Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis
Open accessAbstract
Purpose To determine the linearity, bias, and precision of hepatic proton density fat fraction (PDFF) measurements by using magnetic resonance (MR) imaging across different field strengths, imager manufacturers, and reconstruction methods. Materials and Methods This meta-analysis was performed in accordance with Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A systematic literature search identified studies that evaluated the linearity and/or bias of hepatic PDFF measurements by using MR imaging (hereafter, MR imaging–PDFF) against PDFF measurements by using colocalized MR spectroscopy (hereafter, MR spectroscopy-PDFF) or the precision of MR imaging–PDFF. The quality of each study was evaluated by using the Quality Assessment of Studies of Diagnostic Accuracy 2 tool. De-identified original data sets from the selected studies were pooled. Linearity was evaluated by using linear regression between MR imaging–PDFF and MR spectroscopy-PDFF measurements. Bias, defined as the mean difference between MR imaging–PDFF and MR spectroscopy-PDFF measurements, was evaluated by using Bland-Altman analysis. Precision, defined as the agreement between repeated MR imaging–PDFF measurements, was evaluated by using a linear mixed-effects model, with field strength, imager manufacturer, reconstruction method, and region of interest as random effects. Results Twenty-three studies (1679 participants) were selected for linearity and bias analyses and 11 studies (425 participants) were selected for precision analyses. MR imaging–PDFF was linear with MR spectroscopy-PDFF (R2 = 0.96). Regression slope (0.97; P < .001) and mean Bland-Altman bias (−0.13%; 95% limits of agreement: −3.95%, 3.40%) indicated minimal underestimation by using MR imaging–PDFF. MR imaging–PDFF was precise at the region-of-interest level, with repeatability and reproducibility coefficients of 2.99% and 4.12%, respectively. Field strength, imager manufacturer, and reconstruction method each had minimal effects on reproducibility. Conclusion MR imaging–PDFF has excellent linearity, bias, and precision across different field strengths, imager manufacturers, and reconstruction methods. © RSNA, 2017 Online supplemental material is available for this article. An earlier incorrect version of this article appeared online. This article was corrected on October 2, 2017.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Preferred Reporting Items for Systematic Reviews and Meta-Analyses: the PRISMA statement
Moher, D., Liberati, A., Tetzlaff, J., Altman, D. G., & for the PRISMA Group (2009) · BMJ · doi:10.1136/bmj.b2535 · Cited by 83,823
Bias in meta-analysis detected by a simple, graphical test
Egger, M., Smith, G. D., Schneider, M., & Minder, C. (1997) · BMJ · doi:10.1136/bmj.315.7109.629 · Cited by 56,722
QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies
Whiting, P. F., Rutjes, A. W. S., Westwood, M. E., Mallett, S., Deeks, J. J., Reitsma, J. B., Leeflang, M. M. G., Sterne, J. A. C., Bossuyt, P. M. M., & the QUADAS-2 Group* (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-8-201110180-00009 · Cited by 14,199
Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement
Moher, D., Liberati, A., Tetzlaff, J., & Altman, D. G. (2010) · International Journal of Surgery · doi:10.1016/j.ijsu.2010.02.007 · Cited by 12,156
Nonalcoholic Fatty Liver Disease
Angulo, P. (2002) · New England Journal of Medicine · doi:10.1056/nejmra011775 · Cited by 5,228
Chalasani, N., Younossi, Z., Lavine, J. E., Diehl, A. M., Brunt, E. M., Cusi, K., Charlton, M., & Sanyal, A. J. (2012) · Hepatology · doi:10.1002/hep.25762 · Cited by 3,803
Whiting, P., Rutjes, A. W., Reitsma, J. B., Bossuyt, P. M., & Kleijnen, J. (2003) · BMC Medical Research Methodology · doi:10.1186/1471-2288-3-25 · Cited by 3,592
Hematological Problems in Pediatric Intensive Care
Revel-Vilk, S., Cox, P., Robitaille, N., & Blanchette, V. (2014) · Pediatric and Adolescent Medicine · doi:10.1159/000350346 · Cited by 3,508
Nonalcoholic fatty liver disease: a spectrum of clinical and pathological severity.
