Bone Mineral Density Quantification from Localizer Radiographs: Accuracy and Precision of Energy-integrating Detector CT and Photon-counting Detector CT
Open accessAbstract
Background Bone mineral density (BMD) could be derived from CT localizer radiographs and could potentially enable opportunistic osteoporosis screening. Purpose To assess the accuracy and precision of BMD measurement using two localizer radiographs obtained with energy-integrating detector CT and a single localizer radiograph obtained with photon-counting detector CT. Materials and Methods A calibration phantom and a porcine phantom with lumbar vertebrae were imaged with a dual-energy x-ray absorptiometry (DXA) scanner, a clinical energy-integrating detector CT scanner, and a prototype photon-counting detector CT scanner. Two localizer radiographs at different combinations of tube voltages were obtained with energy-integrating detector CT, and one localizer radiograph was obtained with photon-counting detector CT using different energy thresholds. BMD was calculated for all three approaches and compared with the known specifications in the calibration phantom. In the animal phantom, BMDs from both CT systems were compared with those from the DXA scanner (the reference standard). Accuracy was defined as the measurement error of BMD (ΔBMD), and precision was defined as the coefficient of variation (in percentage). Radiation doses were estimated. Nonparametric tests were applied. Results In the calibration phantom, ΔBMD was smaller with both CT systems compared with the DXA scanner (both P P = .65). The dose-area product in the animal phantom was 4.6 cGy ∙ cm 2 for DXA, 3.5-11.5…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Two prior studies on a prototype PCD-CT scanner in a phantom and in an ex vivo animal experiment have already indicated accurate aBMD measurements with spectral localizer radiographs.
“Two prior studies on a prototype PCD-CT scanner in a phantom and in an ex vivo animal experiment have already indicated accurate aBMD measurements with spectral localizer radiographs.9,11”
From In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT (2025) · cites this paper as [9]
With energy-integrating detector CT, 2 localizer radiographs must be obtained at different tube potential.
“With energy-integrating detector CT, 2 localizer radiographs must be obtained at different tube potential.9”
From In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT (2025) · cites this paper as [9]
Cites
Incidence and Economic Burden of Osteoporosis-Related Fractures in the United States, 2005–2025
Burge, R., Dawson-Hughes, B., Solomon, D. H., Wong, J. B., King, A., & Tosteson, A. (2007) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.061113 · Cited by 4,026 · see citing works
Li, X., Li, X., Li, J., Jiao, X., Jia, X., Zhang, X., Fan, G., Yang, J., & Guo, J. (2020) · Clinical Radiology · doi:10.1016/j.crad.2019.11.008 · Cited by 41 · see citing works
DXA-equivalent quantification of bone mineral density using dual-layer spectral CT scout scans
Laugerette, A., Schwaiger, B. J., Brown, K., Frerking, L. C., Kopp, F. K., Mei, K., Sellerer, T., Kirschke, J., Baum, T., Gersing, A. S., et al. (2019) · European Radiology · doi:10.1007/s00330-019-6005-6 · Cited by 29 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Moser, L. J., Klambauer, K., Diaz Machicado, M. C., Frey, D., Mergen, V., Eberhard, M., Nowak, T., Schmidt, B., Flohr, T., Distler, O., et al. (2025) · Investigative Radiology · doi:10.1097/rli.0000000000001159 · Cited by 8 · see citing works
Self-citation
- Authors
- Nowak, T., Eberhard, M., Schmidt, B., Frey, D., Distler, O., Saltybaeva, N., Alkadhi, H., & Euler, A.
- Affiliations
- University of Zurich (CH)
- Corresponding author
- Euler, André
- ORCID
- Venue
- Radiology
- Published
- 2021-01
- Volume
- 298
- Issue
- 1
- Pages
- 147-152
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W3096943998
- Cited by
- 41
- References
- 3
- Citations by year
- 2026: 4 · 2025: 7 · 2024: 11 · 2023: 9 · 2022: 6 · 2021: 3 · 2020: 1
- FWCI
- 1.46
- Type
- journal article
- Language
- en
- Open access
- green
- www.zora.uzh.ch
- Licence
- cc-by-nc-nd
- Locations
- Radiology (publishedVersion)
- PubMed (publishedVersion)
- Zurich Open Repository and Archive (University of Zurich) (acceptedVersion, cc-by-nc-nd)
- Zurich Open Repository and Archive (University of Zurich) (submittedVersion, cc-by-nc-nd)
- Zurich Open Repository and Archive (University of Zurich)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced X-ray and CT Imaging
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone health and osteoporosis research
- Physical Sciences › Engineering › Biomedical Engineering › Medical Imaging and Analysis
- Keywords
- bone mineral density, energy-integrating detector CT, photon-counting CT, osteoporosis screening, dual-energy X-ray absorptiometry, measurement accuracy, measurement precision, radiation dose, tube voltage, energy threshold, lumbar vertebrae, calibration phantom, opportunistic screening, low-dose imaging, coefficient of variation
- MeSH terms
- Animals, Lumbar Vertebrae, Swine, Tomography, X-Ray Computed, Reproducibility of Results, Absorptiometry, Photon, Bone Density, Photons, Phantoms, Imaging, Models, Animal
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review