Coronary Artery Calcium: Accuracy and Reproducibility of Measurements with Multi–Detector Row CT—Assessment of Effects of Different Thresholds and Quantification Methods

Hong, Cheng, Bae, Kyongtae T., Pilgram, Thomas K. · 2003 · Radiology

Abstract

PURPOSE: To evaluate the effects of different thresholds and quantification methods on the accuracy and reproducibility of coronary calcium measurements with multi-detector row computed tomography (CT). MATERIALS AND METHODS: A cardiac CT phantom containing predetermined calcified cylinders was scanned. Calcium volume and mass were measured at various threshold values ranging from 80 to 230 HU. In 32 patients, two consecutive CT scans were obtained, and the coronary artery calcium score, volume, and mass were measured by one observer at 130- and 90-HU thresholds. Correlation analysis and analysis of variance were performed to evaluate the measurement errors in the phantom study and the interscan variability in the clinical study. RESULTS: In the phantom, mass measurement error varied with threshold and calcium density (P <.01). Mass error was strongly correlated with volume error (r = 0.91, P <.01) but with a much smaller range. In the clinical study, interscan variability of mass measurements was significantly lower than that with other measurement methods for both patients and individual vessels. For the patients, the mean interscan variability of calcium score, volume, and mass at the 130-HU threshold was 20.4%, 13.9%, and 9.3%, respectively. For all methods, interscan variability was not significantly different between the 130- and 90-HU thresholds (P >.05). CONCLUSION: The mass measurement is more accurate, less variable, and more reproducible in coronary calcium quantification than are measurements with other algorithms. Accurate quantification of calcium in each calcified plaque may require that the threshold be set individually, depending on the calcium density.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1148/radiol.2273020369

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 1 references

Cited by

See all 169

Metadata

Authors
Hong, C., Bae, K. T., & Pilgram, T. K.
Venue
Radiology
Published
2003-06
Volume
227
Issue
3
Pages
795-801
ISSN
  • 0033-8419 (ISSN-L)
  • 1527-1315
OpenAlex
W2165088871
Cited by
169
References
1
FWCI
10.60
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Coronary Artery Calcium: Accuracy and Reproducibility of Measurements with Multi–Detector Row CT—Assessment of Effects of Different Thresholds and Quantification Methods. https://app.cochat.ai/papers/10.1148/radiol.2273020369
The paper
Hong, C., Bae, K. T., & Pilgram, T. K. (2003). Coronary Artery Calcium: Accuracy and Reproducibility of Measurements with Multi–Detector Row CT—Assessment of Effects of Different Thresholds and Quantification Methods. Radiology. https://doi.org/10.1148/radiol.2273020369

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.