MATTEONI, C., YOUNOSSI, Z., GRAMLICH, T., BOPARAI, N., LIU, Y., & MCCULLOUGH, A. (1999) · Gastroenterology · doi:10.1016/s0016-5085(99)70506-8 · Cited by 3,345
The Natural History of Nonalcoholic Fatty Liver Disease: A Population-Based Cohort Study
Adams, L. A., Lymp, J. F., St. Sauver, J., Sanderson, S. O., Lindor, K. D., Feldstein, A., & Angulo, P. (2005) · Gastroenterology · doi:10.1053/j.gastro.2005.04.014 · Cited by 2,890
Cited by
Noninvasive assessment of liver disease in patients with nonalcoholic fatty liver disease
Castera, L., Friedrich-Rust, M., & Loomba, R. (2019) · Gastroenterology · doi:10.1053/j.gastro.2018.12.036 · Cited by 1,494
Self-citationShukla‐Dave, A., Obuchowski, N. A., Chenevert, T. L., Jambawalikar, S., Schwartz, L. H., Malyarenko, D., Huang, W., Noworolski, S. M., Young, R. J., Shiroishi, M. S., et al. (2019) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.26518 · Cited by 389
Self-citationQuantification of liver fat content with CT and MRI: state of the art
Starekova, J., Hernando, D., Pickhardt, P. J., & Reeder, S. B. (2021) · Radiology · doi:10.1148/radiol.2021204288 · Cited by 320
Liver fat imaging—a clinical overview of ultrasound, CT, and MR imaging
Zhang, Y. N., Fowler, K. J., Hamilton, G., Cui, J. Y., Sy, E. Z., Balanay, M., Hooker, J. C., Szeverenyi, N., & Sirlin, C. B. (2018) · The British Journal of Radiology · doi:10.1259/bjr.20170959 · Cited by 304
Self-citationStine, J. G., Munaganuru, N., Barnard, A., Wang, J. L., Kaulback, K., Argo, C. K., Singh, S., Fowler, K. J., Sirlin, C. B., & Loomba, R. (2021) · Clinical Gastroenterology and Hepatology · doi:10.1016/j.cgh.2020.08.061 · Cited by 283
Self-citation
Metadata
- Authors
- Yokoo, T., Serai, S. D., Pirasteh, A., Bashir, M. R., Hamilton, G., Hernando, D., Hu, H. H., Hetterich, H., Kühn, J.-P., Kukuk, G. M., Loomba, R., Middleton, M. S., Obuchowski, N. A., Song, J. S., Tang, A., Wu, X., Reeder, S. B., Sirlin, C. B., & For the RSNA-QIBA PDFF Biomarker Committee
- Venue
- Radiology
- Published
- 2018-02
- Volume
- 286
- Issue
- 2
- Pages
- 486-498
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2755909350
- Cited by
- 338
- References
- 87
- FWCI
- 4.39
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis. https://app.cochat.ai/papers/10.1148/radiol.2017170550
- The paper
- Yokoo, T., Serai, S. D., Pirasteh, A., Bashir, M. R., Hamilton, G., Hernando, D., Hu, H. H., Hetterich, H., Kühn, J.-P., Kukuk, G. M., Loomba, R., Middleton, M. S., Obuchowski, N. A., Song, J. S., Tang, A., Wu, X., Reeder, S. B., Sirlin, C. B., & For the RSNA-QIBA PDFF Biomarker Committee. (2018). Linearity, Bias, and Precision of Hepatic Proton Density Fat Fraction Measurements by Using MR Imaging: A Meta-Analysis. Radiology. https://doi.org/10.1148/radiol.2017170550
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.