
W2755909350
Cited by
338
References
87
FWCI
4.39
Type
journal article
Language
en
Open access
green

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis. https://app.cochat.ai/papers/10.1148/radiol.2017170550
The paper
Yokoo, T., Serai, S. D., Pirasteh, A., Bashir, M. R., Hamilton, G., Hernando, D., Hu, H. H., Hetterich, H., Kühn, J.-P., Kukuk, G. M., Loomba, R., Middleton, M. S., Obuchowski, N. A., Song, J. S., Tang, A., Wu, X., Reeder, S. B., Sirlin, C. B., & For the RSNA-QIBA PDFF Biomarker Committee. (2018). Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis. Radiology. https://doi.org/10.1148/radiol.2017170550

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.